N/T Protein Controls SL/L, M, and H are for use as accuracy and precision assayed controls in the determination of the following human serum proteins by immunonephelometry with BN™ Systems: IgG, IgG, IgA, IgM, C3c, C4, Transferrin, in Innonophobionou y ma2-macroglobulin, Haptoglobin, α--acid glycoprotein, Prealbumin, Albumin, c.y antill Jpon) @2 maor J.g/L-chain lambda & kappa, β2-microglobulin, soluble Transferrin Receptor (STFR), Ferritin, IgE, and Total protein; and by immunoturbidimetry with the TurbiTimeSystem: IgG, IgA, IgM, C3c, C4, Transferrin, Albumin, Haptoglobin, α4acid glycoprotein.
Device Story
N/T Protein Control SL is a liquid quality control material derived from human serum with added stabilizers and preservatives. It serves as an assayed control to verify the accuracy and precision of human serum protein measurements. The device is used in clinical laboratory environments by trained personnel. It is processed on BN Systems (via immunonephelometry) and the TurbiTimeSystem (via immunoturbidimetry). By comparing measured values against the control's assigned ranges, healthcare providers assess the performance of their analytical systems, ensuring reliable diagnostic results for various serum proteins, which supports clinical decision-making regarding patient protein levels.
Clinical Evidence
Bench testing only. Stability was evaluated according to manufacturer protocols, confirming the control is stable at +2° to +8° C as originally packaged and for at least 14 days at +2° to +8° C once opened.
Technological Characteristics
Liquid control material prepared from human serum with stabilizers and preservatives. Designed for use with immunonephelometry (BN Systems) and immunoturbidimetry (TurbiTimeSystem) platforms. No specific materials standards or software algorithms are described.
Indications for Use
Indicated for use as accuracy and precision assayed controls for human serum protein determination in clinical laboratory settings using immunonephelometry (BN Systems) or immunoturbidimetry (TurbiTimeSystem).
Regulatory Classification
Identification
A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.
{0}------------------------------------------------
# SEP = 4 2001
## 510(k) Summary For N/T Protein Control SL
### Manufacture's Name, Address, Telephone, and Contact Person, Date of 1. Preparation:
Manufacturer:
Dade Behring Marburg GmbH Emil-von-Behring Str. 76 Marburg/Germany
Contact Information:
Dade Behring Inc. Glasgow Site P.O. Box 6101 Newark, Delaware 19714 Attn: Rebecca S. Ayash Tel: 302-631-6276
Preparation date:
July 31, 2001
#### Device Name/ Classification: 2.
N/T Protein Control SL:
Quality Control Material (assayed)
Classification Number:
Class I (862.1660)
#### Identification of the Legally Marketed Device: 3.
N/T Protein Control SL (K002852)
#### Device Description: 4.
N/T Protein Control SL is a liquid control prepared from human serum with sfabilizers and preservative. It is intended to be used as an accuracy and precision control for the presentation of human serum proteins by immunonephelometry with BN™ Systems and by immunoturbidimetry with the TurbiTimeSystem.
#### 5. Device Intended Use:
N/T Protein Controls SL/L, M, and H are for use as accuracy and precision assayed controls in the determination of the following human serum proteins by immunonephelometry with BN™ Systems: IgG, IgG7, IgA, IgM, C3c, C4, Transferrin, in Innonophobionou y ma2-macroglobulin, Haptoglobin, α--acid glycoprotein, Prealbumin, Albumin, c.y antill Jpon) @2 maor J.g/L-chain lambda & kappa, β2-microglobulin, soluble Transferrin Receptor (STFR), Ferritin, IgE, and Total protein; and by immunoturbidimetry with the TurbiTimeSystem: IgG, IgA, IgM, C3c, C4, Transferrin, Albumin, Haptoglobin, α4acid glycoprotein.
Image /page/0/Picture/22 description: The image shows the word "CONFIDENTIAL" in large, bold, black letters. The letters are closely spaced and fill most of the frame. The background is plain white, providing a stark contrast to the bold text.
000027
{1}------------------------------------------------
### Medical device to which equivalence is claimed and comparison information: 6.
The modified N/T Protein Control SL is substantially equivalent in intended use to N/T The modified NYT T Totell Control SE to Sassianaly Se Sales N/T Control SL, like the Protein Control OL (K002002) ourrolly marketsused as quality control material to current NYT Frotein Oontrol OE is interface to be accuracy in BN™ Systems and the TurbiTimeSystem.
#### Device Performance Characteristics: 7.
## Stability:
Stability.
Stability was evaluated according to Dade Behring protocols and the control was found to Stablity was evaluation as overling to D to +8° C, as originally packaged and for at least 14 days at +2° to +8° C, once opened.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features an abstract image of an eagle with three lines representing its wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
SEP = 4 2001
Ms. Rebecca S. Ayash Director, Regulatory Affairs Dade Behring, Inc. Glasgow Site, PO Box 6101 Newark, DE 19714
K012468 Re:
> Trade/Device Name: N/T Protein Control SL Regulation Number: 21 CFR 862.1660 Regulatory Class: I, reserved Product Code: JJY Dated: July 31, 2001 Received: August 2, 2001
Dear Ms. Ayash:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{3}------------------------------------------------
Page 2 -
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and v additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Steven Sutman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Dade Behring Inc.
N/T Protein Control SL
510(k) Notification
# Indications for Use Statement
N/T Protein Control SL Device Name:
## Indications for Use:
N/T Protein Controls SL/L, M, and H are for use as accuracy and precision N/ 1 Toteln Controls of the following human serum proteins by immunonephelometry with BN™ Systems: IgG, IgG14, IgA, IgM, C3c, C4, immanonopholomosiy wantitrypsin, α2-macroglobulin, Haptoglobin, α;-acid Translermir, Albumin, or antilippensin, Ceruloplasmin, RbP, Ig/L-chain lambda & giyooprotein, 1 Toalbumin, Huble Transferrin Receptor (STTR), Ferritin, IgE, and Kappa, p2-morogiobailin, obluble idimetry with the TurbiTimeSystem: IgG, IgA, IgM, C3c, C4, Transferrin, Albumin, Haptoglobin, α+-acid glycoprotein.
(PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
Over-The-Counter-Use (Optional Format 1-2-96)
Kesia Alexander for Liam Cooper
(Division Sign-Off)
510(k) Number
000025
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.