Customer complaints; Returned product analysis; Post-market feedback
The sponsor utilized post-market surveillance data from the European market (April 2009–2011) to demonstrate that the device performed as intended and met specified requirements.
Post-market performance review; Retrospective review of post-market data; Follow-up/Duration: April 2009 to April 2011; Study Period: April 2009 – 2011
European users of the itouch Sure device; Number of Sites: Not applicable
Not applicable for this study
Product performance and safety as intended
Indications for Use
The itouch Sure is intended to provide electrical stimulation and neuromuscular re-education for the purpose of rehabilitation of weak pelvic floor muscles for the treatment of stress, urge, and mixed urinary incontinence in women
Device Story
TensCare itouch Sure is a battery-powered, single-channel neuromuscular electrical stimulation (NMES) device for home use. It utilizes a vaginal probe with two stainless steel electrodes to deliver bi-phasic, rectangular electrical pulses to pelvic floor muscles. The device features four preset treatment programs, adjustable intensity (200 steps), and an adjustable timer. It operates in intermittent mode to facilitate muscle re-education. The patient controls the device via push buttons; a compliance monitor tracks usage. The device incorporates open circuit detectors and controls output at constant voltage (above 500 Ω) or pseudo-constant current (below 500 Ω) to ensure safety. By stimulating pelvic floor muscles, the device aims to strengthen musculature, thereby assisting in the management of urinary incontinence. Healthcare providers prescribe the device, and patients operate it independently at home.
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by bench testing, including electrical safety (EN60601-1-2:2007), biocompatibility testing (ISO 10993-5:2009 and -10:2002), risk analysis, and FMEA. Post-market feedback from European distribution since 2009 was used to support safety and performance claims.
Technological Characteristics
Battery-powered (2x1.5V AA) NMES device. Vaginal probe constructed of ABS plastic and stainless steel electrodes. Bi-phasic, rectangular waveform. Output: 0-45V (0-90mA over 500 Ω). Constant voltage (>500 Ω) and pseudo-constant current (<500 Ω) control. 4 preset programs. Connectivity: None. Sterilization: Not specified. Software: Embedded firmware for program control and compliance monitoring.
Indications for Use
Indicated for women with weak pelvic floor muscles requiring rehabilitation for the treatment of stress, urge, and mixed urinary incontinence.
Regulatory Classification
Identification
A nonimplanted electrical continence device is a device that consists of a pair of electrodes on a plug or a pessary that are connected by an electrical cable to a battery-powered pulse source. The plug or pessary is inserted into the rectum or into the vagina and used to stimulate the muscles of the pelvic floor to maintain urinary or fecal continence. When necessary, the plug or pessary may be removed by the user. This device excludes an AC-powered nonimplanted electrical continence device and the powered vaginal muscle stimulator for therapeutic use (§ 884.5940).
Predicate Devices
Innova Pelvic Floor Stimulation System by EMPI (K941911)
Kegel 8 Pelvic Muscle Trainer by NE Services (K81480)
Submission Summary (Full Text)
{0}------------------------------------------------
K103698 PG 1/3
# 510(k) Summary
JUN 2 1 2011
Date of Summary prepared: April 27th, 2011
: Submitter Tenscare Ltd 9 Blenheim Road, Longmead Business Park, Epsom, Surrey, KT19 9BE, Address . United Kingdom : +44(0)1372 723 434 Tel Fax : +44(0)1372 745 434 : E-Mail sales@tenscare.co.uk FDA Establishment registration no: 3003446042 Contact person: Andrew Brown
#### Address of the manufacturing facility:
EasyMed Instrument Ltd 5F/6F, Block A, Gupo Gongmao Building, Fengxin Road, Fengxiang Industrial District, Daliang, Shunde, Foshan, Guangdong, China, Zip: 528300
FDA Establishment registration no: 3004049909
#### Address of American Representative:
DEBORAH IAMPIETRO QRC CONSULTING ASSOCIATES 130 Gotzens Rd Conway, NH 03818 Phone: 603 4476086 ext Fax: 734 4236086 Email: grcassoc@aol.com
## Submitted Device:
| Generic name: | Pelvic Floor Stimulator |
|------------------------|-----------------------------------------------------------------------------------------------------------------------------------|
| Trade name: | TensCare itouch Sure |
| Common name: | itouch Sure Pelvic Floor Exerciser |
| Classification name: | Stimulator, Electrical, Non-implantable, for Incontinence –<br>Title 21, Code of Federal Regulations Sec.876.5320 ProCode: 78 KPI |
| Device Classification: | Class II |
| Predicate Devices: | Innova Pelvic Floor Stimulation System by EMPI. K941911<br>Kegel 8 Pelvic Muscle Trainer by NE Services. K81480 |
#### The class of the predicate Devices: Class II
#### Device Description:
The itouch Sure is a small lightweight battery powered single channel neuromuscular stimulation device supplied with a vaginal two electrode stimulation probe. The probe connects to the control unit by cable and plug.
{1}------------------------------------------------
The unit is intended for home use by the patient, and is designed with simplicity and ease of use in mind. It has four preset treatment programs, an adjustable treatment timer, a compliance monitor, and open circuit detectors.
#### The intended use of the device:
The itouch Sure is intended to provide electrical stimulation and neuromuscular re-education for the purpose of rehabilitation of weak pelvic floor muscles for the treatment of stress, urge, and mixed urinary incontinence in women
The intended use and indications for use of the new device are the same as those of the predicate devices.
### Technological Comparison:
| | Itouch Sure | EMPI Innova<br>K941911 | Kegel 8<br>K81480 |
|--------------------------|------------------------------------------------------------------|----------------------------------------------------------------------------------|-----------------------------------------------------------|
| Intended Use | Treatment of urinary<br>incontinence | Treatment of urinary<br>incontinence | Treatment of urinary<br>incontinence |
| Programmable<br>features | Intensity, program<br>session length | Intensity, 12.5/50 Hz,<br>Duty cycle: 5-5/5-10<br>session length | Intensity,<br>session length, Hz, us,<br>Work, Rest, Ramp |
| Preset Programs | 4 | | 9 +manual |
| Power source | 2x1.5V AA | 9V PP3 | 9V PP3 |
| Mode of operation | Intermittent | Intermittent | Intermittent |
| Frequency Hz | 10/20/35/50 | 12.5/50 | 2-100 |
| Pulse Width uS | 200/250/300 | 300 | 50-450 |
| Duty cycle | 5/10 and 3/6 | 5-5/5-10 | Selectable |
| Output current | 0-45V = 0-90mA<br>over 500 Ohm | 0-60mA (100mA<br>override) | 0-90 mA nominal |
| Intensity steps | 200 | Rotary | 100 |
| Timing control | | Continuous/15/30 min | |
| Output type | Constant current<br>160-500 Ω,<br>Constant voltage<br>500-1500 Ω | Nominally constant<br>current up to 1kΩ | Nominally constant<br>current |
| No of electrodes | 1 | 1 (optional anal probe) | 2 |
| Usage conditions | Intermittent | Intermittent | Intermittent |
| Controls | 6 Push buttons +<br>lock button on side | Two rotary intensity<br>Duty cycle slider switch<br>Intensity override<br>switch | 6 Push buttons + 2<br>concealed buttons |
| Probe length mm | 88 | | 87 |
| Probe diameter mm | 28 | | 26 |
| Electrode orientation | Axial | | Axial |
| Electrode material | Stainless Steel | | Stainless Steel |
| Electrode placement | Vaginal | | Vaginal |
| Waveform Type | Bi-phasic | Bi-phasic | Bi-phasic |
| Waveform Shape | Rectangular at<br>positive | Irregular Rectangular | Rectangular at positive |
{2}------------------------------------------------
| Output waveforms: | The Sure output waveform is substantially equivalent to that of the<br>Kegel 8. The difference in shape of negative phase is not clinically<br>significant. |
|-----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Biocompatibility | The materials of the vaginal electrode are exactly the same as the<br>materials of the Kegel 8 probe by NE Systems.<br><br>Body material: Acrylonitrile-Butadiene-Styrene copolymer (ABS)<br>Conductor Material: Stainless steel<br><br>Biocompatibility tests to ISO 10993-5:2009 and -10:2002 on the<br>vaginal electrode have shown no cytotoxicity, negligible vaginal<br>irritation, and no sensitisation. |
| Labelling Comparison: | The Labelling is substantially equivalent to that of the predicate<br>devices. |
| Safety information: | Designed to comply with relevant safety applicable recognized<br>consensus standards; the output energy is controlled well within the<br>safety and effectiveness ranges specified by relevant FDA guidance.<br>Detailed and strictly controlled testing has been carried out.<br>Test results, Risk Analysis, and FMEA analysis show that the new unit<br>itouch Sure is safe with no hazard.<br><br>itouch Sure has been marketed in Europe since April 2009. During this<br>period a review of Customer Complaints, Returned Product and the<br>results of Post Market Feedback, has demonstrated that the product<br>has performed as intended, to its' Specified Requirements.<br>The new device itouch Sure controls the output at constant voltage<br>above 500 Ω load. Below this it controls the output at a pseudo<br>constant current, with output voltage reducing to zero as load reduces.<br>This characteristic reliably makes the device safe. |
| EMC testing: | The Sure has been tested to EN60601-1-2:2007 and found to be<br>suitable for use in domestic establishments and in establishments<br>directly connected to a low voltage power supply network which<br>supplies buildings used for domestic purposes. |
| Submitted times: | This is the first submission to FDA for this new device |
| Conclusion: | The itouch Sure unit is substantially equivalent to the predicate<br>devices and any differences between the devices do not pose any<br>new questions of safety and effectiveness. |
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services (HHS). The logo features a stylized depiction of an eagle with three stripes forming its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002
Mr. Andrew Brown Regulatory Affairs Manager Tenscare Ltd. 9 Blenheim Road Epsom. Surrey UNITED KINGDOM KT19 9BE
#### JUN 2 1 2011
Re: K103698
Trade/Device Name: Tenscare itouch Sure Pelvic Floor Exerciser Regulation Number21 CFR §876.5320 Regulation Name: Non-implanted electrical continence device Regulatory Class: II Product Code: KPI Dated: May 18, 2011 Received: May 18, 2011
Dear Mr. Brown:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related
{4}------------------------------------------------
#### Page 2
adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
Strict are 607) productionDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Hubert Lenoir MD
Herbert P. Lerner, M.D., Director (Acting) Division of Reproductive, Gastro-Renal and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Section 4 Indications for Use
510(k) Number (if known):
Device Name
: Tenscare itouch Sure Pelvic Floor Exerciser
#### Indications for Use
The itouch Sure is intended to provide electrical stimulation and neuromuscular re-education for the purpose of rehabilitation of weak pelvic floor muscles for the treatment of stress, urge, and mixed urinary incontinence in women
Halum
(Division Sign-Off)
Division of Reproductive, Gastro-Renal, and
Urological Devices
510(k) Number K103698
Prescription Use X AND/OR (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.