Andersen EOGas 4 SteriTest

K151585 · Andersen Sterilizers, Inc. · FRC · Nov 18, 2015 · General Hospital

Device Facts

Record IDK151585
Device NameAndersen EOGas 4 SteriTest
ApplicantAndersen Sterilizers, Inc.
Product CodeFRC · General Hospital
Decision DateNov 18, 2015
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.2800
Device ClassClass 2

Indications for Use

The Andersen EOGas 4 SteriTest consists of a self-contained biological indicator inoculated with viable Bacillus atrophaeus bacterial spores that is placed in a dedicated BI receptacle in the sterilizer. It monitors the efficacy of the 3 hour sterilization cycle at 50°C in the EOGas 4 Ethylene Oxide Gas Sterilizer.

Device Story

Andersen EOGas 4 SteriTest is a self-contained biological indicator (SCBI) used to monitor ethylene oxide (EO) sterilization efficacy. The device consists of a plastic vial containing Bacillus atrophaeus spores on a paper carrier and a glass ampoule with growth medium and pH indicator. The operator places the SCBI into a reusable BI receptacle on the sterilizer's purge probe. During the 3-hour, 50°C sterilization cycle, EO gas enters the vial through a filter port. Post-cycle, the operator crushes the ampoule to activate the medium. Incubation at 30-35°C for 48 hours follows. A color change from red-orange to yellow or turbidity indicates sterilization failure; no change indicates a pass. The device provides healthcare providers with a biological verification of sterilization conditions, ensuring the safety of processed instruments.

Clinical Evidence

Bench testing only. Validation performed per FDA 2007 Guidance for Biological Indicator Premarket Notification and AAMI/ANSI/ISO 11138-1. Testing included viable spore population assay, resistance characteristics, growth inhibition evaluation, holding time assessment, and reduced incubation time validation. Results confirmed the device maintains performance specifications over a 2-year shelf life and successfully monitors the 3-hour, 50°C sterilization cycle.

Technological Characteristics

Materials: paper, glass, polypropylene, and aluminum. Sensing principle: biological indicator using Bacillus atrophaeus spores (ATCC 9372) with pH-based colorimetric growth detection. Form factor: SCBI in a reusable BI receptacle. Energy source: none (passive). Sterilization monitoring: ethylene oxide gas. Connectivity: none.

Indications for Use

Indicated for monitoring the efficacy of the 3-hour, 50°C sterilization cycle in the EOGas 4 Ethylene Oxide Gas Sterilizer using a self-contained biological indicator containing Bacillus atrophaeus spores.

Regulatory Classification

Identification

Biological sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor adequacy of sterilization. The device consists of a known number of microorganisms, of known resistance to the mode of sterilization, in or on a carrier and enclosed in a protective package. Subsequent growth or failure of the microorganisms to grow under suitable conditions indicates the adequacy of sterilization. Physical/chemical sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which features a staff with a snake winding around it, and three human figures. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the symbol in a circular fashion. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 November 18, 2015 Andersen Sterilizers, Inc. Dr. William Andersen President 3154 Caroline Drive Haw River, North Carolina 27258 Re: K151585 Trade/Device Name: Andersen EOgas 4 Steritest Regulation Number: 21 CFR 880.2800 Regulation Name: Sterilization Process Indicator Regulatory Class: Class II Product Code: FRC Dated: October 20, 2015 Received: October 22, 2015 Dear Dr. William Andersen: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device {1}------------------------------------------------ related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, # Erin I. Keith -S Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration # Indications for Use 510(k) Number (if known) K151585 Device Name Andersen EOGas 4 SteriTest Indications for Use (Describe) The Andersen EOGas 4 SteriTest consists of a self-contained biological indicator inoculated with viable Bacterial spores that is placed in a dedicated BI receptacle in the sterilizer. It monitors the 3 hour sterilization cycle at 50°C in the EOGas 4 Ethylene Oxide Gas Sterilizer. Critical process parameters for the cycle are summarized below in Table 1. Table 1. Critical parameters for the 3 hour cycle in the EOGas 4 Ethylene Oxide Gas Sterilizer | EO Amount | Temperature | Relative Humidity | EO Exposure Time | Total Cycle Time | |-------------|-------------|-------------------|------------------|------------------| | 17.6 g ± 5% | 50°C ± 3°C | 35-90% | 3 hours | 3.5 hours | Type of Use (Select one or both, as applicable) Prescription Use (Part 21 CFR 801 Subpart D) × Over-The-Counter Use (21 CFR 801 Subpart C) ## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED. ## FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) {3}------------------------------------------------ This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ ## 510(k) Summarv | 5.1 | Applicant's Name and Address | | Andersen Sterilizers, Inc.<br>3154 Caroline Drive<br>Haw River, NC 27258 | | |-----|------------------------------|-------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------|--| | 5.2 | Contact Person | | William K. Andersen, BE, MD, FAAOS<br>President<br>Phone: 336-376-8622, Fax: 336-376-5428 | | | 5.3 | Date of Preparation | | November 17, 2015 | | | 5.4 | Device | Proprietary Name<br>Common Name<br>Classification<br>Product Code | Andersen EOGas 4® SteriTest<br>Biological Sterilization Process Indicator<br>Class II (21 CFR 880.2800)<br>FRC | | | 5.5 | Predicate Device | Device Name<br>510(k) number<br>Manufacturer | EZTest - Gas Biological Indicators<br>K930683<br>Mesa Laboratories Inc. | | This 510(k) submission modifies the indications for use of the predicate device in order to include it as a component in a process challenge device. No modifications were made to the technology or intended use. #### 5.6 Device Description The Andersen EOGas 4® SteriTest consists of a single-use self-contained biological indicator (SCBI) placed in a reusable BI receptacle. It is designed for monitoring the efficacy of the 3 hour sterilization cycle at 50°C in an EOGas 4 Ethylene Oxide Gas Sterilizer. The SCBI, the EZTest - Gas Biological Indicator, consists of a plastic vial that serves as the culture tube and a cap including a filter material port to allow ethylene oxide to enter the vial. The plastic vial contains Bacillus atrophaeus spores inoculated onto a paper carrier, and a glass ampoule containing modified soybean casein digest broth and phenol red acting as a pH indicator. There is a chemical indicator printed on the unit label of the SCBI to indicate EO exposure. Following manufacturer's instructions, the operator inserts the Andersen EOGas 4 SteriTest into the reusable BI receptacle on the purge probe of the EOGas 4 Ethylene Oxide Gas Sterilizer, and initiates a 3 hour cycle at 50°C. After cycle completion, the SCBI is retrieved and activated by crushing the glass ampoule. The chemical indicator on the SCBI changes from blue to a green/brown color depending on the duration of ethylene oxide exposure. The activated SCBI and an unprocessed control are incubated at 30-35℃ for 48 hours, and monitored for any color change and/or turbidity. Evidence of microbial growth by color change from red-orange to {5}------------------------------------------------ vellow and/or turbidity must be interpreted as a failure to meet the conditions necessary for sterilization (cycle failed); no color change or turbidity indicates conditions for sterilization were achieved (cycle passed). #### 5.7 Indications for Use The Andersen EOGas 4 SteriTest consists of a self-contained biological indicator inoculated with viable Bacillus atrophaeus bacterial spores that is placed in a dedicated BI receptacle in the sterilizer. It monitors the efficacy of the 3 hour sterilization cycle at 50°C in the EOGas 4 Ethylene Oxide Gas Sterilizer. Critical process parameters for the cycle are summarized in Table 5-1. | | Table 5-1. Critical parameters for the 3 hour cycle in the EOGas 4 Ethylene Oxide Gas Sterilizer | | | |--|--------------------------------------------------------------------------------------------------|--|--| |--|--------------------------------------------------------------------------------------------------|--|--| | Ethylene Oxide | Temperature | Relative<br>Humidity | Ethylene Oxide<br>Exposure Time | Total Cycle<br>Time | |----------------|-------------|----------------------|---------------------------------|---------------------| | 17.6 g ± 5% | 50°C ± 3°C | 35-90% | 3 hours | 3.5 hours | #### ર જ Device Comparison The technological characteristics of the Andersen EOGas 4 SteriTest are substantially equivalent to the predicate device - EZTest - Gas Biological Indicators, because both indicators are intended for the same use, use the same technology, are designed in the same way, and perform substantially equivalently. The Andersen EOGas 4 SteriTest differs from the predicate device in the configuration used in a load. The difference raises no issues related to safety or effectiveness of the subject device in the sterilization cvcle. A comparison between the devices is listed in Table 5-2. | Element | EZTest - Gas BI | EOGas 4 SteriTest | |------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------| | Intended Use | Sterilization method: EO gas<br>Process parameters: EO concentration, time,<br>temperature, and relative humidity | Identical | | Organism | Bacillus atrophaeus (ATCC 9372) | Identical | | Viable Spore<br>Population | $\geq 1.0 x 10^6$ | Identical | | Device Design | Paper strip containing indicator organism;<br>Glass ampoule containing growth medium;<br>Capped vial serving as a culture tube;<br>A pH indicator in medium for color change;<br>A process indicator indicating EO exposure | Identical for the BI;<br>Additionally, a BI receptacle is<br>added to create a greater<br>challenge to the sterilization<br>process. | | Materials of<br>Construction | Paper, glass, and polypropylene | Paper, glass, polypropylene, and<br>aluminum | | Configuration in<br>Load | SCBI placed within the load | SCBI in a receptacle | Table 5-2. Device Comparison {6}------------------------------------------------ | Indications for Use | Not available | The Andersen EOGas 4 SteriTest consists of a self-contained biological indicator inoculated with viable Bacillus atrophaeus bacterial spores that is placed in a dedicated BI receptacle in the sterilizer. It monitors the efficacy of the 3 hour sterilization cycle at 50°C in the EOGas 4 Ethylene Oxide Gas Sterilizer. | |---------------------|---------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| |---------------------|---------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| #### ર તે Performance Testing The Andersen EOGas 4 SteriTest has been validated using applicable tests in FDA 2007 "Guidance for Industry and FDA Staff: Biological Indicator (BI) Premarket Notification [510(k)] Submissions", and AAMI/ANSI/ISO 11138-1 "Sterilization of health care products - Biological indicators - Part 1: General requirements" (FDA Recognition Number 14-296). For the EZTest - Gas Biological Indicators, tests include viable spore population assay, resistance characteristics study, carrier and primary packaging materials (growth inhibition) evaluation, holding time assessment, and reduced incubation time validation. The results of all studies met the established acceptance criteria when applicable. We also demonstrated that under the same exposure conditions, the resistance characteristics of the EZTest - Gas Biological Indicators were the same when measured in a Biological Indicator Evaluator Resistometer in the presence or absence of vacuum. The Andersen EOGas 4 SteriTest represents a rigorous challenge to the EOGas 4 sterilization process. Its resistance characteristics are greater than the same EZTest-Gas Biological Indicator placed in the worst-case locations of the maximum fabric, metal, plastic, and endoscope validation loads. The performance of the Andersen EOGas 4 SteriTest in the 3 hour cycle at 50 C in an EOGas 4 Ethylene Oxide Gas Sterilizer is summarized in Table 5-3. Table 5-3. Summary of bench tests performed to demonstrate safety and effectiveness of the Andersen EOGas 4 SteriTest | Test | Description | Result | |---------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------| | Functionality | 1) Critical parameters include temperature, time, and gas concentration under a<br>relative humidity of 35-90%;<br>2) Device is appropriate for monitoring the efficacy of the sterilization process<br>claimed | Pass | | Shelf Life | Maintains performance specifications (resistance characteristics and correctly<br>indicate pass/fail in cycles) throughout the stated shelf life of 2 years;<br>Stability demonstrates reasonable assurance for effectiveness | Pass | In conclusion, the Andersen EOGas 4 SteriTest is substantially equivalent to the legally marketed predicate, the EZTest - Gas Biological Indicator.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...