RII INFRARED EAR THERMOMETER, MODEL TH01BN

K070976 · Radiant Innovation, Inc. · FLL · Apr 20, 2007 · General Hospital

Device Facts

Record IDK070976
Device NameRII INFRARED EAR THERMOMETER, MODEL TH01BN
ApplicantRadiant Innovation, Inc.
Product CodeFLL · General Hospital
Decision DateApr 20, 2007
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 880.2910
Device ClassClass 2
AttributesPediatric

Indications for Use

The device is an electronic clinical thermometer using an infrared sensor to detect body temperature from the auditory canal in the neonatal, pediatric and adult population used in the home setting.

Device Story

Infrared ear thermometer; detects natural thermal radiation from tympanic membrane and auditory canal via thermopile detector. User inserts probe into ear canal; presses start button. Device amplifies and calculates electrical signal from detector; displays temperature on LCD. Used in home setting by patients or caregivers. Provides rapid temperature readings to assist in clinical decision-making regarding fever management.

Clinical Evidence

Bench testing only. Compliance with ASTM E1965-98, EN12470-5:2003, IEC 60601-1, and EN 60601-1-2 standards.

Technological Characteristics

Infrared thermopile detector; electronic signal amplification and calculation circuitry; LCD display. Standards: ASTM E1965-98, EN12470-5:2003, IEC 60601-1, EN 60601-1-2.

Indications for Use

Indicated for neonatal, pediatric, and adult populations for body temperature measurement via the auditory canal in a home setting.

Regulatory Classification

Identification

A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.

Special Controls

(1) Device is not a clinical thermometer with telethermographic functions; (2) Device is not a clinical thermometer with continuous temperature measurement functions; and (3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the word "RADIANT" in bold, black letters. Above the word "RADIANT" is the word "INNOVATION" in a smaller font. Below the word "RADIANT" are four Chinese characters. 熱映光電股份有限公司 # EXHIBIT # 1 ## SPECIAL 510(K) SUMMARY This summary of Special 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92. The assigned 510(k) number is: ## 1. Submitter's Identification: APR 2 0 2007 Radiant Innovation Inc., 1F, No.3 Industrial E. 9th Rd., Science-Based Industrial Park, HsinChu, Taiwan Contact: Mr. Frank Lin QA Dept. Manager Date Summary Prepared: Mar/23/2007 ## 2. Name of the Modification Device: Infrared Ear Thermometer TH01BN ## 3. Current Clearance Device: Radiant Innovation Infrared Ear Thermometer, Models TH5XY series (FDA#: K063185), TH00XY series (FDA#: K051269), THXXN series (FDA#: K040377), TH1 series (FDA#: K030324), TH88 series (FDA#: K020504) and TH8 series (FDA#: K011059). ## 4. Device Description: The Radiant Innovation Inc., Infrared Ear Thermometer, Model TH01BN are electronic thermometers using an infrared detector (thermopile detector) to detect body temperature from the auditory canal. Its operation is based on measuring the natural thermal radiation emanating from the tympanic membrane and the adjacent surfaces of the patient. To measure ear temperature, the ear thermometer is inserted into a patient's outer ear canal. A start button is pressed to start the measurement through the radiation exchanges. The electrical signal read out from the detector is fed to the circuit for amplification and calculation. The measured temperature then appears on a LCD display. The total operation takes a few seconds. 1F, No.3, Industrial E. 9th Rd., Science-Based TEL: +886 3 5644185 Industrial Park, HsinChu, Taiwan 300, R.O.C. FAX: +886 3 5644170 {1}------------------------------------------------ Image /page/1/Picture/0 description: The image shows the word "RADIANT" in bold, black letters with the letter "A" replaced by an upside-down "V". Above the word "RADIANT" is the word "INNOVATION" in smaller, black letters. Below the word "RADIANT" are four Chinese characters. 熱映光電股份有限公司 47 4976 12.20P2 ## Intended Use: ട. The device is an electronic clinical thermometer using an infrared sensor to detect body temperature from the auditory canal in the neonatal, pediatric and adult population used in the home setting. ## Tests Performed for Determination of Substantial Equivalence are as follows: 6. Compliance to applicable voluntary standards includes ASTM E1965-98 and EN12470-5:2003, as well as IEC 60601 -1 and EN 60601-1-2 requirements. Guidance Documents included the FDA "Guidance On The Content of Premarket Notification (510(k)) Submissions for Clinical Electronic Thermometers", " How to Prepare A Special 510(k) ", "Deciding When to Submit a 510(k) for a Change to an Existing Device". ## 7. Conclusions: The RII Infrared Ear Thermometer TH01BN, have the same intended use and similar characteristics as the cleared device TH5XY series. Moreover, bench testing contained in this submission supplied demonstrate that the modification of TH01BN do not raise any new questions of safety or effectiveness. Thus, the RII Infrared Ear Thermometer, Model TH01BN is substantially equivalent to the predicate device. Page 14 1F, No.3, Industrial E. 9th Rd., Science-Based TEL:+886 3 5644185 Industrial Park, HsinChu, Taiwan 300, R.O.C. FAX: +886 3 5644170 {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the seal of the Department of Health and Human Services (HHS). The seal features an eagle-like emblem with three stylized lines representing the bird's body and wings. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES - USA" is arranged in a circular pattern around the emblem. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 APR 2 0 2007 Mr. Frank Lin QA Department Manager Radiant Innovation Incorporated 1 F, No. 3 Industrial E. 9th Road, Science-Based Industrial Park, HsinChu TAIWAN Re: K070976 Trade/Device Name: Radiant Innovation Inc. Infrared Ear Thermometer TH01BN Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: II Product Code: FLL Dated: April 6, 2007 Received: April 6, 2007 Dear Mr. Lin: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ Page 2 - Mr. Lin Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Chiu Lin, Ph.D. Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the word "RADIANT" in bold, black letters with the letter "A" replaced by an upside-down "V". Above the word "RADIANT" is the word "INNOVATION" in smaller letters. Below the word "RADIANT" are four Chinese characters. Image /page/4/Picture/1 description: The image shows the text "RADIANT INNOVATION Inc." in English, with the word "RADIANT" above "INNOVATION" and "Inc.". Above the English text is the same text in Chinese characters. The Chinese characters are arranged horizontally. ## INDICATIONS FOR USE ## EXHIBIT # B 510(k) Number (if known): Device Name: Radiant Innovation Inc. Infrared Ear Thermometer TH01BN Indications For Use: : The device is an electronic clinical thermometer using an infrared sensor to detect body temperature from the auditory canal in the neonatal, pediatric and adult population used in the home setting. Prescription Use (Part 21 CFR 801 Subpart D) AND/OR (21 CFR 807 Subpart C) . Over-The-Counter Use (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Anthony D. Orenstein of the Massachusetts General Hospital, Infusion Control, Cancer Devices | 510(k) Number: | K074976 | |----------------|---------| |----------------|---------| Page 1 of 1 Page 12 300 新竹科學園區工業東九路 3 號 1 樓 1F, No.3, Industrial E. 9th Rd., Science-Based TEL:+886 3 5644185 http://www.radiantek.com.tw Industrial Park, HsinChu, Taiwan 300, R.O.C. FAX: +886 3 5644170
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...