TANGO3 WATER STORAGE TANK WITH OZONE DISINFECTION SYSTEM
K140984 · Tango3, LLC · FIN · Jan 12, 2015 · Gastroenterology, Urology
Device Facts
Record ID
K140984
Device Name
TANGO3 WATER STORAGE TANK WITH OZONE DISINFECTION SYSTEM
Applicant
Tango3, LLC
Product Code
FIN · Gastroenterology, Urology
Decision Date
Jan 12, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5820
Device Class
Class 2
Attributes
Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K140984 · Jan 12, 2015
TANGO3 WATER STORAGE TANK WITH OZONE DISINFECTION SYSTEM
Tango3, LLC
Dialysis facility water quality monitoring records
Field testing was conducted in four dialysis facilities with documented histories of water contamination to demonstrate the device's ability to maintain bacterial counts and endotoxin levels below AAMI recommendations during routine clinical use.
Field testing; Dialysis water quality; Routine clinical setting
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Field testing in clinical facilities
Dialysis facilities with histories of water contamination; Sample Size: 4 facilities; Number of Sites: 4
Not applicable for this study
Maintenance of total bacterial counts and endotoxin levels below AAMI RD 62:2006 action levels
Indications for Use
The TANGO3 Water Storage Tank with Ozone Disinfection System is intended to be used for disinfection of the water distribution system of a dialysis facility. The tank of the TANGO3 is also used as the water holding tank of the distribution system. The disinfection process is completely automated. Ozone concentration is between 0.2 ppm and 0.3 ppm. Weekly disinfection cycles should be for four periods of 30 minutes with adequate water flushes between them and after the last cycle. In addition to the weekly disinfection cycle, a 45 minute disinfection cycle with at least two rinse cycles may also be performed multiple times weekly for addition of bioburden as determined by user facility monitoring. At the end of disinfection, the distribution loop will have less than 0.1 mg/l. (0.1 ppm) of ozone, in accordance with AAMI/ISO 26722:2009 (4.2.13.5).
Device Story
TANGO3 Water Storage Tank with Ozone Disinfection System automates disinfection of dialysis facility water distribution loops. System inputs: atmospheric air (processed via desiccant dryer) and water from storage tank. Operation: Corona discharge ozone generator produces 0.2–0.3 ppm ozone; venturi-based injection mixes ozone into water; ozonated water recirculates through distribution loop during non-operational hours. System includes sensors (flow, ozone, level, conductivity, ambient ozone) and PLC-based controller. Output: disinfected water loop with <0.1 ppm residual ozone post-flush. Operated by facility staff; system monitors for ozone presence and triggers alarms if levels exceed safety thresholds. Benefits: maintains water quality per AAMI/ISO 26722:2009; reduces bioburden/biofilm; prevents water contamination in dialysis settings.
Clinical Evidence
No clinical trials on patients; evidence based on bench testing and field performance. Bench testing validated 6-log reduction of microbiological contamination (bacteria, viruses, fungi) and compliance with AAMI RD 62:2006. Reliability testing simulated 10 years of use. Field testing conducted in four dialysis facilities with documented contamination histories confirmed the system maintained water quality below AAMI action levels.
Indicated for disinfection of water distribution systems in dialysis facilities. Used for automated ozone-based bioburden reduction in water holding tanks and distribution loops. Contraindicated for use during patient dialysis treatment.
Regulatory Classification
Identification
A hemodialysis system and accessories is a device that is used as an artificial kidney system for the treatment of patients with renal failure or toxemic conditions and that consists of an extracorporeal blood system, a conventional dialyzer, a dialysate delivery system, and accessories. Blood from a patient flows through the tubing of the extracorporeal blood system and accessories to the blood compartment of the dialyzer, then returns through further tubing of the extracorporeal blood system to the patient. The dialyzer has two compartments that are separated by a semipermeable membrane. While the blood is in the blood compartment, undesirable substances in the blood pass through the semipermeable membrane into the dialysate in the dialysate compartment. The dialysate delivery system controls and monitors the dialysate circulating through the dialysate compartment of the dialyzer.(1) The extracorporeal blood system and accessories consists of tubing, pumps, pressure monitors, air foam or bubble detectors, and alarms to keep blood moving safely from the blood access device and accessories for hemodialysis (§ 876.5540) to the blood compartment of the dialyzer and back to the patient. (2) The conventional dialyzer allows a transfer of water and solutes between the blood and the dialysate through the semipermeable membrane. The semipermeable membrane of the conventional dialyzer has a sufficiently low permeability to water that an ultrafiltration controller is not required to prevent excessive loss of water from the patient's blood. This conventional dialyzer does not include hemodialyzers with the disposable inserts (Kiil type) (§ 876.5830) or dialyzers of high permeability (§ 876.5860). (3) The dialysate delivery system consists of mechanisms that monitor and control the temperature, conductivity, flow rate, and pressure of the dialysate and circulates dialysate through the dialysate compartment of the dialyzer. The dialysate delivery system includes the dialysate concentrate for hemodialysis (liquid or powder) and alarms to indicate abnormal dialysate conditions. This dialysate delivery system does not include the sorbent regenerated dialysate delivery system for hemodialysis (§ 876.5600), the dialysate delivery system of the peritoneal dialysis system and accessories (§ 876.5630), or the controlled dialysate delivery system of the high permeability hemodialysis system § 876.5860). (4) Remote accessories to the hemodialysis system include the unpowered dialysis chair without a scale, the powered dialysis chair without a scale, the dialyzer holder set, dialysis tie gun and ties, and hemodialysis start/stop tray.
Predicate Devices
TANGO3 Water Storage Tank with Ozone Disinfection System (K093641)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the circle is an abstract symbol that resembles three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 12, 2015
TANGO3, LLC C/O Ms. Lori Holder Regulatory Affairs Consultant 141 Citizens Boulevard Simpsonville, Kentucky 40067
Re: K140984
Trade/Device Name: TANGO3 Water Storage Tank with Ozone Disinfection System Regulation Number: 21 CFR 876.5820 Regulation Name: Hemodialysis System and Accessories Regulatory Class: II Product Code: FIN Dated: December 11, 2014 Received: December 12, 2014
Dear Ms. Holder:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Page 2 - Ms. Holder
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Image /page/1/Picture/8 description: The image shows the name "Tejashri Purohit-Sheth, M.D." in a stylized font. The name is written in black ink and appears to be a professional title. The letters are slightly italicized, giving the text a formal and elegant appearance. The "M.D." abbreviation indicates that the person is a medical doctor.
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K140984
Device Name
TANGO3 Water Storage Tank with Ozone Disinfection System
#### Indications for Use (Describe)
The TANGO3 Water Storage Tank with Ozone Disinfection System is intended to be used for disinfection of the water distribution system of a dialysis facility. The tank of the TANGO3 is also used as the water holding tank of the distribution system. The disinfection process is completely automated. Ozone concentration is between 0.2 ppm and 0.3 ppm. Weekly disinfection cycles should be for four periods of 30 minutes with adequate water flushes between them and after the last cycle. In addition to the weekly disinfection cycle, a 45 minute disinfection cycle with at least two rinse cycles may also be performed multiple times weekly for addition of bioburden as determined by user facility monitoring. At the end of disinfection, the distribution loop will have less than 0.1 mg/l. (0.1 ppm) of ozone, in accordance with AAMI/ISO 26722:2009 (4.2.13.5).
Type of Use (Select one or both, as applicable)
- Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K140984 510(K) SUMMARY
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.
| Submitter's Name: | TANGO3, LLC<br>141 Citizens Blvd.<br>Simpsonville, KY 40067<br>Telephone: (502) 722-8794 |
|-------------------|------------------------------------------------------------------------------------------|
| Contact person: | Hugh Doss, Manager |
Date of Summary: January 9, 2015
Device Name: TANGO3 Water Storage Tank with Ozone Disinfection System
Device Classification Name: Hemodialysis System and Accessories (876.5820, FIN)
## Device Classification: Class II
Device Description: The TANGO3 Water Storage Tank with Ozone Disinfection System is specifically designed to facilitate ozone induction into the storage tank, and then distribute the ozonated water through the distribution water loop during non-operational hours of a hemodialysis facility. The storage tank of the TANGO3 system is filled with adequate water and the ozone concentration is increased. The ozonated water is distributed throughout the distribution loop. The solution is recirculated throughout the distribution system. To complete the disinfection process, the system and distribution loop are then rinsed with adequate water. The described process is accomplished with 4 consecutive 30 minute cycles of ozone disinfection weekly. In addition, a disinfection cycle of 45 minutes multiple times per week may also be used for additional reduction of bioburden (as determined by user facility monitoring). After disinfection TANGO3 will leave the distribution loop with less than 0.1 mg/l. (0.1 ppm) of ozone, in accordance with AAMI/ISO 26722:2009 (4.2.13.5).
## System Description and Theory of Operation
The TANGO3 Water Storage Tank with Ozone Disinfection System consists of the following components: a heat regenerative desiccant air dryer at a 20 SCFH of air flow, a Corona discharge ozone generator that produces a working concentration of 0.3ppm of ozone, a venturi based injection system, HDPE or Stainless Steel water storage tank, an off gas ozone destruct system, a multi-stage centrifugal distribution pump (duel distribution pumps are offered as an option), and a main controller and operator's interface. Sensors incorporated into the system include return and output flow sensors, an ozone sensor, a level sensor, a conductivity sensor, and an atmospheric ozone detector. Fluid contacting materials include: Stainless Steel 300 Series, High Density Polyethylene (HDPE), Lexan® (Polycarbonate), Acrylic (PMMA), Teflon® (PTFE), Viton® (FKM), Kynar® (PVDF), and PVC. The System produces a dissolved ozone concentration of 0.20 ppm to 0.30 ppm during the disinfection process.
The TANGO3 System has been designed to meet the applicable requirements of the following voluntary standards:
{4}------------------------------------------------
- AAMI /ANSI RD52:2004 Dialysate for hemodialysis ■
- AAMI/ANSI RD62:2006 Water treatment equipment for hemodialysis applications ■
- AAMI/ISO 26722:2009 Water treatment equipment for hemodialysis applications and ■ related therapies
Ozone is produced in a flat gas stream which results in low energy consumption and high ozone concentration. Energy consumption is less than 7 kW per kilo ozone produced.
The ozone disinfection process is derived through the generator located on the front of the device. Atmospheric air passes through silica beads removing humidity from the air. The air then passes through an air dryer to additionally remove all remaining moisture from the air. The dry air passes between two electrode plates consisting of high current. To prevent the air from breaking down electrically along a simple line of rupture or spark, a sheet of glass is placed between the metal plates. By this means the air is made to break down electrically throughout thus greatly increasing the amount of ozone produced.
The air in the system is not pressurized but operates under a vacuum produced when the ozone recirculation pump is operating. The pump uses the purified water from the storage tank to pressurize the water. Then the water passes through a venturi creating a vacuum which will pull atmospheric air through the acrylic tube with silica gel, through the air dryers, and then through the generating chamber. The ozone gas is mixed with the purified water at the venturi and diffused into the holding tank. The ozonated water is then distributed from the holding tank to the distribution loop through the distribution pump and returns back to the storage tank. The return loop is monitored for dissolved ozone level achieving a level of 0.2 ppm to 0.3 ppm. Upon completion of the disinfection dwell time, a flushing phase is activated to destroy the residual ozone prior to use with patients.
To protect the distribution piping and well as other water contacting materials, a low level of ozone is used. Laboratory testing has proven a 6 log reduction of microbiological contamination when using dissolved ozone concentration levels between 0.2 ppm to 0.3 ppm while showing no immediate effect on PVC distribution piping.
### Operation Overview
Three operational modes are provided: Dialysis, Maintenance, and Disinfection. The system is set to dialysis mode when the water system is being used for dialysis. The system is set to maintenance mode to allow for manual control of the system. Ozone product only occurs in disinfection mode. A description of the disinfection process follows.
The disinfection process is divided into the following phases, as shown above:
- 1) Ozonation: During this phase the ozone generator is on and ozonated water is being recirculated through the loop.
- 2) Loop and Tank Emptying: Once the ozonation phase is complete a valve will be actuated and the returning flow will go to drain instead of returning to the tank. The distribution pump will be kept on until the LOW TANK level is detected. At this point, the distribution pump, ozone generator and recirculation pump will be turned off. The tank drain valve will open. allowing the tank to empty by gravity.
{5}------------------------------------------------
- 3) Tank and Loop Refill: After the loop and tank are empty, TANGO3 will allow adequate water to enter the storage tank. Once the TOP FLOAT level is detected, no more water will be allowed to enter the tank.
- 4) During the weekly disinfection cycle, steps 1) through 3) will be repeated for each cycle.
- 5) Ozone Destruction: A flushing process will be done with non ozonated water until the ozone sensor detects that the level of ozone residual complies with the AAMI requirements
To avoid the possibility of dissolved 03 in the distribution loop while water is being used for other purposes, the system has two (2) flow meters to measure the input and the output of the storage tank. If those two (2) readings are not equal the ozone generator and the distribution pump will be turned off. In addition, daily manual checks may be performed to assure the absence of ozone in the loop prior to the beginning of the dialysis treatment using a properly calibrated digital colorimeter.
The presence of ozone in the room will also be monitored using an ambient ozone sensor and/or ambient ozone detection badges. Visual and audible alarms will be given if the levels are above normal OSHA and EPA exposure values.
Intended Use: The TANGO3 Water Storage Tank with Ozone Disinfection System is intended to be used for disinfection of the water distribution system of a dialysis facility. The tank of the TANGO3 is also used as the water holding tank of the distribution system. The disinfection process is completely automated. Ozone concentration during disinfection is between 0.2 ppm and 0.3 ppm. Weekly disinfection cycles should be for four periods of 30 minutes with adequate water flushes between them and after the last cycle. In addition to the weekly disinfection cycle, a 45 minute disinfection cycle with at least two rinse cycles may also be performed multiple times weekly for additional reduction of bioburden as determined by user facility monitoring. At the end of disinfection, the distribution loop will have less than 0.1 mg/l. (0.1 ppm) of ozone, in accordance with AAMI/ISO 26722:2009 (4.2.13.5).
{6}------------------------------------------------
### Legally Marketed Devices to which Equivalence is Claimed:
- TANGO3 Water Storage Tank with Ozone Disinfection System K0903641 ■
Descriptive Summary of Technological Characteristics and Those of the Predicate Device: The features of TANGO3 are equivalent to the predicate device in terms intended use, technological characteristics, and operational characteristics as summarized in the following table:
| | PREDICATE DEVICE: | |
|----------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | TANGO3, LLC TANGO3 Water Storage | PROPOSED DEVICE: |
| PARAMETER | Tank with Ozone Disinfection System<br>K093641 | TANGO3, LLC TANGO3 Water Storage<br>Tank with Ozone Disinfection System |
| | | |
| Classification Name And<br>Product Code | Hemodialysis System and Accessories<br>FIN | Hemodialysis System and Accessories<br>FIN |
| Intended Use | The TANGO3 Water Storage Tank with<br>Ozone Disinfection System is intended to be<br>used for disinfection of the water<br>distribution system of a dialysis facility. | The TANGO3 Water Storage Tank with<br>Ozone Disinfection System is intended to be<br>used for disinfection of the water distribution<br>system of a dialysis facility. |
| Indications for Use | The TANGO3 Water Storage Tank with<br>Ozone Disinfection System is intended to be<br>used for disinfection of the water<br>distribution system of a dialysis facility. The<br>tank of the TANGO3 is also used as the<br>water holding tank of the distribution<br>system. The disinfection process is<br>completely automated. Ozone concentration<br>during disinfection is between 0.2 ppm and<br>0.3 ppm. The distribution system will be<br>exposed to ozone for one (1) period of 45<br>minutes and three (3) periods of 30 minutes<br>with adequate water flushes between them<br>and at the end, leaving the distribution loop<br>without ozone. | The TANGO3 Water Storage Tank with<br>Ozone Disinfection System is intended to be<br>used for disinfection of the water distribution<br>system of a dialysis facility. The tank of the<br>TANGO3 is also used as the water holding<br>tank of the distribution system. The<br>disinfection process is completely<br>automated. Ozone concentration during<br>disinfection is between 0.2 ppm and 0.3<br>ppm. Weekly disinfection cycles should be<br>for four periods of 30 minutes with adequate<br>water flushes between them and after the last<br>cycle. In addition to the weekly disinfection<br>cycle, a 45 minute disinfection cycle with at<br>least two rinse cycles may also be performed<br>multiple times weekly for additional<br>reduction of bioburden as determined by<br>user facility monitoring. At the end of<br>disinfection, the distribution loop will have<br>less than 0.1 mg/l. (0.1 ppm) of ozone, in<br>accordance with AAMI/ISO 26722:2009<br>(4.2.13.5). |
| | | |
| Materials in Contact With<br>Fluids | Stainless Steel, Teflon, PVC | Stainless Steel, Teflon, PVC, HDPE |
| Water Requirements:<br>Water Quality | AAMI Quality | AAMI Quality |
| Environmental Limits:<br>Ambient Temperature | 40° - 100° F (4.4° - 38° C) | 40° - 100° F (4.4° - 38° C) |
| Relative Humidity | 10 - 80% | 10 - 80% |
| Ambient Ozone | < 0.1ppm (OSHA) | < 0.1ppm (OSHA) |
| | | |
| Contact Time | 30 or 45 minute cycles | 30 or 45 minute cycles |
| Frequency | As required | As required |
| Used during | Non-operational hours | Non-operational hours |
| Rinsing | Automatic with water | Automatic with water |
| Alarms | Audible and visible indicators | Audible and visible indicators |
| Gas Feed | Dried air, with heat regenerative desiccants | Dried air, with heat regenerative desiccants |
| Ozone Generator | Corona discharge | Corona discharge |
| Injection System | Venturi based | Venturi based |
| Storage Tank | Stainless Steel or compatible with ozone | Stainless Steel or compatible with ozone |
| | | |
| Applicable Standards | AAMI RD 62 | AAMI RD 62 |
{7}------------------------------------------------
Performance Data: The following performance data were provided in support of the substantial equivalence determination. The results from these tests show that the TANGO3 performed as expected.
## Reliability Validation Testing
Reliability testing has been conducted to ensure that all components of the system function as designed for the intended life of the TANGO}. This testing included a validation of the software functions, including alarms. The system was tested to simulate 10 years of use.
The following table outlines the tests that were conducted as part of the reliability testing. All tests passed and demonstrated that the device works as intended.
| Test | Evaluation or Test |
|----------------------------|---------------------------------------------------------------------|
| O3 generation capability | • O3 level between 0.2 ppm and 0.3 ppm |
| O3 Calibration | • No variation greater than 5% |
| Valves and dissolved ozone | • No valve related alarms should be observed |
| destruct | • Level of ozone not above 0 ppm |
| Pump | • Loop velocity maintained at 4 ft/sec ± 10% |
| Ambient ozone | • Ambient ozone readings on both automated ozone detector and badge |
| FMEA | • Validate that a failure of critical components will be detected |
## Disinfection Validation
Testing was conducted to validate the disinfection performance of the ozone generator when used for water disinfection for dialysis and to determine the efficacy of O3 to disinfect water to below contamination action levels (bacteria and LAL). Overall, results of the disinfection validation show that the disinfection effectiveness of TANGO3 while keeping the water in the dialysis water distribution loop below AAMI RD 62:2006 action levels.
An overview of each test is provided in the table below.
| Test | Organisms Tested |
|---------------------------------------------------------|----------------------------------------------------------------------------------------------------------|
| Disinfection of Clinically Relevant Waterborne Bacteria | Acidovorax, Pseudomonas Aeruginosa, Burkholderia Cepacia, Brevundimonas Diminuta Mycobacterium Fortuitum |
| Normal Inoculation (Simulated Use) Testing | Acidovorax, Pseudomonas Aeruginosa, Burkholderia Cepacia |
| Biofilm Reduction and Prevention Capability Testing | Acidovorax Pseudomonas Aeruginosa Burkholderia Cepacia |
| Elimination of Enveloped and Non-enveloped Viruses | Poliovirus Type 1 Rotavirus Herpes Simplex Virus Type 1 |
| Fungi Elimination Efficacy | Candida Tropicallis Aspergillus |
## Field Testing
Field testing was performed in four (4) dialysis facilities with histories of water contamination. All four (4) facilities had documented water contamination issues and required periodic disinfection processes in order to maintain total bacterial counts and endotoxin levels below AAMI recommendations. This testing was conducted under a
{8}------------------------------------------------
standard protocol, which was reviewed and approved by the Institutional Review Board (IRB) of the facility at which the testing was conducted. The IRB determined that the testing was to be non-significant risk. Disinfection with the TANGO3 System was performed based on the clinic's policies and procedures. Results show that TANGO3 was able to maintain the water in the distribution loop below AAMI RD 62:2006 action levels.
## Software Verification and Validation Testing
Complete software validation, verification, and performance testing was performed to confirm that the features worked correctly and that the Programmable Logic Controller, the Human Machine Interface, and the Ozone systems hardware functioned together as expected. Documentation was provided as recommended by FDA's Guidance for Industry and Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered a "moderate" level of concern since, prior to mitigation of hazards, a failure of the software could result in Minor Injury, either to a patient or to a user of the device.
### Human Factors Testing
Human factors testing was conducted to verify that the user can safely operate the TANGO3. Different ranges to set up parameters were entered, and the readability of screens and messages was evaluated. Proper recognition of visual and audible alarms was also verified.
Conclusion: The information and data provided in this 510(k) Notification establish that the TANGO3 Water Storage Tank with Ozone Disinfection System is substantially equivalent to the legally marketed predicate device.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.