K110697 · Tiger Medical Products, Ltd. · FMF · Jul 14, 2011 · General Hospital
Device Facts
Record ID
K110697
Device Name
TIGER SYRINGES AND NEEDLES
Applicant
Tiger Medical Products, Ltd.
Product Code
FMF · General Hospital
Decision Date
Jul 14, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5860
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Syringe with or without needle is intended for dispensing/administering fluids, and collecting/ sampling of fluid in medical practice. The insulin syringe (U-100) with fixed needle is a device intended for medical purposes for the manual aspiration of U-100 insulin, and for the injection of U-100 insulin into parts of the body below the surface skin. The Hypodermic Needle for single use is intended for use with syringes and injection devices for general purpose fluid injection/aspiration.
Device Story
Manual syringes and hypodermic needles for fluid administration/aspiration; insulin syringes for U-100 insulin delivery. Devices consist of barrel, plunger, and needle (fixed or detachable). Operated by clinicians or patients (insulin syringes) in clinical or home settings. Mechanical operation; manual aspiration and injection. Benefits include precise fluid delivery and sample collection. Materials include PP, PE, SUS 304 stainless steel, and polyisoprene pistons.
Clinical Evidence
Bench testing only. Compliance with ISO 8537, ISO 7886-1, ISO 594-1/2, and ISO 7864 verified performance specifications. Biocompatibility confirmed per ISO 10993.
Indicated for general fluid injection/aspiration (syringes/needles) and manual aspiration/injection of U-100 insulin (insulin syringes). Patient population includes individuals requiring insulin therapy or general medical fluid administration.
Regulatory Classification
Identification
A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.
Predicate Devices
Becton Dickinson Single Use Hypodermic Syringes (K980987)
Sterile Insulin Syringe for single use, with fixed needle (K090929)
{0}------------------------------------------------
# 510(k) Summary
As Required by 21 CFR 807.92
JUL 1 4 2011
The assigned 510(k) Number is : ______________________________________________________________________________________________________________________________________________
1. Date Prepared: Dec, 17, 2010
### 2. Sponsor Infromation:
Tiger Medical Products Ltd. Liulin Tower, Suit 1910, 1 Huaihai Zhong Road Shanghai 200021, China
#### Contact Person:
David Wang RA Manager Fiona Fu Regulatory Assistant Tiger Medical Products Ltd. Phone Number: 86-21-6386 6300 Fax Number: 86-21-6386 5086 E-mail: davidwang@tigermedicalgroup.com
#### 3. Submission Correspondent
David Wang RA Manager Fiona Fu Regulatory Assistant Tiger Medical Products Ltd. Phone Number: 86-21-6386 6300 Fax Number: 86-21-6386 5086 E-mail: davidwang@tigermedicalgroup.com
#### 4. Device Name and Classification:
#### 1) Syringe (with/without needle)
a. Classification Name: Syringe,Piston
b. Regulation Number: 880.5860
c. Product code: FMF
d. Class: Il
e. Review Panel: General Hospital
#### 2) Insulin syringe (U-100) with fixed needle
f. Classification Name: Syringe,Piston
{1}------------------------------------------------
g. Regulation Number: 880.5860 h. Product Code: FMF i. Class: II j. Review Panel: General Hospital
### 3) Hypodermic Needle for single use
k. Classification Name: Needle, Hypodermic, Single Lumen
1. Regulation Number: 880.5570
m. Product code: FMI
n. Class: II
o. Review Panel: General Hospital
### 5. Predicate Device Identification:
a. K number: K980987 Trade Name: Becton Dickinson Single Use Hypodermic Syringes
b. K number: K090929
Trade Name: Sterile Insulin Syringe for single use, with fixed needle
c. K number: K070440
Trade Name: BD Hypoint
#### 6. Device Description
{2}------------------------------------------------
Premarked Notification 510(k) Submisssion- 510(K) Summar
:
| Device Name | Intended Use | Nozzel | Nominal Capacity / Gauge | Material | Remark |
|-------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------|----------------------------------------------------------------------|------------------------------|------------------------|
| Syringe | Syringe is intended for dispensing/ administering fluids, and collecting/ sampling of fluid in medical practice. | Luer Slip | 1ml、2ml、3ml、5ml、10ml、20ml、30ml、50ml | PP/PE | With or Without Needle |
| Insulin Syringe | Insulin syringe (U-100) with fixed needle is a device intended for medical purposes for the manual aspiration of U-100 insulin, and for the injection of U-100 insulin into parts of the body below the surface skin. | Fixed | 0.3ml、0.5ml、1ml | PP/PE | With Fixed Needle |
| Hypodermic Needle | Hypodermic Needle is intended for use with syringes and injection devices for general purpose fluid injection/aspiration. | Female 6% Luer Conical Socket | 16G, 18G, 19G, 20G, 21G, 22G, 23G, 24G, 25G, 26G, 27G, 28G, 29G, 30G | SUS 304 Stainless Steel / PP | - |
.
{3}------------------------------------------------
### 7. Summary of Comparison in Technological Charateristics
Through comparions between the applicant devices with the predicate devices (see comparison chart in Executive Summary), we belive the applicant devices are substantially equivalent with the predicate devices.
## Technology Comparison to Predicate Device:
| Item | Syringe (with/without needle) | Becton Dickinson Single Use Hypodermic Syringes<br>510(K) #: (K980987) |
|------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | The Syringe is intended for dispensing/ administering fluids, and collecting/ sampling of fluid in medical practice. | The Becton Dickinson Syringes product line consists of single use disposable syringes intended for dispensing/ administering fluids, and collecting/ sampling of fluid in medical practice. Their function is mechanical. |
| Syringe Volume | Nominal capacity: 1,2,3,5,10,20,30,50ml (Luer Slip);<br>2,3,5,10,20,30,50ml (Luer Lock) | Various sizes |
| Needle Gauge | Various sizes | Various sizes |
| Needle Hub Type | Female 6% Luer | Female 6% Luer |
| Gradations Legibility | Legible | Legible |
| Lubricant Composition | Dowcorning 360 silicone (barrelinternal surface)<br>Dowcorning 4159 silicone (needle tip) | Unknown |
| Barrel Transparency | Transparent and clear | Transparent and clear |
| Labeling and Labels | Meet FDA requirements | Meet FDA requirements |
| Plunger Material | PP/PE | Same |
| Barrel Material | PP | Same |
| Piston Material | Kraton IR-307 polyisoprene<br>This component is not made with natural rubber latex. | Unknown |
| lubricant for the syringe Material | Dowcorning 360 (barrel internal surface)<br>Dowcorning 4159 (needle tip) | Unknown |
| Ink on the Barrel Material | Teikuko PPE 911 | Unknown |
| Performance | Conforms to ISO 7886-1 and ISO 7864 | Conforms to ISO 7886-1 and ISO 7864 |
| Biocompatibility | Conforms to ISO 10993 | Conforms to ISO 10993 |
| Sterility | EtO Sterilization | EtO Sterilization |
| Item | Insulin syringe with fixed needle | Sterile Insulin Syringe for single use, with<br>fixed needle<br>510(K) #: K090929 |
| Intended Use | Insulin syringe (U-100) with fixed<br>needle is a device intended for<br>medical purposes for the manual<br>aspiration of U-100 insulin, and for<br>the injection of U-100 insulin into<br>parts of the body below the surface<br>skin. | Sterile Insulin Syringe for single use, with<br>fixed needle is a device intended for medical<br>purpose for the manual injection of insulin. |
| Specific Drug Use | U-100 Insulin | Insulin |
| Nominal Capacity | 0.3ml、0.5ml、1ml | 0.3ml、0.5ml、1ml |
| Needle Hub Type | Fixed | Fixed |
| Gradations Legibility | Legible | Legible |
| Lubricant<br>Composition | Dowcorning 360 (used on internal<br>surface of barrel)<br>Dowcorning 4159 (used on needle<br>tip) | Dowcorning 360 (used on internal surface of<br>barrel)<br>Dowcorning 4159 (used on needle tip) |
| Barrel Transparency | Transparent and clear | Transparent and clear |
| Labeling and Labels | Meets the FDA requirements | Meets the FDA requirements |
| Needle | SUS 304 stainless steel | 304 Stainless Steel |
| Plunger | PE/PP | Medical Grade Polypropylene |
| Barrel | PP | Medical Grade Polypropylene |
| Piston | Kraton IR-307 polyisoprene<br>This compoment is not made with<br>natural rubber latex. | Santoprene TPE (Natural color) |
| Lubricant | Dowcorning 360 (barrel internal<br>surface)<br>Dowcorning 4159 (needle tip) | Polydimethylsioxane (12500 cp) |
| Ink on the Barrel | Teikuko PPE 911 | 50% SSPPNK-911 |
| | | 50% PPE-911 |
| Performance | Conforms to ISO 8537 | Conforms to ISO 8537 |
| Biocompatibility | Conforms to ISO 10993 | Conforms to ISO 10993 |
| Sterility | EtO Sterilization | EtO Sterilization |
| Item | Hypodermic Needle for single use | BD Hypoint<br>510(K) #: K070440 |
| Intended Use | The Hypodermic Needle for single use is intended for use with syringes and injection devices for general purpose fluid injection/aspiration | The BD HypointTM Needle is intended for use with syringes and injection devices for general purpose fluid injection/aspiration. |
| Needle Gauge | Various sizes | Various sizes |
| Needle Hub Type | Female 6% luer conical socket | Female 6% luer conical socket |
| Labeling and Labels | Meet FDA requirements | Meet FDA requirements |
| Needle Material | SUS 304 stainless steel | Unknown |
| Needle Hub Material | PP | Same |
| Needle Sheath Material | PP | Same |
| Lubricant Material | Dowcorning 4159 | Unknown |
| Performance | Conforms to ISO 7864 | Conforms to ISO 7864 |
| Biocompatibility | Conforms to ISO 10993 | Conforms to ISO 10993 |
| Sterility | EtO Sterilization | EtO Sterilization |
{4}------------------------------------------------
{5}------------------------------------------------
#### 8. Non-Clinical Test Summary
Bench tests were conducted to verify that the proposed devices meet all design specifications as are Substantially Equivalent (SE) to the predicate devices. The test results demonstrated that the proposed devices comply with the following standards: ISO 8537:2007 Sterile single-use syringes, with or without needle, for insulin.
ISO 7886 Sterile hypodermic syringes for single use -- Part 1: Syringes for manual use
ISO 594-1:1986 Conical fittings with a 6 % (Luer) taper for syringes, needles and certain other medical equipment -- Part 1: General requirement
ISO 594-2: 1998 Conical fittings with 6 % (Luer) taper for syringes, needles and certain other medical equipment -- Part 2: Lock fittings
ISO 7864:1993 Sterile hypodermic needles for single use.
#### 9. Substantially Equivalent Conclusion
The subject device and predicate device have the same classification information, intended use, sterilization specifications, performance, biocompatibility, chemical specifications, physical and mechanical specifications. Both devices meet requirements of ISO 10993, ISO 7886-1, ISO 7864 and ISO 8537. Even if there is some difference in component material (lubricant, ink, piston), we believe such difference has no adverse effect on safety and efficacy of the subject device, and we can be sure of substantial equivalence between subject device and predicate device.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the seal is an abstract symbol that resembles a stylized bird or eagle, composed of three curved lines.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Mr. David Wang Tiger Medical Products, Limited 1 Huaihai Zhong Road Liu Lin Tower, Suite 1910 Shanghai, China 200021
JUL 1 4 2011
Re: K110697
Trade/Device Name: Syringe (With or Without Needles), Insulin Syringe (U-100) With Fixed Needles, Hypodermic Needle for Single Use Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe Regulatory Class: II Product Code: FMF, FMF, FMI Dated: June 28, 2011 Received: June 28, 2011
Dear Mr. Wang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{7}------------------------------------------------
Page 2 -- Mr. Wang
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/Reporta Problem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
Anthony D. Norton
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital. Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
# Indications for Use
510(k) Number (if known): _ K (10697
Syringe (with or without needle) Device Name: __
Indications for Use:
The Syringe with or without needle is intended for dispensing/administering fluids, and collecting/ sampling of fluid in medical practice.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Division Sign-Off)
Jivision of Anesthesiology, General Hospital nfection Control, Dental Devices
් 10 (k) Number:
Page 1 of 3
{9}------------------------------------------------
# Indications for Use
K110697 510(k) Number (if known):
Device Name: Insulin syringe (U-100) with fixed needle
#### Indications for Use:
The insulin syringe (U-100) with fixed needle is a device intended for medical purposes for the manual aspiration of U-100 insulin, and for the injection of U-100 insulin into parts of the body below the surface skin.
The proposed device of Insulin Syringe (U-100) with fixed needle is available in 0.3ml, 0.5ml, and 1ml in volume.
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use V AND/OR (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
7/1/14
Division Sign-Off) Division of Anesthesiology, General Hospital nfection Control, Dental Devices
:10(k) Number: K110697
{10}------------------------------------------------
Page 2 of 3
# Indications for Use
510(k) Number (if known): _ |< | | 0 6 9 7
Device Name: Hypodermic Needle for single use
Indications for Use:
The Hypodermic Needle for single use is intended for use with syringes and injection devices for general purpose fluid injection/aspiration.
Prescription Use V (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Rild C. Chy 7/13/14
(Division Sign-Off)
് vision of Anesthesiology, General Hospital .fection Control, Dental Devices
510(k) Number: K110697
Page 3 of 3
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.