NMC HEMODIALYSIS ARTERIAL BLOOD TUBING SET

K971313 · National Medical Care, Inc. · FJK · Oct 27, 1997 · Gastroenterology, Urology

Device Facts

Record IDK971313
Device NameNMC HEMODIALYSIS ARTERIAL BLOOD TUBING SET
ApplicantNational Medical Care, Inc.
Product CodeFJK · Gastroenterology, Urology
Decision DateOct 27, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.5820
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Hemodialysis Arterial Blood Tubing Set is intended for use as the extra corporeal blood circuit during hemodialysis. It is intended for single use only. The Hemodialysis Arterial Blood Tubing Set is indicated for use with conventional and high flux negative pressure hemodialyzer equipment.

Device Story

Hemodialysis arterial blood tubing set functions as an extracorporeal blood circuit during hemodialysis procedures. Device connects patient to conventional or high flux negative pressure hemodialyzer equipment. Used in clinical settings by trained healthcare professionals. Tubing facilitates blood flow between patient and dialyzer. Integrity of bonded connections ensures circuit safety. Single-use device.

Clinical Evidence

Bench testing only. No clinical data provided. Evidence includes functional testing of bloodlines and biocompatibility testing of blood-contacting materials.

Technological Characteristics

Extracorporeal blood tubing set. Materials: blood-contacting components tested for biocompatibility. Connectivity: interfaces with conventional and high flux negative pressure hemodialyzer equipment. Sterilization: standard methods employed.

Indications for Use

Indicated for use as an extracorporeal blood circuit during hemodialysis in patients requiring conventional or high flux negative pressure hemodialysis treatment. Single use only.

Regulatory Classification

Identification

A hemodialysis system and accessories is a device that is used as an artificial kidney system for the treatment of patients with renal failure or toxemic conditions and that consists of an extracorporeal blood system, a conventional dialyzer, a dialysate delivery system, and accessories. Blood from a patient flows through the tubing of the extracorporeal blood system and accessories to the blood compartment of the dialyzer, then returns through further tubing of the extracorporeal blood system to the patient. The dialyzer has two compartments that are separated by a semipermeable membrane. While the blood is in the blood compartment, undesirable substances in the blood pass through the semipermeable membrane into the dialysate in the dialysate compartment. The dialysate delivery system controls and monitors the dialysate circulating through the dialysate compartment of the dialyzer.(1) The extracorporeal blood system and accessories consists of tubing, pumps, pressure monitors, air foam or bubble detectors, and alarms to keep blood moving safely from the blood access device and accessories for hemodialysis (§ 876.5540) to the blood compartment of the dialyzer and back to the patient. (2) The conventional dialyzer allows a transfer of water and solutes between the blood and the dialysate through the semipermeable membrane. The semipermeable membrane of the conventional dialyzer has a sufficiently low permeability to water that an ultrafiltration controller is not required to prevent excessive loss of water from the patient's blood. This conventional dialyzer does not include hemodialyzers with the disposable inserts (Kiil type) (§ 876.5830) or dialyzers of high permeability (§ 876.5860). (3) The dialysate delivery system consists of mechanisms that monitor and control the temperature, conductivity, flow rate, and pressure of the dialysate and circulates dialysate through the dialysate compartment of the dialyzer. The dialysate delivery system includes the dialysate concentrate for hemodialysis (liquid or powder) and alarms to indicate abnormal dialysate conditions. This dialysate delivery system does not include the sorbent regenerated dialysate delivery system for hemodialysis (§ 876.5600), the dialysate delivery system of the peritoneal dialysis system and accessories (§ 876.5630), or the controlled dialysate delivery system of the high permeability hemodialysis system § 876.5860). (4) Remote accessories to the hemodialysis system include the unpowered dialysis chair without a scale, the powered dialysis chair without a scale, the dialyzer holder set, dialysis tie gun and ties, and hemodialysis start/stop tray.

Related Devices

Submission Summary (Full Text)

{0} OCT 27 1997 K971313 P(0) # SUMMARY OF SAFETY AND EFFICACY The proposed hemodialysis arterial blood tubing sets raise no new issues of safety or efficacy. Sterilization method and packaging are essentially the same. The arterial bloodlines are substantially equivalent to legally marketed bloodlines. These bloodlines have been functionally tested and data provided. The blood contact materials of the arterial bloodlines have been tested for biocompatibility and data provided. The integrity of the bonded connections of the bloodlines have been tested and data provided. No new claims or indications are made. {1} DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 OCT 27 1997 Mr. Art Eilinsfeld Manager, Regulatory Affairs Fresenius Medical Care North America Renal Products Technologies Two Ledgemont Center 95 Hayden Avenue Lexington, Massachusetts 02173 Re: K971313 Hemodialysis Arterial Blood Tubing Set Dated: July 25, 1997 Received: August 14, 1997 Regulatory class: II 21 CFR §876.5820/Product code: 78 FJK Dear Mr. Eilinsfeld: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, ![img-0.jpeg](img-0.jpeg) Lillian Yin, Ph.D. Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2} Page 1 of 1 510(k) Number (if known): ___________________________ Device Name: Hemodialysis Arterial Blood Tubing Set Indications For Use: The Hemodialysis Arterial Blood Tubing Set is intended for use as the extra corporeal blood circuit during hemodialysis. It is intended for single use only. The Hemodialysis Arterial Blood Tubing Set is indicated for use with conventional and high flux negative pressure hemodialyzer equipment. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Division of Reproductive, Abdominal, ENT, and Radiological Devices 510(k) Number K971313 Prescription Use ☑ (Per 21 CFR 801.109) OR Over-The-Counter Use __________ 000107 (Optional Format 1-2-96)
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...