COMPAT NASOJEJUNAL FEEDING TUBE

K961664 · Sandoz Nutrition · KNT · Oct 4, 1996 · Gastroenterology, Urology

Device Facts

Record IDK961664
Device NameCOMPAT NASOJEJUNAL FEEDING TUBE
ApplicantSandoz Nutrition
Product CodeKNT · Gastroenterology, Urology
Decision DateOct 4, 1996
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.5980
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Sandoz Nutrition #082302 NFG Tube is intended for use with patients who require both enteral feeding and gastric suction simultaneously. The Sandoz Nutrition #082304 NF Tube is intended for use with patients who require only enteral feeding. Indications for such medical and post-surgical treatments may include surgical problems, prolonged artificial ventilation, central nervous system trauma, and others. The placement of the device in patients can be accomplished intraoperatively at laparotomy using standard surgical techniques.

Device Story

Nasojejunal feeding tubes (with/without gastric drainage) designed for intraoperative placement via laparotomy. Features include radiopaque bolus tip, grasping rings, gastric bushing, and tubing stripe to facilitate surgical palpation and radiographic confirmation of placement. Used by surgeons in clinical/surgical settings to deliver enteral formula post-pylorically and/or evacuate gastric contents. Benefits include simultaneous feeding and drainage capabilities for patients with complex medical or post-surgical needs.

Clinical Evidence

Bench testing only; no clinical data provided.

Technological Characteristics

Materials identical to predicate devices. Features radiopaque bolus tip, grasping rings, 9/18 FR gastric bushing, and tubing stripe for radiographic visualization. Designed for intraoperative placement. No energy source or software.

Indications for Use

Indicated for patients requiring enteral feeding and/or gastric suction due to conditions including surgical problems, prolonged artificial ventilation, or central nervous system trauma.

Regulatory Classification

Identification

A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).

Special Controls

*Classification.* (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} K961664 OCT 4 1996 510(K) SUMMARY Submitter Name: Sandoz Nutrition Corporation P.O. Box 370 Minneapolis, MN 55440 Robert J. Lang Director of Quality Operations 612-591-2950 April 22, 1996 Device Name : Classification Name: Gastrointestinal Tube and Accessories (21 CFR 876.5980) Common/Usual Name: Nasojeunal Feeding Tube with or without Gastric Drainage. Trade Name: Sandoz Nutrition COMPAT Nasojejunal Feeding Tube. Identification of the Legally Marketed Device The #082302 NASOJEJUNAL FEEDING TUBE WITH GASTRIC DRAINAGE AND the #082304 NASOJEJUNAL FEEDING TUBE FOR SURGICAL PLACEMENT are Substantially Equivalent (SE) to the Sandoz Nutrition #082301 Nasojejunal Feeding Tube With Gastric Drainage and the Sandoz Nutrition #082303 Nasojejunal Feeding Tube legally marketed under 510(k): K913407/B Description of the device that is the subject of this 510(K). A Nasojejunal Feeding Tube with or without Gastric Drainage. Statement of intended use of the device. The Sandoz Nutrition #082302 NFG Tube is intended for use with patients who require both enteral feeding and gastric suction simultaneously. The Sandoz Nutrition #082304 NF Tube is intended for use with patients who require only enteral feeding. Indications for such medical and post-surgical treatments may include surgical problems, prolonged artificial ventilation, central nervous system trauma, and others. The placement of the device in patients can be accomplished intraoperatively at laparotomy using standard surgical techniques. {1} 46160- 2042 # Summary of technological characteristics in comparison to those of the predicate device. The tubes of this submission and the predicate device are both intended to provide means for delivering enteral formula to the intestine (i.e., post pylorically). The #082302 tube, also provide the means for gastric evacuation or drainage which is the same intended use of the predicate device #082301 The Placement technique and confirmation methods differ slightly from the predicate device. The devices in this submission are designed to facilitate placement into a patient intraoperatively using standard surgical techniques at laparotomy. The predicate devices may be placed using fluoroscopic and endoscopic guidewire techniques; however, these tubes can also be placed surgically, depending on physician preference. Confirmation of placement position for all of the tubes above may be performed using radiographic techniques since certain location features are radiopaque (i.e., Bolus Tip, Grasping Rings, 9/18 FR Gastric Bushing, and tubing stripe). The feeding tubes of this submission are nearly identical to the currently marketed tubes with the exception that **grasping rings and bolus tips** are added to the 9 French feeding portion of the tubes. The grasping rings located at the distal end of the tube are designed to assist in surgical palpation on manipulation of the tube tip during placement of the tube. The feeding tubes and the stylets/stiffener of this submission are made of identical materials to those used in the currently marketed SE devices. The lengths, diameters, gastric suction holes and Y connectors of the devices of this submission are identical to those of the currently marketed SE devices. # Conclusions Drawn that demonstrate that the device is safe, as effective and performs as well as the legally marketed device. Sandoz Nutrition believes that the changes described in this submission do not negatively affect the safety or effectiveness of the product. The material and bonding method for the grasping rings and bolus tip are identical to those used in the SE device. Furthermore, there have been no complaints with respect to the materials in the 5 years that the SE device has been marketed.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...