K172753 · STERIS Corporation · JOJ · Feb 9, 2018 · General Hospital
Device Facts
Record ID
K172753
Device Name
VERIFY VH2O2 Indicator Tape
Applicant
STERIS Corporation
Product Code
JOJ · General Hospital
Decision Date
Feb 9, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.2800
Device Class
Class 2
Indications for Use
The VERIFY™ VH2O2 Indicator Tape is intended to secure items wrapped in synthetic wrap materials until used and to distinguish between processed and unprocessed units through a color change from the start color (pink) to peach, yellow or lighter. The VERIFY™ VH2O2 Indicator Tape is also intended to be used as an external process indicator when applied to synthetic wrap materials and/or Tyvek pouches. The tape may be used in the following sterilization cycles: Non Lumen, Flexible, Lumen and Fast Non Lumen Cycles of the V-PRO 1, 1 Plus, maX, 60 and maX 2 Sterilizers; Standard Cycle of the STERRAD® 100S Sterilizer; Standard and Advanced Cycles of the STERRAD® NX Sterilizer; Express, Flex Scope, Duo and Standard Cycles of the STERRAD® 100 NX Sterilizer
Device Story
VERIFY VH2O2 Indicator Tape is a chemical process indicator used in healthcare settings to secure wrapped surgical items and verify exposure to hydrogen peroxide sterilization. The tape consists of coated crepe paper with a chemical indicator that undergoes a color change (pink to peach/yellow) upon exposure to hydrogen peroxide vapor. It is applied by clinical staff to synthetic wraps or Tyvek pouches before sterilization. The device serves as a visual indicator to distinguish between processed and unprocessed units. It is compatible with various V-PRO and STERRAD sterilization cycles. The output is a simple color change observed by the healthcare provider, which confirms the item has been subjected to the sterilization process, aiding in inventory management and ensuring that only processed items are used in clinical procedures.
Clinical Evidence
Bench testing only. Performance was validated through simulated use testing in the V-PRO maX 2 Fast Non Lumen Cycle under both pass and fail conditions, and 18-month shelf-life stability testing. All tests met the requirements of ANSI/AAMI/ISO 11140-1:2014 for Type 1 chemical indicators.
Technological Characteristics
Substrate: Crepe paper. Indicator: Proprietary chemical formulation conforming to ANSI/AAMI/ISO 11140-1:2014 Type 1. Adhesive: Proprietary. Form factor: Tape. Sterilization compatibility: Vapor-phased hydrogen peroxide. Free of lead, heavy metals, and natural rubber latex.
Indications for Use
Indicated for use in healthcare facilities to secure wrapped surgical items and provide visual confirmation of exposure to hydrogen peroxide sterilization cycles (V-PRO and STERRAD systems) by changing color from pink to peach/yellow.
Regulatory Classification
Identification
Biological sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor adequacy of sterilization. The device consists of a known number of microorganisms, of known resistance to the mode of sterilization, in or on a carrier and enclosed in a protective package. Subsequent growth or failure of the microorganisms to grow under suitable conditions indicates the adequacy of sterilization. Physical/chemical sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a symbol representing the Department of Health & Human Services. To the right, there is the FDA logo in blue, with the words "U.S. FOOD & DRUG" above the word "ADMINISTRATION".
February 9, 2018
Steris Corporation Bill Brodbeck Director, Regulatory Affairs 5960 Heisley Road Mentor, Ohio 44060
Re: K172753
Trade/Device Name: VERIFY VH2O2 Indicator Tape Regulation Number: 21 CFR 880.2800 Regulation Name: Sterilization Process Indicator Regulatory Class: Class II Product Code: JOJ Dated: January 9, 2017 Received: January 10, 2018
Dear Bill Brodbeck:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{2}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Michael J. Ryan -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K172753
Device Name VERIFY VH2O2 Indicator Tape
### Indications for Use (Describe)
The VERIFY™ VH2O2 Indicator Tape is intended to secure items wrapped in synthetic wrap materials until used and to distinguish between processed and unprocessed units through a color change from the start color (pink) to peach, yellow or lighter. The VERIFY™ VH2O2 Indicator Tape is also intended to be used as an external process indicator when applied to synthetic wrap materials and/or Tyvek pouches.
The tape may be used in the following sterilization cycles:
- Non Lumen, Flexible, Lumen and Fast Non Lumen Cycles of the V-PRO 1, 1 Plus, maX, 60 and maX 2 Sterlizers
- · Standard Cycle of the STERRAD® 100S Sterilizer
- · Standard and Advanced Cycles of the STERRAD® NX Sterilizer
- · Express, Flex Scope, Duo and Standard Cycles of the STERRAD® 100 NX Sterilizer
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
STERIS®
Image /page/3/Picture/2 description: The image shows a blue graphic of six horizontal lines that are curved. The lines are stacked on top of each other, with the top line having the smallest curve and the bottom line having the largest curve. The lines are all the same thickness and color, and they are evenly spaced apart.
# 510(k) Summary for VERIFY™ VH2O2 Indicator Tape K172753
STERIS Corporation 5960 Heisley Road Mentor, OH 44060 Phone: (216) 354-2600 Fax No: (216) 639-4459
Contact: Bill Brodbeck Director, Regulatory Affairs Telephone: (440) 392-7690 Fax No: (440) 357-9198 E-mail: William_brodbeck@steris.com
Submission Date: September 11, 2017
STERIS Corporation ■ 5960 Heisley Road ■ Mentor, OH 44060-1834 USA ■ 440-354-2600
{5}------------------------------------------------
### TRADITIONAL PREMARKET NOTIFICATION [510(k)] VERIFYTM VH2O2 Indicator Tape
#### Device Name 1.
| Trade Name: | VERIFY™ VH2O2 Indicator Tape |
|----------------------|-----------------------------------------------------------|
| Common/Usual Name: | Chemical Sterilization Process Indicator |
| Classification Name: | Sterilization Process Indicator<br>(21 CFR 880.2800, JOJ) |
| Classification: | Class II |
#### Predicate Device 2.
VERIFYTM VH2O2 Indicator Tape (K160485)
### 3. Description of Device
The VERIFY VH2O2 Indicator Tape consists of a special coated crepe paper with diagonal lines of a chemical indicator specific for hydrogen peroxide sterilization. It meets the performance specifications for a Type 1 process indicator as defined in ANSI/AAMI/ISO 11140-1:2014.
The VERIFY | VH2O2 Indicator Tape and its packaging are manufactured in the absence of lead, heavy metals, natural rubber latex and dry natural rubber.
#### Intended Use/ Indications for Use 4.
The VERIFY™ VH2O2 Indicator Tape is intended to secure items wrapped in synthetic wrap materials until used and to distinguish between processed and unprocessed units through a color change from the start color (pink) to peach, yellow or lighter. The VERIFY™ VH2O2 Indicator Tape is also intended to be used as an external process indicator when applied to synthetic wrap materials and/or Tyvek pouches.
The tape may be used in the following sterilization cycles:
- Non Lumen, Flexible, Lumen and Fast Non Lumen Cycles of the V-PRO 1, 1 . Plus, maX, 60 and maX 2 Sterilizers
- Standard Cycle of the STERRAD® 100S Sterilizer ●
- Standard and Advanced Cycles of the STERRAD® NX Sterilizer ●
- Express, Flex Scope, Duo and Standard Cycles of the STERRAD® 100 NX Sterilizer
{6}------------------------------------------------
### TRADITIONAL PREMARKET NOTIFICATION [510(k)] VERIFY™ VH2O2 Indicator Tape
| Table 5.1 Indications for Use Comparison | | |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Proposed | Predicate (K160485) | Comparison |
| VERIFY VH2O2 Indicator Tape | VERIFY VH2O2 Indicator Tape | |
| The VERIFYTM VH2O2 Indicator<br>Tape is intended to secure items<br>wrapped in synthetic wrap materials<br>until used and to distinguish<br>between processed and unprocessed<br>units through a color change from<br>the start color (pink) to peach,<br>yellow or lighter. The VERIFYTM<br>VH2O2 Indicator Tape is also<br>intended to be used as an external<br>process indicator when applied to<br>synthetic wrap materials and/or<br>Tyvek pouches.<br><br>The tape may be used in the<br>following sterilization cycles:<br>Non Lumen, Flexible, Lumen<br>and Fast Non Lumen Cycles of<br>the V-PRO 1, 1 Plus, maX, 60<br>and maX 2 Sterilizers Standard Cycle of the<br>STERRAD® 100S Sterilizer Standard and Advanced Cycles<br>of the STERRAD® NX<br>Sterilizer Express, Flex Scope, Duo and<br>Standard Cycles of the<br>STERRAD® 100 NX Sterilizer | The VERIFY VH2O2 Indicator Tape is<br>intended to secure items wrapped in<br>synthetic wrap materials until used and<br>to distinguish between processed and<br>unprocessed units through a color<br>change from the start color (pink) to<br>peach, yellow or lighter.<br><br>The tape may be used in the following<br>sterilization cycles:<br>Non Lumen, Flexible and Lumen<br>Cycles of the V-PRO 1, 1 Plus, maX<br>and 60 Low Temperature Sterilizers Standard Cycle of the STERRAD®<br>100S Sterilizer Standard and Advanced Cycles of the<br>STERRAD® NX Sterilizer Express,<br>Flex Scope, Duo and Standard<br>Cycles of the STERRAD® 100 NX<br>Sterilizer | The Indications for<br>Use has been<br>modified in the<br>proposed device in<br>order to include the<br>Fast Non Lumen<br>Cycle of the V-PRO<br>maX 2 device.<br><br>The proposed device<br>also includes the<br>statement that the<br>VERIFYTM VH2O2<br>Indicator Tape is<br>also intended to be<br>used as an external<br>process indicator<br>when applied to<br>synthetic wrap<br>materials and/or<br>Tyvek pouches. |
#### 5. Description of Safety and Substantial Equivalence
The VERIFY VH2O2 Indicator Tape has equivalent characteristics as compared to its predicate device. Table 5.2 compares the proposed device's design to the predicate.
| Characteristic | Proposed | Predicate | Comparison |
|----------------|----------------------------------------------------------------------------------|---------------------------------------------------------------------------|---------------------------------------------------------|
| Substrate | Crepe Paper | Crepe Paper | Same |
| Indicator | Proprietary -<br>conforms<br>with ISO<br>11140-<br>1:2014 Type<br>1 requirements | Proprietary -<br>conforms with ISO<br>11140-1:2014 Type<br>1 requirements | Same |
| Adhesive | Proprietary | Proprietary | Testing included in this<br>submission<br>demonstrating |
### Table 5.2 Device Comparison
{7}------------------------------------------------
| TRADITIONAL PREMARKET NOTIFICATION [510(k)] |
|---------------------------------------------------------------------------|
| VERIFY™ <span style="border: 1px solid black">VH2O2</span> Indicator Tape |
| Characteristic | Proposed | Predicate | Comparison |
|--------------------|---------------------------|---------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------|
| | | | performance (See Table<br>5.3). |
| Color Change | Pink to<br>Peach/Yellow | Pink to<br>Peach/Yellow | Same |
| Detection | hydrogen<br>peroxide dose | hydrogen peroxide<br>dose | Same |
| Shelf-life | 18 months | 18 months | Same |
| Storage Conditions | 16 - 24°C,<br>30 - 60% RH | 16 - 24°C<br>30 - 60% RH | Testing included in this<br>submission to support<br>proposed device's shelf<br>life when stored under<br>labeled conditions (See<br>Table 5.3). |
Since the VERIFY | VH2O2 Indicator Tape is identical to the predicate device, prior nonclinical testing (K160485) is referenced in this submission to support performance of the chemical indicator in accordance to the vapor-phased hydrogen peroxide, Type 1 chemical indicator requirements in ANSI/AAMI/ISO 11140-1:2014. Furthermore, since the proposed device is identical to the proposed device, testing was provided in K160485 addresses endpoint stability, light stability, biocompatibility and adhesion. To demonstrate that the CI performs appropriately according to the proposed device's expanded claims, performance was tested in the V-PRO maX 2's Fast Non Lumen Cycle with loads under simulated use conditions under pass and fail conditions. Shelf-life testing evaluated CI performance per ANSI/AAMI/ISO 11140-1:2014, endpoint color stability and tape adhesion after various times of storage within the labeled storage conditions. Table 5.3 summarizes the testing performed in the current submission.
### Table 5.3 Summary of Nonclinical Tests:
| Test | Result |
|-----------------------------------------------|--------|
| Simulated Use Testing in Fast Non Lumen Cycle | Pass |
| Stability Testing for 18 month Shelf Life | Pass |
#### 6. Conclusion
Based on the intended uses, technological characteristics and non-clinical performance data, the subject device is as safe, as effective and perform at least as well as the legally marketed predicate device K160485, Class II (21 CFR 880.2800), product code JOJ.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.