RII INFRARED EAR THERMOMETER, MODELS TH00XY SERIES
K051269 · Radiant Innovation, Inc. · FLL · Sep 9, 2005 · General Hospital
Device Facts
Record ID
K051269
Device Name
RII INFRARED EAR THERMOMETER, MODELS TH00XY SERIES
Applicant
Radiant Innovation, Inc.
Product Code
FLL · General Hospital
Decision Date
Sep 9, 2005
Decision
SESE
Submission Type
Special
Regulation
21 CFR 880.2910
Device Class
Class 2
Indications for Use
The device is an electronic clinical thermometer using an infrared sensor to detect body temperature in the ear of the adult population used in the home setting.
Device Story
Infrared ear thermometer; uses thermopile detector to measure infrared radiation from tympanic membrane and adjacent ear canal surfaces. User inserts probe into ear canal; presses start button to initiate measurement. Internal circuitry performs signal amplification and temperature calculation. Measured temperature displayed on LCD. Used in home setting by patients or caregivers. Provides non-invasive temperature monitoring to assist in clinical decision-making regarding health status.
Clinical Evidence
Bench testing only. Compliance with ASTM E1965-98, IEC 60601-1, and EN 60601-1-2 standards demonstrated.
Technological Characteristics
Infrared thermopile detector; electronic clinical thermometer; LCD display; battery-powered; handheld form factor. Standards: ASTM E1965-98, IEC 60601-1, EN 60601-1-2.
Indications for Use
Indicated for adult population for body temperature measurement in the ear in a home setting.
Regulatory Classification
Identification
A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.
Special Controls
(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.
Predicate Devices
Radiant Innovation Infrared Ear Thermometer THXXN series (K04037)
Radiant Innovation Infrared Ear Thermometer TH88 series (K030324)
Radiant Innovation Infrared Ear Thermometer TH88 series (K020504)
Radiant Innovation Infrared Ear Thermometer TH8 series (K011059)
Submission Summary (Full Text)
{0}------------------------------------------------
## SEP - 9 2005
EXHIBIT # 1
## SPECIAL 510(K) SUMMARY
This summary of Special 510(k) safety and effectiveness information is being submitted in accordance of This summary of Special SMDA 1990 and 21 CFR §807.92.
The assigned 510(k) number is:
#### Submitter's Identification: 1.
Radiant Innovation Inc., 1F, No.3 Industrial E. 9th Rd., Science-Based Industrial Park, HsinChu, Taiwan
## Contact:
Mr. Frank Lin QA Dept. Manager
Date Summary Prepared: May/12/2005
### Name of the Modification Device: 2.
Infrared Ear Thermometer TH00XY series (TH005A/B, TH007A/B and TH008A/B)
#### Current Clearance Device: 3.
Radiant Innovation Infrared Ear Thermometer, Models THXXN series (FDA#: K04037), TH11 Radiant Innovation Intrared Ear Themienst, 1000504) and TH8 series (FDA#: K011059).
series (FDA#: K030324), TH88 series (FDA#: K020504) and TH8 series (FDA#: K011059).
#### Device Description: 4.
The Radiant Innovation Inc., Infrared Ear Thermometer, Models TH00XY series are I he Kadian Innovation mos, minantrared detector (thermopile detector) to detect electrome themionelers using an intrared at the operation is based on measuring the body temperature from the auditor from the tympanic membrane and the adjacent surfaces of the patient.
To measure ear temperature, the ear thermometer is inserted into a patient's outcer can canal. A The to measure caremperature, the measurement through the radiation exchanges.
Start button is pressed to start the measurement through the rachanges. The start button 15 pressed to start the the circuit for amplification and calculation, The measured temperature then appears on a LCD display. The total operation takes a few seconds.
{1}------------------------------------------------
#### Intended Use: న.
The device is an electronic clinical thermometer using an infrated sensor to detect body The device is an electronic cinnear themneter ashing and adult population used in the home setting.
KOS 126 9
### Tests Performed for Determination of Substantial Equivalence are as follows: 6.
Compliance to applicable voluntary standards includes ASTM E1965-98, as well as IEC 60601 -1 and EN 60601-1-2 requirements.
Guidance Documents included the FDA "Guidance On The Content of Premarket Notification Guidance Documents Included the TDA ronic The mometers", " "How to Prepare A Special
(510(k)) Submissions for Clinical Elect and Cleck on Casting Davica" (510(k)) Submissions Jor Cimical Electronic r a Change to an Existing Device" ..
#### 9. Conclusions:
The RII Infrared Ear Thermometer TH00XY series, have the same untended use and similar The RII Infrared Ear Thermoneter THXXN series. More are bench testing contained in this characteristics as the cleared device that modification of TH00XY do not raise any new
submission supplicd demonstrate that the Stated For Thormamater Models TH00XY submission supplied demonstrate that the The RI Infrared Ear Thermometer, Models TH00X Y series is substantially equivalent to the predicate device.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the United States Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES, USA" around the perimeter. Inside the circle is a stylized symbol that resembles a caduceus, with three figures representing people. The logo is black and white.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
SEP - 9 2005
Mr. Frank Lin QA Dept. Manager Radiant Innovation, Incorporated 1F, No. 3, Industrial E. 9th Road Science-Based Industrial Park HsinChu, TAIWAN
Re: K051269
Trade/Device Name: Infrared Ear Thermometer TH00XY Series Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: II Product Code: FLL Dated: August 5, 2005 Received: August 10, 2005
Dear Mr. Lin:
We have reviewed your Section 510(k) premarket notification of intent to market the device We have reviewed your betermined the device is substantially equivalent (for the itelered above and not of attennelosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Interstate commerce prior to they been reclassified in accordance with the provisions of Americans, or to do rices that in tic Act (Act) that do not require approval of a premarket the rederal I vou, Drug, and Commay, therefore, market the device, subject to the general approvin application (11-2) The general controls provisions of the Act include controls providions of the restration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting (FMA), it may be subject to back adalaf Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Mr. Lin
Please be advised that FDA's issuance of a substantial equivalence determination does not Please be advised that I DA 3 issualled on that your device complies with other requirements mean that IDA nas made a deceminations administered by other Federal agencies. of the Act of ally I ederal statues and regarments, including, but not limited to: registration You must comply with an the Fee b roquits (21 CFR Part 801); good manufacturing practice and listing (21 CFR Pat 807), labeling (QS) regulation (21 CFR Pat 820); and if requirements as set form in the quality ajon control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) IThis letter will anow you to began mading of substantial equivalence of your device to a premiarket nonfication. - The FDA milling or classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), If you desire specific advice to: your as at (240) 276-0115. Also, please note the regulation i entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the may obtain other general miseturers, International and Consumer Assistance at its toll-free Division (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Chiu Lin, Ph.D.
Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known):
K 05-1269
Device Name: Radiant Innovation Inc. Infrared Ear Thermometer TH00XY series
Indications For Use:
The device is an electronic clinical thermometer using an infrared sensor to detect body
the see l he device is an electronic cimical themionecer doing an innulation in the least in the home setting.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR (21 CFR 807 Subpart C) Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________
(Please Do NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Shule H. Murdich Jr. Anthony Watson GHDB 1/8/05
(Division Sign-Off)
sthesiology, General Hospital, ontrol, Dental Devices
510(k) Number
300 新竹科學園區工業東九路 3 號 1 樓 http://www.radiantek.com.tw
1F, No.3, Industrial E. 9" Rd., Science-Based TEL:+886 3 564418 n , No.J, Industrial Park, HsinChu, Taiwan 300, R.O.C. FAX: +886 3 564417
Page 1 of
Page 17
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.