K962407 · Behring Diagnostics, Inc. · GIL · Jul 29, 1996 · Hematology
Device Facts
Record ID
K962407
Device Name
N/T PROTEIN CONTROL PY
Applicant
Behring Diagnostics, Inc.
Product Code
GIL · Hematology
Decision Date
Jul 29, 1996
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 864.7340
Device Class
Class 2
Indications for Use
N/T Protein Control PY is used for control of accuracy and precision in the quantitative immunochemical determination of alpha₁-Antitrypsin, fibrinogen, antithrombin III, prothrombin, plasminogen, fibronectin*, C₁ Inhibitor with the Behring Nephelometer Systems and with the TurbiTimeSystem. *not available in the U.S.
Device Story
N/T Protein Control PY is a lyophilized, human citrate plasma-based quality control material. It contains specific proteins: fibrinogen, antithrombin III, prothrombin, plasminogen, alpha₁-Antitrypsin, and C₁ Inhibitor. Used in clinical laboratories to monitor accuracy and precision of quantitative immunochemical determinations performed on Behring Nephelometer Systems and TurbiTimeSystem. Healthcare providers use the provided lot-specific table of values to verify instrument performance. The control helps ensure reliable clinical decision-making by validating the consistency of coagulation and protein assay results.
Clinical Evidence
Bench testing only. Precision and reproducibility studies conducted on Behring Nephelometer using 10 vials of one lot, tested in triplicate. %CVs ranged from 1.11% to 2.72%. Stability testing confirmed 12-month shelf life for lyophilized product and 14-day stability post-reconstitution.
Technological Characteristics
Lyophilized human citrate plasma matrix. Multi-constituent control containing fibrinogen, antithrombin III, prothrombin, plasminogen, alpha₁-Antitrypsin, and C₁ Inhibitor. Designed for use with nephelometric and turbidimetric systems.
Indications for Use
Indicated for use as a quality control material to monitor laboratory coagulation and protein procedures in clinical settings.
Regulatory Classification
Identification
A fibrinogen determination system is a device that consists of the instruments, reagents, standards, and controls used to determine the fibrinogen levels in disseminated intravascular coagulation (nonlocalized clotting within the blood vessels) and primary fibrinolysis (the dissolution of fibrin in a blood clot).
Special Controls
*Classification.* Class II (special controls). A control or fibrinogen standard intended for use with a fibrinogen determination system is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 864.9.
Predicate Devices
Control Plasma N
Related Devices
K964065 — N/T PROTEIN CONTROL SL · Behring Diagnostics, Inc. · Nov 6, 1996
K032237 — N/T PROTEIN CONTROL · Dade Behring, Inc. · Sep 5, 2003
K991182 — MODIFICATION OF N/T PROTEIN CONTROL SL · Dade Behring, Inc. · May 19, 1999
K012468 — MODIFICATION TO N/T PROTEIN CONTROL SL · Dade Behring, Inc. · Sep 4, 2001
K962410 — N PROTEIN STANDARD PY · Behring Diagnostics, Inc. · Jul 16, 1996
Submission Summary (Full Text)
{0}
JUL 29 1996
K 96 2407
# Attachment 1
# 510(k) Summary of Safety and Effectiveness for N/T Protein Control PY
1. Manufactures Name, Address, Telephone, and contact person, date of preparation:
Manufacturer
Behringwerke AG
Postfach 1140
35001 Marburg
Germany
Distributor
Behring Diagnostics Inc.
151 University Avenue
Westwood, MA 02090
617-320-3000
Attn: Kathleen Dray-Lyons
Preparation date: June 20, 1996
2. Device Name/ Classification:
N/T Protein Control PY: Quality Control Material (assayed)
Classification Number: Class I (862.1660)
3. Identification of the legally marketed device:
Behring Diagnostic Inc., Control Plasma N
4. Proposed Device Description:
The N/T Protein Control PY is a lyophilized pool of citrate plasma consisting of one level containing the following proteins:
- fibrinogen
- antithrombin III
- prothrombin
- plasminogen
- alpha₁-Antitrypsin
- C1 Inhibitor
000009
{1}
000010
# 5. Proposed Device Intended Use:
N/T Protein Control PY is used for control of accuracy and precision in the quantitative immunochemical determination of alpha₁-Antitrypsin, fibrinogen, antithrombin III, prothrombin, plasminogen, fibronectin*, C₁ Inhibitor with the Behring Nephelometer Systems and with the TurbiTimeSystem.
*not available in the U.S.
# 6. Medical device to which equivalence is claimed and comparison information:
The N/T Protein Control PY is substantially equivalent in intended use to the Control Plasma N. Both products are in vitro diagnostic reagents intended for use as a quality control material to monitor certain laboratory coagulation and protein procedures. The N/T Protein Control PY like the Control Plasma N is single level protein control from a human blood-based matrix (citrate plasma). Both controls are provided with a lot specific table of values. The N/T Protein Control PY like the Control Plasma N is used for the control of the following parameters; prothrombin, fibrinogen, plasminogen, C₁ Inhibitor and antithrombin III. In addition, the N/T Protein Control PY contains alpha₁-Antitrypsin. See Table I for comparative features and the attached N/T Protein Control PY draft package insert for specific product information.
The Behring Diagnostics N/T Protein Control PY differs from the Behring Control Plasma N in that the N/T Protein Control PY is a control for nephelometric coagulation and protein procedures while the Control Plasma N is a control for opto-mechanical and chromogenic coagulation procedures.
# 7. Proposed Device Performance Characteristics:
## Precision and reproducibility:
Precision studies using multiple vials of the controls were run on the Behring Nephelometer. Ten vials of one lot were tested. Each of the parameters were determined in triplicate with each vial. The %CVs were calculated for each parameter. the %CVs ranged from 1.11 to 2.72%.
## Stability:
Stability was run according to an in-house protocols and the control was found stable for at least 12 months lyophilized and 14 days once reconstituted.
{2}
000006
22
# Table I
## COMPARATIVE FEATURES
### N/T Protein Control PY vs Control Plasma N
| Feature | Control Plasma N | N/T Protein Control PY |
| --- | --- | --- |
| 1. Human Blood-Based Matrix | Yes
(citrate plasma) | Yes
(citrate plasma) |
| 2. Single Level Control | Yes | Yes |
| 3. Multi-Constituent Control | Yes | Yes |
| 4. Quality Control Product | Yes | Yes |
| 5. Lyophilized | Yes | Yes |
| 6. Table of Values Provided | Yes | Yes |
| 7. Coagulation Control | Yes | Yes |
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.