INFRARED EAR THERMOMETER

K103097 · Radiant Innovation, Inc. · FLL · Dec 29, 2010 · General Hospital

Device Facts

Record IDK103097
Device NameINFRARED EAR THERMOMETER
ApplicantRadiant Innovation, Inc.
Product CodeFLL · General Hospital
Decision DateDec 29, 2010
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 880.2910
Device ClassClass 2
AttributesPediatric

Indications for Use

The device is an electronic clinical thermometer using an infrared sensor to detect body temperature from the auditory canal in the neonatal, pediatric and adult population used for intermittent monitoring of human body temperature in the home setting. The probe cover is intended as a disposable cover acting as a sanitary barrier between the thermometer and the ear canal.

Device Story

Infrared ear thermometer (THK09, IRT 3020) measures natural infrared thermal radiation from tympanic membrane and auditory canal; uses thermopile detector to convert radiation to electrical signal; signal amplified and processed by microprocessor; displays temperature on LCD. Used in home setting by patients or caregivers for intermittent body temperature monitoring. Disposable probe cover acts as sanitary barrier. Provides non-invasive temperature reading to assist in health monitoring.

Clinical Evidence

Clinical study compared THK09 accuracy against Omron MC 341 and Braun IRT 3020. Evaluated bias, standard deviation, and repeatability per ASTM E1965:2003. Results showed bias and repeatability within +/- 0.2°C acceptance criteria; standard deviations within 0.5°C. Concluded clinical accuracy is comparable to legally marketed reference devices.

Technological Characteristics

Infrared thermopile sensor; ABS plastic housing; LCD display; 3V CR2032 battery; dimensions 13.1x6.2x3.2 cm; weight 60g. Complies with ASTM E1965-98, EN12470-5:2003, EN 60601-1, and EN 60601-1-2. Microprocessor-based signal processing.

Indications for Use

Indicated for intermittent monitoring of human body temperature in neonatal, pediatric, and adult populations in a home setting via the auditory canal.

Regulatory Classification

Identification

A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.

Special Controls

(1) Device is not a clinical thermometer with telethermographic functions; (2) Device is not a clinical thermometer with continuous temperature measurement functions; and (3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the word "RADIANT" in large, bold, black letters. Above the word "RADIANT" is the word "INNOVATION" in smaller, black letters. Below the word "RADIANT" are four Chinese characters in black. # SPECIAL 510(K) SUMMARY DEC 2 9 2010 This Special 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92. The assigned 510(k) number is: ### Submitter's Identification: 1. Radiant Innovation Inc. 1F. No.3 Industrial E. 9th Rd., Science-Based Industrial Park, HsinChu, Taiwan ## Contact: Mr. James Huang OA Department Manager Radiant Innovation Inc. TEL: +886 3 6111666 Ext. 100 FAX:+886 3 5644170 E-mail: jameshuang@radiantek.com.tw Date Summary Prepared: Oct./12/2010 ### Name of the Device: 2. Infrared Ear Thermometer THK09, IRT 3020 (Under brand name Braun IRT 3020 to be distributed by Kaz USA, Inc) Classification Name: Thermometer, Electronic, Clinical Regulation Number: 21 CFR 880.2910 #### 3. Current Cleared Devices: Radiant Innovation Infrared Ear Thermometer, Models TH60N (510(k)#: K101253), TH01BN (510(k)#: K070976), TH5XY series (510(k)#: K063185), TH00XY series (510(k)#: K051269), THXXN series (510(k)#: K040377), TH1 series (510(k)#: K030324), TH88 series (510(k)#: K020504) and TH8 series (510(k)#: K011059). #### Device Description: 4. The Radiant Innovation Inc., Infrared Ear Thermometer, Models THK09, IRT 3020 are electronic thermometers that use an infrared detector (thermopile detector) to detect Page 1 {1}------------------------------------------------ Image /page/1/Picture/0 description: The image contains two logos side by side. The logo on the left says "RADIANT INNOVATION" in bold, stylized letters. The logo on the right contains Chinese characters above the text "RADIANT INNOVATION Inc." body temperature using infrared radiantion from the auditory canal. Its operation is based on measuring the natural infrared thermal radiation emanating from the tympanic membrane and the adjacent surfaces of the patient. To measure ear temperature, the ear thermometer is inserted into a patient's outer ear canal. A start button is pressed to start the measurement through the radiation exchanges. The electrical signal read out from the detector is amplified by hardware and processed by the microprocessor. The temperature from the auditory canal in the neonatal, pediatric and adult population used for intermittent monitoring of human body temperature in the home setting. ### 5. Intended Use: The device is an electronic clinical thermometer using an infrared sensor to detect body temperature from the auditory canal in the neonatal, pediatric and adult population used for intermittent monitoring of human body temperature in the home setting. The probe cover is intended as a disposable cover acting as a sanitary barrier between the thermometer and the ear canal. ### 6. Technological Characteristics and Substantial Equivalence: Both the subject device (THK09, IRT 3020) and the predicate device (TH8 series) have the same intended use as well as same fundamental technology. The comparison table showing the differences between the subject device (THK09, IRT 3020) and the predicate device (TH8 series) is included in the 510(k) submission. The subject device (THK09, IRT 3020) is substantially equivalent to the predicate device (TH8 series) since the same intended use, indications for use and similar technological characteristics. | Features | Predicate device(TH8 series) | Subject device (THK09, IRT 3020) | |-----------------------|--------------------------------------------|----------------------------------| | 510(k)# | K011059 | K | | Accuracy | $35.5~42°C (95.9~107.6°F)+/-0.2°C (0.4°F)$ | other +/-0.3°C (0.5°F). | | Temp. Range | 34.0-42.2°C | | | Ambient Range | 10-40°C | | | Response Time | 1 sec | | | Read modes | Ear (Oral) | | | Scale Selection | °C/°F | | | Display Type | LCD | | | Probe Cover | With | | | Activation | Scan Button | Start Button | | Memory | 9 sets | 1 set | | Sensor Type | Thermopile | | | Case | ABS | | | Weight | 70g | 60g | | Dimension (LxWxH) | 14*3.8*3 cm | 13.1*6.2*3.2 cm | | Battery | 3V Battery, CR2032 * 1 | | | 2 Phase Battery Alarm | Yes | | The basic technological characteristics between subject device vs. predicate device. ### Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence 7. Page 2 {2}------------------------------------------------ Image /page/2/Picture/0 description: The image shows the word "RADIANT" in large, bold, black letters. Above the word "RADIANT" is the word "INNOVATION" in smaller, black letters. Below the word "RADIANT" are four Chinese characters. - 1 熱映光電股份有限公司 RADIANT INNOVATION Inc. ### are as follows: Compliance to applicable voluntary standards includes ASTM E1965-98 and EN12470-5:2003, as well as EN 60601-1 (IEC 60601-1) and EN 60601-1-2 (IEC 60601-1-2) requirements. Guidance Documents included the FDA "Guidance On The Content of Premarket Notification (510(k)) Submissions for Clinical Electronic Thermometers", "How to Prepare A Special 510(k)", "Deciding When to Submit a 510(k) for a Change to an Existing Device". #### 8. Summary of Clinical Evaluation: The objective of this study was to compare the clinical accuracy of the THK09 Infrared Ear thermometer with that of two legally marketed thermometers in the US i) the MC 341 Omron Digital Thermometer (K091676) and ii) Braun IRT 3020 Ear Thermometer (K983295). In accordance with ASTM E1965:2003, the clinical accuracy is characterized in terms of Bias, Standard Deviation or Limits of Agreement and Repeatability. ### Acceptance Criteria: For bias and repeatability, a clinically acceptable value of +/- 0.2 deg C was used as the acceptance criteria. ### Results The summary results for bias, SD and Repeatability using the Omron MC 341 and the Braun IRT 3020 Ear thermometers as reference devices showed that the bias and repeatability was within the clinically relevant acceptance criteria of +/-0.2 deg C. The standard deviations were within 0.5 deg C. ### Conclusion: The THK09 has a clinical accuracy comparable to the legally marketed MC 341 Omron Digital Thermometer (K091676) and ii) Brain IRT 3020 Ear Thermometer (K983295). #### 9. Conclusions: The RII Infrared Ear Thermometer (models THK09, IRT 3020) has the same intended use, indications for use and technological characteristics as the cleared device TH8 series device. The devices have minor differences of some additional features and corresponding hardware or software changes. Bench testing contained in this submission supplied demonstrates these changes do not raise any new questions of safety or effectiveness. The materials used are identical to the ones used in the previously cleared devices. Clinical Evaluation of the THK09 shows that it is within the acceptance criteria of +/-0.2 deg C when compared to the legally marketed predicate reference devices when conducted in accordance with ASTM E1965:2003. Thus, the RII Infrared Ear Thermometer, Models THK09, IRT 3020 are substantially equivalent to the predicate device. {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is a stylized symbol that resembles three overlapping human figures. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 Mr. James Huang Quality Assurance Department Manager Radiant Innovations Incorporated 1F, No. 3 Industrial E. 9th Rd., Science-Based Industrial Park, HsinChu, Taiwan 300 DEC 20 1000 Re: K103097 Trade/Device Name: Radiant Innovation Inc. Infrared Ear Thermometer THK09, IRT 3020 Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: II Product Code: FLL Dated: December 3, 2010 Received: December 6, 2010 Dear Mr. Huang: . We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ ## Page 2- Mr. Huang Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health ### Enclosure {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the word "RADIANT" in large, bold, black letters. Above the word "RADIANT" is the word "INNOVATION" in smaller, black letters. Below the word "RADIANT" are four Chinese characters in black ink. 103097 EXHIBIT # B # INDICATIONS FOR USE DEC 2 9 2010 510(k) Number (if known): Device Name: Radiant Innovation Inc. Infrared Ear Thermometer THK09, IRT 3020 Indications For Use: The device is an electronic clinical thermometer using an infrared sensor to detect body temperature from the auditory canal in the neonatal, pediatric and adult population used for intermittent monitoring of human body temperature in the home setting. Prescription Use (Part 21 CFR 801 Subpart D) AND/OR (21 CFR 807 Subpart C) Over-The-Counter Use ਨਿੱ (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices 510(k) Number: K103097 Page 1 of 1 Page 12
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...