K150611 · Medsource International · FOZ · Jul 15, 2015 · General Hospital
Device Facts
Record ID
K150611
Device Name
MedSource ClearSafe Safety IV Catheter
Applicant
Medsource International
Product Code
FOZ · General Hospital
Decision Date
Jul 15, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5200
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The MedSource ClearSafe Safety IV Catheter is indicated to sample blood or administer fluids intravenously. The MedSource ClearSafe Safety IV Catheter may be used for any patient population with consideration given to adequacy for the vascular anatomy appropriateness for the solution being administered and duration of the therapy.
Device Story
MedSource ClearSafe Safety IV Catheter is a sterile, single-use intravascular catheter for fluid delivery or blood sampling. Device comprises protective needle cover, color-coded catheter hub, radiopaque catheter tube, catheter holder, needle guide, stainless steel needle, porous plug, and flashback chamber. Used in clinical settings by healthcare professionals. Features integrated safety mechanism: user-activated sliding motion with one hand retracts and encloses needle within chamber post-use to prevent accidental needle stick injuries. Output is established intravenous access. Benefits include reduced risk of sharps injury and exposure to patient body fluids for healthcare personnel.
Clinical Evidence
Bench testing only. Performance validated against ISO 10555-1 and ISO 10555-5 standards, including needle geometry, flow rate, corrosion resistance, tensile strength, leakage, and radiopacity. Activation testing (n=500) showed 0% failure rate. Clinical simulation testing with healthcare professionals confirmed ease of use, reliability, and safety feature effectiveness across various hand sizes and conditions (gloved/wet/dry).
Technological Characteristics
Materials: PUR/PTFE catheter, stainless steel needle, ABS hub/flashback chamber, Polyacetal (POM) holder, Poly-ethylene filter. Standards: ISO 10555-1/5, ISO 594-1/2, ASTM F640-07, ASTM D2240-05, ISO 527-1, BS EN 13868. Features: Radiopaque, color-coded, sliding-activation needle shielding mechanism. Sterile, single-use. No software/electronics.
Indications for Use
Indicated for blood sampling or intravenous fluid administration in any patient population, contingent upon vascular anatomy, solution appropriateness, and therapy duration.
Regulatory Classification
Identification
An intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings and that is inserted into the patient's vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravenously. The device may be constructed of metal, rubber, plastic, or a combination of these materials.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized representation of a human figure, with three overlapping profiles suggesting a sense of community and support. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the figure, emphasizing the department's role and national scope.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 15, 2015
MedSource International Ms. Laura Riggen Director, Quality and Regulatory Affairs 4201 Norex Drive Chaska, Minnesota 55318
Re: K150611
Trade/Device Name: MedSource ClearSafe Safety IV Catheter Regulation Number: 21 CFR 880.5200 Regulation Name: Intravascular catheter Regulatory Class: II Product Code: FOZ Dated: June 5, 2015 Received: June 8, 2015
Dear Ms. Riggen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
Tina
Kiang -S
for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K150611
Device Name MedSource ClearSafe Safety IV Catheter
#### Indications for Use (Describe)
The MedSource ClearSafe Safety IV Catheter is indicated to sample blood or administer fluids intravenously. The MedSource ClearSafe Safety IV Catheter may be used for any patient population with consideration given to adequacy for the vascular anatomy appropriateness for the solution being administered and duration of the therapy.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) Summary
Medsource International, LLC Submitter: 4201 Norex Drive Chaska, MN 55318
Contact Person: Laura Riggen, Quality and Regulatory Affairs Manager 4201 Norex Drive Chaska, MN 55318 Phone: 952-472-0131
Date Prepared: Tuesday, July 14, 2015
General Information:
Regulatory Reference: 21 CFR §880.5200 Classification Name: Intravascular Catheter Product Code: FOZ Common Name: Catheter, intravascular, therapeutic, short-term Classification: Class II Panel: General Hospital Proprietary Name: MedSource ClearSafe Safety IV Catheter Single Use: Yes Sterile: Yes Predicate Device: K131555 MedSource IV Safety Catheter
Indication for Use:
The MedSource ClearSafe Safety IV Catheter is indicated to sample blood or administer fluids intravenously. The MedSource ClearSafe Safety IV Catheter may be used for any patient population with consideration given to adequacy for the vascular anatomy appropriateness for the solution being administered and duration of the therapy.
# Description of the Device:
The MedSource ClearSafe Safety IV Catheter is a medical device used for inserting a catheter into a patient's body for purpose of delivery of fluids or drainage of fluids from the patient's body. The device is equipped with a safety feature intended to prevent needle stick injuries. This device is engineered to protect healthcare workers using it against accidental needle stick injury and exposure to parts of the device that have come in contact with patient's blood or other body fluids, and have hence been potentially contaminated with infectious agents. In addition this device also secures the needle (used to insert the catheter) in an enclosed chamber, thus protecting healthcare and other personnel from exposure to patient body fluids after the device has been discarded postuse.
{4}------------------------------------------------
The MedSource ClearSafe Safety IV Catheter is comprised of the following components: protective needle cover, color-coded catheter hub, radiopaque catheter tube, catheter holder, needle guide, medical grade stainless steel needle hub, porous plug and gauge chamber. The components combine for an ergonomic design that incorporate a safety feature that when the device is withdrawn, the needle is completely enclosed within the chamber. The retraction mechanism is activated by the user using only one hand by means of a sliding motion.
The device is available in six gauges identified also by specific colors.
| Part Number | Model Description | Color |
|-------------|-------------------------------------------------|--------|
| MS-84114 | MedSource ClearSafe Safety IV Catheter 14 Gauge | Orange |
| MS-84116 | MedSource ClearSafe Safety IV Catheter 16 Gauge | Gray |
| MS-84118 | MedSource ClearSafe Safety IV Catheter 18 Gauge | Green |
| MS-841182 | MedSource ClearSafe Safety IV Catheter 18 Gauge | Green |
| MS-84120 | MedSource ClearSafe Safety IV Catheter 20 Gauge | Pink |
| MS-841202 | MedSource ClearSafe Safety IV Catheter 20 Gauge | Pink |
| MS-84122 | MedSource ClearSafe Safety IV Catheter 22 Gauge | Blue |
| MS-84124 | MedSource ClearSafe Safety IV Catheter 24 Gauge | Yellow |
Predicate Device Comparison:
The MedSource ClearSafe Safety IV Catheter is substantially equivalent the MedSource Safety Catheter (K131555).
Section 1 Comparison of Technological Characteristics:
| | Submission Device ↓ | Predicate Device ↓ | Comparison ↓ |
|--------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|
| Comparison Point ↓ | MedSource<br>ClearSafe Safety IV<br>Catheter (K150611) | MedSource IV Safety<br>Catheter (K131555) | |
| | Technology | | |
| Distal End<br>configuration | Beveled | Beveled | Same |
| Proximal End<br>configuration | Cupper | Cupper | Same |
| Intended Use | The MedSource<br>ClearSafe Safety IV<br>Catheter is<br>indicated to sample<br>blood or administer<br>fluids<br>intravenously. The<br>MedSource<br>ClearSafe Safety IV<br>Catheter may be<br>used for any patient<br>population with<br>consideration given<br>to adequacy for the<br>vascular anatomy<br>appropriateness for<br>the solution being<br>administered and<br>duration of the<br>therapy. | The MedSource IV<br>Safety Catheter is<br>indicated to sample<br>blood or administer<br>fluids intravenously.<br>The MedSource IV<br>Safety Catheter may<br>be used for any<br>patient population<br>with consideration<br>given to adequacy for<br>the vascular anatomy<br>appropriateness for<br>the solution being<br>administered and<br>duration of the<br>therapy. | Same |
| Needle Stick<br>Prevention Feature | Sliding activation<br>needle<br>Shielding | Push-button needle<br>shielding | The safety feature<br>activation is different<br>but acts in the same<br>fashion. |
| Catheter Length<br>Catheter O.D.<br>Needle Gauge | Gauge Length OD<br>mm<br>mm<br>24G<br>0.7<br>19<br>22G<br>0.9<br>ટર્ટ<br>20G<br>ટર્ટ<br>1.1<br>18G<br>32<br>1.3<br>32<br>16G<br>1.7<br>14G<br>32<br>2.1 | Gauge Length OD<br>mm<br>mm<br>24G<br>0.7<br>19<br>22G<br>25<br>0.9<br>20G<br>ટર્ટ<br>1.1<br>18G<br>ટર્ટ<br>1.3<br>16G<br>30<br>1.7<br>14G<br>2.1<br>તર | Same |
| Catheter | Materials<br>PUR/PTFE | PUR/PTFE | Same |
| Needle | Stainless Steel | Stainless Steel | Same |
| Catheter Body<br>K-Resin | | K-Resin | Same |
| Catheter Holder | Polyacetal (POM) | Polyacetal (POM) | Same |
| Needle Hub<br>ABS | | ABS | Same |
| Activation Lever | NA | Polyacetal (POM) | There is no<br>Activation lever on |
| | | | the Submission<br>Device. |
| Flashback<br>Chamber | ABS | ABS | Same |
| Hydrophobic Filter | Poly-ethylene | Poly-ethylene | Same |
{5}------------------------------------------------
{6}------------------------------------------------
Section 2 Comparison of Non-Clinical Data:
# Non Clinical Tests:
The following tests were performed to show that the MedSource ClearSafe Safety IV Catheter meets the standards set forth in ISO 10555-1: Sterile, single-use intravascular Catheter- General Requirements and ISO 10555-5: Sterile, single-use, intravascular Catheter-Over needle peripheral catheter.
# Needle Point Geometry (Bevel)
Tested per ISO 10555-5:1996(E) Conclusion: Based on the test conditions contained in the Standard, the cannula is subsequently complying with the requirements of ISO 10555-1.
### Gravity Flow
Determination of flow rate through catheter (using water 1, 000kg/m3) measure the efflux not less than 30s, express the average flow rate Tested per ISO10555-5:1996(E) Annex B Conclusion:
As per the test conditions contain herein, the flow rate observed from the catheter is under the specified limit 55 ml/min.
#### Corrosion Resistance Test
Tested per ISO10555-1:1995 Conclusion:
As per the test conditions contain herein, the cannula/needles are checked visually for signs of corrosion and the results are complies the condition as per EN ISO 10555-1.
#### Method for Determining Force at Break
(Hub-tube tensile test, strain rate: 20mm/min/mm. test speed: 400mm/min-min force at break 5N OD 0.75-1.15) Tested Per ISO10555-1:1995 Annex B. Conclusion: As per the test conditions contain herein, the observed value at break is under the specified limit. i.e. Average force at break - 22N.
{7}------------------------------------------------
# Catheter Collapse
Air leakage during aspiration.(Visual inspection made for ingress of air bubbles to the syringe)
Tested per ISO10555-1 Annex D
Conclusion:
As per the test conditions contain herein, the whole assembly is checked for ingress of air bubbles and there was no air bubble formation observed from the hub assembly up to 30sec.
# Gauging Test
Tested per ISO594/1-1986(E) Sec. 5.1 Conclusion: As per the test conditions contain herein, the catheter luer (body) is complies with the requirement of EN 20594-1 as per the steel gauge.
# Liquid Leakage from the Conical Fitting Assembly under Pressure
Tested per ISO594/1-1986(E) Sec. 5.2 Conclusion:
As per the test conditions contain herein, there is no liquid leakage observed from the assembly at 3bar up to 15min.
# Air Leakage from the Conical Fitting Assembly during Aspiration
Tested per EN 20594-1, ISO 10555-1 Annex D
Conclusion:
As per the test conditions contain herein, the whole assembly is checked for ingress of air bubbles and there was no air bubble formation observed from the hub assembly up to 30sec.
# Liquid Leakage under Pressure
(300 to 320 kPa to be generated or 30s) Tested per EN 20594-1 Sec. 5.2. ISO 10555-5:1995(E) Annex C Conclusion:
As per the test conditions contain herein, there is no liquid observed from the assembly at 3bar up to 15min.
# Separation Force of Conical Fitting Assembly
Tested per ISO594-1 Sec. 5.4 Conclusion: As per the test conditions contain herein, the average force 7.4N is observed for the separation of connection, which is under the specified limit.
{8}------------------------------------------------
## Stress Cracking
Tested per ISO594-1 Sec. 5.5, ISO594-2:1998(E) Sec. 5.8 Conclusion: As per the test conditions contain herein, there is no sign of any stress or cracking on the device material as well as no leakage was observed from the conical fitting up to 1min.
#### Unscrewing Torgue of Fitting Assembly
Tested per ISO594-2:1998(E) Conclusion: As per the test conditions contain herein, the average force 0.06 Nm is observed to unscrew the male luer assembly from the female luer assembly.
### Ease of Assembly
Tested Per ISO594-2:1998(E) Sec. 5.6 Conclusion: As per the test conditions contain herein, the male reference fitting inserts without any jerk and fits securely.
### Resistance to Overriding
Tested Per ISO594-2:1998 (E) Sec. 5.7 Conclusion: As per the test conditions contain herein, there are no damage observed during the fitting of male reference fitting to the female luer.
# Radiopacity of Catheter
Tested per ASTM F640-07 Test Result: Passed X- Ray image is as attached.
Image /page/8/Picture/10 description: The image is a blurry, out-of-focus shot of what appears to be a light source or reflection. The background is a muted gray, and the light source is a bright, diffused white. The light is stretched horizontally across the frame, creating a sense of movement or distortion.
# Catheter Elongation Testing
(Determination of Elastic Modulus) Tested per ISO 527-1
{9}------------------------------------------------
Acceptance Criteria: The strength at break should be 25 N for the tube with outer diameter > 1. Conclusion: Passed
### Catheter Flexural Fatigue
Strength Tolerance Tested per BS EN 13868:2002 Conclusion: As per the test conditions contain herein, average kink length observed is as above and there is no kink mark observed while the flow rate is observed <50% from the straight tube.
#### Catheter Stiffness
(Durometer) Tested per ASTM D2240-05:2010 (R) Conclusion: As per the test conditions contain herein, the value observed is complies the hardness property of the material.
#### Resistance to Disinfection
(1-Alcohol 70%, 2- Mixture of Povidone Iodine 10%, 3-Chlorhexidine Gluconate 0.5%) Conclusion:
As per the test conditions contain herein, there is no damage observed during the visual inspection of the catheter.
#### Activation Testing:
Activation Testing was completed on the MedSource ClearSafe Safety IV Catheter. 500 Units of the MedSource ClearSafe Safety IV Catheter were obtained. Each unit was opened and the safety feature was activated and observed, making note of any deviation and the success or failure of the anti-needle stick feature. Failure is defined as a needle stick injury or a significant problem with the safety feature that may lead to an injury.
Testing was completed on April 7, 2014 in 5 Test Cycles of 100.
#### Conclusion of Activation Testing:
Out of the 500 MedSource ClearSafe Safety IV Catheters that were tested, 0 units failed to activate the safety feature. Based on that, it can be determined there is no issue outside of the confidence interval as stated in the Clinical Simulation Testing. (that 97.5% of the
{10}------------------------------------------------
time the failure rate would be no higher than 0.7% and 99.5% confident that the failure rate is not higher than 1.1% )
# Simulation Testing:
Clinical Simulation Testing was completed over the course of April 15 -- April 9th, 2015 for the MedSource ClearSafe Safety IV Catheter. The test was conducted using a variety of Health Care Professionals.
Each participant was able to use the MedSource ClearSafe Safety IV Catheter with little instruction and had no issues being able to activate the safety feature.
### Conclusion of Simulation Testing:
Based on the results of the Safety Feature Evaluation form it can be determined that the Medsource ClearSafe Safety IV Catheter's safety feature:
- 1. Is no more difficult to use then other similar devices
- 2. Is easy to activate and does not require extensive training
- 3. Does not interfere with normal use of the product.
- 4. Is required to be used when correctly operating the device
- 5. Eliminates the need for exposed needles in connections
- Works well with a variety of hand sizes and under various, common uses 6. (gloved, wet, dry hands)
- 7. Operates reliably- 0 units failed to activate the safety feature. Based on that, it can be determined that 97.5% of the time the failure rate would be no higher than 0.7% and 99.5% confident that the failure rate is not higher than 1.1%
- 8. Exposed sharp is covered after use and prior to disposal.
Conformity to the Following Recognized Standards:
- 1. ISO 10555-5: Sterile, single use intravascular catheters
- 2. ISO 11607-1/2 Packaging for terminally sterilized medical devices
- 3. BS EN 868-5 Packaging materials and systems for medical devices which are to be sterilized. Heat and self-sealable pouches and reels of paper and plastic film construction-Requirements and test methods
- 4. DIN 58953-6 "Sterilization paper for bags and tubing pickings ".Testing for germ proof ness in moisture with passage of air
- 5. ASTM F 2054 Standard test method for Burst Testing of Flexible Package Seals Using Internal Air Pressurization within Restraining Plates.
- 6. ASTM F 1980-02 Standard Guide for Accelerated Aging of Sterile Medical Device Packages
- 7. ASTM D638 Plastic Tensile Testing
{11}------------------------------------------------
- 8. ISO 10993: Part 10 Biological evaluation of medical devices: Test of irritation and skin sensitization.
## Conclusion:
As shown by data in the table above, there are no significant performance specification differences between the MedSource ClearSafe Safety IV Catheter and the predicate device. Testing was done to confirm the adequate performance of the device and safety feature. Therefore, we conclude that the performance specifications demonstrate that the subject device is substantially equivalent to the legally marketed predicate device.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.