K970708 · Terumo Medical Corp. · FJI · Dec 19, 1997 · Gastroenterology, Urology
Device Facts
Record ID
K970708
Device Name
CLIRANS T-SERIES HOLLOW FIBER DIALYZERS
Applicant
Terumo Medical Corp.
Product Code
FJI · Gastroenterology, Urology
Decision Date
Dec 19, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5820
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The CLIRANS® T-Series Hollow Fiber Dialyzers are devices used as part of an artificial kidney system for the treatment of patients with renal failure or toxemic conditions and that consists of an extracorporeal blood system, a conventional dialyzer, a dialysate delivery system, and accessories. This device is indicated for single and multiple use.
Device Story
CLIRANS T-Series Hollow Fiber Dialyzer functions as component of artificial kidney system; treats renal failure or toxemic conditions. Device utilizes cellulose hollow fibers (9μ wall thickness) housed in acrylonitrile-styrene copolymer. During operation, patient blood is pumped via roller pump into arterial end; blood flows through fibers while dialysate circulates externally. Waste products diffuse across membrane into dialysate; purified blood returns to patient via venous end. Used in clinical settings by healthcare professionals. Performance metrics include urea, creatinine, phosphate, and Vitamin B12 clearance, plus UFR and priming volume. Biocompatible materials ensure safety during extracorporeal circulation.
Clinical Evidence
Bench testing only. Performance evaluated via in vitro and in vivo clearance tests for urea, creatinine, phosphates, and Vitamin B12. Additional testing included UFR, priming volume, and biocompatibility per ISO-10993. Sterilization validated to SAL 10^-6 per AAMI guidelines.
Technological Characteristics
Hollow fiber dialyzer; cellulose fibers (9μ wall); acrylonitrile-styrene copolymer housing; polypropylene blood ports; silicone rubber O-rings; polyurethane adhesive. Sterilized via ethylene oxide. Biocompatible per ISO-10993. No software or electronic components.
Indications for Use
Indicated for patients with renal failure or toxemic conditions requiring hemodialysis treatment using an artificial kidney system.
Regulatory Classification
Identification
A hemodialysis system and accessories is a device that is used as an artificial kidney system for the treatment of patients with renal failure or toxemic conditions and that consists of an extracorporeal blood system, a conventional dialyzer, a dialysate delivery system, and accessories. Blood from a patient flows through the tubing of the extracorporeal blood system and accessories to the blood compartment of the dialyzer, then returns through further tubing of the extracorporeal blood system to the patient. The dialyzer has two compartments that are separated by a semipermeable membrane. While the blood is in the blood compartment, undesirable substances in the blood pass through the semipermeable membrane into the dialysate in the dialysate compartment. The dialysate delivery system controls and monitors the dialysate circulating through the dialysate compartment of the dialyzer.(1) The extracorporeal blood system and accessories consists of tubing, pumps, pressure monitors, air foam or bubble detectors, and alarms to keep blood moving safely from the blood access device and accessories for hemodialysis (§ 876.5540) to the blood compartment of the dialyzer and back to the patient. (2) The conventional dialyzer allows a transfer of water and solutes between the blood and the dialysate through the semipermeable membrane. The semipermeable membrane of the conventional dialyzer has a sufficiently low permeability to water that an ultrafiltration controller is not required to prevent excessive loss of water from the patient's blood. This conventional dialyzer does not include hemodialyzers with the disposable inserts (Kiil type) (§ 876.5830) or dialyzers of high permeability (§ 876.5860). (3) The dialysate delivery system consists of mechanisms that monitor and control the temperature, conductivity, flow rate, and pressure of the dialysate and circulates dialysate through the dialysate compartment of the dialyzer. The dialysate delivery system includes the dialysate concentrate for hemodialysis (liquid or powder) and alarms to indicate abnormal dialysate conditions. This dialysate delivery system does not include the sorbent regenerated dialysate delivery system for hemodialysis (§ 876.5600), the dialysate delivery system of the peritoneal dialysis system and accessories (§ 876.5630), or the controlled dialysate delivery system of the high permeability hemodialysis system § 876.5860). (4) Remote accessories to the hemodialysis system include the unpowered dialysis chair without a scale, the powered dialysis chair without a scale, the dialyzer holder set, dialysis tie gun and ties, and hemodialysis start/stop tray.
K992565 — A-18 HEMODIALYZER · Althin Medical AB · Sep 23, 1999
K040255 — POLYFLUX 14L, 17L, 21L HEMODIALYZER/FILTER LABELED FOR SINGLE USE · Gambro Renal Products, Inc. · May 24, 2004
Submission Summary (Full Text)
{0}
p.1/4
SECTION II Summary and Certification
DEC 19 1997
# SUMMARY OF SAFETY AND EFFECTIVENESS INFORMATION PERTAINING TO SUBSTANTIAL EQUIVALENCE
Proprietary Device Name: CLIRANS® T-Series Hollow Fiber Dialyzer
Classification Name: Dialyzer, Capillary, Hollow Fiber
Common Name: Hollow Fiber Dialyzer
## INTENDED USE
The CLIRANS® T-Series Hollow Fiber Dialyzers are devices used as part of an artificial kidney system for the treatment of patients with renal failure or toxemic conditions and that consists of an extracorporeal blood system, a conventional dialyzer, a dialysate delivery system, and accessories. This device is indicated for single and multiple use.
## DESCRIPTION
The CLIRANS® T-Series Hollow Fiber Dialyzers consists of cellulose hollow fibers with 9μ wall thickness. Blood is pumped via a roller pump from the artery of the patient into the arterial end of the dialyzer. The blood travels down through the dialyzer fibers where waste products pass through the membrane of the dialyzer into the dialysate, which is constantly circulating through the dialyzer on the outside of the hollow fibers. Blood then exits the venous end of the dialyzer back to the patient.
## SUBSTANTIAL EQUIVALENCE
The CLIRANS® T-Series Hollow Fiber Dialyzers submitted in this 510(k) are substantially equivalent in intended use, design, technology/principles of operation, materials and performance to the largest effective surface area dialyzer (CL*M151L) cleared in the CLIRANS® TAF Hollow Fiber Dialyzer 510(k) K854280.
009
{1}
K110708
p. 2/4
SECTION II: Summary and Certification
## PRINCIPLE OF OPERATION/TECHNOLOGY
The CLIRANS® T-Series Hollow Fiber Dialyzers operate in the following manner. Blood is taken from the artery of a patient and pumped via a roller pump into the arterial end of the dialyzer. The blood travels down through the dialyzer fibers, and waste products then pass through the membrane of the dialyzer into the dialysate which is constantly circulating through the dialyzer on the outside of the hollow fibers. Blood then exits the venous end of the dialyzer, and back into the veins of the patient.
## DESIGN/MATERIALS
| Parts | CLIRANS T-Series Hollow Fiber Dialyzer | Cleared CLIRANS TAF Hollow Fiber Dialyzer K854280 |
| --- | --- | --- |
| Hollow Fiber | Cellulose | Cellulose |
| Housing | Acrylonitrile-Styrene copolymer | Acrylonitrile-Styrene copolymer |
| Blood Port | Polypropylene | Polypropylene |
| Screw ring | Acrylonitrile-Styrene copolymer | Acrylonitrile-Styrene copolymer |
| O-ring | Silicone rubber | Silicone rubber |
| Adhesive | Polyurethane | Polyurethane |
| Blood/Dialysate Port cap | Polyethylene | Polyethylene |
{2}
470708
p 3/4
# SECTION II Summary and Certification
## SPECIFICATIONS
| Parts | CLIRANS T-Series Hollow Fiber Dialyzer | | | Cleared CLIRANS TAF Hollow Fiber Dialyzer K854280 - Largest Effective Surface Area (CL*M151L) |
| --- | --- | --- | --- | --- |
| Hollow Fiber | CL*T150L | CL*T175L | CL*T220L | CL*M151L |
| Inside Diameter | 200μ | 200μ | 200μ | 220μ |
| Wall Thickness | 9μ | 9μ | 9μ | 12μ |
| Effective Length | 235mm | 235mm | 235mm | 235mm |
| Quantity of Fiber | 10,000pcs | 11,800pcs | 15,000pcs | 9300pcs |
| Priming Blood Vol | 101ml | 120ml | 148ml | 115ml |
| Eff. Surface Area | 1.5m² | 1.75m² | 2.2m² | 1.5m² |
## PERFORMANCE
The performance of the CLIRANS® T-Series Hollow Fiber Dialyzers (T-Series) are substantially equivalent to the performance of the currently marketed largest effective surface area CLIRANS® TAF Hollow Fiber Dialyzer (CL*M151L). The T-Series dialyzers exhibit a somewhat higher clearance and UFR than the M151L which is to be expected with the thinner fiber walls in the T-Series dialyzer. However, none of the data raises any issues of safety or effectiveness. Therefore, the CLIRANS® T-Series Hollow Fiber Dialyzers performance are substantially equivalent to the predicate devices.
The following clearance tests were performed demonstrating the substantial equivalence of the CLIRANS® T-Series Hollow Fiber Dialyzers submitted in this 510(k) to the largest effective surface area dialyzer (CL*M151L) cleared in the CLIRANS® TAF Hollow Fiber Dialyzer 510(k) K854280.
- Urea
- Creatinine
- Phosphates
- Vitamin B12
- UFR (In Vitro & In Vivo)
- Priming Volume
011
{3}
p 4/4
SECTION II: Summary and Certification
## ADDITIONAL SAFETY INFORMATION
Sterilization conditions have been validated according to the AAMI guidelines to provide a Sterility Assurance Level (SAL) of 10 to the negative sixth.
Ethylene oxide residuals will not exceed the maximum residue limits proposed for Part 821 of Title 21 in the Federal register of June 23, 1978 (or as finalized or amended).
Manufacturing control test methods include: functional, extraction and sterility tests.
Blood contacting materials were tested in accordance with the tests recommended in the FDA General Program Memorandum #G95-1 (5/1/95): Use of International Standard ISO-10993, "Biological Evaluation of Medical Devices Part-1: Evaluation and Testing." [External Communicating Devices, Circulating Blood, Prolonged (24 hours-30 days) contact duration]. The blood contacting materials were found to be biocompatible.
The expiration dating for the CLIRANS® T-Series Hollow Fiber Dialyzers will be 36 months. This dating period is adopted from legally marketed CLIRANS® TAF Hollow Fiber Dialyzer 510(k) K854280. The sterilization process and packaging materials are the same for these products. Verification testing of aged product consists of package permeability, sterility and shelf life functional testing.
## CONCLUSION
The CLIRANS® T-Series Hollow Fiber Dialyzers submitted in this 510(k) are substantially equivalent in intended use, design, technology/principles of operation, materials and performance to the largest effective surface area dialyzer (CL*M151L) cleared in the CLIRANS® TAF Hollow Fiber Dialyzer 510(k) K854280. Differences between the devices cited in this section do not raise any new issues of safety or effectiveness.
Date Prepared February 23, 1997
Prepared by Keith M. Smith
Senior Regulatory Affairs Specialist
Regulatory Affairs
Prepared for Terumo Medical Corporation
125 Blue Ball Road
Elkton, MD 21921
Phone (410) 392-7375 or (410) 392-7231
Fax (410) 398-6079
012
{4}
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration
9200 Corporate Boulevard
Rockville MD 20850
DEC 19 1997
Mr. Keith M. Smith
Senior Regulatory Affairs Specialist
Terumo Medical Corporation
Regulatory Affairs Department
125 Blue Ball Road
Elkton, Maryland 21921
Re: K970708
CLIRANS® T-Series Hollow Fiber Coventional
Dialyzers for Single and Multiple Use
Dated: November 26, 1997
Received: November 28, 1997
Regulatory Class: II
21 CFR §876.5820/Product Code: 78 FJI and MSE
Dear Mr. Smith:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,

Lillian Yin, Ph.D.
Director, Division of Reproductive,
Abdominal, Ear, Nose and Throat,
and Radiological Devices
Office of Device Evaluation
Center for Devices and
Radiological Health
Enclosure
{5}
510(k) Number (if known):
Device Name: CLIRANS® T-Series Hollow Fiber Dialyzers
## Indications For Use:
The CLIRANS® T-Series Hollow Fiber Dialyzers are devices used as part of an artificial kidney system for the treatment of patients with renal failure or toxemic conditions and that consists of an extracorporeal blood system, a conventional dialyzer, a dialysate delivery system, and accessories. This device is indicated for single and multiple use.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Reproductive, Abdominal, ENT, and Radiological Devices
510(k) Number K 970708
Prescription Use ☑
(Per 21 CFR 801.109)
OR
Over-The-Counter Use ☐
(Optional Format 1-2-96)
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.