QUICKPIN is a Luer-lock spike used in manual or automated pharmacy compounding for addition and/or extraction of IV substances, including antineoplastics and substances for chemotherapy, from rubber-stoppered containers including multi-dose vials.
Device Story
QUICKPIN is a Luer-lock spike device used in pharmacy settings for the addition or extraction of IV substances, including chemotherapy agents, from rubber-stoppered vials. It features a 0.2-micron hydrophobic air filter designed to minimize aerosol formation during the preparation and dispensing process. The device is intended for use by trained healthcare personnel in manual or automated compounding workflows. By facilitating safe access to multi-dose vials, it assists in the preparation of IV medications. The device is provided sterile and is for single-use only.
Clinical Evidence
Bench testing only. Functional laboratory testing under foreseeable operating conditions verified correct operation. Materials underwent chemical and biological testing per ISO 10993-1 and USP standards.
Technological Characteristics
Luer-lock spike; materials: ABS (spike body), PP (cap), PVC (filter housing), PTFE (filter). Features 0.2-micron hydrophobic air filter. Conforms to ISO 8536-4 (except spike dimensions). Sterile, non-pyrogenic, single-use. No software or electronic components.
Indications for Use
Indicated for trained healthcare personnel to add or extract IV substances, including antineoplastics and chemotherapy agents, from rubber-stoppered vials. Restricted to prescription use.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
{0}------------------------------------------------
SEP 2 9 2008
K082752 (P. 1 of 4)
| GRIFOLS | TECHNICAL EVALUATION DOCUMENTATION |
|-------------|------------------------------------|
| # Document: | TED-QUICKPIN-05 |
SECTION 5 -- QUICKPIN: 510(k) SUMMARY#### INTENDED USE:
QUICKPIN is a Luer-lock spike used in manual or automated pharmacy compoundinq for addition and/or extraction of IV substances, including antineoplastics and substances for chemotherapy, from rubber-stoppered containers including multi-dose vials.
#### SUMMARY OF COMPARISON WITH PREDICATE DEVICE:
In the establishment of substantial equivalence, QUICKPIN is compared with another vial access device (Chemo-Aide Dispensing Pin – Baxter Healthcare Corporation) and with an IV fluid transfer set previously marketed by Laboratorios Grifols (Fleboset Multiple). The following fable summarizes the similarities of the principal technological characteristics and features of both predicate and new devices.
| # | Characteristic / Feature | QUICKPIN | PREDICATES | |
|----|-------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | | | BAXTER<br>Chemo Aide<br>Dispensing Pin<br>K003730 | LAB.GRIFOLS<br>Fleboset Multiple<br>K040456 |
| 1. | Intended use / Claims | QUICKPIN is a Luer-lock<br>spike used in manual or<br>automated pharmacy<br>compounding for addition<br>and/or extraction of IV<br>substances, including<br>antineoplastics and<br>substances for chemotherapy,<br>from rubber-stoppered<br>containers including multi-<br>dose vials. | The Baxter CHEMO-AIDE<br>Dispensing Pin is intended<br>for use in the preparation<br>and dispensing of<br>chemotherapeutic<br>medications from rubber-<br>stoppered vials. | FLEBOSET MULTIPLE<br>is an ancillary device<br>used as fluid pathway<br>through which<br>substances from 6<br>glass source flasks<br>containing the same<br>solution may be<br>continuously delivered<br>for: (a) Pharmacy<br>compounding, when<br>used in conjunction with<br>the GRI-FILL 2.0<br>pharmacy compounding<br>device and associated<br>transfer sets, and (b)<br>I.V. administration,<br>when used in<br>conjunction with a<br>gravity or pump infusion<br>set to channel the<br>solution from the source<br>containers to the<br>infusion set.<br>The device should not<br>be used with lipids. |
| # | Characteristic /<br>Feature | QUICKPIN | PREDICATES<br>BAXTER<br>Chemo Aide<br>Dispensing Pin<br>K003730 | LAB.GRIFOLS<br>Fleboset Multiple<br>K040456 |
| 2. | Indications for use | QUICKPIN is a Luer-lock spike used in manual or automated pharmacy compounding for addition and / or extraction of IV substances, including antineoplastics and substances for chemotherapy, from rubber-stoppered containers including multidose vials. It is equipped with a 0.2 micron hydrophobic air-filter that minimizes the formation of aerosols when preparing and dispensing the substances.<br>This device is intended to be used by trained healthcare personnel. It is restricted to sale by or on the order of a physician | The proposed Baxter CHEMO-AIDE Dispensing Pin is intended for use in the preparation and dispensing chemotherapeutic medications from rubber stoppered vials. | Fleboset multiple is an ancillary device used as fluid pathway through which substances from 6 glass source flasks containing the same solution may be continuously delivered for:<br>(a) Pharmacy compounding, when used in conjunction with the GRI-FILL 2.0 pharmacy compounding device and associated transfer sets, and<br>(b) I.V. administration when used in conjunction with a gravity or pump infusion set to channel the solution from the source containers to the infusion set.<br>The device should not be used with lipids.<br>This device is intended to be used by trained health-care personnel. It is restricted to sale by or on order of a physician. |
| 3. | Technological<br>features:<br>- Sterilization<br>- Contact with patient | Ethylene Oxide<br>NO - Intended for use in manual or automated pharmacy compounding of IV substances including | Radiation - Gamma<br>NO - Intended for use by the pharmacist in the preparation and dispensing of chemotherapeutic | Ethylene Oxide<br>YES - May be used on-line with patient upstream of the gravity administration set or |
| # | Characteristic /<br>Feature | QUICKPIN | PREDICATES<br>BAXTER<br>Chemo Aide<br>Dispensing Pin<br>K003730 | LAB.GRIFOLS<br>Fleboset Multiple<br>K040456 |
| | | antineoplastics and<br>substances for chemotherapy. | medications. | infusion pump. |
| | Filter | YES - 0.2 micron hydrophobic<br>air-filter to minimize the<br>formation of aerosols during<br>preparation and<br>administration of the<br>substances. | YES - 0.22 micron<br>hydrophopic / oleophobic<br>air-venting filter that<br>minimizes potential for<br>aerosolization of<br>chemotherapeutic<br>medications during<br>reconstitution and<br>dispensing process. | YES - 1.2 micron air-<br>venting filter. |
| 4. | Physical, Mechanical<br>and Biological<br>Specifications | Sterile / Non-pyrogenic /<br>Single-use only | Sterile / Non pyrogenic /<br>Single-use only | Sterile / Non pyrogenic<br>Single-use only |
| 5. | Materials | Spike body: ABS (Acrylonitrile<br>butadiene styrene)<br>Luer Protective cap: PP<br>(Polypropylene)<br>Filter housing: PVC<br>Filter: PTFE (Polytetrafluoro-<br>ethylene) | Spike body: ABS + PE<br>(Acrylonitrile butadiene<br>styrene and polyester)<br>Filter housing: ABS + PE<br>(Acrylonitrile butadiene<br>styrene and polyester)<br>Filter: PVDF (Polyvinyldene<br>fluoride) | Spike body: ABS<br>(Acrylonitrile butadiene<br>styrene)<br>Spikeable twist-off<br>connection: PVC<br>(Polyvinylchloride)<br>Filter housing: PP<br>Filter: Acrylic copolymer<br>+ PA |
| 6. | Conformance<br>standards | The spike conforms to<br>applicable requirements of<br>ISO 8536-4 except regarding<br>spike dimensions which have<br>been adapted specifically for<br>small container access.<br><br>Materials used meet the<br>requirements of USP<br>physicochemical tests for<br>plastics and biological tests<br>outlined in standard ISO<br>10993-1. | Materials used meet the<br>requirements of USP<br>physicochemical tests for<br>plastics and biological tests<br>outlined in standard ISO<br>10993-1. | The spike conforms to<br>applicable requirements<br>of ISO 8536-4.<br><br>Materials used meet the<br>requirements of USP<br>physicochemical tests<br>for plastics and<br>biological tests outlined<br>in standard ISO 10993-<br>1. |
{1}------------------------------------------------
K482752 (P. 20P4)
GRIFOLS
# Document:
DOCUMENTATION
TECHNICAL EVALÚATION
ted-quickpin-05
# SECTION 5 - QUICKPIN: 510(k) SUMMARY
{2}------------------------------------------------
K482052 (P.3084)
### GRIFOLS
# Document:
DOCUMENTATION
TECHNICAL EVALUATION
TED-QUICKPIN-05
## SECTION 5- QUICKPIN: 510(k) SUMMARY
From the above table, it can be established that the new device and the predicate device Baxter Chemo-Aide Dispensing Pin have very similar intended uses and indications. Also, the materials used in the construction of the proposed spike and those used in spike component of the predicate device Fleboset Multiple are identical. Some differences may be noted in the
{3}------------------------------------------------
| | K482752 (P.4 of 4) | |
|--------------------------------------|----------------------|-----------------|
| GRIFOLS | TECHNICAL EVALUATION | # Document: |
| | DOCUMENTATION | TED-QUICKPIN-05 |
| SECTION 5 - QUICKPIN: 510(k) SUMMARY | | |
dimensions and type of materials used in the filter component and air-inlet on the spike. These differences have been addressed in the different bench tests performed on the proposed QUICKPIN device.
### SUMMARY DISCUSSION OF NON-CLINICAL DATA:
All materials used in the construction of QUICKPIN have been subject to chemical and biological testing in accordance with the applicable requirements taking account of its intended use.
Functional laboratory testing performed in foreseeable operating conditions showed correct operation of the device as per its intended use,
#### CONCLUSIONS:
We believe the intended use, the indications for use, the functionality and the operation of QUICKPIN and the CHEMO-AIDE predicate device as access devices for rubber stoppered containers are essentially the same. Also, the materials used in the construction the proposed device and those used in the spike component of the FLEBOSET MULTIPLE predicate device are identical. Technological differences including the use of different filter materials and different dimensions on the air-inlet channel on the spike have been addrt fined and verified by bench-testing to have no adverse influence on the safety and performance of the proposed device. Hence, substantial equivalence of QUICKPIN with the legally marketed devices may be established.
{4}------------------------------------------------
Image /page/4/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three stripes forming its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
SEP 2 9 2008
Laboratorios Grifols, S.A. C/O Mr. Norbert Stuiber Responsible Third Party Officer TUV SUD America, Incorporated 1775 Old Highway 8 N.W. New Brighton, Minnesota 55112-1891
Re: K082752
Trade/Device Name: QUICKPIN Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: II Product Code: LHI Dated: September 17, 2008 Received: September 19, 2008
Dear Mr. Stuiber:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
#### Page 2 - Mr. Stuiber
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Chiu Lin, Ph.D.
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
GRIFOLS
TECHNICAL EVALUATION DOCUMENTATION
# Document: TED-QUICKPIN-04
### SECTION 04 - QUICKPIN:
### INDICATIONS FOR USE STATEMENT
#### PREMARKET NOTIFICATION INDICATIONS FOR USE STATEMENT (as required by ODE for all 510(k) received after Jan. 1, 1996)
510(k) Number:
K082752
Device Name:
QUICKPIN
Indications for Use:
QUICKPIN is a Luer-lock spike used in manual or automated pharmacy compounding for addition and / or extraction of IV substances, including antineoplastics and substances for chemotherapy, from rubber-stoppered containers including multidose vials. It is equipped with a 0.2 micron hydrophobic air-filter that minimizes the formation of aerosols when preparing and dispensing the substances.
This device is intended to be used by trained healthcare personnel. It is restricted to sale by or on the order of a physician.
(Do not write below this line. Continue on another page if needed) Concurrence of CDRH, Office of Device Evaluation (ODE)
| Prescription Use <span style="text-decoration: underline;">✓</span><br>(21 CFR 801 Subpart D) | OR | Over-The-Counter Use <span style="text-decoration: underline;">____</span><br>(21 CFR 801 Subpart C) |
|-----------------------------------------------------------------------------------------------|----|------------------------------------------------------------------------------------------------------|
|-----------------------------------------------------------------------------------------------|----|------------------------------------------------------------------------------------------------------|
__________________________
(Division Sign-Off)
Division of Anesthesiology, General Hospital
Infection Control, Dental Devices
| 510(k) Number: | <span style="text-decoration: underline;">K982752</span> |
|----------------|----------------------------------------------------------|
|----------------|----------------------------------------------------------|
9
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.