K183336 · Wilson-Cook Medical, Inc. · KNT · May 29, 2019 · Gastroenterology, Urology
Device Facts
Record ID
K183336
Device Name
Entuit PEG, Entuit PEGJ
Applicant
Wilson-Cook Medical, Inc.
Product Code
KNT · Gastroenterology, Urology
Decision Date
May 29, 2019
Decision
SESK
Submission Type
Traditional
Regulation
21 CFR 876.5980
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Entuit PEG is intended for percutaneous endoscopic gastrostomy placement to provide enteral nutrition and/or medication to patients requiring nutritional support, and gastric decompression. This device is indicated for adult use only. The Entuit PEGJ is intended for percutaneous endoscopic placement to provide enteral nutrition and/or medication to patients requiring nutritional support, and gastric decompression. This device is indicated for adult use only.
Device Story
Entuit PEG/PEGJ are gastrointestinal feeding tubes for enteral nutrition, medication delivery, and gastric decompression. PEG tubes (silicone, 20-24 Fr) are placed percutaneously; PEGJ tubes (polyurethane, 9-12 Fr) are inserted through existing PEG tubes to access the jejunum. Kits include site preparation components, introduction tools, and feeding adapters (ENFit or non-ENFit compatible). Used in clinical settings by healthcare professionals. Output is the delivery of nutrition/medication or decompression of the stomach. Benefits include reliable access for patients unable to ingest food orally. Device design facilitates both push and pull placement techniques.
Clinical Evidence
Bench testing only. Testing included functional validation, tensile testing, leak testing, shelf life, packaging validation, enteral connector misconnection, ENFit performance, biocompatibility (ISO 10993-1), flow rate, radiopacity, corrosion, bolster separation forces, and bolster pull-through force testing.
Technological Characteristics
Materials: Silicone (PEG), Polyurethane (PEGJ). Energy: None (mechanical). Connectivity: None. Sterilization: Sterile. Feeding adapters comply with ISO 80369-1 and ISO 80369-3 (ENFit).
Indications for Use
Indicated for adult patients requiring enteral nutrition, medication administration, or gastric decompression via percutaneous endoscopic gastrostomy or jejunostomy placement.
Regulatory Classification
Identification
A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).
Special Controls
*Classification.* (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a bold, sans-serif font and the full name in a smaller, sans-serif font below it. The overall design is clean and professional, reflecting the agency's role in regulating food and drugs in the United States.
May 29, 2019
Wilson-Cook Medical, Inc. Tiffany A. Thomas Global Regulatory Affairs Specialist 4900 Bethania Station Road Winston-Salem, NC 27105
Re: K183336
> Trade/Device Name: Entuit PEG, Entuit PEGJ Regulation Number: 21 CFR\$ 876.5980 Regulation Name: Gastrointestinal Tube and Accessories Regulatory Class: II Product Code: KNT, PIF Dated: April 25, 2019 Received: April 29, 2019
Dear Tiffany A. Thomas:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. However, you are responsible to determine that the medical devices you use as components in the kit have either been determined as substantially equivalent under the premarket notification process (Section 510(k) of the act), or were legally on the market prior to May 28, 1976, the enactment date of the Medical Device Amendments. Please note: If you purchase your device components in bulk (i.e., unfinished) and further process (e.g., sterilize) you must submit a new 510(k) before including these components in your kit/tray. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
for Shani P. Haugen, Ph.D. Acting, Assistant Division Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K183336
Device Name Entuit PEG Entuit PEGJ
Indications for Use (Describe)
The Entuit PEG is intended for percutaneous endoscopic gastrostomy placement to provide enteral nutrition and/or medication to patients requiring nutritional support, and gastric decompression. This device is indicated for adult use only.
The Entuit PEGJ is intended for percutaneous endoscopic placement to provide enteral nutrition and/or medication to patients requiring nutritional support, and gastric decompression. This device is indicated for adult use only.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------|----------------------------------------------------------------------------|
| <div> <span> </span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> <span> </span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Summary
Entuit PEG /Entuit PEGJ
Traditional 510(k) Premarket Notification November 30, 2018
#### Applicant Information
| Applicant: | Wilson-Cook Medical, Inc. /Cook Endoscopy<br>4900 Bethania Station Road<br>Winston-Salem, North Carolina 27105 |
|------------|----------------------------------------------------------------------------------------------------------------|
| Contact: | Tiffanny A. Thomas, Global Regulatory Affairs Specialist |
| Phone: | (336) 744-0157 |
| Fax: | (336) 201-5994 |
#### Device Information
| Trade Name: | Entuit PEG/ Entuit PEGJ |
|----------------------|----------------------------------------|
| Common Name: | Gastrointestinal Feeding Tube |
| Classification Name: | Gastrointestinal tubes and accessories |
| Regulation Number: | 21 CFR 876.5980 |
| Product Code: | KNT, PIF |
| Device Class: | Class II |
| Review Panel: | Gastroenterology/Urology |
## Predicate Device
K920703 Gastrointestinal Feeding Tube (PEG 24 Fr/ PEGJ 12 Fr)
## Device Description
The Entuit PEG includes a silicone enteral feeding tube (PEG tube) available in 24 Fr or 20 Fr, with site preparation components, introduction components, adapters. The PEG tube is supplied sterile and is intended for single use in adults. Entuit PEGJ includes a polyurethane enteral feeding tube (PEGJ tube) that is available in 12 Fr or 9 Fr. The Entuit PEGJ tube is available for direct access to the jejunum. The 12 Fr PEGJ tube is designed to be inserted through the 24 Fr PEG tube and the 9 Fr tube is designed to be inserted through the 20 Fr PEG tube such that the distal end extends into the jejunum and the proximal end extends out of the existing PEG tube.
There are several feeding adapters for use with the PEG and PEGJ tubes. The PEG kit has two feeding adapters that are ENF it compatible and two feeding adapters that are compatible with Christmas tree or slip tip giving sets and syringes (Non-ENFit). The PEGJ tubes are available with either ENFit compatible or Christmas tree compatible feeding adapters (Non-ENFit). The Entuit PEG is provided as a kit to facilitate either push or pull techniques of placement. In addition, the Entuit PEG are available in standard, safety, and international versions. The PEGJ tubes are sold as a separate kit.
{4}------------------------------------------------
# Intended Use
The Entuit PEG is intended for percutaneous endoscopic gastrostomy placement to provide enteral nutrition and/or medication to patients requiring nutritional support, and gastric decompression. This device is indicated for adult use only.
The Entuit PEGJ is intended for percutaneous endoscopic placement to provide enteral nutrition and/or medication to patients requiring nutritional support, and gastric decompression. This device is indicated for adult use only.
# Comparison to Predicate Device
The subject Entuit PEG and Entuit PEGJ are substantially equivalent to the predicate device, the Gastrointestinal Feeding Tube (K920703) with respect to the indications for use, methods of placement, designs, and fundamental technological characteristics. The minor differences between the subject devices and the predicate devices include: feeding tube outer diameter, updated feeding adapters to meet ISO 80369-1 and ISO 80369-3 with ENFit connector design, dilator design, and packaging configuration.
# Summary of Non-Clinical Testing:
The following non-clinical testing was conducted to demonstrate the performance of the subject device and confirmed that the subject device performs as intended.
- Functional Validation
- Tensile Testing
- Leak testing
- Shelf Life Testing ●
- . Packaging Validation
- Enteral Connector Misconnection
- ENFit Performance Testing
- Feeding Adapters Tensile Testing ●
- Biocompatibility Testing ●
- Flow Rate ●
- Radiopacity ●
- Corrosion
- Bolster Separation Forces
- Force required to Pull the Bolster Through the Stoma Testing ●
{5}------------------------------------------------
Biocompatibility testing was performed in accordance with the FDA Guidance, Use of International Standard ISO 10993-1 "Biological evaluation of medical devices-Part 1: Evaluation and testing within a risk management process."
## Conclusions:
We believe that the subject device is substantially equivalent to the predicate device in terms of intended use, key operating principles, materials and technological characteristics. We consider the risks associated with the modifications to the subject device to have been adequately addressed through our Design Control Processes and do not affect safety or effectiveness of the device.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.