Polysafety BC, Polysafety BC Adva, Polywin Safety, Polywin Safety Adva
Applicant
Poly Medicure Limited
Product Code
FOZ · General Hospital
Decision Date
Sep 21, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5200
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Polysafety BC IV Catheter are indicated for short-term use (less than 30 days) for insertion into a patient's vascular system to sample blood or administer fluids such as solutions, parenteral nutrition, and administration of other drugs. The catheters may be used for any patient population with consideration given to patient size, appropriateness for the solution being infused and duration of therapy. The 18G-24G catheters may be used intravascularly with power injectors at a maximum pressure of 300psi. The Polysafety BC Adva IV Catheter are indicated for short-term use (less than 30 days) for insertion into a patient's vascular system to sample blood or administer fluids such as solutions, parenteral nutrition, and administration of other drugs. The catheters may be used for any patient population with consideration given to patient size, appropriateness for the solution being infused and duration of therapy. The 18G-24G catheters may be used intravascularly with power injectors at a maximum pressure of 300psi. The Polywin Safety IV Catheter are indicated for short term use (less than 30 days) for insertion into a patient's vascular system to sample blood or administer fluids such as solutions, parenteral nutrition and administration of other drugs. The catheters may be used for any patient population with consideration given to patient size, appropriateness for the solution being infused and duration of therapy. The 14G-24G catheters may be used intravascularly with power injectors at a maximum pressure of 300psi. The Polywin Safety Adva IV Catheter are indicated for short term use (less than 30 days) for insertion into a patient's vascular system to sample blood or administer fluids such as solutions, parenteral nutrition and administration of other drugs. The catheters may be used for any patient population with consideration given to patient size, appropriateness for the solution being infused and duration of therapy. The 14G-24G catheters may be used intravascularly with power injectors at a maximum pressure of 300psi.
Device Story
Peripheral passive safety IV catheters for vascular access; includes blood control valve to reduce exposure during placement/sampling; features passive safety clip to prevent needlestick injuries. Input: vascular access; Output: fluid administration/blood sampling. Used in clinical settings by healthcare providers. Includes quick flashback visualization. Passive safety mechanism deploys automatically upon needle withdrawal; prevents re-insertion and catheter shearing. Benefits: reduced blood exposure, prevention of needlestick injuries, and safe power injection capability.
Clinical Evidence
No clinical data. Bench testing only. Compliance demonstrated via ISO 10555-1/5, ISO 80369-7, ISO 23908, and biocompatibility testing per ISO 10993 series. Simulated clinical use testing performed per FDA guidance for sharps injury prevention.
Indicated for short-term ( <30 days) vascular access for blood sampling or fluid/drug/parenteral nutrition administration. Suitable for all patient populations based on size, therapy duration, and solution appropriateness. 14G-24G models compatible with power injectors up to 300 psi.
Regulatory Classification
Identification
An intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings and that is inserted into the patient's vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravenously. The device may be constructed of metal, rubber, plastic, or a combination of these materials.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The FDA text logo is in blue and includes the words "U.S. Food & Drug Administration".
September 21, 2023
Poly Medicure Limited % Sunita Teekasingh Regulatory Consultant GSA 2 Group LLC 8049 Hayes Street North East Spring Lake Park, Minnesota 55432
Re: K231401
Trade/Device Name: Polysafety BC, Polysafety BC Adva, Polywin Safety, Polywin Safety Adva Regulation Number: 21 CFR 880.5200 Regulation Name: Intravascular Catheter Regulatory Class: Class II Product Code: FOZ Dated: August 18, 2023 Received: August 22, 2023
Dear Sunita Teekasingh:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Danil Walloschek
David Wolloscheck, Ph.D. Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
# 510(k) Number (if known) K231401
### Device Name
Polysafety BC, Polysafety BC Adva, Polywin Safety, Polywin Safety Adva
## Indications for Use (Describe)
The Polysafety BC IV Catheter are indicated for short-term use (less than 30 days) for insertion into a patient's vascular system to sample blood or administer fluids such as solutions, parenteral nutrition, and administration of other drugs. The catheters may be used for any patient population with consideration given to patient size, appropriateness for the solution being infused and duration of therapy. The 18G-24G catheters may be used intravascularly with power injectors at a maximum pressure of 300psi. maniliani probat v o r vopsi
The Polysafety BC Adva IV Catheter are indicated for short-term use (less than 30 days) for insertion into a patient's vascular system to sample blood or administer fluids such as solutions, parenteral nutrition, and administration of other drugs. The catheters may be used for any patient population with consideration given to patient size, appropriateness for the solution being infused and duration of therapy. The 180-24G catheters may be used intravascularly with power injectors at a maximum pressure of 300psi.
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------The Polywin Safety IV Catheter are indicated for short term use (less than 30 days) for insertion into a patient's vascular system to sample blood or administer fluids such as solutions, parenteral nutrition and administration of other drugs. The catheters may be used for any patient population with consideration given to patient size, appropriateness for the solution being infused and duration of therapy. The 14G-24G catheters may be used intravascularly with power injectors at a maximum pressure of 300psi. ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
The Polywin Safety Adva IV Catheter are indicated for short term use (less than 30 days) for insertion into a patient's vascular system to sample blood or administer fluids such as solutions, parenteral nutrition and administration of other drugs. The catheters may be used for any patient population with consideration given to patient size, appropriateness for the solution being infused and duration of therapy. The 14G-24G catheters may be used intravascularly with power injectors at a maximum pressure of 300psi.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image is a logo for Polymed Medical Devices. The logo features the word "POLYMED" in teal, with a red cross inside a yellow circle in place of the "O". Below the word "POLYMED" are the words "MEDICAL DEVICES" in a smaller font. Below that is the slogan "We Care As We Cure".
# K 231401 - 510(K) SUMMARY
### SUBMISSION ADMINISTRATIVE INFORMATION 1
| Device Name: | Polysafety BC, Polysafety BC Adva, Polywin Safety, Polywin Safety Adva IV Catheters |
|-----------------------------------------|----------------------------------------------------------------------------------------------------------------------------|
| Type of 510(k) submission: | Traditional Bundled 510(k) |
| Manufacturer: | Poly Medicure Ltd<br>Plot No. 115-117, Sector- 59<br>HSIIDC Industrial Area, Ballabgarh Faridabad-121004, Haryana<br>INDIA |
| Phone: | +91-129-3355070 |
| FDA Establishment Reg. Number: | 9616991 |
| Subject Device | |
| FDA Product Code: | FOZ |
| FDA Regulation Number: | 21 CFR 880.5200 |
| FDA Classification Name: | Intravascular catheter |
| Classification Panel: | General Hospital |
| Common Name: | Catheter, Intravascular, Therapeutic, Short-Term Less Than 30 Days |
| FDA Classification: | Class II |
| Submission Type: | 510(k) |
| Date | September 19, 2023 |
| Official Primary Correspondent | Sunita Teekasingh RN, BSN, CCRN, MSc Regulatory Consultant<br>GSA2 Group LLC 8049 Hayes St NE<br>Minneapolis, MN 55432 |
| Email | GSA2Groupllc@gmail.com |
| Phone: | 612-814-7999 |
| Polysafety Blood Control (BC) Predicate | |
| Predicate Manufacturer: | B. Braun Medical Inc. |
| Predicate Trade Name: | Introcan Safety® 2 IV Catheter |
| Predicate 510(k) Number: | K192676 |
| Class | II |
| Predicate Product Code | FOZ |
| FDA Classification Name: | 21 CFR 880.5200 |
| Classification Panel: | General Hospital |
| Common Name: | Catheter, Intravascular, Therapeutic, Short-Term Less Than 30 Days |
| Polywin Safety IV Catheter Predicate | |
| Predicate Manufacturer: | B. Braun Medical Inc |
| Predicate Trade Name: | Introcan Safety® IV Catheter |
| Predicate 510(k) Number: | K020785 |
| Class | II |
| Predicate Product Code | FOZ |
| FDA Classification Name: | 21 CFR 880.5200 |
| Classification Panel: | General Hospital |
| Common Name: | Catheter, Intravascular, Therapeutic, Short-Term Less Than 30 Days |
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for Polymed Medical Devices. The logo features the word "POLYMED" in teal, with a red plus sign inside a yellow circle in place of the "O". Below the company name, the text "MEDICAL DEVICES" is written in a smaller font. The slogan "We Care As We Cure" is written in a smaller font below the company name.
### 2 POLYSAFETY BLOOD CONTROL SUBSTANTIAL EQUIVALANCE
The Polysafety BC predicate is B. Braun Medical Inc, Introcan Safety® 2 IV Catheter (K192676). Both devices are under the same product code FOZ and classification (880.5200) and intended for short term use less than 30 days.
| Feature | Subject device<br>Polysafety BC<br>Polysafety BC Adva | Predicate device<br>Introcan Safety® 2 IV Catheter | Comparison |
|-----------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------|
| Device<br>Manufacturer | Poly Medicure, India | B. Braun Medical Inc,<br>Pennsylvania | N/A |
| 510(k) Reference | This submission | K192676 | N/A |
| Device<br>Description: | The Polysafety-BC safety IV Catheter is an<br>over-the-needle blood control. Peripheral<br>passive Safety IV Catheter incorporates a<br>safety clip to help prevent needle-stick<br>injuries & blood control valve to reduce<br>blood exposure during initial catheter<br>placement, blood sampling, withdrawal of<br>the needle, and connection of the luer. The<br>Polysafety BC Safety IV Catheter is<br>provided with a detachable or integrated,<br>flexible perforated soft wing for providing<br>ventilation & skin maceration. The device<br>also incorporates a 'quick flash back'. | The Introcan Safety® 2 IV Catheter<br>consists of an over-the-needle, peripheral<br>catheter made of radiopaque<br>polyurethane, an integrated one<br>directional septum that controls the flow<br>of blood during and after cannulation, and<br>a passive safety needle-shielding<br>mechanism. Introcan Safety® 2 is<br>designed to reduce blood exposure at<br>insertion until first connection of an<br>infusion line or luer device to protect<br>clinicians and patients from blood<br>exposure. During needle withdrawal, the<br>needle is withdrawn through a septum that<br>seals after the needle has been removed,<br>blood is thus contained within the<br>Introcan Safety® 2 device. The pressure<br>exerted on the needle as it passes through<br>the septum wipes blood from the needle<br>further reducing potential blood<br>exposure. The passive safety needle<br>shielding mechanism of the Introcan<br>Safety® 2 is located inside the catheter<br>hub. Upon withdrawal of the needle, the<br>safety shield engages as the needle passes<br>through the catheter hub and deploys<br>automatically to shield the needle tip. The<br>safety shield protects during disposal,<br>aiding in the prevention of needlestick<br>injuries. Once the safety shield engages<br>and shields the needle tip, the user is<br>unable to re-insert the needle which aids<br>in the prevention of catheter shearing.<br>This device may be used for any patient<br>population with consideration given to<br>adequacy of vascular anatomy and<br>appropriateness for the solution being<br>infused and duration of therapy. The<br>catheters may be used intravascularly<br>with power injectors for which the<br>maximum pressure setting is 300 psi with<br>a luer lock connection only. The devices<br>will be available in 18, 20, 22, and 24- | Different see<br>2.1.1<br>(Substantial<br>equivalent) |
| Feature | Subject device<br>Polysafety BC<br>Polysafety BC Adva | Predicate device<br>Introcan Safety® 2 IV Catheter | Comparison |
| | | gauge versions with and without a<br>stabilization platform. | |
| Indications for<br>use | The Polysafety BC IV Catheter are indicated<br>for short-term use (less than 30 days) for<br>insertion into a patient's vascular system to<br>sample blood or administer fluids such as<br>solutions, parenteral nutrition, and<br>administration of other drugs. The catheters<br>may be used for any patient population with<br>consideration given to patient size,<br>appropriateness for the solution being<br>infused and duration of therapy. The 18G-<br>24G catheters may be used intravascularly<br>with power injectors at a maximum pressure<br>of 300psi.<br>The Polysafety BC Adva IV Catheter are<br>indicated for short-term use (less than 30 days) for insertion into a patient's vascular<br>system to sample blood or administer fluids<br>such as solutions, parenteral nutrition, and<br>administration of other drugs. The catheters<br>may be used for any patient population with<br>consideration given to patient size,<br>appropriateness for the solution being<br>infused and duration of therapy. The 18G-<br>24G catheters may be used intravascularly<br>with power injectors at a maximum pressure<br>of 300psi. | The Introcan Safety® 2 IV Catheter<br>is inserted into a patient's vascular<br>system for short term use to sample<br>blood, monitor blood pressure, or<br>administer fluids and blood<br>intravascularly. The catheters may be<br>used intravascularly with power injectors<br>at a maximum pressure of 300 psi with a<br>luer lock connection only. | Different see<br>2.1.2<br>(Substantial<br>equivalent) |
| Sharps injury<br>protection<br>feature? | Yes - passive, tested in accordance with ISO<br>23908 and FDA 'Guidance for Industry and<br>FDA Staff Medical Devices with Sharps<br>Injury Prevention Features', August 2005. | Yes - passive | Same |
| Wings | With, without and Detachable wings with | Yes- and without wings | Same |
| Catheter tube<br>material | Polyurethane + Barium Sulfate | Polyurethane + Barium Sulfate | Same |
| X-ray visible | Yes | Yes | Same |
| Needle material | Stainless steel | Stainless steel | Same |
| Needle distal end<br>configuration | Back cut ground beveled needle | Back cut ground beveled needle | Same |
| Flashback<br>visualization | Yes | Yes | Same |
| Blood control | Yes | Yes | Same |
| Gauge sizes | 18 - 24G | 18- 24G | Same |
| Needle lengths | 19-45 mm | 19-45 mm | Same |
| Color-coding | Yes, according to ISO 10555-5 | Yes, according to ISO 10555-5 | Same |
| Proximal end<br>configuration | Female 6 % Luer | Female 6 % Luer | Same |
| Single use | Yes | Yes | Same |
| Sterile | Yes, SAL 10-6…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.