KC 4A MICRO COAGULATION ANALYZER SET/SYSTEM

K955735 · Sigma Diagnostics, Inc. · JPA · Mar 11, 1996 · Hematology

Device Facts

Record IDK955735
Device NameKC 4A MICRO COAGULATION ANALYZER SET/SYSTEM
ApplicantSigma Diagnostics, Inc.
Product CodeJPA · Hematology
Decision DateMar 11, 1996
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 864.5425
Device ClassClass 2

Intended Use

The Amelung KC 4 A Micro Coagulation Analyzer is a semi-automated mechanical clot detection system designed for the determination of prothrombin times (PT), activated partial thromboplastin times (APTT), fibrinogen concentrations and other clotting tests. Any clotting time test that has fibrin formation as its endpoint may be performed on the KC 4 A.

Device Story

Semi-automated mechanical clot detection system; determines PT, APTT, fibrinogen, and other clotting tests with fibrin formation endpoints. Used in physician office laboratories; operated by laboratory personnel. Device measures clotting time; provides quantitative results for coagulation assays. Results assist clinicians in monitoring patient coagulation status and managing therapy. Benefits include standardized, reliable coagulation testing in clinical settings.

Clinical Evidence

Performance studies conducted at three physician office laboratories (n > 45 samples) and in-house (n > 100 samples). Compared KC 4 A against reference laboratory and predicate devices (Fibrometer, CA-5000). PT/APTT correlation coefficients ranged 0.927-0.991. Fibrinogen/Factor IX/Factor X correlation coefficients ranged 0.897-0.974. Precision studies showed CV < 9%.

Technological Characteristics

Semi-automated mechanical clot detection system. Benchtop form factor. No specific materials, connectivity, or software algorithm class details provided.

Indications for Use

Indicated for patients requiring coagulation testing, including PT, APTT, fibrinogen, and other clotting assays where fibrin formation is the endpoint. Used in physician office laboratories and clinical settings.

Regulatory Classification

Identification

A multipurpose system for in vitro coagulation studies is a device consisting of one automated or semiautomated instrument and its associated reagents and controls. The system is used to perform a series of coagulation studies and coagulation factor assays.

Special Controls

*Classification.* Class II (special controls). A control intended for use with a multipurpose system for in vitro coagulation studies is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 864.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} 510(k) NOTIFICATION Sigma Diagnostics 545 South Ewing Avenue St. Louis, MO 63103 KC 4 A Micro Coagulation Analyzer December 15, 1995 K.G. 5735 # ATTACHMENT 1 ## Summary of Safety and Effectiveness {1} 510(k) NOTIFICATION Sigma Diagnostics 545 South Ewing Avenue St. Louis, MO 63103 KC 4 A Micro Coagulation Analyzer December 15, 1995 # ATTACHMENT 1 ## 510(k) Summary of Safety and Effectiveness The Amelung KC 4 A Micro Coagulation Analyzer is a semi-automated mechanical clot detection system designed for the determination of prothrombin times (PT), activated partial thromboplastin times (APTT), fibrinogen concentrations and other clotting tests. Any clotting time test that has fibrin formation as its endpoint may be performed on the KC 4 A. In beta sites performance studies, three physician office laboratories (POL) analyzed more than 45 samples using the KC 4 A. Split samples from the same specimens were also analyzed using the KC 4 A at Sigma Diagnostics, the study’s reference laboratory. PT and APTT assays were performed at all sites. The correlation coefficient and the regression equation for these comparisons were as follows: | **Site #1:** | PT: r = 0.991 y = 0.981x + 0.492 | | --- | --- | | | APTT: r = 0.960 y = 1.066x + 0.379 | | **Site #2:** | PT: r = 0.989 y = 1.019x - 0.248 | | | APTT: r = 0.965 y = 1.029x + 1.021 | | **Site #3:** | PT: r = 0.974 y = 1.012x + 0.326 | | | APTT: r = 0.927 y = 0.786x + 9.470 | Precision studies were also performed at all sites. Results were within acceptable limits. In in-house performance studies, PT and APTT were performed on more than one hundred samples using both the KC 4 A and the Fibrometer. The CA-5000 was used as the predicate device for Fibrinogen, Factor IX, and Factor X assays. The correlation coefficient and the regression equation were as follows: | PT: | r = 0.988 | y = 1.051x - 0.241 | | --- | --- | --- | | APTT: | r = 0.896 | y = 1.235x + 0.873 | | Fibrinogen: | r = 0.930 | y = 1.067x + 30.749 | | Factor IX: | r = 0.897 | y = 0.958x - 3.403 | | Factor X: | r = 0.974 | y = 1.010x - 0.166 | Precision studies were performed in-house. The coefficient of variation (CV) for all studies was less than 9%. These data clearly demonstrate that the performance of the KC 4 A Micro Coagulation Analyzer is substantially equivalent to the performance of the predicate device.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...