The CoaguChek® Vantus System measures an INR (International Normalized Ratio) based on a prothrombin time (PT) response to monitor the effect of a therapy with vitamin K antagonists by using the CoaguChek XS PT test strips. The CoaguChek Vantus System uses fresh capillary whole blood from a finger stick. The system is intended for properly selected and suitable trained users on the prescription of the treating doctor. Users should be stabilized on anticoagulation with vitamin K antagonists for at least 6 weeks prior to single patient self-testing with the CoaguChek Vantus System. The CoaguChek Vantus System is intended for single patient self-testing only for adults, age 22 years and older.
Device Story
Handheld meter for quantitative INR measurement; uses CoaguChek XS PT test strips containing lyophilized human recombinant thromboplastin and peptide substrate. User applies finger-stick capillary blood to strip; meter measures time until thrombin cleaves substrate, generating electrochemical signal. Algorithm converts signal to INR. Used by patients for self-testing; results stored in memory (300 tests) and transferable via Bluetooth/USB. Provides on-board quality control for each strip. Enables home monitoring of vitamin K antagonist therapy; assists in clinical decision-making regarding dosage adjustments.
Clinical Evidence
Clinical evaluation included method comparison and precision studies. Method comparison (N=207) against predicate showed Pearson r=0.99. Comparison against reference Sysmex CA-1500 (N=200) showed Pearson r=0.91. Precision studies (repeatability and reproducibility) across 4 sites using 3 strip lots demonstrated CVs ranging from 1.5% to 4.2% across clinically relevant INR ranges (0.8-6.0). Normal range study (N=121) established 95% of healthy subjects between 0.9-1.1 INR.
Technological Characteristics
Handheld meter; electrochemical amperometric detection of thrombin activity. Uses CoaguChek XS PT test strips with human recombinant thromboplastin. Measuring range 0.8-6.0 INR. Connectivity: Bluetooth, USB. On-board quality control checks reagent and strip integrity. Factory calibrated via code chip.
Indications for Use
Indicated for adults age 22+ stabilized on vitamin K antagonist therapy for at least 6 weeks; intended for single-patient self-testing of INR via finger-stick capillary whole blood on prescription of a treating doctor.
Regulatory Classification
Identification
A prothrombin time test is a device used as a general screening procedure for the detection of possible clotting factor deficiencies in the extrinsic coagulation pathway, which involves the reaction between coagulation factors III and VII, and to monitor patients receiving coumarin therapy (the administration of one of the coumarin anticoagulants in the treatment of venous thrombosis or pulmonary embolism).
K183255 — Coag-Sense Prothrombin Time (PT) / INR Monitoring Device · Coagusense, Inc. · Feb 21, 2019
K062925 — COAGUCHEK XS SYSTEM · Roche Diagnostics Corp. · Jan 29, 2007
K093460 — COAGUCHEK XS PRO SYSTEM · Roche Diagnostics · Mar 18, 2010
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 25, 2017
Roche Diagnostics Justin Davis Regulatory Affairs Principal 9115 Hague Road Indianapolis, Indiana 46250
Re: K170960
Trade/Device Name: CoaguChek Vantus System Regulation Number: 21 CFR 864.7750 Regulation Name: Prothrombin time test Regulatory Class: Class II Product Code: GJS Dated: November 24, 2017 Received: November 27, 2017
Dear Justin Davis:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR
{1}------------------------------------------------
Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Leonthena R. Carrington -S
Lea Carrington Director Division of Immunology and Hematology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K170960
Device Name CoaguChek Vantus System
### Indications for Use (Describe)
The CoaguChek® Vantus System measures an INR (International Normalized Ratio) based on a prothrombin time (PT) response to monitor the effect of a therapy with vitamin K antagonists by using the CoaguChek XS PT test strips. The CoaguChek Vantus System uses fresh capillary whole blood from a finger stick.
The system is intended for properly selected and suitable trained users on the treating doctor.
Users should be stabilized on anticoagulation with vitamin K antagonists for at least 6 weeks prior to single patient selftesting with the CoaguChek Vantus System.
The CoaguChek Vantus System is intended for single patient self-testing only for adults, age 22 years and older.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <svg fill="none" height="16" viewbox="0 0 16 16" width="16" xmlns="http://www.w3.org/2000/svg"> <path d="M13.3333 2.66663H2.66663C1.9333 2.66663 1.3333 3.26663 1.3333 3.99996V13.3333C1.3333 14.0666 1.9333 14.6666 2.66663 14.6666H13.3333C14.0666 14.6666 14.6666 14.0666 14.6666 13.3333V3.99996C14.6666 3.26663 14.0666 2.66663 13.3333 2.66663ZM13.3333 13.3333H2.66663V3.99996H13.3333V13.3333ZM12 5.3333H4V6.66663H12V5.3333ZM12 8H4V9.3333H12V8ZM9.3333 10.6666H4V11.9999H9.3333V10.6666Z" fill="#000000"></path> </svg> </span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> <span> <svg fill="none" height="16" viewbox="0 0 16 16" width="16" xmlns="http://www.w3.org/2000/svg"> <path d="M13.3333 2.66663H2.66663C1.9333 2.66663 1.3333 3.26663 1.3333 3.99996V13.3333C1.3333 14.0666 1.9333 14.6666 2.66663 14.6666H13.3333C14.0666 14.6666 14.6666 14.0666 14.6666 13.3333V3.99996C14.6666 3.26663 14.0666 2.66663 13.3333 2.66663ZM13.3333 13.3333H2.66663V3.99996H13.3333V13.3333ZM12 5.3333H4V6.66663H12V5.3333ZM12 8H4V9.3333H12V8ZM9.3333 10.6666H4V11.9999H9.3333V10.6666Z" fill="#000000"></path> </svg> </span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# CoaguChek Vantus System 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92.
In accordance with 21 CFR 807.87, Roche Diagnostics hereby submits official notification as required by Section 510(k) of the Federal Food, Drug and Cosmetics Act of our intention to market the device described in this Premarket Notification 510(k).
The purpose of this Traditional 510(k) Premarket Notification is to obtain FDA review and clearance for the CoaguChek Vantus System.
| Submitter Name | Roche Diagnostics |
|-----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Address | 9115 Hague Road<br>P.O. Box 50416<br>Indianapolis, IN 46250-0457 |
| Contact | Justin Davis<br>Phone: (317) 521-6024<br>FAX: (317) 521-2324<br>Email: justin.davis@roche.com |
| Date Prepared | March 30, 2017 |
| Proprietary Name | CoaguChek Vantus System |
| Common Name | CoaguChek Vantus System |
| Classification Name | Test Time Prothrombin<br>Prothrombin Time Test |
| Product Codes, Regulation Numbers | GJS, 21 CFR 864.7750 |
| Predicate Devices | CoaguChek XS System, K062925 |
| Establishment Registration | For the CoaguChek Vantus System, the establishment registration<br>number for Roche Diagnostics GmbH in Mannheim, Germany is<br>9610126.<br>The establishment registration number for Roche Diagnostics in the<br>United States is 1823260. |
{4}------------------------------------------------
#### DEVICE DESCRIPTION 1.
#### 1.1. CoaguChek Vantus System
The CoaguChek Vantus consists of a handheld meter and associated test strips.
The CoaguChek Vantus meter is a small handheld instrument intended for the quantitative measurement of INR (International Normalized Ratio) based on a prothrombin time (PT) response, by using a single electrochemical test strip. It is designed for ease of use and is intended for patient self-testing only.
#### 1.2. CoaquChek XS PT Test Strip
The CoaguChek XS PT Test strip contains a lyophilized reagent in dried form). The reactive components of this reagent consist of thromboplastin and a peptide substrate. When a sample is applied, thromboplastin activates coagulation, which leads to the formation of thrombin. At the same time, the meter starts to measure the time. The enzyme thrombin cleaves the peptide substrate, generating an electrochemical signal. Depending on the time elapsed when it first appears, this signal is then converted by means of an algorithm into customary coagulation units and the result is displayed.
#### 2. INTENDED USE
The CoaguChek® Vantus System measures an INR (International Normalized Ratio) based on a prothrombin time (PT) response to monitor the effect of a therapy with vitamin K antagonists by using the CoaguChek XS PT test strips. The CoaguChek Vantus System uses fresh capillary whole blood from a finger stick.
The system is intended for properly selected and suitable trained users on the prescription of the treating doctor.
Users should be stabilized on anticoagulation with vitamin K antagonists for at least 6 weeks prior to single patient self-testing with the CoaguChek Vantus System.
The CoaguChek Vantus System is intended for single patient self-testing only for adults, age 22 years and older.
{5}------------------------------------------------
#### TECHNOLOGICAL CHARACTERISTICS 3.
The following tables compare the CoaguChek Vantus System with its predicate device,
CoaguChek XS System (K062925).
| Candidate Device Name | Predicate Device Name | K-Number |
|-------------------------|-----------------------|----------|
| CoaguChek Vantus System | CoaguChek XS System | K062925 |
#### Table 1: Instrument and Assay Comparison
| Feature | CoaguChek XS System (K062925) | CoaguChek Vantus System |
|---------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | The CoaguChek XS PT test strips are part of the<br>CoaguChek XS System. The CoaguChek XS System<br>measures blood clotting time for people who are<br>taking anticoagulation medications such as<br>Coumadin® or warfarin. The CoaguChek XS System<br>uses blood from a finger stick. The system is<br>intended for properly selected and suitably trained<br>users or their caregivers on the prescription or<br>other order of the treating doctor. Users should be<br>stabilized on anticoagulation medications such as<br>Coumadin® or warfarin prior to self- testing with<br>the CoaguChek XS System. | The CoaguChek® Vantus System measures an INR<br>(International Normalized Ratio) based on a<br>prothrombin time (PT) response to monitor the<br>effect of a therapy with vitamin K antagonists by<br>using the CoaguChek XS PT test strips. The<br>CoaguChek Vantus System uses fresh capillary<br>whole blood from a finger stick.<br>The system is intended for properly selected and<br>suitable trained users on the prescription of the<br>treating doctor.<br>Users should be stabilized on anticoagulation with<br>vitamin K antagonists for at least 6 weeks prior to<br>single patient self-testing with the CoaguChek<br>Vantus System.<br>The CoaguChek Vantus System is intended for<br>single patient self-testing only for adults, age 22<br>years and older. |
| Indications for Use | Same as Intended Use | Same as Intended Use |
| Test Strip | CoaguChek XS PT Test Strip | Same |
| Sample Type | Capillary Whole Blood | Same |
| Operating Principle | Electrochemical technology with amperometric<br>(electric current) detection of thrombin activity | Same |
| Reagent Test Strip<br>Principle | Human recombinant thromboplastin | Same |
| Sample Volume | The blood drop must be a minimum of 8 µl | Same |
| Measuring Range | 0.8 to 8.0 INR | 0.8 to 6.0 INR |
| On-Board Control | Built into each CoaguChek XS PT Test Strip | Same |
| Hematocrit Range | Hematocrit ranges between 25-55% have no<br>significant effect on test results | Same |
| Bilirubin | Bilirubin up to 30 mg/dL have no significant effect<br>on test results | Same |
{6}------------------------------------------------
| Feature | CoaguChek XS System (K062925) | CoaguChek Vantus System |
|---------------------------------|------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------|
| Triglyceride | Lipemic samples containing up to 500 mg/dL of<br>triglycerides have no significant effect on test<br>results | Same |
| Hemolysis | Hemolysis up to 1000 mg/dL have no significant<br>effect on test results | Same |
| Heparin | Test results are unaffected by heparin<br>concentrations up to 0.8 U/mL | Same |
| Low Molecular<br>Weight Heparin | The CoaguChek XS PT Test is insensitive to low<br>molecular weight heparins (LMWH) up to 2 IU anti-<br>factor Xa activity/mL | Same |
| Memory Capacity | 300 test results with date and time<br>No test strip lot information stored (code chip<br>data) | 300 test results with date and time<br>Code chip data from up to 5 test strip lots can be<br>stored |
| Communication<br>Interface | Infrared | Bluetooth, USB |
#### NON-CLINICAL PERFORMANCE EVALUATION 4.
The following internal performance data were provided in support of the substantial equivalence determination:
- Endogenous Interferences Hematocrit/Bilirubin/Triglycerides/Hemolysis/Heparin .
- Exogenous Interferences Drugs .
- Stability .
#### 4.1. Linearity/Assay Reportable Range
- 4.1.1. -Linearity
A linearity study is not applicable for the CoaguChek Vantus System.
#### 4.1.2. Assay Reportable Range
The assay reportable range (0.8 - 6.0 INR) of the CoaguChek Vantus System was established through method comparison studies against both the predicate (Roche CoaguChek XS System) and the reference device (Sysmex CA-1500 laboratory analyzer). Dade Innovin Reagent was used for prothrombin time (PT) determinations.
{7}------------------------------------------------
#### Test Strip Stability 4.2.
#### 4.2.1. Traceability
Each lot of CoaguChek XS PT Test Strips is factory calibrated to a reference lot of human recombinant thromboplastin traceable to the World Health Organization International Reference Preparation.
#### 4.2.1.1. Closed Vial and Transport Stability
The CoaguChek XS PT Test strips were stored at +2-8 °C and +26-30°C and tested after a storage time of 3, 6, 16, and 25 months. The 25 month time point captures the component of transport stability, as the 25 months includes a 5-day period of stressed storage at +45 °C (±2 °C).
#### 4.2.1.2. Open Vial Stability
Open vial stability simulates the repeated opening and closing of the vial by the customer when strips are removed from the vial for measurement. Open vial stability was tested by opening the vials once per day for at least 1 minute at +32 °C, 85 % relative humidity (RH) over a period of 30 days.
#### 4.2.1.3. Out of Vial Stability
Out of vial stability was tested to prove that the test strips can be kept outside the vial before measurement for at least 10 minutes when directly exposed to +32 ℃, 85 % RH. This was verified by storing the test strip outside the vial under the specified environmental conditions for 14 minutes before measurement
#### Detection Limit/Factor Sensitivity 4.3.
For Factor Sensitivity studies, four CoaguChek XS PT Test strip lots were used.
Standard human plasma was mixed with varying amounts of factor II, V, VII or X deficient plasma to obtain plasma samples with different factor activities (0, 1, 10, 20, 30, 40, 50, 60, 70, 80, 90, and 100 %). For each test strip lot, at least 4 measurements were performed per coagulation factor and dilution level.
{8}------------------------------------------------
Factor sensitivities were calculated according to CLSI guidelines (CLSI H47-A2; Vol.28, Appendix D). The study verifies Factor Sensitivity at the following levels: Factor II <31%; Factor V <46%; Factor VII <44%; and Factor X <50%.
#### Analytical Specificity 4.4.
#### Endogenous and Exogenous Interferences 4.4.1.
The interference studies were performed for the following interferents: bilirubin, hemolysis, heparin, low molecular weight heparin, triglycerides. In addition, an interference study was conducted to characterize the levels of other known drugs: clopidogrel , fondaparinux, rivaroxaban, apixaban, dabigatran, edoxaban.
Interference limits were established using up to four CoaguChek XS PT Test strip lots, using fresh citrated venous whole blood from both normal and warfarin blood samples spiked separately with the interferents.
| Interferent | Concentration |
|------------------------------|---------------------------------------|
| Bilirubin | Up to 30 mg/dL |
| Hemolysis | Up to 1000 mg/dL |
| Heparin | Up to 0.8 U/mL |
| Low Molecular Weight Heparin | Up to 2 IU anti-factor XA activity/mL |
| Triglycerides | Up to 500 mg/dL |
| Clopidogrel | Up to 20 mg/dL |
| Fondaparinux | Up to 0.5 mg/L |
| Rivaroxaban | Cannot be tested with system |
| Apixaban | Cannot be tested with system |
| Dabigatran | Cannot be tested with system |
| Edoxaban | Cannot be tested with system |
#### Assay Cut-Off 4.5.
Not applicable.
{9}------------------------------------------------
#### EXTERNAL (CLINICAL) TESTING 5.
#### 5.1. Precision
#### 5.1.1. Repeatability
A total of 688 specimens from subjects on oral anticoagulation therapy with vitamin K antagonists and coagulation healthy subjects were collected. In total 344 repeatability series were performed at four PoC sites. Data analysis was carried out by PoC site, by test strip lot and by clinically relevant ranges (<2.0, 2.0-3.5, >3.5-4.5, >4.5 INR).
| Blood | < 2.0 INR | 2.0-3.5 INR | > 3.5 - 4.5 INR | > 4.5 - 6.0 INR |
|------------|-----------|-------------|-----------------|-----------------|
| N | 200 | 394 | 70 | 24 |
| Mean (INR) | 1.1 | 2.6 | 4.0 | 4.9 |
| SD (INR) | 0.04 | 0.08 | 0.12 | 0.07 |
| CV (%) | 3.8 | 3.1 | 3.2 | 1.5 |
Summary of Repeatability for CoaguChek Vantus Table 2:
#### 5.1.2. Reproducibility
The intermediate precision experiment according to the CLSI Guideline EP05-A3 was conducted at all 4 PoC sites, with three lots of CoaguChek XS PT Test strips. The data was obtained from HCP measurements of three lots of CoaguChek XS PT controls covering four levels of controls.
Table 3: Summary of Reproducibility for CoaguChek Vantus
| | | Repeatability<br>(Within Run) | | Between<br>Run | | Between<br>Strip Lot | | Between<br>Day | | Between<br>Site | | Reproducibility | | |
|------------------|------|-------------------------------|------|----------------|------|----------------------|------|----------------|------|-----------------|------|-----------------|------|------|
| Control<br>Level | N | Mean | SD | % CV | SD | % CV | SD | % CV | SD | % CV | SD | % CV | SD | % CV |
| 1 | 1040 | 1.32 | 0.03 | 2.4 | 0.01 | 0.6 | 0.04 | 2.9 | 0.01 | 0.7 | 0.01 | 1.0 | 0.05 | 4.0 |
| 2 | 1040 | 2.79 | 0.11 | 4.0 | 0.00 | 0.0 | 0.02 | 0.7 | 0.00 | 0.0 | 0.03 | 1.0 | 0.12 | 4.2 |
| 3 | 708 | 5.85 | 0.17 | 2.9 | 0.07 | 1.2 | 0.14 | 2.3 | 0.04 | 0.8 | 0.02 | 0.4 | 0.24 | 4.0 |
| 4 | 712 | 3.39 | 0.09 | 2.7 | 0.04 | 1.2 | 0.03 | 1.0 | 0.00 | 0.0 | 0.03 | 0.9 | 0.11 | 3.2 |
{10}------------------------------------------------
#### Method Comparison 5.2.
#### 5.2.1. Method Comparison versus Predicate
Accuracy was evaluated by comparing the CoaguChek Vantus System against the predicate device, the CoaguChek XS System. The clinical method comparison study was performed using capillary whole blood samples from subjects not receiving warfarin or any other anticoagulant and from subjects currently on warfarin therapy. The method comparison study was conducted across 4 PoC sites using three 3 CoaguChek XS PT Test strip lots.
Table 4: CoaguChek Vantus System vs CoaguChek XS System
| Site | N | Slope<br>(95% CI) | Intercept<br>(95% CI) | Pearson<br>r |
|------|-----|----------------------|-----------------------|--------------|
| All | 207 | 1.00<br>(1.00, 1.03) | 0.1<br>(0.1, 0.1) | 0.99 |
#### 5.2.2. Method Comparison versus Reference System
Accuracy was also evaluated by comparing the INR results of capillary samples measured on the CoaguChek Vantus System against the INR of venous plasma samples measured on Sysmex CA 1500 laboratory analyzer using Dade Innovin recombinant human tissue thromboplastin reagent (reference device). The method comparison study was performed s across four (4) sites using CoaguChek XS PT Test strip lots. Results of INR values measured on CoaguChek Vantus System fingerstick capillary whole blood samples were compared to the INR measured on Innovin using venous plasma samples. The data from individual sites were combined and a Passing-Bablok regression analysis was performed.
Table 5: CoaguChek Vantus System vs Innovin
| Site | N | Slope<br>(95% CI) | Intercept<br>(95% CI) | Pearson r |
|------|----|---------------------|-----------------------|-----------|
| 1 | 67 | 0.94<br>(0.85,1.02) | 0.2<br>(-0.0,0.4) | 0.86 |
| 2 | 43 | 1.02<br>(0.92,1.15) | 0.1<br>(-0.2,0.3) | 0.97 |
| 3 | 47 | 1.03<br>(0.94,1.12) | 0.0<br>(-0.2,0.2) | 0.97 |
{11}------------------------------------------------
| Site | N | Slope<br>(95% CI) | Intercept<br>(95% CI) | Pearson r |
|------|-----|----------------------|-----------------------|-----------|
| 4 | 43 | 0.95<br>(0.82,1.13) | 0.2<br>(-0.3,0.5) | 0.93 |
| All | 200 | 0.98<br>(0.93, 1.03) | 0.1<br>(0.0, 0.3) | 0.91 |
#### 5.3. Sample Matrix Comparison
Not applicable, as the CoaguChek Vantus System is intended for use with capillary whole blood samples only.
#### 5.4. Expected Values / Reference Range
A normal range study was conducted on 121 healthy subjects not on anticoagulation therapy. Capillary whole blood sample testing performed on the subjects not using vitamin K antagonist drugs demonstrated that 95% of the INRs ranged between 0.9 - 1.1.
#### 6. SYSTEM DESCRIPTIONS
#### 6.1. Modes of Operation
The CoaguChek Vantus System is a closed system, which only uses the CoaguChek XS PT Test Strip; other test strips will not work with the instrument.
#### 6.2. Software
The user interface of the CoaguChek Vantus instrument guides the user through the test procedure step by step. The user only needs to insert the code chip, turn the meter on, insert the test strip, and apply a blood sample. The CoaguChek Vantus meter measures the coagulation time and displays the result. After the test is completed, the meter automatically saves the test result.
The system also includes Bluetooth connectivity to allow transfer of INR results.
{12}------------------------------------------------
#### Specimen Sampling and Handling 6.3.
The CoaguChek XS PT Test strip is intended for single-use only. Once the test strip is inserted into the analyzer, a drop of fresh capillary whole blood sample collected by fingerstick is manually applied to the test strip target area and analyzed by the CoaguChek Vantus instrument.
#### 6.4. Calibration
Each lot of CoaguChek XS PT Test strips is factory calibrated to a reference lot of human recombinant thromboplastin traceable to the World Health Organization International Reference Preparation. This lot-specific calibration information is embedded within the code chip, which is required by the CoaguChek Vantus instrument prior to using the specific lot of the CoaguChek XS PT Test strip.
#### Quality Control 6.5.
The CoaguChek Vantus System provides On-Board Controls (OBC), which provide a quality control check for each individual CoaguChek XS PT Test strip used with the CoaguChek Vantus instrument. There are no additional steps required by the user to activate the OBC.
When a test strip is inserted, the CoaguChek Vantus instrument, the first check assesses potential issues with the reagent. The second check ensures that the test strip has not been exposed to environmental conditions or physical stresses, like bending. If either of these checks does not pass, then no PT/INR result is provided.
#### CONCLUSIONS 7.
The submitted information in this premarket notification supports a substantial equivalence decision.
The results of these studies demonstrate that the CoaguChek Vantus System is similar to the predicate. The data presented are a summary of external clinical evaluation, internal laboratory evaluation, and software development information.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.