IL TEST LAC SCREEN (CLAIMS ADDED TO ACL FUTURA), IL TEST LAC CONFIRM (CLAIMS ADDAD TO ACL FUTURA)

K990302 · Instrumentation Laboratory CO · GIR · Apr 8, 1999 · Hematology

Device Facts

Record IDK990302
Device NameIL TEST LAC SCREEN (CLAIMS ADDED TO ACL FUTURA), IL TEST LAC CONFIRM (CLAIMS ADDAD TO ACL FUTURA)
ApplicantInstrumentation Laboratory CO
Product CodeGIR · Hematology
Decision DateApr 8, 1999
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 864.8950
Device ClassClass 1

Indications for Use

IL Test™ LAC Screen and IL Test™ LAC Confirm are in vitro diagnostic products for the detection of lupus anticoagulants (a type of phospholipid interfering antibody) in human citrated plasma on IL Coagulation Systems. These tests are indicated for use with patients who have prolonged APTT test of undetermined origin. This 510(k) is intended to extend the use of these reagents onto another member in the IL family of coagulation analyzers, the ACL Futura (K951891).

Device Story

IL Test™ LAC Screen and IL Test™ LAC Confirm are in vitro diagnostic reagents used to detect lupus anticoagulants in human citrated plasma. The device functions by extending the use of existing reagents to the ACL Futura coagulation analyzer. The system processes plasma samples to identify phospholipid-interfering antibodies. Results are used by clinicians to investigate the cause of prolonged APTT tests. The device is intended for professional use in clinical laboratory settings.

Clinical Evidence

Bench testing only. Comparison study between ACL 300 (reference) and ACL Futura (test) using 53 samples showed a correlation (r) of 0.988 for normalized LAC ratio. Within-run precision assessed using 3 levels of plasma yielded CVs of 2.52%, 6.32%, and 2.36%.

Technological Characteristics

In vitro diagnostic reagents for coagulation analysis. Utilizes Russell Viper Venom reagent chemistry. Designed for use on the ACL Futura coagulation analyzer. No specific materials of construction or software algorithm class provided beyond standard coagulation analyzer operation.

Indications for Use

Indicated for detection of lupus anticoagulants in human citrated plasma for patients with prolonged APTT of undetermined origin.

Regulatory Classification

Identification

Russell viper venom reagent is a device used to determine the cause of an increase in the prothrombin time.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{1}------------------------------------------------ 4/8/99 K990302 #### Section 3 IL Test™ LAC Screen and IL Test™ LAC Confirm (Extension onto the ACL Futura) 510(k) Summary (Summary of Safety and Effectiveness) #### Submitted by: Carol Marble Regulatory Affairs Manager Instrumentation Laboratory Company 113 Hartwell Avenue Lexington, MA 02421 Phone: (781) 861-4467 (781) 861-4464 Fax: #### Contact Person: Carol Marble Phone: (781) 861-4467 #### Summary Prepared: January 29, 1999 #### Name of the device(s): IL Test™ LAC Screen (extension of reagent onto the ACL Futura) IL Test™ LAC Confirm (extension of reagent onto the ACL Futura) #### Classification name(s): 864.8950 Russell Viper Venom Reagent 81GIR Reagent, Russell Viper Venom #### Identification of predicate device(s): K922156 LUPO-TEST II Reagent (included ACL Series performance data) LUCOR Confirmatory Reagent (included ACL Series performance data) K922326 ## Description of the device/intended use(s): IL Test™ LAC Screen and IL Test™ LAC Confirm are in vitro diagnostic products for the detection of lupus anticoagulants (a type of phospholipid interfering antibody) in human citrated plasma on IL Coagulation Systems. This 510(k) is intended to extend the use of these reagents onto another member in the IL family of coagulation analyzers, the ACL Futura (K951891). ## Statement of How the Technological Characteristics of the Device Compare to the Predicate device: The performance, intended use and safety and effectiveness of IL Test™ LAC Screen (LUPO-Test II) and IL Test™ LAC Confirm (Lucor Confirmatory Reagent) on the ACL Futura Coagulation System is substantially equivalent to that on the ACL Hundred/Thousand Series Coagulation System. #### Summary of Performance Data: In a comparison study of IL Test™ LAC Screen and IL Test™ LAC Confirm on an ACL 300 (reference instrument) versus an ACL Futura (test instrument), the correlation (r) of the normalized LAC ratio was 0.988 (n = 53). On an ACL Futura, within run precision of IL Test™ LAC Screen and IL Test™ LAC Confirm assessed over multiple runs using 3 levels of plasma gave a CV of 2.52% (at a mean normalized ratio of 0.98), 6.32% (at a mean ratio of 2.03) and 2.36% (at a mean normalized ratio of 1.51). {2}------------------------------------------------ Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized depiction of a human figure, represented by three overlapping profiles, suggesting a sense of community and support. The profiles are positioned to the right of the department's name, which is arranged in a circular fashion around the logo. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA". Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 8 1999 APR Ms. Carol Marble Regulatory Affairs Manager Instrumentation Laboratory Company 113 Hartwell Avenue Lexington, Massachusetts 02173-3190 Re: K990302 Trade Name: IL Test™ LAC Screen and IL Test™ LAC Confirm on ACL Futura Regulatory Class: II Product Code: GIR Dated: January 29, 1999 Received: February 1, 1999 Dear Ms. Marble: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {3}------------------------------------------------ Page 2 Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html". Sincerely yours, Steven Sutman Steven I. Gutman, M.D. M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # Indications for Use Statement | 510(k) Number (if known): | K990302 | |---------------------------|---------| |---------------------------|---------| Device Names: IL Test™ LAC Screen and IL Test™ LAC Confirm (Extension of Reagents onto the ACL Futura) ## Indications for Use: ) IL Test™ LAC Screen and IL Test™ LAC Confirm are in vitro diagnostic products for the detection of lupus anticoagulants (a type of phospholipid interfering antibody) in human citrated plasma on IL Coagulation Systems. These tests are indicated for use with patients who have prolonged APTT test of undetermined origin. This 510(k) is intended to extend the use of these reagents onto another member in the IL family of coagulation analyzers, the ACL Futura (K951891). (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Peter E. Makin (Division Sign-Off) Division of Clinical Laboratory Devices K990302 510(k) Number. | Prescription Use | | |----------------------|--| | (Per 21 CFR 801.019) | | OR Over-The-Counter Use IL Test™ LAC Screen and LAC Confirm on ACL Futura 510(k) Section 2 Page 1 of 1
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...