K093182 · Laboratorios Grifols, S.A. · LHI · Jan 27, 2010 · General Hospital
Device Facts
Record ID
K093182
Device Name
SET GRI-FILL 3.0
Applicant
Laboratorios Grifols, S.A.
Product Code
LHI · General Hospital
Decision Date
Jan 27, 2010
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 880.5440
Device Class
Class 2
Indications for Use
SET GRI-FILL 3.0 6 TO 1 is an ancillary device used as a fluid pathway in conjunction with the Gri-fill 3.0 Pharmacy Compounder and associated 1 way or 2 way transfer sets through which the same substance from up to 6 rubber-stoppered glass vials may be delivered into a final IV container. Equipped with a spike on each line and a 0.2 µm hydrophobic air-filter, it minimizes the formation of aerosols when preparing / dispensing the source substances. Facilitates easy puncture of thick rubber stoppers of small diameter. Provides fast fluid addition and extraction due to the large surface area of the air-filter. This device should not be used with lipids, suspensions or solutions that are incompatible with PVC with DEHP plasticizer. Substances that are known to show incompatibility include, but are not limited to, Paclitaxel, Docetaxel, Etoposide, Carmustine, Propofol, Nitroglycerin, Isosorbide Dinitrate and Diazepam. For information concerning compatibility of substances, please consult the information provided with the substance. This device is intended to be used by trained healthcare personnel. It is restricted to sale by or on the order of a physician.
Device Story
Fluid transfer set; connects 6 minispikes to Gri-fill 3.0 Pharmacy Compounder; enables delivery of substance from up to 6 rubber-stoppered glass vials into single final IV container. Features PVC tubing; 0.2 micron hydrophobic air-filter minimizes aerosol formation; facilitates puncture of thick, small-diameter rubber stoppers; large air-filter surface area enables rapid pressure equalization for fast fluid addition/extraction. Used in pharmacy/clinical settings by trained healthcare personnel. Output is transferred fluid; assists in automated compounding workflows. Benefits include reduced aerosol exposure and efficient multi-vial fluid transfer.
Clinical Evidence
Bench testing only. Functional laboratory bench testing performed under foreseeable operating conditions confirmed correct operation. Chemical and biological testing conducted in accordance with applicable requirements.
Technological Characteristics
Fluid transfer set; PVC tubing with DEHP plasticizer; 6 minispikes; 0.2 micron hydrophobic air-filter. Mechanical design facilitates vial puncture and pressure equalization. No electronic components or software.
Indications for Use
Indicated for use by trained healthcare personnel as a fluid pathway for transferring substances from up to 6 rubber-stoppered glass vials into a final IV container, used with the Gri-fill 3.0 Pharmacy Compounder. Contraindicated for use with lipids, suspensions, or solutions incompatible with PVC with DEHP plasticizer (e.g., Paclitaxel, Docetaxel, Etoposide, Carmustine, Propofol, Nitroglycerin, Isosorbide Dinitrate, Diazepam).
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
K212530 — IVX Fluid Transfer Set · Omnicell, Inc. · Apr 24, 2022
K033682 — SET GRI-FILL 2.0-1 WAY; 2.0-2 WAYS; LUER CONNECTION TUBE · Laboratorios Grifols, S.A. · Dec 17, 2003
Submission Summary (Full Text)
{0}------------------------------------------------
K093182
# GRIFOLS
TECHNICAL EVALUATION # Document: DOCUMENTATION
TED-SET GRI-FILL 3.0 6 to 1 -05
## SECTION 5 - SET GRI-FILL 3.0 6-to-1: 510(k) SUMMARY
JAN 2 7 2010
DATE OF SUBMISSION:
2009-09-24
SUBMITTER NAME: SUBMITTER ADDRESS: Laboratorios Grifols, S.A. C/ Can Guasch, 2 08150 PARETS DEL VALLES BARCELONA SPAIN
Contact person:
Phone: e-mail:
DEVICE TRADE NAME: COMMON NAME: CLASSIFICATION NAME:
PREDICATE DEVICE(S):
Sebastián Gascón C/ Can Guasch, 2 08150 PARETS DEL VALLES + 34 93 571 01 40 sebastian.gascon@grifols.com
SET GRI-FILL 3.0 6 to 1 I.V. FLUID TRANSFER SETS I.V. FLUID TRANSFER SETS (21 CFR 880.5440)
QUICKPIN (Laboratorios Grifols - K082752) FLEBOSET MULTIPLE (Laboratorios Grifols - K040456)
DEVICE DESCRIPTION:
Fluid transfer set consisting of PVC tubing linking 6 minispikes to be used in conjunction with SETS GRI-FILL 3.0 1 Way or 2 Way transfer sets through which the same substance from up to 6 rubber-stoppered glass vials may be delivered into a final IV container. It is equipped with a 0.2 micron hydrophobic air-filter that minimizes the formation of aerosols when preparing and dispensing the source substances. The spike on each line facilitates easy puncture of thick rubber stoppers of small diameter. The device provides fast fluid addition and extraction due to the large surface area of the air-filter that quickly equalizes pressures.
{1}------------------------------------------------
## GRIFOLS
### TECHNICAL EVALUATION DOCUMENTATION
# Document:
TED-SET GRI-FILL 3.0 6 to 1 -05
## SECTION 5 - SET GRI-FILL 3.0 6-to-1: 510(k) SUMMARY
#### INTENDED USE:
SET GRI-FILL 3.0 6 to 1 is a fluid transfer set to be used in conjunction with SETS GRI-FILL 3.0 1 WAY or 2 WAY through which the same substance from up to 6 rubberstoppered glass vials may be delivered into a final IV container.
#### INDICATIONS FOR USE (as stated in the Statement provided in Section 4):
SET GRI-FILL 3.0 6 TO 1 is an ancillary device used as a fluid pathway in conjunction with the Gri-fill 3.0 Pharmacy Compounder and associated 1 way or 2 way transfer sets through which the same substance from up to 6 rubber-stoppered glass vials may be delivered into a final IV container. Equipped with a spike on each line and a 0.2 µm hydrophobic air-filter, it minimizes the formation of aerosols when preparing / dispensing the source substances. Facilitates easy puncture of thick rubber stoppers of small diameter. Provides fast fluid addition and extraction due to the large surface area of the air-filter.
This device should not be used with lipids, suspensions or solutions that are incompatible with PVC with DEHP plasticizer. Substances that are known to show incompatibility include, but are not limited to, Paclitaxel, Docetaxel, Etoposide, Carmustine, Propofol, Nitroglycerin, Isosorbide Dinitrate and Diazepam. For information concerning compatibility of substances, please consult the information provided with the substance.
This device is intended to be used by trained healthcare personnel. It is restricted to sale by or on the order of a physician.
#### SUMMARY OF COMPARISON WITH PREDICATE DEVICE:
In the establishment of substantial equivalence, SET GRI-FILL 3.0 6 to 1 is compared with another vial access device (QUICKPIN) and with IV fluid transfer sets (FLEBOSET MULTIPLE) both previously marketed by Laboratorios Grifols.
As described in Section 12 of this submission, features, technological characteristics, mechanical specifications and bench performance of the proposed device have been compared in detail with those of the predicate devices.
On the basis of this data, we believe that the new 6 to 1 device consists of a combination of the predicate devices incorporating small dimension spikes equivalent to the Quickpin on a multiple path fluid transfer set similar to the Fleboset Multiple. Fluid transfer tubing material on the new device is identical to the fluid transfer tubing material of the predicate transfer set. Differences that may be highlighted between the new and predicate devices include the configuration of the output connectors. All differences have been addressed in the different bench tests performed on the proposed device and are discussed further in Section 12 of this submission.
{2}------------------------------------------------
GRIFOLS
TECHNICAL EVALUATION DOCUMENTATION
# Document: TED-SET GRI-FILL 3.0 6 to 1 -05
### SECTION 5 - SET GRI-FILL 3.0 6-to-1: 510(k) SUMMARY
#### SUMMARY DISCUSSION OF NON-CLINICAL DATA:
All materials used in the construction of SET GRI-FILL 3.0 6 to 1 have been subject to chemical and biological testing in accordance with the applicable requirements taking account of its intended use. Functional laboratory bench testing performed in foreseeable operating conditions showed correct operation of the device as per its intended use.
#### CONCLUSIONS:
We believe the intended use, the indications for use, the functionality and the operation of the SET GRI-FILL 3.0 6 to 1 combines the indications for use, functionality and operation of the mentioned predicate devices QUICKPIN and FLEBOSET MULTIPLE. Technological differences including the different output connector configuration on the transfer set have been addressed and verified by bench-testing to have no adverse influence on the safety and performance of the proposed device. Hence, substantial equivalence of SET GRI-FILL 3.0 6 to 1 with the legally marketed predicate devices may be established.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the text "DEPARTMENT OF HEALTH & HUMAN SERVICES". The text is in a sans-serif font and is black. To the left of the text is a symbol that is made up of three lines.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Mr. Sebastián Gascón Technical Director Laboratorios Grifols, S. A. C/ Can Guasch, 2 08150 Parets Del Valles Barcelona SPAIN
### JAN 2 7 2010
Re: K093182
Trade/Device Name: Set Gri-Fill 3.0 6 to 1 Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: II Product Code: LHI Dated: December 18, 2009 Received: December 24, 2009
Dear Mr. Gascón:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
#### Page 2 -- Mr. Gascón
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820): and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/ CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
h fa.
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control. and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
| GRIFOLS | TECHNICAL EVALUATION DOCUMENTATION |
|-------------|---------------------------------------|
| # Document: | TED-SET GRI-FILL 3.0 6 to 1-04 |
| | SECTION 04 – SET GRI-FILL 3.0 6 TO 1: |
| | INDICATIONS FOR USE STATEMENT |
### Indications for Use
K093182 510(k) Number (if known):
Device Name: SET GRI-FILL 3.0 6 TO 1
Indications For Use:
SET GRI-FILL 3.0 6 TO 1 is an ancillary device used as a fluid pathway in conjunction with the Gri-fill 3.0 Pharmacy Compounder and associated 1 way or 2 way transfer sets through which the same substance from up to 6 rubber-stoppered glass vials may be delivered into a final IV container. Equipped with a spike on each line and a 0.2 um hydrophobic air-filter, it minimizes the formation of aerosols when preparing / dispensing the source substances. Facilitates easy puncture of thick rubber stoppers of small diameter. Provides fast fluid addition and extraction due to the large surface area of the air-filter.
This device should not be used with lipids, suspensions or solutions that are incompatible with PVC with DEHP plasticizer. Substances that are known to show incompatibility include, but are not limited to, Pacifiaxel, Docetaxel, Etoposide, Carmustine, Propofol, Nitroglycerin, Isosorbide Dinitrate and Diazepam. For information concerning compatibility of substances, please consult the information provided with the substance.
This device is intended to be used by trained healthcare personnel. It is restricted to sale by or on the order of a physician.
V Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign. 08)
(Division Sign-UR) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: k093182
Page 1 of
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.