For use in prothrombin time determinations and prothrombin time-based assays.
Device Story
Dade® Innovin® is an in vitro diagnostic reagent used to determine prothrombin time (PT). It consists of a lyophilized preparation of purified recombinant human tissue factor (produced in E. coli), synthetic phospholipids, calcium, and stabilizers. The reagent is used in clinical laboratory settings to initiate clotting via the extrinsic common pathway. By measuring the time required for clot formation, the device identifies deficiencies in the extrinsic coagulation system, monitors patients undergoing oral anticoagulant therapy, and assists in the assessment of specific coagulation factors. Healthcare providers use the resulting clotting time—often converted to International Normalized Ratio (INR) values—to guide clinical decisions regarding anticoagulation management and the diagnosis of hereditary or acquired coagulation disorders.
Clinical Evidence
Bench testing only. Comparative performance study conducted using clinical specimens tested with both Dade® Innovin® and the predicate device. Results showed a correlation coefficient of 0.975 for PT seconds (regression Y = 1.32X - 8.3) and 0.98 for INR values (regression Y = 0.99X - 0.09).
Technological Characteristics
Lyophilized reagent containing purified recombinant human tissue factor (E. coli derived), synthetic phospholipids, calcium, and stabilizers. Functions as an extrinsic pathway activator for coagulation screening. In vitro diagnostic use.
Indications for Use
Indicated for the determination of prothrombin time (PT) and PT-based assays to detect single or combined deficiencies of the extrinsic coagulation system, monitor oral anticoagulation therapy, and assay specific coagulation factors.
Regulatory Classification
Identification
A prothrombin time test is a device used as a general screening procedure for the detection of possible clotting factor deficiencies in the extrinsic coagulation pathway, which involves the reaction between coagulation factors III and VII, and to monitor patients receiving coumarin therapy (the administration of one of the coumarin anticoagulants in the treatment of venous thrombosis or pulmonary embolism).
{0}
JUL 22 1998
K974343
# 510(k) Summary of Safety and Effectiveness
## Dade® Innovin®
## November 18, 1997
Dade International Inc.
2173 NW 99th Avenue
Miami, FL 33172
Contact Person: Radames Riesgo at 305-392-5615 or by facsimile at 305-392-5622.
**Trade or Proprietary Name:** Dade® Innovin®
**Common or Usual Name:** Prothrombin time assay
**Classification Name:** Prothrombin Time Test (21 CFR § 864.7750)
**Registration Number:**
Dade International Inc. 1017272
1851 Delaware Pkwy.
Miami, FL 33125
Dade International Inc. 1025506
2173 NW 99th Avenue
Miami, FL 33172
The product is an *in vitro* diagnostic test for the determination of prothrombin time and prothrombin time-based assays. Dade® Innovin® is a lyophilized preparation of purified recombinant human tissue factor produced in *E. coli* combined with synthetic phospholipids, calcium and stabilizers. The reagent initiates clotting via the extrinsic common pathway in a global screening test, the prothrombin time (PT). The obtained clotting time detects single or combined deficiencies of the extrinsic coagulation system indicative of hereditary and acquired coagulation disorders, is a sensitive monitoring test for oral anticoagulation therapy and an assay for specific coagulation factors.
Dade® Innovin® is substantially equivalent in intended use and performance to Dade® Thromboplastin IS (K891169). Both the proposed product and the predicate device are formulated to provide high sensitivity to factor deficiencies and can be used for monitoring patients on oral anticoagulants. In a comparative performance study, specimens were tested using both reagents. The correlation coefficient in PT seconds between the two products was 0.975 and the regression equation was $$Y = 1.32X - 8.3$$. When results were converted into INR (International Normalized Ratio) values, the correlation coefficient was 0.98 and the regression equation was $$Y = 0.99X - 0.09$$.
We believe the documentation included in this premarket notification demonstrates that Dade® Innovin® reagent is substantially equivalent to Dade® Thromboplastin IS.
017
{1}
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration
2098 Gaither Road
Rockville MD 20850
JUL 22 1998
Radames Riesgo
Manager Regulatory Affairs
DADE Behring Inc.
P.O. Box 520672
Miami, Florida 33152-0672
Re: K974343
Dade® Innovin®
Regulatory Class: II
Product Code: GJS
Dated: July 1, 1998
Received: July 2, 1998
Dear Mr. Riesgo:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{2}
Page 2
Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for *in vitro* diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Steven I. Gutman, M.D., M.B.A.
Director
Division of Clinical Laboratory Devices
Office of Device Evaluation
Center for Devices and Radiological Health
Enclosure
{3}
510(k) Number (if known): _________________________
Device Name: DADE® INNOVIN®
Indications For Use:
For use in prothrombin time determinations and prothrombin time-based assays.
(PLEASE DO NOT WITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use ☑ (Per 21 CFR 801.109)
OR
Over-The-Counter Use ☐
(Optional Format 1-2-96)

Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.