The COMPAT® Surgical Gastrostomy/Jejunostomy Feeding Tube Kit is intended for use as a surgically place gastrostomy tube intended to deliver enteral feeding formulas directly into the small bowel (jejunum). The COMPAT® Surgical Gastrostomy/Jejunostomy Feeding Tube is intended for use with patients who require enteral feeding. Indications for such medical and post-surgical treatments may include surgical problems, prolonged artificial ventilation, central nervous system trauma, and others. The placement of the device in patients is accomplished surgically using standard surgical techniques. This method is used to enhance the absorption of nutrients and minimize the risks of gastric feeding including gastric reflux and aspiration. The COMPAT® Surgical Gastrostomy/Jejunostomy Feeding Tube also provides for the evacuation of gastric contents through the gastrostomy tube while simultaneously feeding directly into the small bowel with the jejunal feeding tube. This device is for use under medical supervision only.
Device Story
Sterile kit for intra-operative placement of gastrostomy/jejunostomy feeding tubes; enables simultaneous enteral feeding into jejunum and gastric decompression/drainage. System includes 28 FR silicone gastrostomy tube, 12 FR polyurethane jejunal feeding tube with double-lumen adapter, and disposable balloon-tipped stylet. Physician performs surgical placement via laparotomy/Stamm technique; jejunal tube manipulated through gastrostomy lumen into small bowel using stylet and balloon tip. External retainer immobilizes assembly. Output is delivery of enteral formula via feeding pump and evacuation of gastric contents. Benefits include improved nutrient absorption and reduced aspiration risk. Used in clinical/surgical settings under medical supervision.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Kit includes 28 FR silicone gastrostomy tube and 12 FR polyurethane jejunal feeding tube. Features radiopaque, flow-through tip with 20% barium sulfate filler. Includes disposable balloon-tipped stylet (natural rubber latex). Mechanical system; no energy source or software. Sterilized kit.
Indications for Use
Indicated for patients requiring enteral feeding past the stomach directly into the jejunum, including those with surgical problems, prolonged artificial ventilation, or central nervous system trauma, to enhance nutrient absorption and minimize risks of gastric reflux and aspiration.
Regulatory Classification
Identification
A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).
Special Controls
*Classification.* (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.
{1}------------------------------------------------
Kozy 91 1044
# 510(k) Summary (per 21 CFR 807.87(h))
JUL 2 3 2007
| Common/Usual Name: | Gastrostomy/Jejunostomy Feeding Tube Kit |
|-----------------------------|----------------------------------------------------------------------------------------------------------------------------------|
| Product Trade Name: | COMPAT® Surgical Gastrostomy/Jejunostomy<br>Feeding Tube Kit, Model #088000 |
| Classification Name: | Tubes, Gastrointestinal (and Accessories)<br>Class II per 21 CFR §876.5980<br>Product Code KNT |
| Predicate Device: | Novartis Item 088000, COMPAT® Surgical<br>Gastrojejunostomy Feeding Tube Kit, (K965087) |
| Manufacturer: | Novartis Nutrition Corporation<br>1541 Park Place Blvd<br>St. Louis Park, MN 55416 |
| Establishment Registration: | 2110851 |
| Contact: | Sue Fidler,<br>Manager, Regulatory Affairs NA<br>Telephone: 952-848-6328 / Fax: 952-848-6319<br>Email: susan.fidler@novartis.com |
| Performance Standards: | No performance standards have been developed<br>under section 514 for this device. |
| Date Prepared: | April 21, 2006 |
## Device Description:
The Novartis Nutrition COMPAT® Surgical Gastrostomy/Jejunostomy Feeding Tube is packaged in a sterile kit. It is designed for easy intra-operative placement and facilitates immediate postoperative feeding into the jejunum. The two-tube system allows for easy placement and replacement in the event of a jejunal tube clog. The large gastrostomy tube provides access to the stomach for decompression and
{2}------------------------------------------------
drainage. The system's jejunal feeding tube is easy to place intra-operatively through the gastrostomy tube into the small bowel using the innovative balloon tipped stylet provided in the kit.
The major components of the kit include:
- . 28 FR Silicone Gastrostomy Tube
- 12 FR Polyurethane Jejunal Feeding Tube with Double Lumen FEED/DRAIN . adapter and radiopaque, flow-through tip
- Preassembled disposable balloon tipped stylet1 .
The COMPAT administration set contains or attaches to the formula reservoir/container and provides the conduit to deliver the feeding solution to the patient's feeding tube via a feeding pump. The patient's feeding tube may be a variety of tubes including, but not limited to, a nasogastic tube, a gastrostomy tube, or a jejunostomy tube. The administration set attaches to the patient's feeding tube to provide the formula or feeding solution required.
## Statement of Intended Use:
The COMPAT® Surgical Gastrostomy/Jejunostomy Feeding Tube Kit is intended for use as a surgically place gastrostomy tube intended to deliver enteral feeding formulas directly into the small bowel (jejunum). The COMPAT® Surgical Gastrostomy/Jejunostomy Feeding Tube is intended for use with patients who require enteral feeding. Indications for such medical and post-surgical treatments may include surgical problems, prolonged artificial ventilation, central nervous system trauma, and others. The placement of the device in patients is accomplished surgically using standard surgical techniques. This method is used to enhance the absorption of nutrients and minimize the risks of gastric feeding including gastric reflux and aspiration.
The COMPAT® Surgical Gastrostomy/Jejunostomy Feeding Tube also provides for the evacuation of gastric contents through the gastrostomy tube while simultaneously feeding directly into the small bowel with the jejunal feeding tube. This device is for use under medical supervision only.
## Indications for Use:
The COMPAT® Surgical Gastrostomy/Jejunostomy Feeding Tube Kit provides for the delivery of enteral formula through a surgically placed Gastrostomy/Jejunostomy feeding tube. Jejunal feeding is used when patients require feeding past the stomach directly
<sup>1</sup> The balloon on the disposable stylet is made from natural rubber latex. The stylet is removed after placement.
{3}------------------------------------------------
into the jejunum. This method is used to enhance the absorption of nutrients and minimize the risks of gastric feeding including gastric reflux and aspiration.
# Summary of technological characteristics in comparison to those of the predicate device.
#### 1. Intestinal feeding and gastric drainage.
The COMPAT® Surgical Gastrostomy/Jejunostomy Feeding Tube Kit provides the same ability as the comparative substantially equivalent device for delivery of formula and medications to the small intestine (through "FEED" port) while allowing simultaneous gastric drainage (through "DRAIN" port) via the gastrostomy tube. The COMPAT® Surgical Gastrostomy/Jejunostomy Feeding Tube also provides for the evacuation of gastric contents through the gastrostomy tube while simultaneously feeding directly into the small bowel with the jejunal feeding tube.
#### Placement techniques and confirmation methods. 2.
Initial placement of the COMPAT® Surgical Gastrostomy/Jejunostomy Feeding Tube may be accomplished surgically, as with the predicate substantially equivalent device through a faparotomy midline incision and Stamm (gastrostomy) technique. The jejunal feeding tube can be placed in a similar fashion to that of the substantially equivalent device in that the tube is led through the lumen of the gastrostomy tube into the stomach, then grasped at its tip and manipulated through the gastric pylorus into the small intestine.
This manual manipulation into the bowel is assisted by the presence of the bolus tip on the jejunal tube and by a removable stylet. The inflation of the cuff provides an additional (temporary) feature for grasping and directing the jejunal tube tip.
Both the substantially equivalent and this device contain a Y-port at the connector end of the Jejunal Feeding tube, which may be pressed into the lumen of the Gastrostomy tube to complete the placement and allow for gastric drainage through the gastrostomy tube. The Gastrostomy/Jejunostomy tube combination may be immobilized using the external retainer and the abdomen may be closed in the standard fashion.
Confirmation of placement of both tubes may be accomplished using x-ray or fluoroscopic techniques.
#### 3. Tube and Stylet Construction.
The materials of construction of the Gastrostomy and Jejunostomy tubes are identical to that of the substantially equivalent device. The only difference is the orientation of the radiopaque material in the jejunal tubing. The tubing of the
{4}------------------------------------------------
substantially equivalent device contained 20% barium sulfate in a stripe configuration, whereas the modified tubing is solid filled 20% barium sulfate. The content of the barium sulfate contained in the tube has not changed from that of the substantially equivalent device.
## Summary of Non-Clinical Testing:
No additional non-clinical testing of this product for this use was conducted.
### Summary of Clinical Testing:
No additional clinical evaluations of this product for this use have been conducted.
### Predicate Devices:
The predicate device of the Novartis Nutrition Corporation COMPAT® Surgical Gastrostomy/Jejunostomy Feeding Tube Kit item number 088000, as modified, is the Sandoz Nutrition COMPAT® Surgical Gastro/Jejunostomy Feeding Tube Kit, (K965087). The company name has been changed from Sandoz Nutrition to Novartis Nutrition Corporation.
### Conclusions:
Novartis Nutrition Corporation COMPAT® Surgical Gastrostomy/Jejunostomy Feeding Tube Kit item number 088000, as modified, is substantially equivalent to Sandoz Nutrition COMPAT® Surgical Gastro/Jejunostomy Feeding Tube Kit, (K965087).
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of an eagle or bird-like figure with three curved lines representing its wings or body. The logo is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement. The text is in all capital letters and is evenly spaced around the circle.
Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850
Novartis Nutrition Corporation c/o Ms. Laura Danielson Program Manager TÜV SÜD America, Inc. 1775 Old Highway 8 NW, Ste. 104 NEW BRIGHTON MN 55112-1891
Re: K071191
Trade/Device Name: COMPAT® Surgical Gastrostomy/Jejunostomy Feeding Tube Kit Regulation Number: 21 CFR §876.5980 Regulation Name: Gastrointestinal tube and accessories Regulatory Class: II Product Code: FPD Dated: July 9, 2007 Received: July 10, 2007
2 3 2007
Dear Ms. Danielson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Image /page/4/Picture/9 description: The image is a black and white circular logo. The logo has the text "1906-2006 FDA Centennial" in the center. The text is surrounded by a circle of stars. The logo is likely for the 100th anniversary of the Food and Drug Administration.
Protecting and Promoting Public Health
{6}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
> 21 CFR 876.xxxx 21 CFR 884.xxxx 21 CFR 892 xxxxx Other
(Gastroenterology/Renal/Urology) 240-276-0115 (Obstetrics/Gynecology) 240-276-0115 (Radiology) 240-276-0120 240-276-0100
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html
Sincerely yours,
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
# INDICATIONS FOR USE STATEMENT
Ko71191 Current 510(k) Number:
Device Name: COMPAT® Surgical Gastrostomy/Jejunostomy Feeding Tube Kit
### Indications for Use:
The COMPAT® Surgical Gastrostomy/Jejunostomy Feeding Tube Kit provides for the delivery of enteral formula through a surgically placed Gastrostomy/Jejunostomy feeding tube. Jejunal feeding is used when patients require feeding past the stomach directly into the jejunum. This method is used to enhance the absorption of nutrients and minimize the risks of gastric feeding including gastric reflux and aspiration.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
oductive. Abdominal. Radiological Device
510(k) Number *K871191*
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.