COAG-A-MATE MTX

K962857 · Organon Teknika Corp. · GKP · Jan 22, 1997 · Hematology

Device Facts

Record IDK962857
Device NameCOAG-A-MATE MTX
ApplicantOrganon Teknika Corp.
Product CodeGKP · Hematology
Decision DateJan 22, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 864.5400
Device ClassClass 2

Indications for Use

The Coag-A-Mate MTX is a multipurpose system for use in performing in-vitro coagulation studies, clot based and chromogenic.

Device Story

Automated system for in-vitro coagulation studies; performs clot-based and chromogenic assays. Input: patient plasma specimens. Operation: random access, fully automated processing. Output: coagulation test results. Used in clinical laboratories; operated by laboratory personnel. System facilitates diagnostic and therapeutic monitoring of coagulation status. Benefits: provides automated, precise, and efficient coagulation analysis comparable to existing predicate systems.

Clinical Evidence

Clinical comparison study using patient specimens across normal, diagnostic, and therapeutic ranges. Correlation coefficients (r) for clot-based screening assays: 0.954–0.994; clot-based quantitative assays: 0.988–0.995; chromogenic assays: 0.954. Precision studies followed NCCLS EP5-T2; total CVs <5% (clot-based screening), <8% (clot-based quantitative), and <7% (colorimetric).

Technological Characteristics

Fully automated, random-access coagulation analyzer. Performs clot-based and chromogenic assays. Spectrophotometric detection at 405 nm. Features LIS interface and user-definable assay parameters. Precision testing conducted per NCCLS EP5-T2 standards.

Indications for Use

Indicated for in-vitro coagulation studies, including clot-based and chromogenic assays, for use in clinical laboratory settings.

Regulatory Classification

Identification

A coagulation instrument is an automated or semiautomated device used to determine the onset of clot formation for in vitro coagulation studies.

Special Controls

*Classification.* Class II (special controls). A fibrometer or coagulation timer intended for use with a coagulation instrument is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 864.9.

Predicate Devices

Submission Summary (Full Text)

{1} K962857 5.0 510(k) Summary Coag-A-Mate MTX JAN 22 1997 This summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and the final rule under 21 CFR 807.92 published December 14, 1994. (A)(1) The submitter's name, address, telephone number, a contact person, and the date the summary was prepared: Submitter's Name: Organon Teknika Corporation Submitter's Address: 100 Akzo Avenue, Durham, NC, 27712 Submitter's Telephone: (919) 620-2634 Submitter's Contact: Ann M. Quinn Date 510(k) Summary Prepared: July 19, 1996 (a)(2) The name of the device, including the trade or proprietary name if applicable, the common or usual name, and the classification name, if known. Trade or Proprietary Name: Coag-A-Mate MTX Common or Usual Name: Multipurpose System for In Vitro Coagulation Studies Classification Name: Multipurpose System for In Vitro Coagulation Studies (a)(3) An identification of the legally marketed device to which the submitter claims substantial equivalence. Device Equivalent to: Organon Teknika MDA-180 (K924453) Organon Teknika Coag-A-Mate X2 (K813564) (a)(4) A description of the device Device description: Automated system for in vitro coagulation studies, clot based and chromogenic. (a)(5) A statement of the intended use of the device Device Intended Use: The Coag-A-Mate MTX is a multipurpose system for use in performing in-vitro coagulation studies, clot based and chromogenic. 5 1 {2} (a)(6) A summary of the technological characteristics of the new device in comparison to those of the predicate device. The CAM MTX, MDA-180 and CAM-X2 are multipurpose systems capable of performing in-vitro coagulation studies, all three perform clot based assays. The MDA-180 and CAM MTX are also capable of performing chromogenic assays. Similar to the MDA-180, the Coag-A-Mate MTX is a fully automated system that has features which increase its ease of use. Table 1.1 below outlines similarities/differences between the MDA-180 and the CAM MTX. Table 1.1 Similarities and Differences Between the MDA-180 and the CAM-MTX | MDA-180 | | CAM-MTX | | --- | --- | --- | | Spectrophotometric Detection | 405 - 710 nm | 405 nm | | Random Access | yes | yes | | Stat Mode Capability | yes | yes | | LIS Interface | yes | yes | | Assay Capability | Clot Based Chromogenic Immunoassay | Clot Based Chromogenic | | Ancillary Components | Imidazole Buffer Reagent Water Wash Solution Probe Cleaner VeriCal Calibrators Test Cuvettes | Probe Cleaner Cleaning Solution Test Cuvettes | | User Definable Assay Parameters | yes | yes | (b)(1) A brief discussion of the nonclinical tests submitted, reference, or relied on in the premarket notification submission for a determination of substantial equivalency. Testing was performed to determine normal range, correlation and bias, precision and interfering substances. Interfering substances, and specimen collection and preparation are well documented for these types assays and are in accordance with NCCLS standards and recommendations of the International Society of Thrombosis and Hemostasis. (b)(2) A brief discussion of the clinical tests submitted, reference, or relied on in the premarket notification submission for a determination of substantial equivalency. coagmtx {3} A comparison study was conducted using patient specimens over the normal, diagnostic and therapeutic range. R values ranged from (0.954 - 0.994) for clot based screening assays, (0.988 - 0.995 for clot based quantitative assays and (r = 0.954) for chromogenic assays. Total imprecision for each analyte was well within NCCLS recommendations. (b)(3) The conclusions drawn from the nonclinical and clinical tests that demonstrate that the device is as safe, as effective, and performed as well or better than the legally marketed device identified in (a)(3). The performance characteristics of the new device are comparable to those of the predicate device and typical of these systems in general. The data presented in the premarket notification demonstrate that the new device performs substantially equivalent to the predicate device. Comparison studies were performed to demonstrate the CAM MTX is equivalent to the MDA-180 and CAM-X2 for the performance of in vitro coagulation studies (clotting, chromogenic). Equivalence was demonstrated using currently commercially available reagents along with patient specimens covering the normal, therapeutic and diagnostic range. Correlation coefficients ranged from r = 0.954 to 0.995 for clot based assays. A correlation coefficient of r = 0.954 was obtained for assays requiring spectrophotometric measurement at a specified wavelength such as chromogenic assays. Precision studies were performed following NCCLS EP5-T2 "Evaluation of Precision Performance of Clinical Chemistry Devices." Total precision c.v.'s were less than 5% for clot based screening assays, less than 8% for clot based quantitative assays and less than 7% for colorimetric assays. Data was generated on the CAM MTX using commercially available normal and abnormal control plasma. coagmtx
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...