BRAUN HEMODIALYSIS BLOOD CIRCUITS
Device Facts
| Record ID | K952631 |
|---|---|
| Device Name | BRAUN HEMODIALYSIS BLOOD CIRCUITS |
| Applicant | B.Braun Medical, Inc. |
| Product Code | FJK · Gastroenterology, Urology |
| Decision Date | Apr 30, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 876.5820 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Indications for Use
The configurations can be summarized tubing segments intended for use in transporting blood from a patient's vascular access device to hemodialyzer systems. Also, the tubing segments intended for use in transporting blood to a patient's vascular access device from such a system.
Device Story
Braun Hemodialysis Blood Circuits consist of tubing segments; function as conduit for extracorporeal blood flow between patient vascular access and hemodialyzer. Used in clinical hemodialysis settings; operated by trained healthcare professionals. Device facilitates blood transport during dialysis therapy; ensures closed-loop connection between patient and dialysis machine. Design accounts for various hemodialysis systems and user preferences.
Clinical Evidence
No clinical data. Bench testing only; includes sterility, pyrogenicity (LAL method), physical testing, and visual examination. Performance conforms to AAMI RD17R-6/94 draft standards.
Technological Characteristics
Tubing segments for extracorporeal blood transport. Biocompatibility testing per Tripartite Guidance. Performance standards: AAMI RD17R-6/94. Quality control per Device Master Records and GMPs.
Indications for Use
Indicated for patients requiring hemodialysis treatment; used to transport blood between the patient's vascular access device and hemodialyzer systems.
Regulatory Classification
Identification
A hemodialysis system and accessories is a device that is used as an artificial kidney system for the treatment of patients with renal failure or toxemic conditions and that consists of an extracorporeal blood system, a conventional dialyzer, a dialysate delivery system, and accessories. Blood from a patient flows through the tubing of the extracorporeal blood system and accessories to the blood compartment of the dialyzer, then returns through further tubing of the extracorporeal blood system to the patient. The dialyzer has two compartments that are separated by a semipermeable membrane. While the blood is in the blood compartment, undesirable substances in the blood pass through the semipermeable membrane into the dialysate in the dialysate compartment. The dialysate delivery system controls and monitors the dialysate circulating through the dialysate compartment of the dialyzer.(1) The extracorporeal blood system and accessories consists of tubing, pumps, pressure monitors, air foam or bubble detectors, and alarms to keep blood moving safely from the blood access device and accessories for hemodialysis (§ 876.5540) to the blood compartment of the dialyzer and back to the patient. (2) The conventional dialyzer allows a transfer of water and solutes between the blood and the dialysate through the semipermeable membrane. The semipermeable membrane of the conventional dialyzer has a sufficiently low permeability to water that an ultrafiltration controller is not required to prevent excessive loss of water from the patient's blood. This conventional dialyzer does not include hemodialyzers with the disposable inserts (Kiil type) (§ 876.5830) or dialyzers of high permeability (§ 876.5860). (3) The dialysate delivery system consists of mechanisms that monitor and control the temperature, conductivity, flow rate, and pressure of the dialysate and circulates dialysate through the dialysate compartment of the dialyzer. The dialysate delivery system includes the dialysate concentrate for hemodialysis (liquid or powder) and alarms to indicate abnormal dialysate conditions. This dialysate delivery system does not include the sorbent regenerated dialysate delivery system for hemodialysis (§ 876.5600), the dialysate delivery system of the peritoneal dialysis system and accessories (§ 876.5630), or the controlled dialysate delivery system of the high permeability hemodialysis system § 876.5860). (4) Remote accessories to the hemodialysis system include the unpowered dialysis chair without a scale, the powered dialysis chair without a scale, the dialyzer holder set, dialysis tie gun and ties, and hemodialysis start/stop tray.
Predicate Devices
- Blood Tubing Set (K884243)
Related Devices
- K113023 — STREAMLINE AIRLESS SYSTEM SET WITH PRE-ATTACHED DIALYZER · Nxstage Medical, Inc. · Dec 21, 2012
- K971687 — HEMODIALYSIS VENOUS BLOOD TUBING SET OR SIMILAR TRADE NAME · National Medical Care, Inc. · Jul 29, 1997
- K971313 — NMC HEMODIALYSIS ARTERIAL BLOOD TUBING SET · National Medical Care, Inc. · Oct 27, 1997
- K982340 — EXTRACORPOREAL BLOOD CIRCUIT · Haidylena Medical Egypt · Apr 30, 1999
- K032975 — JMS BLOOD TUBING SETS · Jms North America Corp. · Aug 16, 2004