N PROTEIN STANDARD PY
Device Facts
| Record ID | K962410 |
|---|---|
| Device Name | N PROTEIN STANDARD PY |
| Applicant | Behring Diagnostics, Inc. |
| Product Code | JIX · Clinical Chemistry |
| Decision Date | Jul 16, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 862.1150 |
| Device Class | Class 2 |
Intended Use
N Protein Standard PY is intended to be used for the establishment of reference curves.
Device Story
N Protein Standard PY is a lyophilized human serum-based calibrator containing stabilizers and preservatives. Used with Behring Nephelometer systems (K860894, K892223, K943997) to establish reference curves for coagulation assays. Operates in clinical laboratory settings; utilized by laboratory technicians to calibrate nephelometric instrumentation. Output provides reference values for patient sample quantification; facilitates accurate clinical assessment of coagulation parameters.
Clinical Evidence
Bench testing only. Precision studies on Behring Nephelometer yielded %CVs ranging from 1.01% to 2.29%. Stability testing confirmed product stability for at least 18 months in lyophilized form and 14 days post-reconstitution.
Technological Characteristics
Lyophilized human serum-based calibrator; contains stabilizers and preservatives. Designed for use with nephelometric instrumentation. No specific ASTM standards cited.
Indications for Use
Indicated for use as a calibrator for the establishment of reference curves for coagulation assays including Fibrinogen, Antithrombin III, Prothrombin, Plasminogen, alpha1-Antitrypsin, and C1 Inhibitor.
Regulatory Classification
Identification
A calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. (See also § 862.2 in this part.)
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
Predicate Devices
- N Protein Standard SY
Reference Devices
Related Devices
- K964062 — N PROTEIN STRANDARD SL · Behring Diagnostics, Inc. · Nov 21, 1996
- K991181 — MODIFICATION OF N PROTEIN STANDARD SL · Dade Behring, Inc. · Jun 1, 1999
- K012470 — MODIFICATION TO N PROTEIN STANDARD SL · Dade Behring, Inc. · Sep 28, 2001
- K962406 — N CRP STANDARD SY · Behring Diagnostics, Inc. · Jul 16, 1996
- K962407 — N/T PROTEIN CONTROL PY · Behring Diagnostics, Inc. · Jul 29, 1996