K122785 · Rochester Medical Corp. · EZD · May 14, 2013 · Gastroenterology, Urology
Device Facts
Record ID
K122785
Device Name
HYDROSIL; MAGIC; PERSONAL CATHETER
Applicant
Rochester Medical Corp.
Product Code
EZD · Gastroenterology, Urology
Decision Date
May 14, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5130
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
For urological use only. Intended for use by patients for bladder management including urine drainage, collection and measurement. The devices are passed to the urinary bladder via the urethra.
Device Story
Hydrophilic Intermittent Urinary Catheter; single-lumen silicone elastomer tube; proximal drainage eyes; distal tapered funnel; outer hydrophilic coating binds water to create lubricating film; includes silicone handle for ergonomic grip and no-touch insertion. Used by patients for bladder management; facilitates urine drainage, collection, and measurement. Device passed through urethra into bladder. Supplied sterile for single use.
Clinical Evidence
Bench testing only. Evidence includes design qualification (dimensional, appearance, functional testing per EN 1616; lubricity per modified ASTM D1894), long-term stability/shelf-life testing, and biocompatibility (ISO 10993-5 cytotoxicity, ISO 10993-10 sensitization, ISO 10993-10 vaginal irritation).
Technological Characteristics
All-silicone single-lumen drainage tube; hydrophilic outer coating; silicone handle. Dimensions/form factor: proximal drainage eyes, distal tapered funnel. Sterilization: supplied sterile for single use. Biocompatibility standards: ISO 10993-5, ISO 10993-10. Functional standards: EN 1616, modified ASTM D1894.
Indications for Use
Indicated for patients requiring bladder management, including urine drainage, collection, and measurement via urethral catheterization.
Regulatory Classification
Identification
A urological catheter and accessories is a flexible tubular device that is inserted through the urethra and used to pass fluids to or from the urinary tract. This generic type of device includes radiopaque urological catheters, ureteral catheters, urethral catheters, coudé catheters, balloon retention type catheters, straight catheters, upper urinary tract catheters, double lumen female urethrographic catheters, disposable ureteral catheters, male urethrographic catheters, and urological catheter accessories including ureteral catheter stylets, ureteral catheter adapters, ureteral catheter holders, ureteral catheter stylets, ureteral catheterization trays, and the gastro-urological irrigation tray (for urological use).
Predicate Devices
Rochester Medical Corporation Personal® Catheter (Hydrophilic) (K000723)
{1}------------------------------------------------
K122785 Page 1 of 2
Image /page/0/Picture/1 description: The image shows the logo for Rochester Medical. The logo consists of a circle on the left side, with a line extending from the circle to the right, forming a mountain-like shape. The text "Rochester" is written in a stylized font, with the word "MEDICAL" written in smaller letters below it.
## MAY 1 4 2013
#### 5.0 510(k) Summary
510(k) Submitter:
Rochester Medical Corporation One Rochester Medical Drive Stewartville, MN 55976 USA Phone: (507) 533-9600 Fax: (507) 533-9740
Contact Name:
Robert Anglin Rochester Medical Corporation One Rochester Medical Drive Stewartville, MN 55976 Phone: (507) 533-9613 Fax: (507) 533-9740
Date prepared:
August 28, 2012
Trade or Proprietary Name: HydroSil, Magic, Personal Catheter
Common or Usual Name:
Urological Catheter
Urological catheter and accessories (21 CFR 876.5130, Product Code EZD)
Predicate Device(s):
Classification Name:
K000723 Rochester Medical Corporation Personal® Catheter (Hydrophilic)
K072808 Coloplast Corp. SpeediCath Compact
#### Device Description:
The Hydrophilic Intermittent Urinary Catheter is made from silicone elastomer with a hydrophilic coating. The device is for urological use only. It is intended for use by patients for bladder management including urine drainage, collection, and measurement.
The device consists of an all silicone elastomer single lumen catheter and includes a silicone handle. The catheter has drainage eyes located in the proximal tip and a tapered funnel located at the distal end. The outer surface of the all silicone catheter has a hydrophilic coating which binds water molecules to the surface creating a smooth lubricating film. The handle provides the patient-user an ergonomically designed area for a secure grip and notouch area when inserting the catheter.
{2}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for Rochester Medical. The logo features a stylized mountain range above the word "Rochester" in a bold, italicized font. Below "Rochester" is the word "MEDICAL" in a smaller, sans-serif font. To the left of the word "Rochester" is a circle with a line extending from the left side of the circle to the mountain range.
#### Intended Use of the Device:
For urological use only. Intended for use by patients for bladder management including urine drainage, collection and measurement. The devices are passed to the urinary bladder via the urethra.
### Technological Characteristics:
The device described in the 510(k) is an all silicone single lumen drainage tube with drainage eyes at the proximal end and a tapered funnel at the distal end. The predicate devices are also made from similar materials such as silicone and polyurethane. The 510(k) device has an outer hydrophilic coating which is hydrated with sterile water. The predicate devices also have an outer hydrophilic coating, which is hydrated in the similar manner such as sterife water and saline solution. The 510(k) device has a handle designed to facilitate ease of catheter insertion. The predicate devices may also have a handle. All of the devices are supplied sterile for single use.
### Non-Clinical Performance Data
The following non-clinical performance data demonstrates substantial equivalence of the 510(k) device to the predicate devices:
#### Design Oualification
- Hydrophilic Intermittent Catheter with a Handle VAL439, TR1691, This non-. clinical test data includes: dimensional analysis, appearance, and functional testing according to EN 1616. Lubricity testing was conducted according to modified ASTM D1894.
#### Long Term Stability/Shelf Life
- Hydrophilic Intermittent Catheter with a Handle LTS096. This non-clinical test . data includes: dimensional analysis, appearance, and functional testing according to EN 1616.
#### Biocompatibility
- ISO 10993-5, Cytotoxicity .
- ISO 10993-10, Sensitization GPMT (Klingman), 2 extracts .
- ISO 10993-10, Vaginal Irritation w/Histopathology, 2 extracts ●
#### Conclusion:
Substantial equivalence of the Hydrophilic Intermittent Urinary Catheter is supported by a comparison of the design and intended use compared to the predicate devices which received marketing clearance under Premarket Notification, as well as acceptable results from functional performance and biocompatibility testing.
{3}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the caduceus symbol. The logo is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 14, 2013
Rochester Medical Corporation % Mr. Robert Anglin Vice President of Quality and Regulatory One Rochester Medical Drive STEWARTVILLE MN 55976
Re: K122785
Trade/Device Name: Hydrophilic Intermittent Urinary Catheter Regulation Number: 21 CFR§ 876.5130 Regulation Name: Urological catheter and accessories Regulatory Class: II Product Code: EZD Dated: April 11, 2013 Received: April 19, 2013
Dear Mr. Anglin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{4}------------------------------------------------
Page 2 - Mr. Robert Anglin
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
Herbert P. Lerner -S
for
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation ' Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
#### 4.0 INDICATIONS FOR USE STATEMENT
K122785 510(k) Number (if known):
Hydrophilic Intermittent Urinary Catheter DEVICE NAME:
INDICATIONS FOR USE:
For urological use only. Intended for use by patients for bladder management including urine drainage, collection and measurement. The devices are passed to the urinary bladder via the urethra.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
# Herbert | Pagerner -S
(Division Sign-Off) Division of Reproductive, Gastro-Renal, and Urological Devices 510(k) Number ________________________________________________________________________________________________________________________________________________________________ K122785
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.