LUPOTEK CORRECTIN VL, LUPOTEK DETECTIN VL, PLASMACON LA
K083878 · R2 Diagnostics, Inc. · GIR · Jun 25, 2010 · Hematology
Device Facts
Record ID
K083878
Device Name
LUPOTEK CORRECTIN VL, LUPOTEK DETECTIN VL, PLASMACON LA
Applicant
R2 Diagnostics, Inc.
Product Code
GIR · Hematology
Decision Date
Jun 25, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 864.8950
Device Class
Class 1
Attributes
Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K083878 · Jun 25, 2010
LUPOTEK CORRECTIN VL, LUPOTEK DETECTIN VL, PLASMACON LA
R2 Diagnostics, Inc.
Retrospective clinical patient samples from referring laboratories
Retrospective clinical samples were used to establish the assay cutoff via ROC analysis and to perform a method comparison against the predicate device.
122 patients from referring laboratories with known clinical status (LA positive/negative); Sample Size: 122; Number of Sites: 1
Diagnostica Stago DRVV Screen and DRVV Confirm
Normalized ratio (NR) and clinical agreement (PPA/NPA)
Indications for Use
LupoTek Detectin VL and Correctin VL test kits are qualitative tests intended to aid in the detection of lupus anticoagulants (LA) in citrated human plasma by the dilute Russell's viper venom method in professional clinical laboratories. PlasmaCon LA is intended for use as an LA positive, abnormal quality control plasma to monitor the performance of diagnostic assays, performed in professional clinical laboratories, for the presence of lupus anticoagulants in citrated plasma.
Device Story
LupoTek Detectin VL and Correctin VL are in vitro diagnostic reagents for lupus anticoagulant (LA) detection in citrated human plasma. Unlike traditional Russell's viper venom, these use Vipera lebetina venom to activate Factor X, which, with Factors V, II, calcium, and phospholipid, generates thrombin to clot fibrinogen. Detectin VL (low phospholipid) screens for prolonged clotting times; Correctin VL (high phospholipid) neutralizes LA to normalize clotting times, confirming presence. PlasmaCon LA is a lyophilized, LA-positive abnormal control plasma used to monitor assay performance. Used in clinical laboratories by technicians/pathologists; results aid clinicians in diagnosing LA-related conditions. Benefits include standardized, reliable LA screening and quality control.
Clinical Evidence
Bench testing only. Precision studies (CLSI EP5-A2) on three lots showed within-device CV < 5%. Method comparison study on 155 patient samples across three sites compared to predicate reagents yielded 98% PPA (91.9-99.4% CI) and 96% NPA (88.0-98.5% CI). Interference testing established limits for hemolysis, icterus, lipemia, and heparin. No clinical prospective/retrospective studies were required.
Technological Characteristics
Lyophilized reagents containing Vipera lebetina venom, phospholipids, anti-heparin agents, calcium, buffers, stabilizers, and dye. Sensing principle: dilute venom clotting assay (dRVVT). Connectivity: standalone reagent for use on automated coagulation analyzers (e.g., STA Compact). Sterilization: not applicable (reagents).
Indications for Use
Indicated for use in professional clinical laboratories to aid in the detection of lupus anticoagulants in citrated human plasma samples and as an abnormal quality control plasma for monitoring diagnostic assay performance.
Regulatory Classification
Identification
Russell viper venom reagent is a device used to determine the cause of an increase in the prothrombin time.
Predicate Devices
STA®-Staclot® DRVV Screen and STA®-Staclot® DRVV Confirm (k061805)
American Diagnostica LAtrol Abnormal Control (k935254)
Submission Summary (Full Text)
{1}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
A. 510(k) Number:
k083878
B. Purpose for Submission:
Clearance of a new device
C. Measurand:
Lupus anticoagulant
D. Type of Test:
Dilute venom clotting assay
E. Applicant:
R² Diagnostics, Inc.
F. Proprietary and Established Names:
LupoTek Detectin VL
LupoTek Correctin VL
PlasmaCon LA
G. Regulatory Information:
1. Regulation section:
21 CFR §864.8950, Russell viper venom reagent
21 CFR §864.5425, Multipurpose system for in vitro coagulation studies
2. Classification:
Class II
3. Product code:
GIR, Reagent, Russell viper venom
GGC, Control, plasma, abnormal
4. Panel:
81 Hematology
H. Intended Use:
1. Intended use(s):
LupoTek DetectTin VL and CorrecTin VL test kits are qualitative tests intended to aid in the detection of lupus anticoagulants (LA) in citrated human plasma by the dilute Russell’s viper venom method in professional clinical laboratories.
PlasmaCon LA is intended for use as an LA positive, abnormal quality control plasma to monitor the performance of diagnostic assays, performed in professional clinical laboratories, for the presence of lupus anticoagulants in citrated plasma.
2. Indication(s) for use:
Same as Intended Use
3. Special conditions for use statement(s):
For prescription use only.
4. Special instrument requirements:
Not applicable
I. Device Description:
LupoTek DetectTin VL and LupoTek CorrecTin VL use Vipera lebetina venom rather
{2}
than Vipera russelli (Russell's Viper) venom in the dRVVT assay for lupus anticoagulant. LupoTek DetecTin is a lyophilized reagent containing a low concentration of phospholipid, anti-heparin agents, calcium, buffers, stabilizers and a dye. LupoTek DetecTin VL is available in 2 mL vials. LupoTek CorrecTin is a lyophilized reagent containing a high concentration of phospholipid, anti-heparin agents, calcium, buffers, stabilizers and a dye. LupoTek CorrecTin is available in 1 mL vials.
PlasmaCon LA is a lyophilized reagent prepared from citrated plasma from known LA donors and normal donors and contains buffer and stabilizers.
# J. Substantial Equivalence Information:
1. Predicate device name(s):
STA®-Staclot® DRVV Screen and STA®-Staclot® DRVV Confirm
American Diagnostica LAtrol Abnormal Control
2. Predicate 510(k) number(s):
k061805
k935254
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| | LupoTek DetecTin VL & LupoTek CorrecTin VL | STA®-Staclot® DRVV Screen & DRVV Confirm |
| Intended Use | LupoTek DetecTin VL and CorrecTin VL test kits are qualitative tests intended to aid in the detection of lupus anticoagulants (LA) in citrated human plasma by the dilute Russell's viper venom method in professional clinical laboratories. | Intended for the detection of lupus anticoagulants (LA) in plasma by the dilute Russell's viper venom method (1) performed with analyzers of the STA line suitable to these reagents. |
| Patient sample | Citrated human plasma | Same |
| Analyte | Lupus Anticoagulant | Same |
| Principle | LupoTek DetecTin VL is a Russell's viper venom method performed at low concentrations of phospholipid designed as a screening reagent.LupoTek CorrecTin VL is a Russell's viper venom method performed at high phospholipid concentration designed to neutralize the LA and corrects the clotting time. | STA®-Staclot® DRVV Screen is a Russell's viper venom method performed at low concentrations of phospholipid to screen test plasma.STA®-Staclot® DRVV Confirm is a Russell's viper venom method performed at higher phospholipid concentration to neutralize the LA present in the test plasma. |
{3}
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| | LupoTek DetecTin VL & LupoTek CorrecTin VL | STA®-Staclot® DRVV Screen & DRVV Confirm |
| Venom source | Vipera lebetina | Vipera russelli |
| Reconstituted stability | 24 hr - 2-8°C; 8 hr - room temp | 72 hr on board stability |
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| | PlasmaCon LA | American Diagnostica LAtrol Abnormal Control & LAtrol Normal Control |
| Intended Use | PlasmaCon LA is intended for use as an LA positive, abnormal quality control plasma to monitor the performance of diagnostic assays, performed in professional clinical laboratories, for the presence of lupus anticoagulants in citrated plasma. | Plasmas developed for use as part of daily quality control procedures for Lupus Anticoagulant (LA) testing. |
| Matrix | Lyophilized citrated human plasma positive for LA. | Same |
| Principle | The LA in the abnormal control plasma lengthens the clotting times of LA sensitive diagnostic assays. | Same |
| Analyte | Lupus anticoagulantA | Same |
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| | PlasmaCon LA | American Diagnostica LAtrol Abnormal Control & LAtrol Normal Control |
| Reconstituted stability | 8 hr - 2-8°C; 4 hr - room temp | 8 hr - 2-8°C |
# K. Standard/Guidance Document Referenced (if applicable):
CLSI EP5-A2, Evaluation of Precision Performance of Quantitative Measurement Methods, Approved Guideline, 2004
CLSI EP12-A2, User Protocol for Evaluation of Qualitative Test Performance Approved Guideline, $2^{\mathrm{nd}}$ Ed., 2008
CLSI C28-A2, Defining, Establishing and Verifying Reference Intervals in the Clinical Laboratory; Approved Guideline, 2004
{4}
CLSI H21-A5 Collection, Transport and Processing of Blood Samples for Testing Plasma-Based Coagulation Assays and Molecular Hemostatsis Assays; Approved Guideline, 2008
Guidance for Industry and FDA Staff: Statistical Guidance on Reporting Results from Studies Evaluating Diagnostic Tests, 2007
# L. Test Principle:
Vipera lebetina venom, like Vipera russelli (Russell's viper) venom, directly activates Factor X without requiring Factor VII. Factor Xa in the presence of calcium, phospholipid, and Factor V/Va will activate prothrombin to thrombin. Once present, thrombin will cleave fibrinogen to fibrin and the fibrin polymerizes into a solid clot.
Lupus anticoagulants are phospholipid dependent antibodies. Low levels of phospholipid are inadequate to neutralize the antibody/inhibitor, whereas, high phospholipid levels will. The low level of phospholipid in LupoTek DetecTin VL is sensitive to the present of lupus anticoagulants (LA). The high level of phospholipid in LupoTek CorrecTin VL renders the reagent insensitive to LA.
PlasmaCon LA is used to monitor testing variable in laboratory quality control system for assays sensitive to the presence of LA.
# M. Performance Characteristics (if/when applicable):
# 1. Analytical performance:
# a. Precision/Reproducibility:
Repeatability (within-run precision) studies were performed on three lots of LupoTek DetecTin VL, three lots of LupoTek CorrecTin VL and three lots of PlasmaCon LA. Results (clotting time in seconds) were determined from testing two runs per day over 20 days on the STA Compact analyzer according to CLSI EP5-A2. The mean, SD and $95\%$ CI were calculated for each analyte. The $\% \mathrm{CV}$ of the average clotting times was $< 5\%$ and was within the acceptance criteria of $\leq 5\%$ for repeatability and $\leq 10\%$ for within-device precision.
DetecTin VL
| Detectin VL lot # | Repeatability, %CV | | Within-device, %CV | |
| --- | --- | --- | --- | --- |
| | #NP2460 | #LA0560 | #NP2460 | #LA0560 |
| DT2330 | 1.7% | 1.2% | 3.2% | 3.3% |
| DT2340 | 1.1% | 1.5% | 4.2% | 4.4% |
| DT2350 | 1.0% | 1.4% | 3.3% | 3.3% |
| Root mean sq. | 1.3% | 1.4% | 3.6% | 3.7% |
CorrecTin VL
| Correctin VL lot # | Repeatability, %CV | | Within-device, %CV | |
| --- | --- | --- | --- | --- |
| | #NP2460 | #LA0560 | #NP2460 | #LA0560 |
| CT4210 | 1.4% | 1.5% | 3.5% | 2.8% |
| CT4220 | 0.9% | 1.1% | 4.0% | 2.4% |
| CT4250 | 1.1% | 1.3% | 3.4% | 2.6% |
| Root mean sq. | 1.1% | 1.3% | 3.7% | 2.6% |
{5}
5
# PlasmaCon LA
| PlasmaCon LA lot# | Repeatability, %CV | | Within-device, %CV | |
| --- | --- | --- | --- | --- |
| | #DetecTin VL DT2350 | | #CorrecTin VL CT4250 | |
| LA0540 | 1.2% | 3.3% | 1.0% | 3.2% |
| LA0550 | 1.5% | 3.1% | 1.3% | 2.8% |
| LA0560 | 1.4% | 3.3% | 1.3% | 2.3% |
| Root mean sq. | 1.4% | 3.3% | 1.2% | 2.8% |
# Pooled precision data clotting time (secs)
| | Detectin VL (3 lots) | | Correctin VL (3 lots) | |
| --- | --- | --- | --- | --- |
| | PlasmaCon N #NP2460 | PlasmaCon LA #LA0560 | PlasmaCon N #NP2460 | PlasmaCon LA #LA0560 |
| Mean secs | 38.1 | 82.4 | 32.9 | 37.6 |
| SD | 1.4 | 5.8 | 3.7 | 4.2 |
| 95% CI | 35.3 – 41.0 | 70.8 – 94.1 | 25.6 – 40.3 | 29.1 – 46.1 |
## b. Linearity/assay reportable range:
Not applicable
## c. Traceability, Stability, Expected values (controls, calibrators, or methods):
The reconstituted stability of the LupoTek DetecTin VL, CorrecTin VL and PlasmaCon LA were assessed by longitudinal studies. With criteria of an age-related trend and a maximum shift of 10% of the zero point clotting time, the predicted reconstituted stability of the LupoTek DetecTin VL and CorrecTin VL reagents is 24 hours when stored at 2-8°C and 8 hours when stored at room temperature (23-25°C).
The reconstituted stability of PlasmaCon LA is 8 hours when stored at 2-8°C and 4 hours at room temperature.
## d. Detection limit:
Not applicable
## e. Analytical specificity:
Interference studies of LupoTek DetecTin VL and CorrecTin VL were determined on a Stago Compact analyzer. Interferent was spiked into pooled normal plasma and serial dilutions were prepared. The maximum concentration tolerated in the assay was defined as the highest concentration of interferent relative to any change in the recovered value of the based PNP clotting time was less than 10%. The maximum concentrations were:
| Interferent | Maximum concentration tested | Maximum tolerated concentration |
| --- | --- | --- |
| Hemolysis | 500 mg/dL | 500 mg/dL |
| Icterus | 20 mg/dL | 1 mg/dL |
| Lipemia | 2,000 mg/dL triglyceride | 2,000 mg/dL triglyceride |
| Unfractionated Heparin | 2.0 U/mL | 0.6 U/mL |
{6}
f. Assay cut-off:
The cutoff for the LupoTek DetecTin VL and CorrecTin VL was determined by ROC analysis. Plasma from 122 patients was obtained from referring laboratories. These plasmas were de-identified, but a diagnosis was provided by the laboratory. For those samples diagnosed as LA positive, the samples were tested by ISTH guidelines and included two or more assays of different principles, one of which demonstrates phospholipid dependence (i.e., the ability of excess phospholipid to neutralize the antibody).
Each sample was evaluated with LupoTek DetecTin VL, CorrecTin VL, and Diagnostica Stago DRVV Screen and DRVV Confirm in $\mathbf{R}^{2}$ 's laboratory.
Normalized ratios (NR) were calculated according to the following equation:
$$
\mathrm {N R} = \frac {\left[ \text {S c r e e n r e a g e n t t i m e o f s a m p l e / S c r e e n r e a g e n t t i m e o f P N P} \right]}{\left[ \text {C o n f i r m r e a g e n t t i m e o f s a m p l e / C o n f i r m r e a g e n t t i m e o f P N P} \right]}
$$
Each sample was classified as TP, FP, TN or FN in reference to the clinical status (diagnosis) provided by the referring lab for the LupoTek and predicate reagents. Results were classified according to the following:
True positive (TP): positive by both the NR and clinical status
False positive (FP): positive by the NR but negative by clinical status
True negative (TN): negative by both the NR and clinical status
False negative (FN): negative by NR but positive by clinical status
# 2. Comparison studies:
a. Method comparison with predicate device:
Method comparison was conducted on 122 patients analyzed on the STA Compact in an internal study to determine the cutoff for the LupoTek DetecTin VL/CorrecTin VL normalized ratio. Subjects included a mix of known LA patients and other miscellaneous clinical conditions.
Another study using 155 patient samples were analyzed in three laboratories using LupoTek DetecTin VL and CorrecTin VL, and Diagnostica Stago DRVV Screen and DRVV Confirm on STA Compact analyzers. Samples were a mix of known LA patients and other miscellaneous clinical conditions.
Each sample was classified as being positive or negative for LA for each of the reagents (assays) by calculating the normalized ratio. Percent agreement was reported as follows:
| Percent Agreement LupoTek reagents vs. Stago reagents | | | | |
| --- | --- | --- | --- | --- |
| | Pooled data | Site 1 | Site 2 | Site 3 |
| PPA (95% CI) | 98% (91.9-99.4%) | 100% (89.3-100%) | 96% (80.5-99.3%) | 97% (82.8-99.4) |
| NPA (95% CI) | 96% (88.0-98.5%) | 96% (80.5-99.3%) | 94% (80.9-98.4%) | 100% (72.3-100%) |
| Overall Agreement | 97% | 98% | 95% | 97% |
{7}
b. Matrix comparison:
Fresh/frozen study: Patient samples were obtained from known normals (n=20) and known LA positive (n=11) donors. Each sample was tested before freezing at -70°C, and after freezing and storage, the samples were thawed in a 37°C water bath. The difference between the fresh and frozen samples was calculated and expressed in raw seconds and in percent of the fresh value. The normalized ratios (NR) were calculated for each pair of fresh sample and frozen sample and examined for any change that would result in changing the diagnosis of LA. The study found random shifts in fresh versus frozen samples; however, there was no change in NR which prompted a change in diagnosis of the 31 samples. The normalized ratios of the LA positive samples were essentially unchanged.
3. Clinical studies:
a. Clinical Sensitivity:
Not applicable
b. Clinical specificity:
Not applicable
c. Other clinical supportive data (when a. and b. are not applicable):
None
4. Clinical cut-off:
Not applicable
5. Expected values/Reference range:
A study of 131 non-hospitalized healthy donors (as determined by screening with a PT and APTT) were analyzed with LupoTek DetecTin VL and CorrecTin VL on a Stago STA Compact analyzer. Samples were double centrifuged and testing was performed in duplicate. The geometric mean and standard deviation of the clotting times were calculated and the ranges were calculated as the mean ±2 SD.
The mean PlasmaCon LA clot times (as determined in the precision study) with DetecTin VL and CorrecTin VL were 83.7 and 39.7, respectively. Results demonstrate an abnormally prolonged clotting time with DetecTin VL and a corrected time within the normal range with CorrecTin VL.
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.