K022441 · Advanced Sterilization Products · JOJ · Oct 8, 2002 · General Hospital
Device Facts
Record ID
K022441
Device Name
STERRAD SEALSURE CHEMICAL INDICATOR TAPE
Applicant
Advanced Sterilization Products
Product Code
JOJ · General Hospital
Decision Date
Oct 8, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.2800
Device Class
Class 2
Indications for Use
STERRAD® SealSure™ Chemical Indicator Tape is a process indicator intended for use by health care providers to secure non-woven sterilization packs and wraps to be sterilized in the STERRAD® Sterilization System. The color of the STERRAD® SealSure™ Chemical Indicator Tape changes from red to gold (or lighter) when exposed to hydrogen peroxide and is intended to differentiate between processed and unprocessed loads. Reference the color comparator bars on the package box label.
Device Story
STERRAD® SealSure™ Chemical Indicator Tape is a physical/chemical process indicator used by healthcare providers to secure non-woven sterilization packs/wraps. The tape functions as a through-put indicator; it changes color from red to gold (or lighter) upon exposure to hydrogen peroxide within the STERRAD® Sterilization System. It serves to differentiate processed from unprocessed loads. The device does not signify sterilization; it is used in addition to biological indicators. It is applied manually to sterilization packs before processing. The color change provides a visual cue to clinicians to verify that items have undergone the sterilization cycle.
Clinical Evidence
Bench testing only. Studies included functionality testing in STERRAD® systems (color change at half-cycle), chemical/plasma/temperature resistance, shelf-life stability (pre- and post-processing), and adhesion strength. Efficacy testing confirmed that the use of the indicator tape does not interfere with the sterilization efficacy of the STERRAD® system.
Technological Characteristics
Class 1 chemical indicator (ISO 11140-1) and Class A (EN867-1). Tape form factor; 55 meters per roll. Operates via chemical reaction to hydrogen peroxide resulting in a visual color change from red to gold. Adhesion strength meets minimum requirements. No electronic components or software.
Indications for Use
Indicated for use by healthcare providers to secure non-woven sterilization packs and wraps for sterilization in the STERRAD® Sterilization System; intended to differentiate between processed and unprocessed loads via color change.
Regulatory Classification
Identification
Biological sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor adequacy of sterilization. The device consists of a known number of microorganisms, of known resistance to the mode of sterilization, in or on a carrier and enclosed in a protective package. Subsequent growth or failure of the microorganisms to grow under suitable conditions indicates the adequacy of sterilization. Physical/chemical sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.
{0}------------------------------------------------
# K022441 510(k) SUMMARY
#### Applicant's Name, Address, Telephone, FAX, Contact Person
Advanced Sterilization Products Division of Ethicon, Inc. 33 Technology Drive Irvine, CA 92618
### Contact Person
Natalie Bennington Senior Regulatory Affairs Specialist Tel: (949) 453-6482 Fax: (949) 789-3900
September 12, 2002
This summary of 510(k) substantial equivalence information is being submitted in accordance with the requirements of 21 CFR §807.92.
### CLASSIFICATION, COMMON OR USUAL NAME, DEVICE NAME
Common/Usual Name: Chemical Indicator Product Classification: Physical / Chemical Sterilization Process Indicator, Class II Proprietary Name: STERRAD® SealSure™ Chemical Indicator Tape
### PREDICATE DEVICE
The predicate device is the STERRAD® Chemical Indicator Tape for use in the STERRAD® Sterilization Systems, currently manufactured and distributed by Advanced Sterilization Products.
### INDICATIONS-FOR-USE
STERRAD® SealSure™ Chemical Indicator Tape is a process indicator intended for use by health care providers to secure non-woven sterilization packs and wraps to be sterilized in the STERRAD® Sterilization System.
The color of the STERRAD® SealSure™ Chemical Indicator Tape changes from red to gold (or lighter) when exposed to hydrogen peroxide and is intended to differentiate between processed and unprocessed loads. Reference the color comparator bars on the package box label.
### DEVICE DESCRIPTION
STERRAD® SealSure™ Chemical Indicator Tape (Class A per EN867-1 and Class 1 ISO 11140-1) is a through-put process indicator tape to be used with ASP's STERRAD® Sterilization System. STERRAD® SealSure™ Chemical Indicator Tape offers an additional way to verify processing in the sterilization cycle. It should be used in addition to, not in place of, the biological indicator. STERRAD® SealSure™ Chemical Indicator Tape does not signify sterilization; it only indicates that the indicator has been exposed to hydrogen peroxide.
{1}------------------------------------------------
# KO 22441
There are six rolls of STERRAD® SealSure™ Chemical Indicator Tape per box. Each roll contains 55 meters (60 yards) of indicator tape.
# COMPARISON TO THE PREDICATE DEVICE
STERRAD® SealSure™ Chemical Indicator Tape for use in the STERRAD® Sterilization System is substantially equivalent to STERRAD® Chemical Indicator Tape also for use in the STERRAD® Sterilization System. These devices are intended for use by health care providers to secure non-woven sterilization wraps used on medical devices to be sterilized in the STERRAD® Sterilization System and to differentiate processed items. In addition, both of these devices share similar design and appearance.
# VALIDATION STUDIES
- 1. Functionality Study of the STERRAD® SealSure™ Chemical Indicator Tape in STERRAD® Sterilization Systems The functionality of STERRAD® SealSure™ Chemical Indicator Tape was tested in each of the STERRAD® Sterilization Systems. Functionality was verified as all tape samples
changed from red to yellow at half-cycle conditions.
- 2. Effects of Chemicals, Air Plasma and Shipping Temperature on STERRAD® SealSure™ Chemical Indicator Tape for use in the STERRAD® Systems The effects of chemicals, air plasma and shipping temperature on STERRAD® SealSure™ Chemical Indicator Tape for use in the STERRAD® Sterilization Systems were tested.
The test results demonstrated that STERRAD® SealSure™ Chemical Indicator Tape has adequate resistance to chemicals, with exception of organic solvents with high polarity as well as strong acids. STERRAD® SealSure™ Chemical Indicator Tape also has the ability to distinguish between a processed load vs. a load in which the cycle was cancelled prior to exposure to the hydrogen peroxide. Lastly, it was shown that STERRAD® SealSure™ Chemical Indicator Tape maintained its stability in a simulated shipping environment.
3. Color and Functionality Studies of Pre and Post-processing STERRAD® SealSure™ Chemical Indicator Tape Samples (Unopened Package) Shelf life color stability studies were conducted for both pre-processed and post-processed
samples of STERRAD® SealSure™ Chemical Indicator Tape. In addition, a functionality stability study was also performed on the pre-processed samples.
The test results demonstrated that the pre-processed samples maintained the pre-processing color and functionality at both ambient and accelerated aging conditions for four months. In addition, the post-processed tape samples retained their post-processing color after aged in both ambient and accelerated aging conditions for two months.
- 4. Color and Functionality Stability Studies of Pre and Post-processing STERRAD® SealSure™ Chemical Indicator Tape Samples (Opened Package)
The color and functionality stability of opened-package STERRAD® SealSure™ Chemical Indicator Tape samples stored in two ambient conditions were tested. The first ambient
{2}------------------------------------------------
# KO22441
condition was used to simulate a hospital's sterilizer room and the other was used to simulate the storage room of the central supply area (CS Department).
The stability of post-processed tape samples stored in a storage room condition was also evaluated.
The results of this study demonstrated that the color and functionality stability of the preprocessed tape samples were maintained in each ambient condition for one month. In addition, the post-processing color was stable for one month. The pre-processed color and functionality stability studies will continue for a total of two months.
- 5. Efficacy Testing of STERRAD® Sterilization System with excess amount of STERRAD® SealSure™M Chemical Indicator Tape
The efficacy of using an extreme amount of STERRAD® SealSure™ Chemical Indicator Tape on each validation tray in each of the STERRAD® Sterilization Systems was tested. Complete inactivation of challenge microorganisms in the defined test configuration was achieved in each System.
The test results demonstrated that the effect on sterilization efficacy of each of the STERRAD® Systems is insignificant, regardless of using an extreme amount of chemical indicator tape on each validation tray.
# 6. Adhesion Strength and Stability of the STERRAD® SealSure™ Chemical Indicator Tape for use in the STERRAD® Systems
The adhesion strength and stability of STERRAD® SealSure™ Chemical Indicator Tape was tested.
The test results demonstrated that STERRAD® SealSure™ Chemical Indicator Tape meets the minimum adhesion strength requirements. Stability testing results also demonstrated the maintenance of the minimum adhesion strength over three months in accelerated and room temperature aging environments.
# CONCLUSION
Results of the validation studies demonstrate that the STERRAD® SealSure™ Chemical Indicator Tape meet or exceed all functional requirements and support its suitability for use.
{3}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like figure with three curved lines representing its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular pattern around the bird-like figure.
#### Public Health Service
Food and Drug Administration 200 Corporate Boulevard Rockville MD 20850
OCT 0 8 2002
Ms. Natalie Bennington Senior Regulatory Affairs Specialist Advanced Sterilization Products Division of Ethicon, Incorporated 33 Technology Drive Irvine, California 92618
- Re: K022441
Trade/Device Name: STERRAD® Sealsure™ Chemical Indicator Tape Regulation Number: 21 CFR 880.2800 Regulation Name: Chemical Sterilization Process Indicator Regulatory Class: II Product Code: JOJ Dated: July 24, 2002 Received: July 25, 2002
Dear Ms. Bennington:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Ms. Bennington
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4618. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.qov/cdrh/dsma/dsmamain.html
Sincerely,
Timothy A. Ulatowski
Timothy A. Ulatowski Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows a logo with the letters "ASP" inside of a rounded rectangle. The letters are white and the background is black. The "A" is shaped like a triangle, and the "S" and "P" are stylized.
ADVANCED STERILIZATION PRODUCTS a Johnon Johnion company
REGULATORY AFFAIRS DEPARTMENT
# 510(k) Number (if known): K022441
STERRAD® SealSure™ Chemical Indicator Tape Device Name:
# Indications-For-Use:
STERRAD® SealSure™ Chemical Indicator Tape is a process indicator intended for use by health care providers to secure non-woven sterilization packs and wraps to be sterilized in the STERRAD® Sterilization System.
The color of the STERRAD® SealSure™ Chemical Indicator Tape changes from red to gold (or lighter) when exposed to hydrogen peroxide and is intended to differentiate between processed and unprocessed loads. Reference the color comparator bars on the package box label.
### (PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| Prescription Use<br>(PER 21 CFR 801.109) | | OR | Over-The-Counter Use<br>(Optional Format 1-2-96) |
|------------------------------------------|--|----|--------------------------------------------------|
|------------------------------------------|--|----|--------------------------------------------------|
DIVISION OF ETHICON, INC. • 33 TECHNOLOGY DRIVE • IRVINE, CA 92618 • (949) 581-5799 FAX (949) 789-3900
(Division Sign-Off)
Division of Anesthesiology, General Hospital,
Infection Control, Dental Devices
| 510(k) Number: | K022441 |
|----------------|---------|
|----------------|---------|
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.