EZTEST-GAS

K930683 · Sgm Biomedical, Inc. · FRC · Mar 4, 1997 · General Hospital

Device Facts

Record IDK930683
Device NameEZTEST-GAS
ApplicantSgm Biomedical, Inc.
Product CodeFRC · General Hospital
Decision DateMar 4, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.2800
Device ClassClass 2

Indications for Use

EZTest® - Gas self contained biological indicators are for monitoring the efficacy of ethylene oxide gas sterilization processors. Performance characteristics are established in accordance with USP XXIII for the 600mg/l process.

Device Story

EZTest®-Gas is a self-contained biological indicator for ethylene oxide (EO) sterilization. Device consists of a polypropylene vial containing a paper carrier inoculated with Bacillus subtilis spores and a crushable glass ampule filled with soybean casein digest culture medium and a pH indicator (bromothymol blue or phenol red). During use, the device is placed in an EO sterilizer load. Post-cycle, the operator manually flexes the vial to crush the ampule, releasing medium onto the spores. The device is incubated at 35°C–39°C. If sterilization is ineffective, surviving spores metabolize the medium, producing acid that lowers pH and triggers a color change (to yellow) and/or cloudiness. If sterilization is effective, the medium remains the original color. This provides healthcare personnel with a visual confirmation of sterilization efficacy.

Clinical Evidence

Bench testing only. Performance evaluated across three spore lots and multiple media lots. Testing included: D-values at 600 mg/l EO, 54°C, 60% RH; survival/kill times; recovery of injured spores post-exposure; validation of 48-hour incubation readout; effect of pH indicator on recovery; and 18-month shelf-life stability. Results confirmed functional conformance to USP XXIII specifications.

Technological Characteristics

Self-contained biological indicator. Materials: polypropylene vial/cap, Type I pharmaceutical grade glass ampule, chromatography-grade filter paper carrier. Sensing: biological (Bacillus subtilis spores). Detection: pH-sensitive colorimetric (bromothymol blue or phenol red). Dimensions: 1.75" length, 0.34" OD. Sterilization: ethylene oxide compatible. No software or electronic components.

Indications for Use

Indicated for monitoring the efficacy of ethylene oxide gas sterilization processors in healthcare settings.

Regulatory Classification

Identification

Biological sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor adequacy of sterilization. The device consists of a known number of microorganisms, of known resistance to the mode of sterilization, in or on a carrier and enclosed in a protective package. Subsequent growth or failure of the microorganisms to grow under suitable conditions indicates the adequacy of sterilization. Physical/chemical sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.

Predicate Devices

Submission Summary (Full Text)

{0} k930683 MAR - 4 1997 "510(k) SUMMARY" [as required by Section 807.92(c)] 5-1 {1} 10 Evergreen Dr. Suite E · Bozeman, MT 59715 · 406 / 585-9535 FAX: 406 / 585-9219 John R. Gillis, Ph.D. President # "510(k) Summary" [as required by Section 807.92(c)] ## 510(k) Notification **EZTest® - Gas** biological indicator monitor for ethylene oxide gas sterilizers ### Submitted by: SGM Biotech, Inc. 10 Evergreen Drive, Suite E Bozeman, MT 59715 ### Contact: John R. Gillis, Ph.D. President Phone: (406) 585-9535 FAX: (406) 585-9219 This summary was prepared on October 22, 1996 [as required by Section 807.92(a)(1)] ### DEVICE: Trade name: **EZTest® - Gas** biological indicator Common name: Self-contained biological indicator for ethylene oxide gas [as required by Section 807.92(a)(2)] 5-2 {2} SUBSTANTIALLY EQUIVALENT TO THE FOLLOWING LEGALLY MARKETED DEVICES: ATITEST® - PyMaH Corp. (K895704) ATTEST® - 3M Co. (pre 1976 device) PROOF® - AMSCO (K801109) [as required by Section 807.92(a)(3)] DESCRIPTION OF DEVICE: EZTest® - Gas [as required by Section 807.92(a)(4)] The biological indicator consists of a self-contained unit which includes: bacterial spores inoculated onto a paper carrier, a small glass ampule containing sterile culture medium and color indicator, and a plastic vial that serves as the culture tube. This unit complies to the performance characteristics described in the USP XXIII. ## Spores Bacterial endospores of *Bacillus subtilis*, ATCC 9372 or equivalent, are inoculated onto a 591 Schleicher & Schuell, Inc. filter paper carrier* cut into strips of a convenient size to be placed inside the plastic body of the device. The population of spores on each individual piece of paper will be between $0.5 \times 10^{00}$ and $5 \times 10^{00}$. The spore concentration will be labeled as a $10^{6}$ concentration. The guideline in the general section on sterilization, USP XXI, p. 121, indicates that the population of spores for a $10^{6}$ biological indicator should be between $0.5 \times 10^{6}$ and $5.0 \times 10^{6}$. This is also the SGM Biotech, Inc. specification. The population data is based on the heat shock population assay. The heat shock procedure is $65^{\circ}\mathrm{C}$ to $70^{\circ}\mathrm{C}$ for 15 minutes as outlined in USP XXIII official monographs, P. 204. * reference USP XXIII p. 202 5-3 {3} 5-4 # Culture Tube The culture tube is made of a polypropylene. The plastic culture tube is a) 1.75" in length and 0.34" in OD. The 0.02" wall allows the culture tube to be flexible so that the culture media glass ampule can be broken to activate the test. # Culture Medium The culture medium, consisting of a formulated soybean casein digest base, is filled into ampules of Type I pharmaceutical grade glass, flame sealed and sterilized. The sealed ampules are of convenient size to be placed into the plastic body with the spore paper. Following filling, the ampules are exposed to a steam processing cycle to render them sterile. The ampule is a thin wall glass which allows it to be easily crushed when the plastic body is compressed. This provides the spores with nutrient medium for growth. The culture medium has a pH indicator (bromothymol blue or phenol red) added to it which appears as a blue green or orange red color respectively prior to crushing for incubation. When the spores grow, they produce acid and alter the pH. This causes the medium to change to a yellow color. If the medium changes to yellow, viable spores are present. If the medium remains the original blue green or orange red color, the spores did not grow, which means they are not viable. Therefore, if the ethylene oxide gas sterilization process was effective, the medium will stay clear and maintain the original color. If the process was not effective, the spores will turn the medium yellow and/or cloudy. # Filter Paper Vent The plastic body is closed with a ½ disc filter vent material that allows the exchange of gases from the sterilizer or from the environment. # Cap A polypropylene plastic cap is used to hold the vent material in place. This cap has a small hole in it to allow gas to freely move in and out of the body. These materials are all compatible with ethylene oxide gas sterilization. These materials have no effect on the resistance of the biological indicator. They are similar to those used in the 3M Attest®. {4} 5-5 # INTENDED USE OF DEVICE: [as required by Section 807.92(a)(5)] **EZTest® - Gas** self contained biological indicators are for monitoring the efficacy of ethylene oxide gas sterilization processors. Performance characteristics are established in accordance with USP XXIII for the 600mg/l process. The operational principles of the device are as follows. The **EZTest® - Gas** indicator is placed into a sterilizer load and an appropriate sterilization cycle is run. As sterilization conditions are met, the bacterial spores are destroyed by logarithmic chemical death principles. Following the sterilization cycle, the BI is removed from the load, the operator activates the self-contained ampule by mechanically flexing the side wall of the plastic vial or by using the plastic crusher provided to break the glass ampule containing the culture medium. The activated **EZTest® - Gas BI** is then placed in an incubator which will provide temperatures 35°C to 39°C which is conducive for growth of any bacterial spores which might have survived the sterilization cycle. Microbial growth of the indicator organisms will produce acid in the microbiological culture medium containing the pH indicator, thus lowering the pH of the solution and invoking a yellow and/or cloudy color change of the media components. If all spores have been killed the media will remain clear and maintain its original color. # SUMMARY OF THE TECHNOLOGICAL CHARACTERISTICS OF EZTest® - Gas COMPARED TO THE CHARACTERISTICS OF THE PREDICTATE DEVICE: [as required by Section 807.92(a)(6)] **EZTest® - Gas**, being similar in type to other products presently in the marketplace whose useful intent is the monitoring of sterilizer efficacy, has been determined to be safe and effective. Active components of the indicator have been in routine use more than 20 years. They predate the medical device amendment of the Food and Drug Cosmetic Act and are recognized by the United States Pharmacopoeia XXIII. {5} 5-6 Spores of *Bacillus subtilis* var *niger* are the microorganisms of choice for the evaluation of ethylene oxide gas sterilization processes because of their high degree of resistance to this type of processes. The USP XXIII specifically indicates that this species of microorganism be used to monitor ethylene oxide gas sterilization process. Numerous biological indicators on the marketplace, such as Proof®, Attest®, and ATITest®, use these microorganisms and have proved to be both safe and efficacious. EZTest® - Gas biological indicators are produced under a manufacturing in-process quality control system to specifications defining exacting parameters for production to assure a consistent product and good manufacturing procedures. Upon completion of production, each lot is inspected and tested to a quality assurance specification to affirm functional conformance to product specifications. A significant portion of the quality assurance specification is testing to the performance characteristics of biological indicators as outlined in USP XXIII. Such compliance is required on each lot prior to release. ## DETERMINATION OF SUBSTANTIALLY EQUIVALENT BASED ON PHYSICAL CHARACTERISTICS: [as required by Section 807.92(b)(1)] A comparison is provided of the legally marketed predicate devices describing similarities and differences such as technology and other important characteristics. A. ATITest® PyMaH/ATI (K902344): This device is exactly the same as the PyMaH ATITest® disclosed in 510(k) notification K902344. The ATITest® was developed by SGM Biotech for private label to PyMaH/ATI. B. Attest® 3M - Pre '76 Product: This device is virtually identical in size and shape to the 3M Attest®. It has virtually the same carrier material for the spores with the same concentration of *Bacillus subtilis* spores (10⁶). The volume of culture medium in the crushable glass ampule is similar. The pH indicator is bromothymol blue in the Attest®, as well as the EZTest®. Therefore, when viable spores grow, the color changes from blue green to yellow. {6} The glass of the crushable glass ampule in Attest® is virtually the same as that used in the EZTest®. The culture medium in the 3M Attest® is a soybean casein digest derived culture medium. EZTest® uses a soybean casein digest derived culture medium. The specific formulation of the SGM Biotech medium is considered proprietary and confidential. Activation of the 3M Attest® biological indicator devices are performed by compressing the flexible sides of the plastic culture tubes to break the crushable glass ampule containing the culture medium. This is exactly the way EZTest®-Gas is activated. ## C. PROOF® - AMSCO (K801109): The EZTest® - Gas is similar to the AMSCO PROOF® in that: 1) similar concentration *B. subtilis* spores are used, i.e. $10^{6}$. 2) spores are carried on chromatography grade pure cotton derived filter paper. 3) culture medium is contained in crushable glass ampule. 4) culture medium is a soybean casein digest derived medium. 5) spore growth is detected by lowering the pH of the culture medium. This shift in pH is detected by the addition of phenol red to the medium which is similar to the EZTest - Gas. ## DETERMINATION OF SUBSTANTIALLY EQUIVALENCE BASED ON IN USE PERFORMANCE: [as required by Section 807.92(b)(2)] The performance of EZTest® - Gas was evaluated for the following characteristics (Certificate of Performance - attached): a. Bacterial spore species b. Bacterial spore concentration c. Sample purity d. Resistance characteristics including: 1. D values at 600 mg/l EO, 54°C, 60% RH. 2. Survival/Kill times at 600 mg/l EO, 54°C, 60% RH. {7} e. Claims were evaluated using at least three (3) spore lots using different spore crops and different media lots. f. Resistance characteristics were evaluated to determine the effect of holding time after the exposure to the sterilant until incubation, upon the recovery of injured spores. g. Validation of 48 hours incubation read out time following the CDRH guideline. h. Evaluation of the effect of the pH indicator upon the recovery of injured spores. i. Validation of the effect of the sterilization process on the ability of the media to support the growth of injured spores. j. Stability of the product over the eighteen (18) month shelf life claim. ## CONCLUSION: [as required by Section 807.92(b)(3)] EZTest® - Gas is appropriate for monitoring ethylene oxide gas sterilization processes of 600 mg/l EO, 54°C, 60% RH and EZTest® - Gas meets the USP XXIII requirements.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...