K033137 · Wada, Inc. · JPA · Dec 8, 2003 · Hematology
Device Facts
Record ID
K033137
Device Name
POTENS +
Applicant
Wada, Inc.
Product Code
JPA · Hematology
Decision Date
Dec 8, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 864.5425
Device Class
Class 2
Indications for Use
The POTENS+ is a multipurpose system for use in performing clot based, photo-optical in-vitro coagulation studies including Prothrombin Time (PT) derived Fibrinogen (FBG), and the International Normalization Ratio (INR).
Device Story
POTENS+ is a semi-automated, photo-optical system for in vitro coagulation studies. It accepts citrated human plasma samples; utilizes a potentiophotometer, reagent dispenser, pipettor, and heating block to perform clot-based assays. The device detects changes in light transmittance at 660nm via an LED and silicon photovoltaic cell during clot formation. It processes signals using linear digital electronics to calculate PT, FBG, and INR. Operated by medical professionals in laboratory settings, the system provides quantitative results to clinicians. These outputs assist in monitoring anticoagulation therapy, assessing clotting pathway status, and evaluating cardiovascular risk. The device benefits patients by providing rapid, accurate coagulation profiles for clinical decision-making.
Clinical Evidence
Clinical comparison study using patient specimens across normal, diagnostic, and therapeutic ranges. Correlation coefficients (r-values): Fibrinogen 0.973-0.980; PT 0.95-0.974; INR 0.919-0.963. Precision studies followed NCCLS EP5A; total CVs <6% for fibrinogen, <5% for PT, and <8% for INR using human/bovine control plasmas. Results demonstrate performance comparable to predicate devices.
Technological Characteristics
Semi-automated potentiophotometer; optical detection via 660nm LED and silicon photovoltaic cell. Operates at 37 ± 0.3°C. Components include reagent dispenser, pipettor, heating block, computer, and printer. Performs clot-based assays using linear digital signal processing. Requires citrated plasma; no sample dilution needed. Complies with NCCLS standards for precision and interference testing.
Indications for Use
Indicated for medical professionals in laboratory settings to determine the coagulation profile of citrated human plasma specimens, including Prothrombin Time (PT), derived Fibrinogen (FBG), and International Normalization Ratio (INR).
Regulatory Classification
Identification
A multipurpose system for in vitro coagulation studies is a device consisting of one automated or semiautomated instrument and its associated reagents and controls. The system is used to perform a series of coagulation studies and coagulation factor assays.
Special Controls
*Classification.* Class II (special controls). A control intended for use with a multipurpose system for in vitro coagulation studies is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 864.9.
K183390 — QNext and DG-PT · Diagnostic Grifols, S.A. · Aug 21, 2019
K955735 — KC 4A MICRO COAGULATION ANALYZER SET/SYSTEM · Sigma Diagnostics, Inc. · Mar 11, 1996
K955724 — KC 1A COAGULATION ANALYZER/SET · Sigma Diagnostics, Inc. · Mar 12, 1996
Submission Summary (Full Text)
{0}
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
DEVICE AND INSTRUMENT TEMPLATE
A. 510(k) Number:
K033137
B. Analyte:
Prothrombin Time, Derived Fibrinogen, INR
C. Type of Test:
Semi-quantitative, Fibrin clot based and photo-optical
D. Applicant:
WADA, Inc.
E. Proprietary and Established Names:
POTENS+
F. Regulatory Information:
1. Regulation section:
21 CFR 864.5425
Multipurpose System for In Vitro Coagulation Studies
2. Classification:
Class II
3. Product Code:
JPA
4. Panel:
Hematology (81)
G. Intended Use:
1. Intended use:
The POTENS+ is a multipurpose system for use in performing clot based, photo-optical in-vitro coagulation studies including Prothrombin Time (PT) derived Fibrinogen (FBG), and the International Normalization Ratio (INR).
2. Indication for use:
The POTENS+ an in vitro medical device to determine the coagulation profile of citrated human plasma specimens to be used by a medical professional in a laboratory setting. This profile includes Prothrombin Time (PT) derived Fibrinogen (FBG), and the International Normalization Ratio (INR).
3. Special condition for use statement(s):
Not applicable
4. Special instrument Requirements:
Not applicable
{1}
Page 2 of 8
# H. Device Description
POTENS+ is a semi-automated system for *in vitro* coagulation studies, utilizing a clot based, photo-optical method. It is an arithmetically linear, digital coagulation device that determines the Prothrombin Time (PT) and calculates the International Normalization Ratio (INR) of a patient’s plasma sample. It determines the derived Fibrinogen (FBG) in which the Optical Density (O.D.) change of the optically clear FBG in plasma converts to turbid fibrin clot to determine the FBG content of the specimen. Also, included in this system are two Fibrinogen standards and three Fibrinogen controls.
# I. Substantial Equivalence Information:
1. Predicate device name(s):
Organon Tecknika MDA-180 for PT and INR
Organon Tecknika Coag-A-Mate (CAM) MTX for Fibrinogen
2. Predicate K number(s):
K962857, K924453
3. Comparison with predicate:
| Similarities | | | |
| --- | --- | --- | --- |
| Item | POTENS+ | MDA-180 | CAM -MTX |
| Intended Use | Determine coagulation profile of clinical specimens | Determine coagulation profile of clinical specimens | Determine coagulation profile of clinical specimens |
| Analytes | PT, FBG, INR | PT, FBG, INR | FBG |
| Methodology | Fibrin Clot based assay; optical measurement of direct light transmission during clot formation. | Clauss method: Fibrin Clot based assay; optical measurement of relative light transmission to the time of clot formation | Clauss method: Fibrin Clot based assay; optical measurement of relative light transmission to the time of clot formation |
| Principles of operation | Detects change in light transmittance at 660nm | Detects change in light transmittance in the spectral range 405 to 710nm. FBG @425 & PT @ 580nm | Detects change in light transmittance at 405nm |
| Method of Detection | Optical (LED with silicon photovoltaic cell | Optical (tungsten lamp) using photodiode detectors | Optical using a photometer and detectors |
| Target User | Clinical Laboratory | Clinical Laboratory | Clinical Laboratory |
| Operating Temp | 37±0.3°C | 37±1°C | 37 |
| Test Sample | Citrate plasma | Citrate plasma | Citrate plasma |
| Expression of Results | Fibrinogen: mg/dl
PT: seconds
INR: INR units | Fibrinogen: mg/dl
PT: seconds
INR: INR units | Fibrinogen: mg/dl |
| Differences | | | |
| Item | POTENS+ | MDA-180 | CAM -MTX |
{2}
Page 3 of 8
| Instrument System | Potentiophotometer, Reagent Dispenser, Pipettor, Heating block, computer and printer | Multi-Channel Discrete Analyzer | Analyzer (measuring rotor and photometer), computer and printer |
| --- | --- | --- | --- |
| Signal Processing | Linear handling of light transmission using digital electronics | Logarithmic handling of light transmission using analog electronics | Logarithmic handling of light transmission using analog electronics |
| Reagents/ Accessories | Calibration standards (2)
Controls (3)
Reagent Water
Test Cuvetts
Cuvette rack
Normal Saline
Thromboplastin (Dade® Thromboplastin C Plus) | VeriCal Calibrator Set
Verify Controls
Reagent Water
Buffer
Test Cuvetts
Clauss Reagents
Thrombin, Imidiazole buffer) | Reference Plasma (for calibration)
Verify Controls: Veronal Buffer
Sample cups
Cuvette Rotor
Fibriquik (Thrombin) |
| Test Sample Dilution | None | Required | Required |
| Reportable Range | FBC: linear range of FBG
This is from high to low standards (HS & LS),
beyond this to be established by user (99-851 mg/dl obtained in comparative study)
PT: 7.1 to 90 seconds
INR: ≥0.5 with limited clinical utility >4.5. Upper limit dictated by ISI of thromboplastin used. | FBG II: 35-700 mg/dl depending on dilution set up
PT: 8-150
INR 0.8-8.0 with limited clinical utility >4.5 | FBG: 40-650 mg/dl depending on dilution set up |
| Quality Control | Use control plasmas to monitor performance; run once every 55 samples and every 8 hours, whichever occurs first as specified in operators manual. | Use control plasmas and other parameters at frequency and limits specified in operators manual | Use control plasmas and other parameters at frequency and limits specified in operators manual |
J. Standard/Guidance Document Referenced (if applicable):
WHO Expert Committee on Biological Standardization 33rd Report; WHO
Technical Report Series 687, Geneva, 1983.
EP5-A “Evaluation of Precision Performance of Clinical Chemistry Devices,” NCCLS
EP6-P “Evaluation of the Linearity of Quantitative Analytical Methods,” NCCLS
EP7-P “Interference Testing in Clinical Chemistry,” NCCLS
{3}
Page 4 of 8
K. Test Principle:
POTENS+ is an arithmetically linear, digital coagulation instrument that simultaneously determines the Prothrombin time (PT), the International Normalization Ratio (INR) and Fibrinogen (FBG) level in plasma samples.
POTENS+ measures PT and Fibrinogen following the principles of PT-derived methods. When Thromboplastin-Calcium reagent is added to a citrated plasma sample, enzymatic reactions occur leading to the conversion of Prothrombin to Thrombin. Thrombin then converts Fibrinogen to a fibrin clot. POTENS+ detects the beginning of the fibrin clot formation that indicates the PT of the sample. The change in optical density (OD) during fibrin clot formation is detected and compared to the results from similarly treated Fibrinogen standards. POTENS+ detects these OD changes using patented potentiophotometric technology.
POTENS+ directly displays the concentration of the substance being measured because the output of the POTENS+ is inherently linear, in the true digital sense and does not require a logarithmic-linear converting device, circuit, or computers which are required
L. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
Within-run precision studies were performed according to NCCLS Guideline EP5-A using three levels of control plasma (Control 1, Control 2, and Control 3) and two instruments.
| Analyte | Sample | Inst.# 1/2 | N | Mean | CV |
| --- | --- | --- | --- | --- | --- |
| Fibrinogen | CTL 1 | combined | 40 | 322.2 | 2.8 % |
| | CTL 2 | combined | 40 | 189.7 | 3.2 % |
| | CTL 3 | combined | 40 | 101.4 | 5.7 % |
| PT | CTL 1 | combined | 40 | 20.4 | 10.7 % |
| | CTL 2 | combined | 40 | 11.5 | 2.8 % |
| | CTL 3 | combined | 40 | 16.7 | 6.2 % |
| INR | CTL 1 | combined | 40 | 3.1 | 18.4 % |
| | CTL 2 | combined | 40 | 1.2 | 6.6 % |
| | CTL 3 | combined | 40 | 2.2 | 12.3 % |
{4}
Page 5 of 8
Day to day precision studies were performed according to NCCLS Guideline EP5-A using three levels of control plasma (Control 1, Control 2, and Control 3) and two instruments on at least 20 days.
| Analyte | Sample | Inst.# 1/2 | Number of days | Number of results | Mean | CV |
| --- | --- | --- | --- | --- | --- | --- |
| Fibrinogen | CTL 1 | combined | 41 | 82 | 320.9 | 4.5% |
| | CTL 2 | combined | 41 | 82 | 191.5 | 4.4% |
| | CTL 3 | combined | 41 | 82 | 106.9 | 5.6% |
| PT | CTL 1 | combined | 41 | 82 | 20.5 | 6.4% |
| | CTL 2 | combined | 41 | 82 | 11.63 | 2.3% |
| | CTL 3 | combined | 41 | 82 | 17.2 | 4.6% |
| INR | CTL 1 | combined | 41 | 82 | 3.07 | 11.0% |
| | CTL 2 | combined | 41 | 82 | 1.18 | 5.5% |
| | CTL 3 | combined | 41 | 82 | 2.28 | 7.6% |
b. Linearity/assay reportable range:
The linearity was established according to NCCLS Guideline EP6-P. POTENS+ is designed to be linear from the Low Standard through the High Standard. A plasma pool was spiked with Fibrinogen cryoprecipitate and a portion diluted 1:9 with normal saline.
Mixtures of these two solutions were made to obtain a total of five levels. Four replicates of the five levels were evaluated. The results were compared against the theoretical results. These results indicate that the POTENS+ was linear from 99 to 851 mg/dl. Each laboratory should verify the linearity of the instrument below and above the Standard for each Lot of Standards and Thromboplastin used. Only patient results in the linear range should be reported.
Reportable range:
Method comparison studies were used to verify that the instrument reports results within the range of:
Fibrinogen 136-851 mg/dl
PT 8.8-41.2 seconds
INR 0.7-9.9
However, each laboratory should verify the reportable range of each analyte.
c. Traceability (controls, calibrators, or method):
Indirect traceability was demonstrated by including plasma fibrinogen (98/614) standard in the Ware-Ratnoff method currently used to determine the Fibrinogen concentration of the POTEN+ Standards. Communication from the NIBSC indicated that the 1st International Standard (IS) (98/614) could be used in the Ratnoff method.
{5}
Page 6 of 8
d. Detection limit:
The lower detection limit was measured in comparative studies:
Lowest Level measured:
Fibrinogen 79 mg/dl
PT 8.8 seconds (software limit 7.1)
INR 0.7 (INR calculations are limited by the lower PT limit of 7.1 and by the ISI of the thromboplastin used)
e. Analytical specificity:
Interference studies were performed by spiking human plasma or plasma pools with bilirubin, hemoglobin, heparin triglycerides, and fibrinogen degradation products. Although bilirubin, hemoglobin and triglyceride showed no affect above the normal range typically found in plasma, visibly icteric, hemolyzed or lipemic specimens should not be used. Heparin at the upper therapeutic range (0.5 Units/ml) affected PT and INR but did not affect Fibrinogen result at the highest level tested (1.7 Units/ml). Fibrinogen Degradation Product digest did not affect the PT or INR results, however, precautions should be taken when analyzing specimens with this elevation until additional studies have been completed determined the level of Fibrinogen Degradation Products that does not cause interference.
f. Assay cut-off:
Not applicable
2. Comparison studies:
a. Method comparison with predicate device:
Comparison studies were performed at four sites to demonstrate that POTENS+ is equivalent to the MTX for determining Fibrinogen concentration and to the MDA-180 for determining PT and INR results using a minimum of 200 specimens. Equivalence was demonstrated using commercially available methods/product along with patient specimens covering the normal, therapeutic and diagnostic ranges. Correlation coefficients (r-value) ranged from 0.973 to 0.980 for Fibrinogen, 0.95 to 0.974 for PT and 0.919 to 0.963 for INR results.
b. Matrix comparison:
Not reported
3. Clinical studies:
a. Clinical sensitivity:
Not provided.
b. Clinical specificity:
Not provided.
c. Other clinical supportive data (when a and b are not applicable):
4. Clinical cut-off:
Not applicable
{6}
Page 7 of 8
5. Expected values/Reference range:
From the literature, the generally accepted fibrinogen reference interval is 200mg/dl to 400mg/dl. This is only a guide to the user. Each laboratory should establish specific fibrinogen reference intervals.
## M. Instrument Name: POTENS+
## N. System Descriptions:
1. Modes of Operation:
Semi-automatic
2. Software:
The POTENS+ software is PC based with the primary responsibility of providing an interface to the user, performing the testing, and handling the results data.
FDA has reviewed applicant’s Hazard Analysis and software development processes for this line of product types:
Yes ☐ X or No ☐
3. Sample Identification:
Sample identification is by manual entry. The software accepts input from both a scanner and a keyboard.
4. Specimen Sampling and Handling:
The POTENS+ uses citrate plasma samples and the software controls interface with the pipettor so that the user can transfer the sample to a cuvette.
5. Assay Types:
Clot based assays
6. Reaction Types:
Photo-optical
7. Calibration:
The system includes two Fibrinogen standards. A four standards test is required (2 “high” and 2 “low”). The software shall recognize via the Specimen ID that the standard is being run. If it is the first standard, POTENS+ shall reset all parameters to indicate that a new batch is being started. When the instrument has successfully completed the fourth standard (second low standard) test it will determine the mean values of the PT and the Fibrinogen (FBG) calibration curve. All future FBG calculations for the batch uses this equation.
8. Quality Control:
This system includes three Fibrinogen controls. Whenever the POTENS+ encounters a Specimen ID of CTL1 (Control 1), CTL2 (Control 2) OR CTL3 (Control 2), it regards the test as a control test. The software will use the appropriate maximum/minimum control values from the parameter table to assure that the instrument is properly calibrated. If the checks fail, the software will report the value as “OUT of control.” Otherwise, the software will report the value as “In control.”
{7}
Page 8 of 8
POTENS+ Control 1 is not suitable for use as a control for Prothrombin Time (PT) or INR assays
O. Other Supportive Instrument Performance Characteristics Data Not Covered In The "L. Performance Characteristics" Section Of the SE Determination Decision Summary.
P. Conclusion:
Based upon review of the information and labeling provided, this device is found to be substantially equivalent (SE) to 21 CFR 864.5425 JPA, System, Multipurpose for In Vitro Coagulation Studies, Class II.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.