K980987 · Bd Becton Dickinson Vacutainer Systems Preanalytic · FMF · Jun 11, 1998 · General Hospital
Device Facts
Record ID
K980987
Device Name
BECTON DICKINSON SYRINGE
Applicant
Bd Becton Dickinson Vacutainer Systems Preanalytic
Product Code
FMF · General Hospital
Decision Date
Jun 11, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5860
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Intended Use
These syringes are intended for use by health care professionals for general purpose fluid aspiration/injection.
Device Story
Becton Dickinson single-use hypodermic syringes; intended for fluid aspiration/injection in medical practice. Device consists of barrel and plunger with synthetic rubber plunger tip; replaces natural rubber (Hevea brasiliensis) to eliminate potential for natural rubber protein allergic reactions. Mechanical operation; used by healthcare professionals in clinical settings. Syringes compatible with syringe pumps, including low flow rate (0.01 ml/hr) neonatal drug administration. Synthetic tip maintains leak-proof seal, withstands sterilization (radiation, ethylene oxide, autoclave), and exhibits actuation forces equivalent to natural rubber predicate. Benefits include reduced risk of latex allergy while maintaining functional performance and safety standards.
Clinical Evidence
Bench testing only. No clinical data. Functional testing confirmed equivalence to predicate in leak-proof seal integrity, sterilization resilience, and actuation forces. Syringe pump application testing verified performance at low flow rates (0.01 ml/hr to 1.0 ml/hr) for neonatal administration. Biocompatibility verified via chemical and biological assays per ISO guidance.
Technological Characteristics
Single-use hypodermic syringe; synthetic rubber plunger tip (latex-free). Mechanical actuation. Compatible with radiation, ethylene oxide, and autoclave sterilization. Designed for syringe pump applications including low flow rate neonatal use.
Indications for Use
Indicated for health care professionals for general purpose fluid aspiration and injection.
Regulatory Classification
Identification
A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.
Predicate Devices
Becton Dickinson Single Use Hypodermic Syringes
Related Devices
K980580 — BECTON DICKINSON SYRINGES · Bd Becton Dickinson Vacutainer Systems Preanalytic · Jun 25, 1998
K162081 — BD 1ml Luer-lok Hypodermic Syringe, BD 1 mL Luer-Lok Hypodermic Syringe with BD Hypodermic Needle or BD Eclipse Hypodermic Needle, BD 1ml Luer-lok Insulin Syringe · Becton, Dickinson and Company · Dec 19, 2016
K241821 — Luer Lock Syringe with Safety Needle; Luer Lock Syringe with Exchangeable Needle; Luer Lock Syringe with Blunt Fill Needle · Sol-Millennium Medical, Inc. · Sep 20, 2024
K213811 — Sterile Hypodermic Syringes for Single Use · Berpu Medical Technology Co., Ltd. · May 12, 2022
K231727 — Sterile Auto-Disable Syringes with/without Needle for Single Use · Wepon Medical Technology Co., Ltd. · Sep 8, 2023
Submission Summary (Full Text)
{0}------------------------------------------------
JUN | | 1998
# SUMMARY OF SAFETY AND EFFECTIVENESS DATA
## 1.0 B-D Contact Person
Gregory W. Morgan Manager, Regulatory Affairs Division Quality Assurance Becton Dickinson & Company 1 Becton Drive, Building 2 Franklin Lakes, NJ 07417-1884 (201) 847-4344 - Phone (201) 847-4855 - FAX
## 2.0 Device Name
Becton Dickinson Single Use Hypodermic Syringes
## 3.0 Predicate Device
Becton Dickinson Single Use Hypodermic Syringes
## Product Description/Function 4.0
## 4.1 Description
Single use sterile and non-sterile disposable hypodermic syringes manufactured by Becton Dickinson.
## 4.2 Function
The Becton Dickinson hypodermic syringe product line consists of single use disposable syringes intended for dispensing/administering fluids, and collecting/sampling of fluids in medical practice. Their function is mechanical.
## . 2.0 Comparison of Modified and Predicate Devices
## 5.1 Design Changes
No design changes are being made.
{1}------------------------------------------------
### 5.2 Material Changes
Becton Dickinson intends to change the medical grade rubber formulation used for its molded syringe plunger tip.
Becton Dickinson manufactures its predicate plunger tip using dry gum natural rubber from the Hevae Braziliensis tree. While this material has proven safe through a long history of use, its composition suggests the possibility of natural rubber allergic reactions in susceptible patients.
The proposed plunger tip contains synthetic rubber. No natural rubber or other naturally occurring protein containing material is used in its formulation.
#### 5.3 Manufacturing Process Changes
No manufacturing process changes are being made.
#### 5.4 Manufacturing Site Changes
No manufacturing site changes are being made.
## 5.5 Packaging Component Changes
No packaging components are being changed.
## 6.0 Equivalence
The following data demonstrates functional equivalence to Becton Dickinson's predicate plunger tip and fitness for use.
## 6.1 Design Change
No design changes are being made.
## 6.2 Material Change
The new synthetic "latex/natural rubber free" syringe plunger tip formulation, and syringes made from this material, have been proven equivalent to existing product in functional performance (efficacy). Biological safety has been proven through a series of chemical and biological assays.
{2}------------------------------------------------
## 6.2.1 Mechanical Function/Efficacy
Syringes manufactured using the new synthetic "latex/natural rubber free" stopper material have demonstrated equivalence to syringes manufactured with natural rubber stoppers for:
- . Ability to maintain a leak-proof seal.
- . Perform after exposure to all expected sterilization conditions svringes will experience. This includes radiation, ethylene oxide, and autoclave methods.
- . Exhibit actuation forces equivalent to current natural rubber containing svringes.
- . Meet the stringent requirements for use in syringe pump application including low flow rate (0.01 mi/hr) neonatal administration of fast acting drugs.
## 6.2.2 Safety
.
- Functional Safety Hypodermic syringes can be used in syringe pumps. Post market surveillance has identified the use of svringes in svringe pumps for administration of dopamine to neonates at low flow rates (0.1 ml/hr to 1.0 ml/hr) as the most stringent requirement. At the resultant rates of translation, the consistency of forces of movement are critical. Inconsistency in movement of the plunger and slow start-up/alarm response is undesirable.
Becton Dickinson's predicate device meets the stringent neonatal pump application functional requirements. The proposed synthetic "latex/natural rubber free" plunger tip also meets this requirement as demonstrated in syringe pump application testing covering the range of use including low flow rate use.
{3}------------------------------------------------
- Biocompatibility Syringes manufactured using the proposed synthetic "latex/natural rubber free" plunger tip have proven to be safe over specified manufacturing process and formulation variables. They pass all biological and chemical evaluations recommended in the ISO biocompatibility guidance.
- Functional Safety Syringes manufactured using the proposed synthetic "latex/natural rubber free" plunger tip have proven to be functionally safe. Thev demonstrate functional equivalence to the predicate device in all applications, including neonatal syringe pump use.
## Manufacturing Process Changes 6.3
No manufacturing process changes are being made.
- 6.4 Manufacturing Site Changes
No manufacturing site changes are being made.
## 6.5 Packaging Component Changes
No packaging components are being changed.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract image of an eagle with its wings spread, rendered in a stylized, minimalist design.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUN 11 1998
Mr. Gregory W. Morgan · Manager, Regulatory Affairs Division Quality Assurance Becton Dickinson & Company 1 Becton Drive 07417-1886 Franklin Lakes, New Jersey
K980987 Re : Becton Dickinson Syringe Trade Name: Regulatory Class: II Product Code: FMF March 13, 1998 Dated: Received: March 17, 1998
Dear Mr. Morgan:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual reqistration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical General regulation (21 CFR Part 820) and that, Devices: through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531
{5}------------------------------------------------
Page 2 - Mr. Morgan
through 542 of the Act for devices under the Electronic enroags and of on Control provisions, or other Federal laws or requlations.
This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4692. Additionally, for questions on the promotion and advertising of your device, please contact Also, please note the Office of Compliance at (301) 594-4639. the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address
"http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Timothy A. Ulatowski
Director Division of Dental, Infection Control, and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure ... ... ..
{6}------------------------------------------------
Page 1 이 1
510(K) Number (if known) K980987
Device Name Syringe Plunger Tip Formulation Change
Indications for Use:
ﻬﺎ . . . .
> These syringes are intended for use by health care professionals for general purpose fluid aspiration/injection.
G., May 3/20/88
(Please DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence f CDRH, Office of Device Evaluation (ODE)
Helugo Lucerite
(Division Sian-Off) Division of Dental, Infection Control, and General Hospital Devices
K980987 510(k) Number _
Prescription Use 1 OR
Over-The Counter Use
(Per 21 CFR 801.109)
(Division Sign-Off) Division of Dental, Infection Control, and General Hospital Devices (Optional Format 1-2-96) 510(k) Number _
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.