K091676 · Omron Healthcare, Inc. · FLL · Sep 1, 2009 · General Hospital
Device Facts
Record ID
K091676
Device Name
MC-245, MC-246, MC-247, MC-341, MC-343, MC-344
Applicant
Omron Healthcare, Inc.
Product Code
FLL · General Hospital
Decision Date
Sep 1, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.2910
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Omron electronic thermometer Models MC-245, MC-341, MC-341, MC-343, and MC-344 are intended to measure the body temperature either oral, axillaries (under arm) and rectal and Model - MC-247 measure the body temperature oral (basal) and to be used by medical professionals in clinical and hospital environments and consumers in a home environment. It is intended for use on people of all ages.
Device Story
Electronic clinical thermometers; thermistor-based temperature sensing. Input: thermal energy from patient body site (oral, axillary, rectal, or basal). Transformation: thermistor detects temperature; internal circuitry calculates reading; buzzer signals peak temperature. Output: digital temperature reading on LCD. Used in clinics, hospitals, and home environments by medical professionals or patients. Provides objective temperature data to assist healthcare providers in clinical decision-making; enables patient monitoring of body temperature.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on bench testing and comparison of technological characteristics and specifications to the predicate device.
Technological Characteristics
Thermistor-based temperature sensing; battery-operated (LR41); LCD digital display; 4-digit resolution (0.1 increments); accuracy ±0.2°F/±0.1°C. Materials tested per ISO 10993. Models feature rigid or flexible tips. Connectivity: None. Sterilization: Not specified.
Indications for Use
Indicated for measuring body temperature (oral, axillary, rectal) in patients of all ages; Model MC-247 indicated for oral (basal) temperature only. For use by medical professionals in clinical/hospital settings and consumers in home environments.
Regulatory Classification
Identification
A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.
Special Controls
(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.
{1}------------------------------------------------
Premarket Notification 510(k) Section 5 - 510(k) Summary
Omron Thermometers
KC916676
| Omron Healthcare, Inc.<br>1200 Lakeside Dr.<br>Bannockburn, IL 60015 | Tel - 847-247-5678<br>Fax - 847-680-6269 | SEP - 1 2009 |
|----------------------------------------------------------------------|------------------------------------------------------------|--------------|
| Official Contact: | Ronald Dudek - Director of Quality | |
| Proprietary or Trade Name: | Models – MC-245, MC-246, MC-247, MC-341, MC-343,<br>MC-344 | |
| Common/Usual Name: | Clinical electronic thermometer | |
| Classification Name/Code: | FLL – clinical electronic thermometer<br>21 CFR 880.2910 | |
| Device: | Models – MC-245, MC-246, MC-247, MC-341, MC-343,<br>MC-344 | |
#### Device Description:
These are simple electronic thermometer which is a thermistor to measure temperature.
- . All the functions, except posture symbol, are the same as the predicate device, MC-3B, K881446.
- Power source battery operated .
- Calculation method (system) is the same and identical to the predicate, K881449 �
The differences between the various models are:
- Some models have a flexible or rigid or rigid / flexible tip .
- Offered with a probe cover .
- Temperature scale offered in C° & F° or F° only .
- Readout time is different .
#### Indications for Use:
The Omron electronic thermometer Models MC-245, MC-341, MC-341, MC-343, and MC-344 are intended to measure the body temperature either oral, axillaries (under arm) and rectal and Model - MC-247 measure the body temperature oral (basal) and to be used by medical professionals in clinical and hospital environments and consumers in a home environment. It is intended for use on people of all ages.
Environment of Use: Clinics, hospital and home environments
Contraindications: None
{2}------------------------------------------------
# 510(k) Summary Page 2 of 3 5-Jun-09
#### Summary of substantial equivalence .
| | MC-3B<br>K881446 | MC-245/246/247<br>MC-341/343/344 |
|-----------------------|---------------------------------------------------------------------|---------------------------------------------------------------------|
| Measurement<br>Method | Human temperature is detected by<br>thermistor and calculated | Human temperature is detected by<br>thermistor and calculated |
| Display | Reading value is displayed on LCD<br>with 4 digits.(0.1 increments) | Reading value is displayed on LCD<br>with 4 digits.(0.1 increments) |
| Buzzer | Beeper sound at peak temperature | Beeper sound at peak temperature |
| Power Source | Battery(LR41 type) | Battery(LR41 type) |
| SPECIFICATION | MC-3B<br>K881446 | MC-245, MC-246, MC-247<br>MC-341, MC-343, MC-344 |
|---------------------------------------|----------------------------------------|-------------------------------------------------------------------------------------------------------|
| Display | LCD Digital Display | LCD Digital Display |
| Measurement<br>Range | 89.6 °F to 107.6 °F | 89.6 °F to 107.6 °F, 32.0 °C to 42.0 °C<br>These models switch the display unit (°F, °C) alternately. |
| Measurement area | Oral, rectum, under arm | Oral, rectum, under arm, except Model MC-247 is for oral (basal) only |
| Accuracy | ±0.2 °F | ±0.2 °F, ±0.1 °C |
| Display Resolution | 4 Digits<br>(0.1 °F increments) | 4 Digits<br>(0.1 °F / 0.1 °C increments) |
| Weight | Approx. 1/3 oz /(11g) | Approx. 1/3 oz / (11g) |
| Length | Approx. 130mm | Approx. 130mm |
| Battery | 1 each LR41 Alkali<br>1.5V D.C | 1 each LR41 Alkali 1.5V D.C |
| Battery Life | Approx. 300 hours.<br>(Continuous use) | Approx. 2 years (3 times measurements / day ) |
| Ambient<br>Temperature for<br>use | 50 °F to 104 °F<br>(10 °C to 40 °C) | 50 °F to 104 °F<br>(10 °C to 40 °C) |
| Ambient<br>Temperature for<br>use | (10 ° C to 40 ° C) | (10 ° C to 40 ° C) |
| Ambient<br>Temperature for<br>storage | 4 °F to 140 °F<br>(-20 °C to 60 °C) | 4 °F to 140 °F<br>(-20 °C to 60 °C) |
| Ambient<br>Temperature for<br>storage | (-20 ° C to 60 ° C) | (-20 ° C to 60 ° C) |
| Ambient Humidity<br>for use | 30 to 85%RH | 30 to 85%RH |
| Ambient Humidity<br>for storage | 30 to 95%RH | 10 to 95%RH |
.
{3}------------------------------------------------
#### 510(k) Summary Page 3 of 3
5-Jun-09
#### Differences Between Other Legally Marketed Predicate Devices
The Omron thermometer models MC-245, MC-247, MC-341, MC-341, MC-343, and MC-344 are viewed as substantially equivalent to the predicate device because:
# Indications -
- . Identical to predicate - K881446
### Technology -
- . Identical algorithms and means of measuring temperature to predicate - K881446
# Operating specifications -
- Identical to predicate K881446 .
# Materials -
The materials have been tested in accordance to ISO 10993.
# Environment of Use -
- Identical to predicate K881446 �
#### Patient Population -
- Identical to predicate K881446 .
{4}------------------------------------------------
Image /page/3/Picture/11 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a stylized symbol. The symbol consists of four abstract shapes that resemble a stylized human figure or a flame-like design. The logo is in black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002
Omron Healthcare, Incorporated C/O Mr. Paul Dryden President ProMedic, Incorporated 24301 Woodsage Drive Bonita Springs, Florida 34134
Re: K091676
Trade/Device Name: Omron Thermometer Models MC-245, MC-246, MC-247,
MC-341, MD-343, MC-344 Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: II Product Code: FLL Dated: July 25, 2009 Received: July 29, 2009
Dear Mr. Dryden:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
# Page 2- Mr. Dryden
Enclosure
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH /CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance,
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Anthony L., Amato Jr.
Susan Runner, D.D.S., M.A. Acting Division Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{6}------------------------------------------------
# Indications for Use Statement
Page 1 of 1
### 510(k) Number:
1. 0 91676 (To be assigned)
Device Name:
Omron Thermometers Models MC-245, MC-246, MC-247, MC-341, MC-343, MC-344
# Indications for Use:
The Omron electronic thermometer Models MC-245, MC-341, MC-341, MC-343, and MC-344 are intended to measure the body temperature either oral, axillaries (under arm) and rectal and Model - MC-247 measure the body temperature oral (basal) and to be used by medical professionals in clinical and hospital environments and consumers in a home environment. It is intended for use on people of all ages.
| Prescription Use | or | Over-the-counter use XX |
|-----------------------------|----|-------------------------|
| (Part 21 CFR 801 Subpart D) | | (21 CFR 807 Subpart C) |
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Y. Schacter
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K091676
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.