K182404 · Advanced Sterilization Products (Asp) · FRC · Dec 27, 2018 · General Hospital
Device Facts
Record ID
K182404
Device Name
STERRAD VELOCITY Biological Indicator
Applicant
Advanced Sterilization Products (Asp)
Product Code
FRC · General Hospital
Decision Date
Dec 27, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.2800
Device Class
Class 2
Indications for Use
The STERRAD VELOCITY Biological Indicator, in conjunction with the STERRAD VELOCITY Reader, is intended to be used as a standard method for frequent monitoring and periodic testing of the following STERRAD Sterilization Systems: - STERRAD® 100NX (STANDARD, FLEX, EXPRESS, and DUO Cycles) with and without ALLClear™ Technology - STERRAD NX® (STANDARD and ADVANCED Cycles) with and without ALLClear™ Technology - STERRAD® 100S
Device Story
Self-contained biological indicator (BI) used with STERRAD VELOCITY Reader; monitors hydrogen peroxide gas plasma sterilization. Input: Geobacillus stearothermophilus spores on glass fiber disc. Operation: Spore germination releases α-glucosidase enzyme; enzyme hydrolyzes non-fluorescent α-MUG substrate into fluorescent 4-MU. Reader excites 4-MU with UV light; fluorescence intensity indicates spore inactivation status. Used in clinical settings by healthcare staff. Output: 30-minute fluorescence-based result (positive/negative) or optional 5-7 day visual pH-based color change. Provides rapid verification of sterilization cycle efficacy; ensures medical devices are safe for patient use.
Clinical Evidence
No clinical data. Bench testing only. Performance qualification demonstrated that BI resistance is greater than biological models for all claimed STERRAD cycles; all fluorescence-negative results obtained in triplicate runs.
Technological Characteristics
Self-contained BI; glass fiber carrier; Geobacillus stearothermophilus (ATCC 7953) spores (≥1x10^6 CFU/BI). α-glucosidase enzyme system for rapid fluorescence readout. Type 1 chemical process indicator (red/pink to yellow). Incubation 57±2°C. Complies with ISO 11138-1.
Indications for Use
Indicated for frequent monitoring and periodic testing of STERRAD sterilization systems (100NX, NX, 100S) to confirm sterilization efficacy.
Regulatory Classification
Identification
Biological sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor adequacy of sterilization. The device consists of a known number of microorganisms, of known resistance to the mode of sterilization, in or on a carrier and enclosed in a protective package. Subsequent growth or failure of the microorganisms to grow under suitable conditions indicates the adequacy of sterilization. Physical/chemical sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
December 27, 2018
Advanced Sterilization Products (ASP) Ms. Sun Choi Principal Regulatory Affairs Specialist 33 Technology Drive Irvine, California 92618
Re: K182404
Trade/Device Name: STERRAD VELOCITY Biological Indicator Regulation Number: 21 CFR 880.2800 Regulation Name: Sterilization Process Indicator Regulatory Class: Class II Product Code: FRC Dated: November 29, 2018 Received: December 31, 2018
Dear Sun Choi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You mav, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be avare that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Elizabeth F. Claverie -S
For Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K182404
Device Name
STERRADVELOCITY™ Biological Indicator
Indications for Use (Describe)
The STERRAD VELOCITY™ Biological Indicator, in conjunction with the STERRAD VELOCITY Reader, is intended to be used as a standard method for frequent monitoring and periodic testing of the following STERRAD Sterilization Systems:
- · STERRAD® 100NX (STANDARD, FLEX, EXPRESS, and DUO Cycles) with and without ALLClear™ Technology
- STERRAD NX® (STANDARD and ADVANCED Cycles) with and without ALLClear™ Technology
· STERRAD® 100S
Type of Use (Select one or both, as applicable)
| <span style="font-family: sans-serif;"> <input type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</span> | <span style="font-family: sans-serif;"> <input checked="" type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
|----------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------|
|----------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDED.This section applies onlyto requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data neededand complete and review the collection of information. Send comments regarding this burden estimate or anyother aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act(PRA) Staff PRAStaff(@fda.hhs.gov
"An aqency may notconductor sponsor, and a person is not required to respond to, a collection of informationunless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows a logo with a triangle and the letters 'SP' inside a rounded rectangle. The logo is light blue, and there is a registered trademark symbol in the upper right corner. The logo appears to be for a company or organization with the initials 'ASP'.
## 510(k) Summary
## Advanced Sterilization Products STERRAD VELOCITY™ Biological Indicator
#### General Information
| Submitter Name: | Advanced Sterilization Products<br>Division of Ethicon, Inc., a Johnson & Johnson company |
|-----------------|---------------------------------------------------------------------------------------------------------|
| Address: | 33 Technology Drive<br>Irvine, CA 92618 |
| Contact Person: | Sun Choi<br>Principal Regulatory Affairs Specialist<br>Phone: 949-453-6378<br>Email: sjchoi@its.jnj.com |
| Date Prepared: | December 20, 2018 |
| 510(k) Number: | K182404 |
#### Device Name
| Proprietary Name: | STERRAD VELOCITY™ Biological Indicator |
|----------------------|--------------------------------------------|
| Common Name: | Biological Indicator |
| Classification Name: | Biological Sterilization Process Indicator |
| Device Class: | Class II |
| Product Code: | FRC |
| CFR Section: | 21 CFR 880.2800 |
#### Predicate Device
- Primary Predicate Device: STERRAD VELOCITY™ Biological Indicator, K170039 cleared ● June 20, 2017
Note: The subject device is identical to the predicate STERRAD VELOCITY Biological Indicator (K170039). Only the indications for use have been modified to incorporate the periodic testing indication.
- . Reference Device: STERRAD® 100NX DUO Cycle Test Pack, K111391 cleared September 26, 2012
#### Device Description
The STERRAD VELOCITY Biological Indicator is a self-contained biological indicator, used in conjunction with the STERRAD VELOCITY Reader, that is intended for frequent monitoring and periodic testing of the STERRD Sterilization Cycles, using rapid readout technology that provides a final fluorescence result in 30 minutes at the incubation temperature of 57 ± 2ºC.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows a logo with the letters ASP in white against a light blue background. The letters are stylized, with the A being a triangle and the S and P being connected. There is a registered trademark symbol in the upper right corner of the logo. The logo has rounded corners.
The STERRAD VELOCITY BI can also be determined as growth-positive or growth-negative via an optional visual pH-based color change result (using bromocresol purple) if used for frequent monitoring purposes. When using this method, the biological indicator must be cultured in an incubator at 55-60℃ for 5 to 7 days to get a final visual result.
The STERRAD VELOCITY BI consists of a glass fiber disc containing a minimum of 1 x 100 Geobacillus stearothermophilus (ATCC 7953) spores and a glass ampoule containing nutrient growth medium and non-fluorescent substrate, as well as a vial, cap, cap label, insert, and chemical indicator. The spore disc, growth media ampoule, and insert are contained in a clear plastic vial with a vented cap. The cap is designed with sterilant ingress openings which allow for penetration of hydrogen peroxide vapor into the vial during the sterilization process. The chemical indicator (CI), placed on the top of the cap, is a Type 1 process indicator that changes color from red/pink to yellow or yellow with some red/orange/brown dots when exposed to hydrogen peroxide.
The STERRAD VELOCITY BI has the same α-glucosidase enzyme system for the fundamental scientific technology as the predicate device cleared under K170039. The a-glucosidase enzyme, which is generated naturally during growth of G. stearothermophilus and released during spore germination, hydrolyzes the bond between the glucose and 4-methylumbelliferyl (4-MU) moieties of 4-methylumbelliferyl a-D-glucopyranoside (α-MUG). In the combined state, α-MUG is not fluorescent. Once the bond between the glucose and 4-MU is hydrolyzed, the 4-MU component becomes fluorescent when excited with UV light. Therefore, the a-glucosidase enzyme in its active state can be detected by measuring the fluorescence produced by the enzymatic hydrolysis of a-MUG.
The resultant fluorescent by-product (4-MU), is detected by the Reader and the fluorescent signal is used to determine the positive or negative result of the biological indicator. The measured enzyme activity is reduced upon exposure to hydrogen peroxide. As the enzyme activity is directly correlated with the spore outgrowth. the reduction of the enzyme activity below a certain level indicates that all spores have been inactivated. The level of the fluorescence response is determined using the algorithm developed for the STERRAD VELOCITY BI and is used to distinguish between the positive and negative responses.
#### Intended Use/Indications for Use
The STERRAD VELOCITY Biological Indicator, in conjunction with the STERRAD VELOCITY Reader, is intended to be used as a standard method for frequent monitoring and periodic testing of the following STERRAD Sterilization Systems:
- . STERRAD® 100NX (STANDARD, FLEX, EXPRESS, and DUO Cycles) with and without ALLClear™ Technology
- STERRAD NX® (STANDARD and ADVANCED Cycles) with and without ALLClear ● Technology
- STERRAD 100S .
The Indications for Use statement for the STERRAD VELOCITY Biological Indicator has been expanded to incorporate the periodic testing indication. The subject and its predicate device share the same intended use for monitoring of hydrogen peroxide gas plasma sterilization processes.
#### Comparison of Technological Characteristics with Predicate Device/Reference Device
The subject device and predicate device have the same technological characteristics based on their use in the monitoring of hydrogen peroxide gas plasma sterilization processes. Refer to the following
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows a logo with the letters "ASP" in a stylized, interconnected design. The letters are white and are set against a rounded, light blue-green background. A registered trademark symbol is visible in the upper right corner of the logo.
table for a comparison of the subject device and predicate device/reference device characteristics, showing the similarities and differences.
| | Reference Device | Predicate Device | Subject device |
|--------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device<br>Characteristics | STERRAD 100NX DUO<br>Cycle Test Pack<br>(K111391) | STERRAD VELOCITY BI<br>(K170039) | STERRAD VELOCITY BI<br>with Expanded Indications<br>(K182404) |
| Intended Use | Routine monitoring of the<br>sterilizer cycle (hydrogen<br>peroxide gas plasma) | Monitoring of hydrogen<br>peroxide gas plasma<br>sterilization processes | Same |
| Indications for Use | The STERRAD 100NX DUO<br>Cycle Test Pack is used for<br>routine monitoring of the<br>STERRAD 100NX DUO<br>Sterilization Cycle and is also<br>used for the periodic testing<br>of a STERRAD 100NX<br>System DUO Cycle, using<br>hospital-defined loads<br>containing devices that do not<br>exceed claims of the cycle.<br>The STERRAD 100NX DUO<br>Cycle Test Pack consists of a<br>STERRAD® CYCLESURE®<br>24 Biological Indicator, vial<br>and cap to hold the BI. | Used as a standard method<br>for frequent monitoring of the<br>following STERRAD<br>Sterilization Systems:<br>• STERRAD 100NX<br>(STANDARD, FLEX,<br>EXPRESS, and DUO<br>Cycles) with and without<br>ALLClear Technology<br>• STERRAD NX<br>(STANDARD and<br>ADVANCED Cycles) with<br>and without ALLClear<br>Technology<br>• STERRAD 100S | Used as a standard method<br>for frequent monitoring and<br>periodic testing of the<br>following STERRAD<br>Sterilization Systems:<br>• STERRAD 100NX<br>(STANDARD, FLEX,<br>EXPRESS, and DUO<br>Cycles) with and without<br>ALLClear Technology<br>• STERRAD NX<br>(STANDARD and<br>ADVANCED Cycles) with<br>and without ALLClear<br>Technology<br>• STERRAD 100S |
| Organism<br>(Spore, Species,<br>Strain) | <i>Geobacillus<br/>stearothermophilus</i><br>ATCC 7953 | <i>Geobacillus<br/>stearothermophilus</i><br>ATCC 7953 | Same |
| Viable Spore<br>Population | $≥ 1x10^6$ CFU/BI | $≥ 1x10^6$ CFU/BI | Same |
| Carrier Material | Glass Fiber | Glass Fiber | Same |
| Device Design | Self-contained biological<br>indicator in test pack<br>configuration | Self-contained biological<br>indicator | Same as predicate device |
| Resistance<br>Characteristics | Equal to or greater than the<br>most difficult item routinely<br>processed (biological model). | Tested at 5 mg/L hydrogen<br>peroxide:<br>D-value 5mg/L: $≥$ 1 second,<br>using two D-value methods<br>(Survivor Curve and Fraction<br>Negative) per ISO 11138-1. | Tested at 5 mg/L hydrogen<br>peroxide:<br>D-value 5mg/L: $≥$ 1 second,<br>using two D-value methods<br>(Survivor Curve and<br>Fraction Negative) per<br>ISO 11138-1.<br>Equal to or greater than the<br>most difficult item<br>routinely processed<br>(biological model). |
| | Reference Device | Predicate Device | Subject device |
| Device<br>Characteristics | STERRAD 100NX DUO<br>Cycle Test Pack<br>(K111391) | STERRAD VELOCITY BI<br>(K170039) | STERRAD VELOCITY BI<br>with Expanded Indications<br>(K182404) |
| Rapid Readout<br>Technology | N/A | The α-glucosidase enzyme<br>system is generated naturally<br>during growth of Geobacillus<br>stearothermophilus. The α-<br>glucosidase enzyme in its<br>active state is detected by<br>measuring the fluorescence<br>produced by the enzymatic<br>hydrolysis of a non-<br>fluorescent substrate. The<br>resultant fluorescent by-<br>product is detected in the<br>reader. The fluorescence<br>signal is used to determine a<br>positive or negative result. | Same as predicate device |
| Incubation<br>Temperature | 55 - 60°C | 57 +/- 2°C | Same as predicate device |
| Reduced<br>Incubation Time | 24-hour visual pH color<br>change results | 30-minute fluorescence<br>results read in STERRAD<br>VELOCITY Reader and<br>optional visual pH color<br>change results at 5 to 7 days. | Same as predicate device |
| Holding Time | Immediately after<br>sterilization cycle | Within 2 hours after<br>sterilization cycle | Same as predicate device |
| Carrier growth<br>inhibition/media<br>growth promotion | No bacteriostatic effects that<br>inhibit the growth of the<br>indicator microorganism | No bacteriostatic effects that<br>inhibit the growth of the<br>indicator microorganism | Same |
| Chemical Indicator | Throughput indicator,<br>changes color from red to<br>golden yellow or bronze to<br>indicate exposure to hydrogen<br>peroxide | Type 1 process indicator<br>changes color from red/pink<br>to yellow or yellow with<br>some red/orange/brown dots<br>to indicate exposure to<br>hydrogen peroxide. | Same as predicate device |
| Shelf Life | 12 months | 9 months | Same as predicate device |
### Substantial Equivalence Comparison Table
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows a logo with the letters "ASP" in a stylized font. The letters are white and are set against a light blue background. The logo is enclosed in a rounded rectangle shape. There is a registered trademark symbol in the upper right corner of the logo.
The subject device and its predicate device (STERRAD VELOCITY BI) have the same intended use and indications for use with one modification to the subject device, which is to incorporate the periodic testing indication. Further, the subject device and predicate device share the same technological characteristics.
#### Non-Clinical Data
Performance testing was conducted to satisfy the requirements for the STERRAD VELOCITY BI with periodic testing indication that is the subject of this 510(k), as outlined in the Guidance for Industry and FDA Staff: Biological Indicator (BI) Premarket Notification [510(k)] Submissions, Section 10, Test Pack, issued on October 4, 2007.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows a logo with the letters "ASP" in a stylized font. The "A" is represented by a triangle, and the "S" and "P" are connected. The logo is enclosed in a rounded rectangle with a light blue background. There is a registered trademark symbol in the upper right corner of the logo.
The subject device was evaluated for its challenge to all claimed STERRAD Cycles and the resistance of the subject device was compared to the biological models for all claimed STERRAD Cycles.
The study results indicated that the resistance of the subject device was greater than the biological model for all claimed STERRAD Cycles, and all fluorescence-negative results were obtained in triplicate runs. The results demonstrated that the STERRAD VELOCITY BI is qualified for its design and performance for periodic testing indication in all claimed STERRAD Cycles.
All testing yielded passing results. This testing is summarized in the following table.
| Summary of Performance Testing | | |
|---------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|
| Performance Testing | Description | Pass / Fail |
| Performance Study in<br>STERRAD 100NX | Design Evaluation and Performance Qualification for Periodic<br>Testing in STERRAD 100NX STANDARD and FLEX Cycles<br>with and without ALLClear Technology | Pass |
| Performance Study in<br>STERRAD 100NX | Design Evaluation and Performance Qualification for Periodic<br>Testing in STERRAD 100NX EXPRESS Cycle with and without<br>ALLClear Technology | Pass |
| Performance Study in<br>STERRAD 100NX | Design Evaluation and Performance Qualification for Periodic<br>Testing in STERRAD 100NX DUO Cycle | Pass |
| Performance Study in<br>STERRAD NX | Design Evaluation and Performance Qualification for Periodic<br>Testing in STERRAD NX STANDARD and ADVANCED<br>Cycles with and without ALLClear Technology | Pass |
| Performance Study in<br>STERRAD 100S | Design Evaluation and Performance Qualification for Periodic<br>Testing in STERRAD 100S Cycle | Pass |
#### Summary of Performance Testing
#### Clinical Data
No clinical data was generated in support of this Premarket Notification.
#### Summary
The subject device and its reference device (STERRAD 100NX DUO Cycle Test Pack) have the same periodic testing indication; however, there are differences in their device designs. The subject device is a self-contained biological indicator, whereas the reference device is a self-contained biological indicator in a test pack configuration. Non-clinical data demonstrates that the use of a test pack configuration is not necessary for the subject device.
#### Conclusion
The subject device (STERRAD VELOCITY Biological Indicator) is substantially equivalent to the claimed predicate and reference devices.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.