STERILE HYPODERMIC SYRINGE AND NEEDLE FOR SINGLE USE, STERILE INSULING SYRINGE FOR SINGLE USE WITH FIXED NEEDLE
K090929 · Shanghai Double Dove Industrial Co., Ltd. · FMF · Jun 12, 2009 · General Hospital
Device Facts
Record ID
K090929
Device Name
STERILE HYPODERMIC SYRINGE AND NEEDLE FOR SINGLE USE, STERILE INSULING SYRINGE FOR SINGLE USE WITH FIXED NEEDLE
Applicant
Shanghai Double Dove Industrial Co., Ltd.
Product Code
FMF · General Hospital
Decision Date
Jun 12, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5860
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Sterile Hypodermic Syringe for Single Use is intended for dispensing/ administering fluids, and collecting/ sampling of fluid in medical practice. Their function is mechanical. The sterile Insulin Syringe for single use with fixed needle is a device is a device intended for medical purposes for the manual aspiration of insulin, and for the injection of insulin, and for the injection of insulin into parts of the body below the surface skin. The Sterile Hypodermic Needle for single use is intended for use with syringes and injection devices for general purpose fluid injection/aspiration.
Device Story
Double-Dove sterile syringes and needles are manual, single-use medical devices. Syringes consist of a piston-driven barrel (polypropylene) for fluid aspiration and injection; needles (stainless steel) attach via Luer slip or Luer lock mechanisms, or are fixed to the insulin syringe. Operated by clinicians or patients (for insulin) in clinical or home settings. Devices function mechanically to deliver or collect fluids. No electronic or software components. Benefit includes sterile, single-use delivery of medication or fluid sampling.
Clinical Evidence
Bench testing only. No clinical data provided.
Technological Characteristics
Materials: Medical grade polypropylene (syringes), stainless steel (needles). Energy source: Manual/mechanical. Form factor: Various volumes (0.3ml to 50ml) and needle gauges. Connectivity: None. Sterilization: Sterile, single-use.
Indications for Use
Indicated for patients requiring fluid administration, fluid sampling, or insulin injection. Insulin syringes are for subcutaneous insulin delivery. Hypodermic needles are for general fluid injection or aspiration. No specific age or gender restrictions provided.
Regulatory Classification
Identification
A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.
Predicate Devices
Becton Dickinson Single Use Hypodermic Syringes (K980987)
{0}------------------------------------------------
## JUN 1 2 2009 Chapter III 510(k) Summary
(As required by 21 CFR 807.92)
The assigned 510(k) Number is:
- 1. Date Prepared: March 17, 2009
- 2. Sponsor Information
Shanghai Double-Dove Industry Co.,Ltd
No.1888 Huhang Road FengXian Economic Zone Shanghai, 201400, China
Contact Person: Mr. Sanba Yang, Quality Manager Tel: +86-21-67104888 Fax: +86-21-67104666 E-Mail: Yangsbmaster@hotmail.com
### 3. Submission Correspondent
Ms. Diana Hong Mr. Lee Fu Shanghai Mid-Link Business Consulting Co., Ltd Suite 8D, Zhongshan Zhongxin Mansion No.19, Lane 999, Zhongshan No.2 Road(S) Shanghai, 200030, China
4. Device Name and Classification:
a. Sterile Hypodermic Syringe for single use
(1) Classification Name: Syringe, Piston
(2) Regulation Number: 880.5860
- (3) Product Code: FMF
- (4) Class: II
{1}------------------------------------------------
Kogor29
2
b. Sterile Insulin Syringe for single use, with fixed needle
(1) Classification Name: Syringe, Piston
(2) Regulation Number: 880.5860
(3) Product Code: FMF
(4) Class: II
(5) Review Panel: General Hospital
c. Sterile Hypodermic Needle for single use
(1) Classification Name: Needle, Hypodermic, Single Lumen
(2) Regulation Number: 880.5570
(3) Product Code: FMI ·
(4) Class: II
(5) Review Panel: General Hospital
#### 5. Predicate Device Identification:
#### a. K number: K980987
Trade Name: Becton Dickinson Single Use Hypodermic Syringes
b. K number: K071630 Trade Name: TERUMO 31G ThinPro Insulin Syringe
c. K number: K070440 Trade Name: BD Hypoint
{2}------------------------------------------------
| | Device Name | Intended Use | Nozzel | Volume | Material | Remark |
|--|-----------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------|------------------------|
| | Sterile Hypodermic<br>Syringe for single use | The Sterile Hypodermic Syringe for Single<br>Use is intended for dispensing/<br>administering fluids, and collecting/<br>sampling of fluid in medical practice. Their<br>function is mechanical. | Luer Slip<br>Luer Lock | 1ml、3ml、5ml、10ml、20ml、<br>30ml、50ml | Medical Grade<br>Polypropylene | With or Without Needle |
| | Sterile Insulin<br>Syringe for single<br>use, with fixed needle | The sterile Insulin Syringe for single use<br>with fixed needle is a device intended for<br>medical purposes for the manual aspiration<br>of insulin, and for the injection of insulin<br>into parts of the body below the surface<br>skin. | Fixed | 0.3ml、0.5ml、1ml | Medical Grade<br>Polypropylene | With Fixed Needle |
| | Sterile Hypodermic<br>Needle for single use | The Sterile Hypodermic Needle for single<br>use is intended for use with syringes and<br>injection devices for general purpose fluid<br>injection/aspiration | Luer Slip<br>Luer Lock | 0.3*12.7; 0.3*25; 0.33*12.7;<br>0.33*12.7; 0.33*25;<br>0.4*12.7; 0.4*25; 0.45*16;<br>0.45*25; 0.5*16; 0.5*25;<br>0.55*16; 0.55*25; 0.6*25;<br>0.6*30; 0.7*25; 0.7*32;<br>0.8*25; 0.8*38; 0.9*25;<br>0.9*38; 1.1*25; 1.1*38;<br>1.2*25; 1.2*38 | Stainless Steel | |
i
.
K090929
III-3
{3}------------------------------------------------
K090929
4
Premarket Notification Submission - Chapter III 510(K) Summary Report No.: A2008-034-061
### 7. Test Conclusion
Laboratory testing was conducted to validate and verify that Double-Dove Syringes and Needle met all design specifications and was substantially equivalent to the predicate device.
### 8. Substantially Equivalent Conclusion:
The subject device, Double-Dove Syringes and Needle, is substantially equivalent to the predicate device.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" arranged in a circular pattern around the eagle. The eagle is black, and the text is also black. The logo is simple and recognizable.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Shanghai Double-Dove Industry Company, Limited C/O Ms. Diana Hong General Manager Shanghai Mid-Link Business Consulting Company, Limited Suite 8D, No. 19, Lane 999, Zhongshan No.2 Road (S) Shanghai CHINA 200030
JUN 1 2 2009
Re: K090929
Trade/Device Name: Sterile Insulin Syringe for Single Sue Needle, with Fixed Needle Regulation Number: 880.5860 Regulation Name: Piston Syringe Regulatory Class: II Product Code: FMF Dated: March 31, 2009 Received: April 2, 2009
Dear Ms. Hong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
#### Page 2 - Ms. Hong
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/ CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Susan Roane
Susan Runner, D.D.S., M.A. Acting Director Division of Anesthesiology, General Hospital, Infection_Control_and_Dental_Devices_ Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# Indication for Use
510(k) Number:
Device Name: Sterile Hypodermic Syringe for Single Use
## Indications for Use:
The Sterile Hypodermic Syringe for Single Use is intended for dispensing/ administering fluids, and collecting/ sampling of fluid in medical practice. Their function is mechanical.
Prescription Use V (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
sta Carlsson
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K090929
Page 1 of 3
{7}------------------------------------------------
## Indication for Use
510(k) Number:
Device Name: Sterile Insulin Syringe for single sue needle, with fixed needle
Indications for Use:
The sterile Insulin Syringe for single use with fixed needle is a device is a device intended for medical purposes for the manual aspiration of insulin, and for the injection of insulin, and for the injection of insulin into parts of the body below the surface skin.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use V (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Adall
(Division Sign-on)
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K090939
Page 2 of
{8}------------------------------------------------
## Indication for Use
510(k) Number:
Device Name: Sterile Hypodermic Needle for single use
## Indications for Use:
The Sterile Hypodermic Needle for single use is intended for use with syringes and injection devices for general purpose fluid injection/aspiration.
Prescription Use N (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K090229
Page 3 of 3
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.