STA - STACLOT DRVV SCREEN; STA - STACLOT DRVV CONFIRM
K061805 · Diagnostica Stago, Inc. · GIR · Dec 6, 2006 · Hematology
Device Facts
Record ID
K061805
Device Name
STA - STACLOT DRVV SCREEN; STA - STACLOT DRVV CONFIRM
Applicant
Diagnostica Stago, Inc.
Product Code
GIR · Hematology
Decision Date
Dec 6, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 864.8950
Device Class
Class 1
Indications for Use
The STA®-Staclot® dRVV Screen and STA®-Staclot® dRVV Confirm kits are intended for the detection of lupus anticoagulants (LA) in plasma by the dilute Russell's viper venom method (1) performed with analyzers of the STA® line suitable to these reagents.
Device Story
STA®-Staclot® dRVV Screen and Confirm are in vitro diagnostic reagents used on STA® line analyzers to detect lupus anticoagulants (LA) in citrated plasma. The assay utilizes the dilute Russell’s viper venom method. The Screen reagent contains low phospholipid concentrations; presence of LA prolongs clotting time. The Confirm reagent contains high phospholipid concentrations to neutralize LA, shortening the clotting time compared to the Screen. The kit includes a heparin inhibitor to neutralize heparin up to 0.8 IU/mL. Results are used by clinicians to aid in the identification of LA, which are antibodies directed against phospholipid/protein complexes. The device is intended for use in clinical laboratory settings by trained personnel.
Clinical Evidence
Bench testing only. Precision/reproducibility studies showed within-run CV < 1.0% and inter-assay CV ≤ 4.1%. Analytical specificity testing confirmed heparin neutralization up to 0.8 IU/mL and LMWH (Fragmin) up to 0.8 anti-Xa IU/mL. Method comparison study at two sites demonstrated 92% agreement with the predicate device. Reference range established using 27 normal plasmas.
Technological Characteristics
Reagents consist of freeze-dried Russell’s viper venom, phospholipids, calcium, and heparin inhibitor. Sensing principle is clotting time measurement via STA® line analyzers. Reconstituted reagent stability is 72 hours. Storage at 2-8 °C.
Indications for Use
Indicated for the detection of lupus anticoagulants (LA) in human citrated plasma. Intended for professional use in clinical laboratory settings using STA® line coagulation analyzers.
Regulatory Classification
Identification
Russell viper venom reagent is a device used to determine the cause of an increase in the prothrombin time.
K083878 — LUPOTEK CORRECTIN VL, LUPOTEK DETECTIN VL, PLASMACON LA · R2 Diagnostics, Inc. · Jun 25, 2010
K132076 — REAGENT, RUSSEL VIPER VENOM - LA CONFIRM · Dsrv, Inc. · Jan 10, 2014
K110031 — HEMOSIL DRVVT SCREEN & HEMOSIL DRVVT CONFIRM · Instrumentation Laboratory CO · Aug 24, 2011
Submission Summary (Full Text)
{0}
1
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
ASSAY ONLY TEMPLATE
A. 510(k) Number:
K061805
B. Purpose for Submission:
New assay
C. Measurand:
Lupus Anticoagulant
D. Type of Test:
Clotting
E. Applicant:
Diagnostica Stago
F. Proprietary and Established Names:
Staclot® DRVV Screen
Staclot® DRVV Confirm
G. Regulatory Information:
1. Regulation section:
21 CFR 864.8950
2. Classification:
Class I
3. Product code:
GIR
4. Panel:
{1}
81 Hematology
H. Intended Use:
1. Intended use(s):
The STA®-Staclot® dRVV Screen and Confirm are intended for the detection of lupus anticoagulants (LA) in plasma by the dilute Russell’s viper venom method.
2. Indication(s) for use:
3. Special conditions for use statement(s):
4. Special instrument requirements:
Analyzers of the STA® line
I. Device Description:
The STA®-Staclot® dRVV Screen and Confirm assay Kit consists of freeze-dried Russell’s viper venom, phospholipids, calcium and heparin inhibitor (UHF). The STA®-Staclot® dRVV Screen is available in 2ml and 5 ml vials. The STA®-Staclot® dRVV Confirm is available in 2 ml vials.
J. Substantial Equivalence Information:
1. Predicate device name(s):
HemosIL LAC Screen and Confirm
2. Predicate 510(k) number(s):
K990302
3. Comparison with predicate:
{2}
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Intended Use | Detection of lupus anticoagulant in plasma by the dilute Russell’s viper venom method | Same |
| Sample Requirements | Citrated plasma | same |
| Storage requirement | 2-8 °C | same |
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Reconstituted reagent stability | 72 hours | 48 hours |
## K. Standard/Guidance Document referenced (if applicable):
## L. Test Principle:
Lupus anticoagulants are antibodies directed against phospholipid/protein complexes.
Staclot® dRVV Screen test is performed with a low phospholipids concentration reagent to screen samples. If lupus anticoagulants are present, the clotting time will be prolonged.
The Staclot® dRVV Confirm Reagent contains a higher phospholipids concentration to neutralize the LA present in the plasma to be tested. The clotting time obtained with the Staclot® dRVV Confirm Reagent will be shorter than the one observed with the Staclot® dRVV Screen reagent.
## M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
3 lots of Staclot® dRVV Screen and Confirm were used to determine within-run precision using LA positive and negative control plasmas. Each control was assayed 21 times on each lot. A CV of less than 1.0% was obtained for all lots.
{3}
Inter-assay precision was assessed using lyophilized positive and negative LA samples on 1 lot of Staclot® dRVV Screen and Confirm. Positive and negative samples were assayed on the Staclot® dRVV Screen, and a positive sample was assayed on the Staclot® dRVV Confirm. Samples were assayed in 10 runs over 6 days. CV’s ≤ 4.1% was obtained for the Screen and Confirm reagents.
b. Linearity/assay reportable range:
c. Traceability, Stability, Expected values (controls, calibrators, or methods): Stability data supported the reagent shelf life.
d. Detection limit:
e. Analytical specificity:
Samples from patients on heparin may give a falsely elevated clotting time using the Staclot® dRVV Screen and Confirm assays. The Staclot® dRVV Screen and Confirm Assay contains a heparin inhibitor that will neutralize the effects of heparin up to 0.8 IU/mL. To verify this claim a normal plasma pool and a LA positive plasma were spiked with known quantities of UF. The plasmas were then tested with three lots of Staclot® dRVV Screen and Confirm reagents. No significant difference was seen in the Screen, Confirm, and Normalized ratios up to 0.8 IU/mL.
To assess the effect of LMWH on the Staclot® dRVV Screen and Confirm assays, a normal plasma pool and a LA positive plasma were spiked with known quantities of Fragmin. The plasmas were then tested with three lots of Staclot® dRVV Screen and Confirm reagents. No significant difference was seen in the Screen, Confirm, and Normalized ratios up to 0.8 anti-Xa IU/mL fragmin.
f. Assay cut-off:
2. Comparison studies:
a. Method comparison with predicate device:
The device was compared to the predicate at a 2 site study. At site one samples were collected from several hospitals in France, and assayed at Diagnostica Stago. Site 2 testing was performed at a hospital reference
4
{4}
laboratory in Canada. Results demonstrated 92% agreement with the predicate device.
b. Matrix comparison:
3. Clinical studies:
a. Clinical Sensitivity:
b. Clinical specificity:
c. Other clinical supportive data (when a. and b. are not applicable):
4. Clinical cut-off:
5. Expected values/Reference range:
27 normal plasmas were tested with the Staclot® dRVV Screen and Confirm assays. A normal reference sample was also tested with each batch of test plasmas. Testing was performed over multiple days on the STA-R. The mean Staclot Screen and Confirm ratio was 1.04 and 1.08.
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.