STERRAD (R) 100NX DUO CYCLE TEST PACK

K111391 · Advanced Sterilization Products · FRC · Sep 26, 2012 · General Hospital

Device Facts

Record IDK111391
Device NameSTERRAD (R) 100NX DUO CYCLE TEST PACK
ApplicantAdvanced Sterilization Products
Product CodeFRC · General Hospital
Decision DateSep 26, 2012
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.2800
Device ClassClass 2

Indications for Use

The STERRAD® 100NX® DUO Cycle Test Pack is used for routine monitoring of the STERRAD® 100NX® DUO Sterilization Cycle and is also used for the periodic testing of a STERRAD® 100NX® System DUO Cycle, using hospital-defined loads containing devices that do not exceed claims of the cycle. The STERRAD® 100NX® DUO Cycle Test Pack consists of a STERRAD® CYCLESURE® 24 Biological Indicator, vial and cap to hold the BI.

Device Story

The STERRAD® 100NX® DUO Cycle Test Pack serves as a biological challenge for hydrogen peroxide sterilization cycles. It consists of a CYCLESURE® 24 Biological Indicator (BI) housed within a specialized test vial and cap. The device is used by hospital staff to verify the efficacy of the STERRAD® 100NX® DUO sterilization process. By exposing the BI to the sterilization cycle, the user assesses whether the process achieves the required lethality. The test pack is designed to be at least as resistant to the sterilization process as the biological model developed for the DUO cycle. Results are used by healthcare providers to confirm that sterilization parameters are met, ensuring the safety of sterilized medical devices for patient use.

Clinical Evidence

Bench testing only. Performance was validated through design evaluation and performance qualification of the test pack in the STERRAD® 100NX® sterilizer. Resistance was confirmed via survivor curves and fraction negative data compared to the biological model for the DUO cycle. Chemical indicator functionality was also verified.

Technological Characteristics

Biological indicator (CYCLESURE® 24) housed in a test vial and cap. Operates via exposure to hydrogen peroxide vapor in a STERRAD® 100NX® sterilizer. Designed to challenge the sterilization cycle efficacy.

Indications for Use

Indicated for routine monitoring and periodic testing of the STERRAD® 100NX® DUO Sterilization Cycle using hospital-defined loads.

Regulatory Classification

Identification

Biological sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor adequacy of sterilization. The device consists of a known number of microorganisms, of known resistance to the mode of sterilization, in or on a carrier and enclosed in a protective package. Subsequent growth or failure of the microorganisms to grow under suitable conditions indicates the adequacy of sterilization. Physical/chemical sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K111391 1 of 3 ### 510(k) Summary Applicant's Name, Address, Telephone, FAX, Contact Person Advanced Sterilization Products, Division of Ethicon, Inc. 33 Technology Drive Irvine, CA 92618 #### Contact Person Nancy Chu Manager, Regulatory Affairs Tel: (949) 453-6435 Fax: (949) 789-3900 Summary Date: September 10, 2012 # 1. CLASSIFICATION, COMMON OR USUAL NAME, DEVICE NAME Classification Name: Common/Usual Name: Product Classification: Classification Regulation: Proprietary Name: Biological Sterilization Process Indicator Biological Indicator (Test Pack) II 21 CFR 880.2800 STERRAD® 100NX® DUO Cycle Test Pack #### 2. PREDICATE DEVICES STERRAD® 100NX® Test Pack, K071537 cleared on December 18, 2007 ## 3. INDICATIONS FOR USE The STERRAD® 100NX® DUO Cycle Test Pack is used for routine monitoring of the STERRAD® 100NX® DUO Sterilization Cycle and is also used for the periodic testing of a STERRAD® 100NX® System DUO Cycle, using hospital-defined loads containing devices that do not exceed claims of the cycle. The STERRAD® 100NX® DUO Cycle Test Pack consists of a STERRAD® CYCLESURE® 24 Biological Indicator, vial and cap to hold the BI. SEP 26 2012 {1}------------------------------------------------ ## 4. DESCRIPTION OF DEVICE The STERRAD® 100NX® DUO Test Pack consists of a CYCLESURE® 24 Biological Indicator (BI) and a STERRAD® NX® Test Vial with cap for holding the BI during the sterilization cycle. ### 5. SUMMARY OF NONCLINICAL TESTS The STERRAD® 100NX® DUO Test Pack has been evaluated for its resistance to the DUO Cycle in the STERRAD® 100NX® Sterilizer. A comparison of the DUO Test Pack to the biological model developed for the DUO Cycle indicates that the Test Pack is at least as resistant to the sterilization process as the biological model. This is based on both survival curves and fraction negative data as a function of dose. DUO Test Packs assembled from three lots of CYCLESURE® 24 Biological Indicator (BI) were exposed to several doses of hydrogen peroxide in a DUO Cycle. These survivor curves were compared to the survivor curves for the biological model developed for the DUO Cycle. The test data showed that the Test Pack configuration was at least as resistant as the biological model. Additionally, fraction negative data were collected using Test Packs assembled from three lots of CYCLESURE® 24 BI and exposed to increasing volumes of hydrogen peroxide in a DUO Cycle. The results indicated that the Test Pack configuration was at least as resistant as the biological model. Indicative functionality of the chemical indicator in a DUO Test Pack configuration was evaluated using half-cycle parameters of the DUO Cycle and the response was determined to be appropriate for a chemical indicator. The subject device and its predicate device have the same intended use which is for routine monitoring of the sterilizer cycle. Additionally, they have the same technological characteristics, the same operating principles and are subjected to the same sterilant (hydrogen peroxide) and therefore, the subject device is substantially equivalent to the predicate. The Table below lists the tests performed to demonstrate that the DUO Test Pack functions as intended in the STERRAD® 100NX® sterilizer using a DUO Cycle. {2}------------------------------------------------ | Studies Performed | Results | |----------------------------------------------------------------------------------------|---------| | Design Evaluation and Performance Qualification of<br>STERRAD® 100 NX® DUO Test Pack | Passed | | Functionality Study of the Chemical Indicator Disc of<br>CYCLESURE® 24 BI in Test Pack | Passed | # 6. DESCRIPTION OF CHANGE: - Indications for Use Statement has been revised to incorporate the STERRAD® . 100NX® DUO Test Pack. - Instructions for Use have been revised to include STERRAD® 100NX® DUO Test . Pack information. ## 7. OVERALL PERFORMANCE CONCLUSIONS The performance data shows that the STERRAD® 100NX® DUO Test Pack has the necessary resistance relative to the biological model to be an appropriate challenge for testing the DUO Cycle of the STERRAD® 100NX® Sterilizer and it is substantially equivalent to predicate device. {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three stripes forming its wing and tail. The eagle is positioned to the right of a circular arrangement of text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES USA". Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 Ms. Nancy Chu Manager, Regulatory Affairs Advanced Sterilization Products 33 Technology Drive Irvine, California 92618 SEP 2 6 2012 Re: K111391 Trade/Device Name: STERRAD® 100NX® DUO Cycle Test Pack Regulation Number: 21 CFR 880.2800 Regulation Name: Sterilization Process Indicator Regulatory Class: II Product Code: FRC Dated: September 10, 2012 Received: September 12, 2012 Dear Ms. Chu: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ #### Page 2- Ms. Chu Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801): medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours. For Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ #### Indications for Use 510(k) Number (if known): K111391 STERRAD® 100NX® DUO Cycle Test Pack Device Name: Indication for Use: The STERRAD® 100NX® DUO Cycle Test Pack is used for routine monitoring of the STERRAD® 100NX® DUO Sterilization Cycle and is also used for the periodic testing of a STERRAD® 100NX® System DUO Cycle, using hospital-defined loads containing devices that do not exceed claims of the cycle. The STERRAD® 100NX® DUO Cycle Test Pack consists of a STERRAD® CYCLESURE® 24 Biological Indicator, vial and cap to hold the BI. Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use __ V (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) fu uld. Page 1 of 1 (Division Sign-Off) (Division Sign-Only) Division of Anesthesiology, General Hospital Infection Control, Dental Devices i K 11 510(k) Number:_
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...