TANGO3 WATER STORAGE TANK WITH OZONE DISINFECTION SYSTEM
K093641 · Tango3, LLC · FIN · Dec 8, 2010 · Gastroenterology, Urology
Device Facts
Record ID
K093641
Device Name
TANGO3 WATER STORAGE TANK WITH OZONE DISINFECTION SYSTEM
Applicant
Tango3, LLC
Product Code
FIN · Gastroenterology, Urology
Decision Date
Dec 8, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5820
Device Class
Class 2
Attributes
Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K093641 · Dec 8, 2010
TANGO3 WATER STORAGE TANK WITH OZONE DISINFECTION SYSTEM
Tango3, LLC
Field test in dialysis facilities
A field test was conducted in dialysis facilities with known water contamination issues to validate the device's performance in a real-world clinical environment.
field test; dialysis facility; water contamination
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Field test; Field evaluation
Dialysis facilities with known water contamination issues
Not applicable for this study
Validation of device performance in addressing water contamination
Indications for Use
The TANGO3 Water Storage Tank with Ozone Disinfection System is intended to be used for disinfection of the water distribution system of a dialysis facility. The tank of the TANGO3 is also used as the water holding tank of the distribution system. The disinfection process is completely automated. Ozone concentration during disinfection is between 0.2 ppm and 0.3 ppm. The distribution system will be exposed to ozone for one (1) period of 45 minutes and three (3) subsequent periods of 30 minutes with adequate water flushes between them and at the end, leaving the distribution loop without ozone.
Device Story
TANGO3 Water Storage Tank with Ozone Disinfection System; functions as water holding tank and disinfection unit for hemodialysis facility water distribution loops. System uses corona discharge generator to produce ozone from dry air; venturi-based injection system introduces ozone into tank water. Centrifugal pump distributes ozonated water through loop during non-operational hours. Automated cycle: 45-minute ozone exposure followed by three 30-minute cycles with intermediate water flushes; final flush leaves system residual-free. Flow sensors at loop input/output prevent facility water usage during disinfection. Benefits include automated, effective water system disinfection to mitigate contamination. Operated by facility staff; output is a disinfected water distribution system.
Clinical Evidence
Bench testing only. In vitro testing validated disinfection capabilities against waterborne organisms. Material compatibility testing performed for all components in contact with ozone. Field testing validated system performance in facilities with known water contamination issues.
Technological Characteristics
System includes water storage tank, corona discharge ozone generator, heat regenerative desiccant air dryer, venturi injection system, centrifugal pump, and flow sensors. Automated disinfection cycle (0.2-0.3 ppm ozone). Materials tested for ozone compatibility. Standalone system for hemodialysis water loops.
Indications for Use
Indicated for disinfection of water distribution systems in hemodialysis facilities. Used as a water holding tank for the distribution system. Automated process; requires prescription use.
Regulatory Classification
Identification
A hemodialysis system and accessories is a device that is used as an artificial kidney system for the treatment of patients with renal failure or toxemic conditions and that consists of an extracorporeal blood system, a conventional dialyzer, a dialysate delivery system, and accessories. Blood from a patient flows through the tubing of the extracorporeal blood system and accessories to the blood compartment of the dialyzer, then returns through further tubing of the extracorporeal blood system to the patient. The dialyzer has two compartments that are separated by a semipermeable membrane. While the blood is in the blood compartment, undesirable substances in the blood pass through the semipermeable membrane into the dialysate in the dialysate compartment. The dialysate delivery system controls and monitors the dialysate circulating through the dialysate compartment of the dialyzer.(1) The extracorporeal blood system and accessories consists of tubing, pumps, pressure monitors, air foam or bubble detectors, and alarms to keep blood moving safely from the blood access device and accessories for hemodialysis (§ 876.5540) to the blood compartment of the dialyzer and back to the patient. (2) The conventional dialyzer allows a transfer of water and solutes between the blood and the dialysate through the semipermeable membrane. The semipermeable membrane of the conventional dialyzer has a sufficiently low permeability to water that an ultrafiltration controller is not required to prevent excessive loss of water from the patient's blood. This conventional dialyzer does not include hemodialyzers with the disposable inserts (Kiil type) (§ 876.5830) or dialyzers of high permeability (§ 876.5860). (3) The dialysate delivery system consists of mechanisms that monitor and control the temperature, conductivity, flow rate, and pressure of the dialysate and circulates dialysate through the dialysate compartment of the dialyzer. The dialysate delivery system includes the dialysate concentrate for hemodialysis (liquid or powder) and alarms to indicate abnormal dialysate conditions. This dialysate delivery system does not include the sorbent regenerated dialysate delivery system for hemodialysis (§ 876.5600), the dialysate delivery system of the peritoneal dialysis system and accessories (§ 876.5630), or the controlled dialysate delivery system of the high permeability hemodialysis system § 876.5860). (4) Remote accessories to the hemodialysis system include the unpowered dialysis chair without a scale, the powered dialysis chair without a scale, the dialyzer holder set, dialysis tie gun and ties, and hemodialysis start/stop tray.
{0}------------------------------------------------
# K093641 PAGE 1 OF 2
#### 510(K) SUMMARY
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.
## Submitter's Name: TANGO3, LLC
141A Citizens Blvd
Simpsonville, KY 40067
Telephone: (502) 722-8794
DEC 8 - 2010
Contact person: Hugh Doss, Manager
Date of Summary: October X, 2009
Device Name: TANGO3 Water Storage Tank with Ozone Disinfection System
Device Classification Name: Disinfectant, Dialysate Delivery System (876.5860, NII)
Device Description: The TANGO3 Water Storage Tank with Ozone Disinfection System is specifically designed to facilitate ozone induction into the storage tank, and then distribute the ozonated water through the distribution water loop during non-operational hours of a hemodialysis facility. The storage tank of the TANGOs system is filled with adequate water and the ozone concentration is increased. The ozonated water is distributed throughout the the disinfection process, the system and distribution system. To complete the disinfection process, the system and distribution system. To complete The described process is repeated three (3) more times. After the last cycle TANGO3 will leave the system residual free of ozone.
To accomplish this, TANGO3 has a corona discharge generator that generates ozone from a source of dry air. The air dryer consists of two heat regenerative desiccant modules. The ozone if om a injected into the tank by means of a venturi based injection system. Once in the tank, the ozonated water is sent to the distribution loop with a centrifugal pump. Ozone levels are monitored at the return of the loop. Two (2) flow sensors, located at the input and output of the distribution loop will assure that the dialysis facility is not utilizing water while ozone is present in the loop.
Intended Use: The TANGO3 Water Storage Tank with Ozone Disinfection System is intended to be used for disention of the water distribution system of a dialysis facility. The tank of the TANGO3 is also used as the water holding tank of the distribution system. The daintection process is completely automated. Ozone concentration during disinfection is between 0.2 ppm and 0.3 ppm. The distribution system will be exposed to ozone for one (1) period of 45 minutes and three (3) subsequent periods of 30 minutes with adequate water flushes between them and at the end, leaving the distribution loop without ozone.
{1}------------------------------------------------
Devices with equivalent intended use:
- Minneare® (Minntech Corporation) and other Peracetic Acid products such as Renalin® . (K830575) (Minntech Corporation) and Peracidin® (K962959) (AlcavisHDC) whilah are commonly used in hemodialysis clinics to disinfect the reverse osmosis water prification equipment and the water distribution loop
- Sodium hypochlorite at 500 ppm dilution is commonly used in dialysis facilities with . soak times of 30 minutes.
Devices with equivalent intended use, technological characteristics and operational characteristics:
- O3Z (K043207) The O3Z Ozone System is an optional accessory for the Solution 트 Delivery Systems (SDS) manufactured by GE Infrastructure, Water & Process Technologies, and is intended to be used for disinfection of the SDS bicarb mixing and distribution system. This device and the bicarb distribution system have equivalent technological features with TANGO3 and the water distribution loop, respectively. The 03Z Ozone System disinfects with ozone and then flushes the system with adequate water after disinfection. Disinfection must be performed during non-operational hours just as TANGO3.
Devices with equivalent intended use and operational characteristics:
- CWP 100 WRO H (K974899) (heat disinfection unit for the distribution loop only) -. The CWP 100 - WRO H has a heat disinfection unit that produces enough heat to disinfect the water in the water distribution loop. This disinfection process is automatic and usually performed during non-operational hours. The device has automatic flushing programs to be used when the system is not in use just as TANGO2.
Descriptive Summary of Technological Characteristics and Those of Predicate Devices: The features of TANGO3 are equivalent to those of other medical devices, systems or disinfectants currently in distribution in the United States with equivalent intended uses, on technological characteristics, and operational characteristics.
Performance Data: Each function of the TANGO3 System was tested to see if it performed as intended. Any errors or failures detected during testing were corrected. In vitro testing was also performed to validate the disinfection capabilities of TANGO3 with waterborne organisms. All materials found in water distribution loops have been tested for material compatibility with ozone. A field test has been performed to validate that TANGO2 can address the needs of facilities with known water contamination issues. The results from these tests show that the TANGO3 performed as expected.
Conclusion: The information and data provided in this 510(k) Notification establish that the TANGO3 Water Storage Tank with Ozone Disinfection System is substantially equivalent to the legally marketed predicate devices.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes entwined around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002
.
Lori Kleinschrodt Holder, RAC Regulatory Affairs Consultant TANGO3, LLC 141A Citizens Blvd. SIMPSONVILLE KY 40067
DEC 8 2010
Re: K093641
Trade/Device Name: TANGO3 Water Storage Tank with Ozone Disinfection System Regulation Number: 21 CFR §876.5820 Regulation Name: Hemodialysis system and accessories Regulatory Class: II Product Code: FIN Dated: November 26, 2010 Received: November 30, 2010
Dear Ms. Holder:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related
{3}------------------------------------------------
#### Page 2
adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Herbert P. Lerner, M.D., Director (Acting) Division of Reproductive, Gastro-Renal and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
#### I. INDICATIONS FOR USE STATEMENT
November 23, 2010
Page 1 of 1
### 510(k) Number: K093641
Device Name: TANGO3 Water Storage Tank with Ozone Disinfection System
Indications for Use: The TANGO, Water Storage Tank with Ozone Disiafection System is intended to be used for disinfection of the water distribution system of a dialysis facility. The tank of the TANGO, is also used as the water holding tank of the distribution system. The disinfection process is completely automated. Ozone concentration during disinfection is between 0.2 ppm and 0.3 ppm. The distribution system will be exposed to ozone for one (1) period of 45 minutes and three (3) subsequent periods of 30 minutes with adequate water the (1) perfod of 45 milliurs and intee (1) subsequent periods of 30
ozone. ozone.
(Concurrence of CDRH, Office of Device Evaluation (ODE)
Heilut Reum
(Division Sign-Off)
Division of Reproductive, Gastro-Renal, and
ological Devices
510(k) Number K093641-
Prescription Use X (Per 21 CFR 801.109)
OR
Over-the-Counter Use
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.