K112057 · Shanghai Kindly Enterprise Dev. Group Co., · FMF · Dec 14, 2011 · General Hospital
Device Facts
Record ID
K112057
Device Name
KDL SYRINGES AND NEEDLE
Applicant
Shanghai Kindly Enterprise Dev. Group Co.,
Product Code
FMF · General Hospital
Decision Date
Dec 14, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5860
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Sterile Hypodermic Syringe for Single Use with/without Needle is intended to be used for medical purposes to inject fluid into or withdraw fluid from body. The sterile insulin syringe for single use with needle, with the calibration unit of insulin for U-100, is a device intended for medical purposes for the manual aspiration of insulin, and for the injection of insulin into parts of the body below the surface skin. The sterile insulin syringe for single use with needle, with the calibration unit of U-40, is a device intended for medical purposes for the manual aspiration of insulin, and for the injection of insulin into parts of the body below the surface skin. The Sterile Hypodermic Needle for single use is intended for use with syringes and injection devices for general purpose fluid injection/aspiration.
Device Story
Device consists of sterile, single-use hypodermic syringes (with/without needles) and hypodermic needles. Syringes feature Luer slip or Luer lock nozzles; insulin syringes feature fixed needles. Used by clinicians or patients for manual aspiration and injection of fluids or insulin. Operation is purely mechanical; user manually draws fluid into syringe and injects into body. No electronic or software components. Provides a sterile, disposable means for fluid delivery or withdrawal.
Clinical Evidence
Bench testing only. No clinical data provided. Testing verified compliance with ISO 7886-1 (syringes), ISO 8537 (insulin syringes), ISO 7864 (needles), ISO 9626 (tubing), and ISO 594 (Luer fittings).
Technological Characteristics
Materials: Polypropylene (PP) for syringes, stainless steel for needles. Design: Luer slip/lock or fixed needle configurations. Energy: Manual. Connectivity: None. Sterilization: Not specified (implied sterile). Standards: ISO 7886-1, ISO 8537, ISO 7864, ISO 9626, ISO 594-1/2.
Indications for Use
Indicated for patients requiring manual aspiration and injection of insulin (U-40 or U-100) into subcutaneous tissue, or general purpose fluid injection/aspiration via hypodermic syringe/needle.
Regulatory Classification
Identification
A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.
{0}------------------------------------------------
DEC 1 4 2011
Premarket Notification 510(k) Submission
Exhibit #2 510(k) Summary
Project #:M0102011Ab
# Exhibit #2510(k) Summary
This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and 21 CFR 807.92.
The assigned 510(k) Number: K112057
- 】. Date of Preparation: 18 NOV 2011
- 2. Sponsor
Shanghai Kindly Enterprise Development Group Co., Ltd. No. 18 Lanc 668 Huajiang Road, Jiading Shanghai, 201803. China Establishment Registration Number: 3005405146 Contact Person: Cailiang Wang Position: General Manager Tel: +86-21-59140045 Fax: +86-21-59118500-78 Email: wangcl@kdIchina.net
- Submission Correspondent 3. Ms. Diana Hong & Mr. Lee Fu Mid-Link Consulting Co., Ltd P.O. Box 237-023, Shanghai, 200237, China Tel: +86-21-22815850 Fax: 240-238-7587 Email: info@mid-link.net
- Proposed Device Identification 4.
- Proposed Device Name: Sterile Hypodermic Syringe for single use, with/without needle a. Proposed Device Model: Lucr slip (1, 2, 3, 5, 10, 20, 30, 35, 50 (ml)) and Luer lock (1, 2, 3, 5, 10. 20. 30. 35, 50 (ml)) Classification: II Product Code: FMF Regulation Number: 21 CFR 880.5860 Review Panel: General Hospital Intended Use Statement: The Sterile Hypodermic Syringe for Single Use with/without needle is intended to be used for medical purposes to inject fluid into or withdraw fluid from body.
2-1
{1}------------------------------------------------
- b. Proposed Device Name: Sterile Insulin Syringe for single use, with needle Proposed Device Model: Fixed needle (0.3, 0.5, 1 (ml)) Classification: II
Product Code: FMF
Regulation Number: 21 CFR 880.5860
- Review Panel: General Hospital
Intended Use Statement:
The sterile insulin syringe for single use with needle, with the calibration unit of insulin for U-100, is a device intended for medical purposes for the manual aspiration of insulin, and for the injection of insulin into parts of the body below the surface skin.
The sterile insulin syringe for single use with needle, with the calibration unit of U-40, is a device intended for medical purposes for the manual aspiration of insulin, and for the injection of insulin into parts of the body below the surface skin.
- Proposed Device Name: Sterile Hypodermic Needle for single use ن
Proposed Device Model: Luer slip and Luer lock (18G, 19G, 20G, 21G, 22G, 23G, 24G, 25G, 26G, 27G, 28G, 29G, 30G) Classification: 11 Product Code: FMI Regulation Number: 21 CFR 880.5570 Review Panel: General Hospital Intended Use Statement: The Sterile Hypodermic Needle for single use is intended for use with syringes and injection devices for general purpose fluid injection/aspiration
- Predicate Device Identification 5.
- 510(k) Number: K070936 a. Product Name: Welmed Hypodermic Syringe (various sizes) Manufacturer: Welmed Inc.
- b. 510(k) Number: K072739. Product Name: Sterile Hypodermic Syringe for Single Use, without Needle Manufacturer: Shandong Weigao Group Medical Polymer Products Co., Ltd.
- 510(k) Number: K110421 C. Product Name: Disposable Insulin Syringe Manufacturer: Wenzhou Wuzhou Import & Export Co., Ltd.
- 510(k) Number: K070440 d. Product Name: BD Hypoint Manufacturer: BD Medical-Pharmaceutical Systems
{2}------------------------------------------------
Exhibit #2 510(k) Summary
Premarket Notification 510(k) Submission
Project #:M0102011Ab
# 6. Device Description
| Device Name | Device Description<br>Intended Use | Nozzle | Volume | Material | Remark |
|------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------|--------------------------------------------------------------------|-----------------|------------------------|
| Sterile Hypodermic<br>Syringe for single<br>use with/without<br>Needle | The Sterile Hypodermic Syringe for Single<br>Use with/without Needle is intended to be<br>used for medical purposes to inject fluid into<br>or withdraw fluid from body. | Luer Slip<br>Luer Lock | 1,2,3,5,10,20,30,35,50(ml) | PP | With or Without Needle |
| Sterile Insulin<br>Syringe for single<br>use with Needle | The sterile Insulin Syringe for single use with<br>needle is a device intended for medical<br>purposes for the manual aspiration of insulin,<br>and for the injection of insulin into parts of<br>the body below the surface skin. | Fixed Needle | 0.3,0.5,1(ml) | PP | With Fixed Needle |
| Sterile Hypodermic<br>Needle for single use | The Sterile Hypodermic Needle for single use<br>is intended for use with syringes and<br>injection devices for general purpose fluid<br>injection/aspiration | Luer Slip<br>Luer Lock | 18G, 19G, 20G, 21G, 22G, 23G,<br>24G, 25G, 26G, 27G, 28G, 29G, 30G | Stainless Steel | .... |
2-3
:
.
・
{3}------------------------------------------------
- 7. Summary Comparing the Technological Characteristics
The proposed devices and predicate devices have the same classification information, intended use, principle of operation, performance, biocompatibility, materials, and sterilizations. Both the proposed devices and predicate devices meet the reguirements of ISO 10993, ISO 7886-1, ISO 8537, ISO 7864 and ISO 9626. Although there is some difference in syringe volumes and syringe nozzle, such difference will not affect the safety and effectiveness of the proposed devices. So the proposed devices are substantially equivalent to the predicate devices through the comparisons between the proposed devices and predicate devices.
- 8. Non-Clinical Test Conclusion
Bench tests were conducted to verify that the proposed devices met all design specifications as were Substantially Equivalent (SE) to the predicate devices. The test results demonstrated that the proposed devices comply with the following standards:
- For Sterile Hypodermic Syringe for single use, with/without needle a.
ISO 7886-1:1993 Sterile hypodermic syringes for single use -- Part 1: Syringes for manual use; ISO 594:1986 Conical fittings with a 6 % (Luer) taper for syringes, needles and certain other medical equipment -- Part 1: General requirements
ISO 594:1998 Conical fittings with 6 % (Luer) taper for syringes, needles and certain other medical equipment -- Part 2: Lock fittings
- For Sterile Insulin Syringe for single use, with needle b.
ISO 8537:2007 Sterile single-use syringes, with or without needle, for insulin; ISO 594:1986 Conical fittings with a 6 % (Luer) taper for syringes, needles and certain other medical equipment -- Part 1: General requirements ISO 594:1998 Conical fittings with 6 % (Luer) taper for syringes, needles and certain other medical equipment -- Part 2: Lock fittings
- For Sterile Hypodermic Needle for single use C.
ISO 7864:1993 Sterile hypodermic needles for single use;
ISO 9626:1991 Stainless steel needle tubing for the manufacture of medical devices/Amendment: 2001
ISO 594:1986 Conical fittings with a 6 % (Luer) taper for syringes, needles and certain other medical equipment -- Part 1: General requirements
ISO 594:1998 Conical fittings with 6 % (Luer) taper for syringes, needles and certain other medical equipment -- Part 2: Lock fittings
{4}------------------------------------------------
#### 9. Substantially Equivalent Conclusion
The proposed device. Sterile Hypodermic Syringe for single use, with needle, is determined to be Substantially Equivalent (SE) to the predicate device, Welmed Hypodermic Syringe (various sizes), K070936, in respect of safety and effectiveness.
The proposed device, Sterile Hypodermic Syringe for single use, without needle, is determined to be Substantially Equivalent (SE) to the predicate device, Sterile Hypodermic Syringe for Single Use, without Needle, K072739, in respect of safety and effectiveness.
The proposed device, Sterile Insulin Syringe for single use with needle, is determined to be Substantially Equivalent (SE) to the predicate device, Disposable Insulin Syringe, K110421, in respect of safety and effectiveness.
The proposed device, Sterile Hypodermic Needle for single use, is determined to be Substantially Equivalent (SE) to the predicate device. BD Hypoint, K070040, in respect of safety and effectiveness.
{5}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like symbol with three curved lines forming its body and wings. The logo is surrounded by a circular border containing the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" in capital letters. The text is arranged to follow the curve of the circle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Shanghai Kindly Enterprise Division Group Company C/O Ms. Diane Hong MID-Link Consulting Company Limited P.O. Box 237-023 Shanghai China 200237
DEC 1 4 2011
### Re: K112057
Trade/Device Name: Sterile Hypodermic Syringe for Single use with/ without Needle, Sterile Hypodermic Needle for Single use, Sterile Insulin Syringe for Single use with Needle (U-40), Sterile Insulin Syringe for Use with Needle (U-100) Regulation Number: 12 CFR 880.5860 Regulation Name: Piston Syringe Regulatory Class: II Product Code: FMF, FMF, FMI Dated: November 18, 2011 Received: November 29, 2011
Dear Ms. Hong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
Page 2- Ms. Hong
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820): and if applicable, the clectronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please.go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
Anthony V. Nun
Anthony D. Watson, B.S., M.S., M.B.A. Division Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
# Exhibit #1 Indications for Use Statement
510(k) Number: K112057
Device Name: Sterile Hypodermic Syringe for Single Use with/without Needle
Indications for Use:
The sterile hypodermic syringe for single use with/without needle is intended to be used for medical purposes to inject fluid into or withdraw fluid from body.
OPRESCRIPTION USE (Part 21 CFR 801 Subpart D) □OVER-THE-COUNTER USE (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page I of 4
Rill C. Ch 12/09/11
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K112057
{8}------------------------------------------------
# Indications for Use Statement
510(k) Number: K112057
Device Name: Sterile Insulin Syringe for Single Use with Needle (U-100)
Indications for Use:
The sterile insulin syringe for single use with needle, with the calibration unit of insulin for U-100, is a device intended for medical purposes for the manual aspiration of insulin, and for the injection of insulin into parts of the body below the surface skin.
□PRESCRIPTION USE (Part 21 CFR 801 Subpart D) ZOVER-THE-COUNTER USE (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 2 of 4
Al C. Ch (2/14/11
ivision Sign-Off) islon of Anesthesiology, General Hospital ection Control, Dental Devices
510(k) Number: K112057
{9}------------------------------------------------
# Indications for Use Statement
510(k) Number: Ki112057 Device Name: Sterile Insulin Syringe for Single Use with Needle (U-40)
Indications for Use:
The sterile insulin syringe for single use with needle, with the callbration unit of U-40, is a device intended for medical purposes for the manual aspiration of insulin, and for the injection of insulin into parts of the body below the surface skin.
[ PRESCRIPTION USE (Part 21 CFR 801 Subpart D) ZOVER-THE-COUNTER USE (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER I'AGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 2 of 4
Riff C. Chaym 12/14/11
(Division Sign-Off)
Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number:
112057
{10}------------------------------------------------
# Indications for Use Statement
510(k) Number: K112057 Device Name: Sterile Hypodermic Needle for Single Use
Indications for Use:
The sterile hypodermic needle for single use is intended for use with syringes and injection devices for general purpose fluid injection/aspiration.
図PRESCRIPTION USE (Part 21 CFR 801 Subpart D) OOVER-THE-COUNTER USE (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE) '
Page 3 of 4
4
Rill C. Chape 12/14/11
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K112057
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.