K040456 · Laboratorios Grifols, S.A. · LHI · Mar 9, 2004 · General Hospital
Device Facts
Record ID
K040456
Device Name
FLEBOSET MULTIPLE
Applicant
Laboratorios Grifols, S.A.
Product Code
LHI · General Hospital
Decision Date
Mar 9, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5440
Device Class
Class 2
Attributes
3rd-Party Reviewed
Indications for Use
FLEBOSET MULTIPLE is an ancillary device used as fluid pathway through which substances from 6 glass source flasks containing the same solution may be continuously delivered for: (a) Pharmacy compounding, when used in conjunction with the GRI-FILL 2.0 pharmacy compounding device and associated transfer sets, and (b) I.V. administration, when used in conjunction with a gravity or pump infusion set to channel the solution from the source containers to the infusion set. The device should not be used with lipids. This device is intended to be used by trained health-care personnel. It is restricted to sale by or on order of a physician.
Device Story
Fluid transfer tubing set; enables continuous delivery of drug solutions from 6 glass source containers. Device consists of 6 spikes connected in series with flexible tubing segments; each segment includes individual clamp; terminal features spikeable twist-off valve connector. Used in pharmacy compounding (with GRI-FILL 2.0 device) or I.V. administration (with gravity/pump infusion sets). Operated by trained healthcare personnel in clinical or pharmacy settings. Minimizes tubing manipulation during small volume source container handling. Provides fluid pathway to infusion sets; facilitates uninterrupted solution delivery. Benefits patient by reducing manual intervention and maintaining continuous fluid supply.
Clinical Evidence
Bench testing only. No clinical data provided. Materials underwent chemical and biological testing per requirements. Functional laboratory testing confirmed correct operation under foreseeable conditions.
Technological Characteristics
Materials: PVC with DEHP plasticizer. Configuration: 6 spikes in series with flexible tubing, individual clamps, and twist-off valve connector. Sterilization: Ethylene Oxide. Connectivity: Standalone fluid pathway. Energy: None (gravity/pump driven).
Indications for Use
Indicated for continuous delivery of identical solutions from 6 glass source flasks for pharmacy compounding or I.V. administration. Intended for use by trained healthcare personnel. Contraindicated for use with lipids.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
Merit Medical Contrast Management System (K961794)
Submission Summary (Full Text)
{0}------------------------------------------------
K040456
3/9/04
## TECHNICAL EVALUATION DOCUMENTATION multiple-01 GRIFOLS SECTION 1 - FLEBOSET MULTIPLE: 510(k) SUMMARY
: 上一
| DATE OF SUBMISSION: | 2004-02-06 |
|---------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| SUBMITTER NAME:<br>SUBMITTER ADDRESS: | Laboratorios Grifols, S.A.<br>C/ Can Guasch, 2<br>08150 PARETS DEL VALLES<br>BARCELONA<br>SPAIN |
| TELEPHONE:<br>FAX:<br>e-mail: | + 34 93 571 01 00<br>+ 34 93 573 09 12<br>sebastian.gascon@grifols.com |
| CONTACT: | Sebastián Gascón<br>Technical Director |
| DEVICE TRADE NAME:<br>COMMON NAME: | FLEBOSET MULTIPLE<br>I.V. FLUID TRANSFER SETS<br>FLUID DELIVERY TUBING |
| CLASSIFICATION NAME: | SET, I.V. FLUID TRANSFER (21 CFR 880.5440, LHI)<br>TUBING. FLUID DELIVERY (21 CFR 880.5440, FPK) |
| PREDICATE DEVICE: | Automix 3+3 Compounder Transfer Set (BAXTER<br>HEALTHCARE CORP.)<br>Medrad Transfer Set (MEDRAD INC.)<br>Merit Medical Contrast Management<br>System<br>(MERIT MEDICAL SYSTEMS, INC.) |
| DEVICE DESCRIPTION: | FLEBOSET MULTIPLE is a fluid transfer tubing set<br>used to enable continuous (uninterrupted) delivery<br>of drug solutions from 6 glass source containers. It<br>is used in pharmacy compounding or for I.V. fluid<br>transfer to minimize tubing manipulation when<br>working with small volume source containers. The<br>device is made up of 6 spikes connected in series<br>with flexible tubing segments, each segment with an<br>individual clamp. The set terminal, for connection to<br>pharmacy compounding device sets or<br>the<br>administration sets (gravity or pump) consists of a<br>spikeable twist-off valve connector. |
a marka mana mana mana marka masa masa mana mana mara mara mar
ાવ 1
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image is a document header with the word GRIFOLS on the left side. The header also contains the words TECHNICAL EVALUATION DOCUMENTATION. The document number is TED-FLEBOSET MULTIPLE-01. The section is titled SECTION 1 - FLEBOSET MULTIPLE: 510(k) SUMMARY.
#### INTENDED USE:
FLEBOSET MULTIPLE is an ancillary device used as fluid pathway through which substances from 6 glass source flasks containing the same solution may be continuously delivered for.
Pharmacy compounding, (a) when used in conjunction with the GRI-FILL 2.0 pharmacy compounding device and associated transfer sets, and
I.V. administration, when used in conjunction with (b) a gravity or pump infusion set to channel the solution from the source containers to the infusion set. The device should not be used with lipids.
#### SUMMARY OF COMPARISON WITH PREDICATE DEVICE:
In the establishment of substantial equivalence, FLEBOSET MULTIPLE is compared with other fluid delivery tubing used in pharmacy compounding and I.V. administration. The following table summarizes the similarities of the principal technological characteristics and features of both predicate and new devices.
| | | | PREDICATES | | |
|----|-----------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------|
| # | Characteristic /<br>Feature | FLEBOSET<br>MULTIPLE | BAXTER<br>Automix 3+3<br>(transfer set<br>presented<br>together with<br>compounder)<br>K894827 | MEDRAD<br>Transfer Set<br>K022431 | MERIT MEDICAL<br>Transfer Sets<br>K961794 |
| 1. | Intended use / Claims | FLEBOSET MULTIPLE is<br>an ancillary device used<br>as fluid pathway through<br>which substances from 6<br>glass<br>source<br>flasks<br>containing<br>the<br>same<br>solution<br>may<br>be<br>continuously delivered for.<br>(a) Pharmacy compoun-<br>ding, when<br>used<br>in<br>conjunction with the GRI-<br>FILL<br>2.0 pharmacy<br>compounding device and<br>associated transfer sets,<br>and (b) I.V. administration,<br>when used in conjunction<br>with a gravity or pump<br>infusion set to channel the<br>solution from the source<br>containers to the infusion | Positive<br>displacement<br>fluid<br>compounding<br>system employing the<br>3+3<br>compounder,<br>TRANSFER<br>SETS,<br>empty<br>Viaflex<br>or<br>Travamulsion containers<br>with<br>connectors and<br>multitask<br>computer<br>software to provide<br>compounding of a wide<br>variety of fluids | The Medrad Transfer<br>Set is medical<br>disposable device used<br>to transfer intravascular<br>contrast media and<br>saline from a spikeable<br>container to a power<br>injector syringe. The<br>device components<br>consist of a vented<br>spike, connector tube, a<br>means of manually<br>stopping flow, female<br>luer, and individually<br>packaged sterile caps. | Contrast media delivery<br>systems used to transfer<br>I.V. contrast media from<br>a spikeable container to<br>a syringe. |
કી
{2}------------------------------------------------
# 230SET MULTIPLE: 510(k) S
| # | Characteristic /<br>Feature | FLEBOSET<br>MULTIPLE | PREDICATES | | |
|----|-----------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------|
| | | | BAXTER<br>Automix 3+3<br>(transfer set<br>presented<br>together with<br>compounder)<br>K894827 | MEDRAD<br>Transfer Set<br>K022431 | MERIT MEDICAL<br>Transfer Sets<br>K961794 |
| | | set. The device should not<br>be used with lipids. | | | |
| 2. | Technological<br>features:<br>-Sterilization<br>-Contact with patient | Ethylene Oxide<br>NO- Intended for use<br>with the pharmacy<br>compounding device or<br>INDIRECT - May be used<br>on-line with patient,<br>upstream of the gravity or<br>pump administration sets. | Radiation<br>NO - Intended for use<br>with the pharmacy<br>compounding device. | Ethylene Oxide<br>NO- Must be removed<br>from the syringe before<br>the syringe can be<br>connected to the<br>patient. | Ethylene Oxide<br>INDIRECT - May be<br>used on-line with the<br>patient through a<br>syringe with stopcock or<br>a manifold port. |
| | -Source solutions | 6 | 3 or 6. | 2 | 1 |
| 3. | Principle Materials | PVC with DEHP plasticizer | PVC with DEHP<br>plasticizer | PVC | Flexible plastic |
| 4. | Sterility | Sterile | Sterile | Sterile | Sterile |
From the above table, it can be established that the new device and the predicate devices are very similar.
# SUMMARY DISCUSSION OF NON-CLINICAL DATA:
All materials used in the construction of FLEBOSET MULTIPLE have been subject to chemical and biological testing in accordance with the applicable requirements taking account of its intended use.
Functional laboratory testing performed in foreseeable operating conditions showed correct operation of the device as per its intended use.
#### CONCLUSIONS:
We believe the intended use, the indications for use, the functionality and the operation of FLEBOSET MULTIPLE and the predicate devices for fluid transfer both in pharmacy compounding and for 1.V. administration are essentially the same. Hence, substantial equivalence of FLEBOSET MULTIPLE with the legally marketed devices may be established.
: :
ਟੀ 3
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of a human figure, represented by three overlapping profiles, with flowing lines suggesting movement or energy. The profiles are arranged to resemble a bird in flight. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular pattern around the figure.
MAR - 9 2004
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Laboratorios Grifols, S.A. C/O Ms. Susan Gill Responsible Third Party Official Underwriters Laboratories, Incorporated 12 Laboratory Drive P.O. Box 13995 Research Triangle Park, North Carolina 27709-3995
Re: K040456
Trade/Device Name: Fleboset Multiple Regulation Number: 880.5440 Regulation Name: Set, Intravenous Fluid Transfer Regulatory Class: II Product Code: LHI Dated: February 20, 2004 Received: February 23, 2004
Dear Ms. Gill:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations offecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 -Ms. Gill
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please contact the Office of Compliance at (301) 594-4618. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours.
Chiu Lin, Ph.D.
Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## ECHNICAL EVALUATION DOCUMENTATION ON 09 FLEBOSET MULTIPLE: INDICATIONS FO A CHE CHENE
### PREMARKET NOTIFICATION INDICATIONS FOR USE STATEMENT
(as required by ODE for all 510(k) received after Jan. 1, 1996)
510(k) Number:
K040456
Device Name:
FLEBOSET MULTIPLE
Indications for Use:
FLEBOSET MULTIPLE is an ancillary device used as fluid pathway through which substances from 6 glass source flasks containing the same solution may be continuously delivered for:
- (a) Pharmacy compounding, when used in conjunction with the GRI-FILL 2.0 pharmacy compounding device and associated transfer sets, and
- (b) I.V. administration, when used in conjunction with a gravity or pump infusion set to channel the solution from the source containers to the infusion set.
The device should not be used with lipids.
This device is intended to be used by trained health-care personnel. It is restricted to sale by or on order of a physician.
(Do not write below this line. Continue on another page in needed)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Arne Nannestad, Interim Branch Chief
(Division Sign-Off)
Division of Anesthesiology, General Hospital, Infection Control, Dental Devices
510(k) Number: K040456
Prescription Use X (Per 21 CFR 801.109)
OR
Over-The-Counter Use
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.