K142369 · Guangzhou Finecure Medical Equipment Co.,Ltd · KPI · Jun 17, 2015 · Gastroenterology, Urology
Device Facts
Record ID
K142369
Device Name
Urinary Incontinence System
Applicant
Guangzhou Finecure Medical Equipment Co.,Ltd
Product Code
KPI · Gastroenterology, Urology
Decision Date
Jun 17, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5320
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Urinary Incontinence System with various treatment modalities is intended to be used by males and females for muscle contraction in the treatment of stress incontinence, urge incontinence, and mixed incontinence.
Device Story
Non-implanted electrical pelvic floor muscle stimulator; strengthens pelvic floor muscles via electrical stimulation. Device comprises enclosure, display, PCB, rotary switch, and electronic components; used with pelvic floor or anorectal probes. Operates via preset modes corresponding to incontinence conditions. Used in clinical or home settings by patients or clinicians. Device delivers dual-phasic, rectangular electrical pulses to pelvic floor musculature; output parameters (voltage, frequency, pulse width) are adjustable based on selected mode. Healthcare providers use device to facilitate neuromuscular re-education and muscle contraction, potentially improving urinary control and reducing incontinence symptoms.
Clinical Evidence
No clinical data. Bench testing only. Testing included performance, software verification/validation, electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), and home healthcare environment safety (IEC 60601-1-11). Biocompatibility confirmed via material identity with predicate K130691.
Technological Characteristics
Non-implanted electrical stimulator; 2 output channels; dual-phasic rectangular pulse waveform. Materials: ABS (shell), PET (membrane switch). Power: 9V battery. Output: 0-60V (500 ohm load), max current 120mA. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 60601-2-10. Firmware-controlled with automatic shut-off.
Indications for Use
Indicated for males and females with stress, urge, or mixed urinary incontinence requiring pelvic floor muscle contraction.
Regulatory Classification
Identification
A nonimplanted electrical continence device is a device that consists of a pair of electrodes on a plug or a pessary that are connected by an electrical cable to a battery-powered pulse source. The plug or pessary is inserted into the rectum or into the vagina and used to stimulate the muscles of the pelvic floor to maintain urinary or fecal continence. When necessary, the plug or pessary may be removed by the user. This device excludes an AC-powered nonimplanted electrical continence device and the powered vaginal muscle stimulator for therapeutic use (§ 884.5940).
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing right, with a flowing ribbon-like element connecting them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 17, 2015
Guangzhou Finecure Medical Equipment Co., Ltd % Field Fu Shenzhen Joyantech Consulting Co., Ltd. 4th Floor, Jinhui Building, Nanhai Blvd. Nanshan District Shenzhen, Guangdong China
Re: K142369
> Trade/Device Name: Urinary Incontinence System Regulation Number: 21 CFR 876.5320 Regulation Name: Nonimplanted electrical continence device Regulatory Class: II Product Code: KPI Dated: Mav 4, 2015 Received: May 15, 2015
Dear Field Fu,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{2}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Image /page/1/Picture/8 description: The image shows the text "Herbert P. Lerner -S" in a large, bold font. The text is arranged horizontally, with the name "Herbert P. Lerner" appearing prominently. The "-S" is slightly smaller and separated by a hyphen. The background is plain and white.
for
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
### Indications for Use
510(k) Number (if known) K142369
Device Name Urinary Incontinence System
Indications for Use (Describe)
The Urinary Incontinence System with various treatment modalities is intended to be used by males and females for muscle contraction in the treatment of stress incontinence, and mixed incontinence.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
] Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
{4}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
## SECTION 05 510(K) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR§807.92
| The assigned 510(k) number is: | (applicant leave blank) |
|--------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submission Date: | May 7, 2015 |
| Submitter: | Guangzhou Finecure Medical Equipment Co., Ltd.<br>Address: F19, No.1 Kesheng Road, Baiyun Torch Building,<br>No.1633 Beitai Road, Baiyun District, Guangzhou, CHINA. |
| Submitter Contact: | Shenzhen Joyantech Consulting Co., Ltd.<br>4th Floor, Jinhui Building, Nanhai BLVD, Nanshan District,<br>Shenzhen, Guangdong, China.<br>Contact person: Mr. Field.Fu<br>E-Mail: cefda13485@163.com<br>Image: [logo] |
| Manufacturing Site: | Guangzhou Finecure Medical Equipment Co., Ltd.<br>Address: F19, No.1 Kesheng Road, Baiyun Torch Building,<br>No.1633 Beitai Road, Baiyun District, Guangzhou, CHINA. |
| Device Name: | Urinary Incontinence System |
| Common Name: | Pelvic floor muscle stimulator |
| Classification Name: | Nonimplanted electrical continence device |
| Regulation: | 21 CFR §876.5320 |
| Product Code: | KPI |
| Type/Model of the device | P3-9632E1 and P3-9632E2 |
| Substantially<br>Equivalent Devices: | The Finecure Urinary Incontinence System is substantially<br>equivalent to: the InControl InTone (K110179) and the I touch |
{6}------------------------------------------------
Intended Use The Urinary Incontinence System with various treatment Indications for Use: modalities is intended to be used by males and females for muscle contraction in the treatment of stress incontinence, urge incontinence, and mixed incontinence.
Device Description: The Finecure Urinary Incontinence System is a non-implanted, electrical, pelvic floor muscle stimulator. It is intended to strengthen the pelvic floor muscles by electrical stimulation in the treatment for urinary incontinence. The system has several preset modes that correspond to different conditions associated with urinary incontinence. The Urinary Incontinence System can be used in the clinical setting and the home care environment. The System is composed of enclosure, display screen, PCB, rotary switch, and other electronic components. The system is to be used with pelvic floor probes and anorectal probes.
Technology Comparison to predicate device
The Finecure Urinary Incontinence System is employ the same technological characteristics compare to InControl InTone (K110179) and Itouch Sure ( K103698) in the following table for determine the device whether substantially equivalent to the predicate devices.
{7}------------------------------------------------
# Equivalence Comparison to the Predicate
| | Finecure Urinary Incontinence<br>System (new device) | InControl InTone K110179 | Itouch Sure<br>K103698 | Comparison |
|------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended use &<br>Indications for<br>Use | The Urinary Incontinence System with<br>various treatment modalities is<br>intended to be used by males and<br>females for muscle contraction in the<br>treatment of stress incontinence, urge<br>incontinence, and mixed incontinence. | The InControl device is<br>non-implantedelectrical stimulator<br>indicated for use in the treatment<br>of female urinary incontinence. It<br>applies electrical stimulation to the<br>pelvic floor musculature and<br>surrounding structures. It is intended<br>for acute and ongoing treatment of<br>mixed urinary incontinence where<br>the following results may improve<br>urinary control: strengthening of<br>pelvic floor muscles, inhibition of the<br>detrus or muscle through reflexive<br>mechanisms. The biofeedback<br>feature can be used for muscle | The Itouch Sure is intended to<br>provide electrical stimulation<br>and neuromuscular re-education<br>for the purpose of rehabilitation<br>of weak pelvic floor muscles for<br>the treatment of stress, urge,<br>and mixed urinary incontinence<br>in women. The intended use and<br>indications for use of the new<br>device are the same as those of<br>the predicate devices. | Similar1 |
| | Finecure Urinary Incontinence<br>System (new device) | InControl InTone K110179 | Itouch Sure<br>K103698 | Compariso<br>n |
| | | re-education purposes. | | |
| Power Source | 6LR61/AM9V Alkalescent battery | 4/5 AA nickel metal hydride battery | 2 ×1.5AA | Equivalent |
| Output voltage | Adjustable, from 0~60V ( 500 ohm<br>load). | 50Vdc (500 ohm) | 0~45V ( 500 ohm load). | Different2 |
| Output current | Max Output current (500 Ohm)<br>120mA | Max Output current (500 Ohm )<br>100mA | Max Output current (500 Ohm)<br>90mA | Identical |
| Maximum Phase<br>Charge, (mC) | 36uC@500Ω | 50uC@500Ω | unknown | Equivalent |
| Maximum<br>Current Density | 7.64mA/cm2 , for vaginal probe;<br>16.3 mA/cm2, fro anal probe. | 4.7mA/cm2 | unknown | Different3 |
| Maximum<br>Average Power<br>Density | Pre2, 2.14mW/ cm2;<br>Pre4, 14.6mW/ cm2 ;<br>Pre5, 1.96mW/ cm2 ;<br>Pre7, 9.78mW/ cm2 | 14.3mW/cm2@500Ω | unknown | Equivalent |
| | Finecure Urinary Incontinence<br>System (new device) | InControl InTone K110179 | Itouch Sure<br>K103698 | Compariso<br>n |
| Electrode<br>Surface Area | For vaginal probe: $7.85×2=15.7 cm²$<br>For Anal probe: $3.68×2=7.36 cm²$ | $10.5 cm² x2$ | unknown | Equivalent |
| Number of output<br>channels | 2 channels | 1 channel | unknown | Different4 |
| Waveform | Dual phasic, rectangular pulse. | Dual phasic, rectangular pulses | Dual phasic, rectangular pulses | Same |
| Session Duration<br>(min) | Mode:<br>Time(min)<br>PFS-Pre2, 0min/20min;<br>(Pelvic floor muscle<br>contraction )<br>PFS-Pre4, 0min/20min;<br>(Stress incontinence)<br>PFS-Pre5, 0min/32min;<br>(Urge incontinence) | 12 mins | Continuous/15/30 min | Equivalent:<br>For Pre2,<br>Pre4, Pre7,<br>the duration<br>is within the<br>predicate<br>K103698;<br>For Pre5,<br>the duration<br>is close to<br>the<br>predicate<br>K103698. |
| | Finecure Urinary Incontinence<br>System (new device) | InControl InTone K110179 | Itouch Sure<br>K103698 | Comparison |
| | PFS-Pre7<br>0min/20min.<br>(Mixed incontinence) | | | |
| Pulse Rate (PR,<br>Hz) | | 50Hz | 10/20/35/50Hz | Equivalent: |
| | Mode: | | | For Pre2,<br>Pre4, Pre5: |
| | PFS-Pre2<br>(muscle contraction) 35Hz; | | | the<br>frequency is<br>respectively<br>same as the |
| | PFS-Pre4(Stress UI):50Hz; | | | predicate<br>k103698; |
| | PFS-Pre5(Urge UI):10Hz; | | | For Pre7,<br>the<br>frequency |
| | PFS-Pre7 (Mixed UI): 50Hz | | | same as<br>the<br>predicate<br>k110179; |
| | Finecure Urinary Incontinence<br>System (new device) | InControl InTone K110179 | Itouch Sure<br>K103698 | Compariso<br>n |
| Pulse Width | Mode:<br>PW(µS)<br>PFS-Pre2(muscle contraction) :<br>250us;<br>PFS-Pre4(Stress UI):300us;<br>PFS-Pre5(Urge UI) :200us;<br>PFS-Pre7 (Mixed):200us; | 200µS/phase | 200/250/300/phase | Equivalent |
| Time<br>ON and OFF | Mode: ON and OFF time<br>PFS-Pre2, ON:15secs, OFF:5secs ;<br>PFS-Pre4,ON:10secs, OFF:10secs;<br>PFS-Pre5, ON:10secs, OFF:10secs;<br>PFS-Pre7 ON:10secs, OFF:10secs; | ON:20secs<br>OFF:10secs | unknown | Equivalent<br>(Within<br>range) |
| Applicable<br>Patients | Male and female | Female | Female | Equivalent |
| Regulated<br>current<br>of<br>voltage? | Regulated voltage | Regulated voltage | Regulated voltage | Same |
| Firmware<br>controlled? | Yes | Yes | Yes | Same |
| Automatic<br>Shut<br>Off? | Yes | Yes | Yes | Same |
{8}------------------------------------------------
510k Summary
Version: A/0
{9}------------------------------------------------
Version: A/0
510k Summary
{10}------------------------------------------------
## Guangzhou Finecure Medical Equipment Co., Ltd.
Product: Urinary Incontinence System
Version: A/0
## 510k Summary
{11}------------------------------------------------
Product: Urinary Incontinence System
Version: A/0
510k Summary
{12}------------------------------------------------
### Testing Summary
The testing for Urinary Incontinence System included performance, software, electrical safety, EMC, biocompatibility and bench. Urinary Incontinence System successfully passed all testing.
| Biocompatibility Safety | The Host shell (ABS material) and Membrane Touch<br>Switch (PET material) of the Urinary Incontinence<br>System(P3-9632E1 and P3-9632E2) are respectively<br>identical with the enclosure (ABS) and LCD cover (PET)<br>of Combo Electrical Stimulation (P0-9612) as it was<br>approved in [K130691,12/20/2013] in formulation,<br>processing, and no other chemicals have been added<br>(e.g., plasticizers, fillers, color additives, cleaning agents,<br>mold release agents, etc.). |
|------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Electrical Safety | Finecure Urinary Incontinence System were tested in<br>accordance with applicable clause of IEC 60601-1: 2005<br>+ CORR. 1 (2006) + CORR. 2 (2007), Medical electrical<br>equipment –Part 1:General requirements for basic safety<br>and essential performance.<br>Test results indicated that the Urinary Incontinence<br>System comply with the applicable clauses of the<br>Standards. |
| Electromagnetic<br>Compatibility Testing | Finecure Urinary Incontinence System were tested in<br>accordance with applicable clause of IEC<br>60601-1-2:2007, Medical electrical equipment - Part 1-2:<br>General requirements for basic safety and essential<br>performance - Collateral standard: Electromagnetic<br>compatibility - Requirements and tests.<br>Test results indicated that the Urinary Incontinence<br>System comply with the applicable clauses of the<br>Standards. |
| Home healthcare<br>environment Testing | Finecure Urinary Incontinence System were tested in<br>accordance with applicable clause of IEC 60601-1-11:<br>2010 MEDICAL ELECTRICAL EQUIPMENT - Part 1-11:<br>General requirements for basic safety and essential<br>performance - Collateral Standard: Requirements for<br>medical electrical equipment and medical electrical<br>systems used in the home healthcare environment.<br>Test results indicated that the Urinary Incontinence<br>System comply with the applicable clauses of the<br>Standards. |
| Bench testing | The bench test had been conducted on Urinary<br>Incontinence System, the test results indicated that the<br>device meet design requirements. |
| Software verification and<br>validation | The software verification and validation was performed<br>on Urinary Incontinence System, the test results<br>indicated that the Urinary Incontinence System meet the<br>requirements of design. |
| Non-Clinical Summary | Non-clinical tests consist of Electrical safety (IEC<br>60601-1), EMC(IEC 60601-1-2), Performance (IEC<br>60601-2-10), safety in the home healthcare<br>environment (IEC 60601-1-11) and bench test<br>(manufacturer specification), the test results show that<br>the proposed device complied with the requirements of<br>standards aforesaid. The materials contacting with<br>body is identical with the materials used in Combo<br>Electrical Stimulation (P0-9612) as it was approved in<br>[K130691,12/20/2013], the test for compatibility was<br>not conducted. |
{13}------------------------------------------------
### Conclusion
Based on the comparison and performance testing, the proposed device is Substantially Equivalent (SE) to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.