DOVER SILVER HYDROGEL COATED SILICONE FOLEY CATHETER

K024010 · The Kendall Company Div. of Tyco Healthcare Group · EZL · Jun 11, 2003 · Gastroenterology, Urology

Device Facts

Record IDK024010
Device NameDOVER SILVER HYDROGEL COATED SILICONE FOLEY CATHETER
ApplicantThe Kendall Company Div. of Tyco Healthcare Group
Product CodeEZL · Gastroenterology, Urology
Decision DateJun 11, 2003
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.5130
Device ClassClass 2
AttributesTherapeutic

Intended Use

The proposed device is intended for use in the drainage and/or collection and/or measurement of urine. Generally drainage is accomplished by inserting the catheter through the urethra and into the bladder. However, drainage is sometimes accomplished by suprapubic or other placement of the catheter, such as a nephrostomy tract.

Device Story

Sterile, single-patient-use urinary drainage catheter; extruded from 100% vulcanized silicone; coated with lubricious hydrophilic topcoat containing inorganic silver-releasing polymer. Inserted into bladder via urethra, suprapubic, or nephrostomy tract. Facilitates urine drainage, collection, or measurement. Used in clinical settings. No claims regarding microbial adherence reduction or infection prevention permitted.

Clinical Evidence

Bench testing only. Biocompatibility testing performed in accordance with ISO 10993, Part 1, and FDA memorandum G95-1.

Technological Characteristics

Material: 100% vulcanized silicone. Coating: Lubricious hydrophilic topcoat with inorganic silver-releasing polymer. Form factor: Foley catheter. Sterilization: Sterile. Biocompatibility: ISO 10993-1 compliant.

Indications for Use

Indicated for patients requiring urinary drainage, collection, or measurement via urethral, suprapubic, or nephrostomy tract placement.

Regulatory Classification

Identification

A urological catheter and accessories is a flexible tubular device that is inserted through the urethra and used to pass fluids to or from the urinary tract. This generic type of device includes radiopaque urological catheters, ureteral catheters, urethral catheters, coudé catheters, balloon retention type catheters, straight catheters, upper urinary tract catheters, double lumen female urethrographic catheters, disposable ureteral catheters, male urethrographic catheters, and urological catheter accessories including ureteral catheter stylets, ureteral catheter adapters, ureteral catheter holders, ureteral catheter stylets, ureteral catheterization trays, and the gastro-urological irrigation tray (for urological use).

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ JUN 1 1 2003 ### 510(k) Summary ## Kendall DOVER Silver Hydrogel Coated Silicone Foley Catheter In accordance with section 513(I) of the SMDA and as defined in 21 CFR Part 807.3 final rule dated December 14, 1994, this summary is submitted by: Tyco Healthcare/Kendall 15 Hampshire Street Mansfield, MA 02048 Date Prepared: November 20, 2002 - 1. Contact Person David A. Olson Vice President, Regulatory Affairs (508) 261-8530 #### 2. Name of Medical Device Urological catheter and accessories Classification Name: Urinary Drainage Catheter, Silicone Foley Common or Usual Name: #### 3. Identification of Legally Marketed Device The proposed Kendall DOVER Silver Hydrogel Coated Silicone Foley Catheter is substantially equivalent in intended use, function and mode of operation to the DOVER Silicone Foley Catheter, which was marketed prior to May 28, 1976. #### 4. Device Description The Kendall DOVER Silver Hydrogel Coated Silicone Foley Catheter is a sterile, single patient use, urinary drainage catheter extruded from 100% vulcanized silicone material. It is coated with a lubricious hydrophilic topcoat containing an inorganic silver releasing polymer. #### 5. Device Intended Use The proposed device is intended for use in the drainage and/or collection and/or measurement of urine. Generally drainage is accomplished by inserting the catheter through the urethra and into the bladder. However, drainage is sometimes accomplished by suprapubic or other placement of the catheter, such as a nephrostomy tract. {1}------------------------------------------------ K024010 Page 2 of 2 ### Product Comparison The Kendall DOVER Silver Hydrogel Coated Silicone Foley Catheter is substantially equivalent to the Kendall DOVER Foley Catheter predicate device as each product is intended to pass fluids to or from the urinary tract. ### Nonclinical Testing Biocompatibility testing of the proposed device has demonstrated that it meets the requirements of guidelines presented in the 10993 ISO Standard, Part 1, with the FDA modified matrix presented in memorandum G95-1. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter of the circle. Inside the circle is a stylized symbol that resembles three curved lines, possibly representing a stylized human figure or a symbol of health and well-being. The logo is simple and clean, with a focus on the text and the central symbol. Public Health Service JUN 1 1 2003 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Mr. David A. Olson Vice President, Regulatory Affairs Tyco Healthcare 15 Hampshire Street MANSFIELD MA 02048 Re: K024010 Trade/Device Name: Kendall Dover Silver Hydrogel Coated Silicone Foley Catheter Regulation Number: 21 CFR §876.5130 Regulation Name: Urological catheter and accessories Regulatory Class: II Product Codes: 78 EZL and KOD Dated: March 13, 2003 Received: March 14, 2003 Dear Mr. Olson: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). Specifically, your device is substantially equivalent to Kendall DOVER Silicone Foley Catheter. Given that you have not provided any microbial adherence data, you may not make any claims regarding the effectiveness of the inorganic silver ion releasing polymer to reduce microbial adherence to the device or to reduce urinary tract infections in the patients treated with the device. You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ Page 2 - Mr. David Olson Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820): and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to continue marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter. | 8xx.1xxx | (301) 594-4591 | |----------------------------------|----------------| | 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 | | 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 | | 892.2xxx, 3xxx, 4xxx, 5xxx, | (301) 594-4654 | | Other | (301) 594-4692 | Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at their toll free number (800) 638-2041 or at (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html. Sincerely yours, Nancy C. Brogdon Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # Indications for Use Statement KO24010 510(k) Number (if known): Device Name: Kendall Dover Silver Hydrogel Coated Silicone Foley Catheter ### Indications for Use: The proposed device is intended for use in the drainage and/or collection and/or measurement of urine. Generally drainage is accomplished by inserting the catheter through the urethra and into the bladder. However, drainage is sometimes accomplished by suprapubic or other placement of the catheter, such as a nephrostomy tract. Please Do Not Write Below This Line - Continue On Another Page If Needed Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use Use (Per 21 CFR 801.109) OR Over-The-Counter Caulyn Y. Neuland for N. Brogdon (Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number K024010
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...