The InControl InTone device is a non-implanted electrical stimulator indicated for use in the treatment of female urinary incontinence. It applies electrical stimulation to the pelvic floor musculature and surrounding structures. It is intended for acute and ongoing treatment of mixed urinary incontinence where the following results may improve urinary control: strengthening of pelvic floor muscles, inhibition of the detrusor muscle through reflexive mechanisms. The biofeedback feature can be used for muscle re-education purposes. Federal (USA) law restricts this device to sale by or on the order of a physician.
Device Story
InControl InTone is a non-implanted electrical stimulator for female urinary incontinence. System comprises an intra-vaginal insertion probe with stainless steel electrodes and a hand-held control unit. Device delivers electrical stimulation to pelvic floor musculature and surrounding structures; provides air-pressure-based biofeedback for muscle re-education. Operated in clinic or home settings under physician direction. Patient controls session start/stop; clinician locks in current parameters. Device records and stores contraction data to assist in determining necessary stimulation levels. Output helps patients strengthen pelvic floor muscles and inhibit detrusor muscle activity, potentially improving urinary control.
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by bench testing, including EMC, electrical, mechanical, safety (operator and patient), temperature/humidity, and software testing, confirming compliance with applicable standards.
Technological Characteristics
Non-implanted electrical stimulator. Materials: ABS plastic (housing), silicone/plastic (insertion probe), stainless steel (electrodes). Energy: Nickel metal hydride battery. Output: 1 channel, regulated voltage, dual-phase rectangular pulses (50 Hz, 200 µs/phase). Biofeedback: Air pressure (0-2 psi). Connectivity: Standalone. Software: Firmware-controlled with automatic overload/no-load trip and shut-off features. Moderate level of concern.
Indications for Use
Indicated for female patients with mixed urinary incontinence. Used for acute and ongoing treatment to strengthen pelvic floor muscles and inhibit detrusor muscle activity via reflexive mechanisms. Includes biofeedback for muscle re-education.
Regulatory Classification
Identification
A nonimplanted electrical continence device is a device that consists of a pair of electrodes on a plug or a pessary that are connected by an electrical cable to a battery-powered pulse source. The plug or pessary is inserted into the rectum or into the vagina and used to stimulate the muscles of the pelvic floor to maintain urinary or fecal continence. When necessary, the plug or pessary may be removed by the user. This device excludes an AC-powered nonimplanted electrical continence device and the powered vaginal muscle stimulator for therapeutic use (§ 884.5940).
{0}------------------------------------------------
K110179
pg. 1 of 7
## 510(k) Summary for InControl InTone
| Submitter: | | InControl Medical, LLC | | FEB 2 2 2012 |
|-------------------------------|------------------------------------------------------------------------------------------|--------------------------------------------------------------------------|-------------------------------------------------------------|-----------------|
| Address: | | | 3225 Gateway Road, Suite 250<br>Brookfield, Wisconsin 53045 | |
| Corporate Contact: | | InControl Medical, LLC | Herschel Peddicord, President | |
| Telephone: | | Ph: (414) 429-2625 | | |
| Establishment Registration #: | | | to be registered following 510(k) | |
| Submission Contact: | | Michael J. Leigh, consultant<br>12715 Falcon Drive<br>Ph: (262) 957-6797 | Brookfield, Wisconsin 53005 | |
| Trade Name: | InControl InTone | | | |
| Predicate Device: | Otto Bock STIWELL med, K080950<br>MyoTrac Infiniti, K053434<br>Hollister evadri, K050483 | | | |
| Common Name: | | | Stimulator, Electrical, Non-Implantable, for Incontinence | |
| Classification Name: | | | | |
| | | | | Device<br>Class |
| | | | Jon-implanted electrical continen | II |
### Device Description:
The device includes three parts; an insertion probe and a hand-held control unit. The insertion The device includes three parts, an inoortion provides more adaptable fit for intra-vaginal use. The probe includes stainless steel electrodes to deliver electro-stimulation. The hand-held control probe includes stainless steel clours and displays the results of electro-stimulation and patient and ulsplay device records, storos, and alspiary one to select and lock in the correct amount generated pervic noor excressor from to the contraction. It also stores data on contraction of current necessary to stimulate pervice noor ochination to select and lock in the correct amount of current necessary to stimulate pelvic floor contraction.
device
Intended Use:
d Use:
The InControl InTone device is a non-implanted indicated for use in the treatment of The InControl In One device is a non-inplanted electrical stimulation has and manufacture and female urinaly incontinence. It applies electrial only of mixed urinary incontinent surrounding structures. It is intended for active and strengthening of pelvic floor muscles and where the following results may improve unflary bonnery bonnel of the biofeedback feature can be used for muscle re-education purposes.
{1}------------------------------------------------
K110179
. pg. 2 of 7
Performance Data:
Complete EMC, electrical, mechanical, safety (operator and patient), temperature/humidity, and software testing demonstrate compliance with applicable standards. The results demonstrate that the InControl InTone device is in compliance with the guidelines and standards referenced in the FDA reviewer's guides, and that it performed within its specifications and functional requirements for the device.
This summary of 510(k) safety and effectiveness information is being submitted in accordance with FDA requirements.
{2}------------------------------------------------
| InControl InTone<br>(new device) | Otto Bock STIWELL med4<br>(K080950) | MyoTrac Infiniti<br>(K053434) | Hollister evadri<br>(K050483) | Comparison |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| The InControl device is a non-implanted<br>electrical stimulator indicated for use in<br>the treatment of female urinary<br>incontinence. It applies electrical<br>stimulation to the pelvic floor musculature<br>and surrounding structures. It is intended<br>for acute and ongoing treatment of mixed<br>urinary incontinence where the following<br>results may improve urinary control:<br>strengthening of pelvic floor muscles,<br>inhibition of the detrusor muscle through<br>reflexive mechanisms. The biofeedback<br>feature can be used for muscle re-<br>education purposes. | The STIWELL med4 is a neuromuscular<br>electronic stimulator for use under<br>medical supervision for adjunctive<br>therapy in the treatment of medical<br>diseases and conditions.<br><br>As a nonimplanted electrical continence<br>device the STIWELL med4 is indicated<br>for the following conditions:<br>Acute and ongoing treatment of<br>stress, urge, or mixed urinary<br>incontinence and where the<br>following results may improve<br>urinary control: Inhibition of the<br>detruser muscles through reflexive<br>mechanisms and strengthening of<br>the pelvic floor muscles. Incontinence treatment for<br>assessing EMG activity of the<br>pelvic floor and accessory<br>muscles such as the abdominal<br>and the gluteus muscles. As a biofeedback device the STIWELL<br>med4 is indicated for the following<br>conditions:<br>Biofeedback, relaxation and<br>muscle re-education purposes. | Treatment of urinary incontinence<br>EMG biofeedback<br><br>The MyoTrac Infiniti system is indicated for<br>acute and ongoing treatment of stress, urge,<br>or mixed urinary incontinence and where the<br>following results may improve urinary<br>control: Inhibition of the detruser muscle<br>through reflexive mechanisms, strengthening<br>of pelvic floor muscle. It is also indicated<br>during incontinence treatment for assessing<br>EMG activity of the pelvic floor and<br>accessory muscles such as the abdominal or<br>gluteal muscles. | The evadri Bladder Control<br>System is intended to<br>provide electrical<br>stimulation or<br>electromyographic or<br>pressure biofeedback for the<br>treatment of urinary or fecal<br>(electromyographic<br>biofeedback) incontinence. | Equivalent |
| (see product labeling) | (see product labeling) | (see product labeling) | (see product labeling) | Equivalent |
Section 5 - 510(k) Summary.doc
K110179
pg. 3 of 7
{3}------------------------------------------------
Section 5 - 510(k) Summary.doc
K110179
Pg. 4 of 7
{4}------------------------------------------------
| 1 |
|------------------|
| |
| 1<br>P |
| œ<br>0 |
| t<br>1 |
| |
| *** *** |
| 1 |
| |
| |
| 4<br>11 |
| œ |
| 7<br>1 |
| œ<br>- |
| |
| 2<br>7 |
| œ<br>œ |
| |
| ><br>0 |
| )<br>œ<br>4<br>1 |
| Method of line current<br>Solation | n/a (battery) | n/a (battery) | n/a (battery) | | Identical<br>(to Otto Bock,<br>MyoTrac) |
|-----------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------|
| Patient leakage<br>current | n/a (battery) | n/a (battery) | n/a (battery) | | Identical<br>(to Otto Bock,<br>MyoTrac) |
| Number of output<br>modes | 1 | 1 | 1 | | Identical<br>(to all) |
| Number of output<br>channels | 1 | 1 | 1 | | Identical<br>(to all) |
| Regulated current or<br>Voltage? | Regulated voltage | Regulated current | Unknown | | Equivalent |
| Firmware controlled? | Yes | Yes | Yes | | Identical<br>(to all) |
| Automatic Overload<br>Trip? | Yes | Yes | Unknown | | Identical<br>(to Otto Bock) |
| Automatic No-Load<br>Trip? | Yes | Yes | Unknown | | Identical<br>(to Otto Bock) |
| Automatic Shut Off? | Yes | Yes | Unknown | | Identical<br>(to Otto Bock) |
| Indicator Display<br>On/Off Status<br>Low Battery | Yes | Yes | Yes | | Identical<br>(to Otto Bock,<br>MyoTrac) |
| Waveform, shape | dual phase, rectangular pulses | Biphasic symmetrical rectangular | asymmetric, balanced pulse | Balanced biphasic | Identical |
| Frequency<br>Mixed<br>Stress<br>Urge | 50 Hz | 35 Hz<br>10 Hz | 12.5, 50, 100, 200 Hz | 10, 12.5, 20, 50, 100, 200<br>Hz | Within range<br>(of MyoTrac,<br>Hollister) |
| Pulse width<br>Mixed<br>Stress<br>Urge | 200 µs/phase | 300 µs/phase<br>500 µs/phase | 0.2 ms | 0.3 or 1 ms | Within range |
| Time<br>On<br>Off | 20 secs<br>10 secs | 9 - 12 s for Stress, 13 s for Urge<br>7 - 9 s for Stress, 3 s for Urge | 2- 20 secs<br>2- 50 secs | 1 - 80 secs<br>0 - 80 secs | Equivalent |
| | | 5 - 25 mins | 0 - 120 mins | 1 - 30 mins | Within range |
| Total Session Time | 12 mins | | | | Identical<br>(to Otto Bock) |
| Max output voltage<br>(500Ω) | 50 Vdc | 50 Vdc | 30 Vdc | 30 Vdc | Identical |
| Max output current<br>(500Ω) | 100 mA | 100 mA | 100 mA | 100 mA | Identical<br>(to Otto Bock) |
| Maximum phase<br>charge (500Ω) | 50 µC | 50 μC | 60 µC | Unknown | Equivalent |
| Electrode surface area | 10.5 cm² x 2 | 21.2 cm² | Various, 1.53 cm² to 10.5 cm² | unknown | Identical<br>(to Otto Bock) |
| Max current density | 4.7 mA/cm² | 4.7 mA/cm² | Various, 4.7 mA/cm² to 32 mA/cm² | | Equivalent. |
| Maximum power<br>density (500Ω) | 14.3 µW/cm² | 23.5 µW/cm² | Various, 11.32 µW/cm² to 22.84 μW/cm² | Unknown | Equivalent |
| Biofeedback | Air pressure, 0-2 psi | EMG | EMG | EMG<br>pressure, 0 - 350 cm H₂O<br>or combination | Equivalent<br>(to Hollister) |
| Materials | | | | | |
| | | 175 x 95 x 30 mm | 102 x 152 x 51 mm | 100 x 70 x 130 mm | Equivalent |
| Dimensions | 8 x 5 x 4 inches | | | | |
| Control housing | ABS plastics | plastics | plastics | plastics | Equivalent |
| Insertion material | Silicone, plastics | plastics | plastics | plastics | Equivalent |
| Packaging or<br>Expiration Dating | 1 year for insertion unit | N/A | N/A | N/A | |
| Sterilization | N/A | N/A | N/A | N/A | |
| User Interaction and Software | | | | | |
| Operational Method:<br>Clinical Use<br>(e.g., ambulatory use, home use) | Clinic or Home use, under direction of physician | Clinic or Home use, under direction of physician | Clinic or Home use, under direction of physician | Clinic or Home use, under direction of physician | Identical<br>(to all) |
| Patient Interaction:<br>Functions<br>Controllable | The patient can control the starting and stopping of each session. However, the device will stop on its own once the session in normally completed. | | | | Equivalent |
| An explanation of how the device interacts with the patient. | | | | | |
| Patient Interaction:<br>Programming<br>Capability | None, programming can only be changed by clinician | None, programming is locked by clinician | None, programming is locked by clinician | | Equivalent<br>(to Otto Bock, MyoTrac) |
| Whether the device can be programmed and to what extent | | | | | |
| Override | Yes | Yes | | | Identical<br>(to Otto Bock) |
| Patient Interaction:<br>Operator<br>Requirements | Intended as part of a complete therapy program with physician coaching. No special knowledge or training; instruction manual provided | Intended as part of a complete therapy program with physician coaching. No special knowledge or training; instruction manual provided | Intended as part of a complete therapy program with physician coaching. No special knowledge or training; instruction manual provided | Intended as part of a complete therapy program with physician coaching. No special knowledge or training; instruction manual provided | Identical<br>(to all) |
| Knowledge or training required of the operator | | | | | |
| Software Level of Concern | Moderate | | | | |
| A statement indicating the Level of Concern and a description of the rationale for that level | | | | | |
| Feature/ Function | InControl InTone<br>(new device) | Otto Bock STIWELL med4<br>(K080950) | MyoTrac Infiniti<br>(K053434) | Hollister evadri<br>(K050483) | Comparison |
| Technology, Engineering, and Performance | | | | | |
| Environmental<br>Specifications | For indoor use only | For indoor use only | For indoor use only | For indoor use only | Identical<br>(to all) |
| Power Source | Nickel metal hydride battery | Li-Ion battery | Four AAA 1.5V alkaline batteries | AC power supply | Equivalent |
| Number of output<br>modes | 1 | 1 | 1 | 1 | Identical<br>(to all) |
| Number of output<br>channels | 1 | 1 | 1 | 1 | Identical<br>(to all) |
| Regulated current of<br>voltage? | Regulated voltage | Regulated current | Unknown | Unknown | Equivalent |
| Firmware controlled? | Yes | Yes | Yes | Yes | Identical<br>(to all) |
| Automatic Overload<br>Trip? | Yes | Yes | Unknown | Unknown | Identical<br>(to Otto Bock) |
| Automatic Shut Off? | Yes | Yes | Unknown | unknown | Identical<br>(to Otto Bock) |
| Indicator Display<br>On/Off Status<br>Low Battery | Yes<br>Yes | Yes<br>Yes | Yes<br>Yes | unknown | Identical<br>(to Otto Bock,<br>MyoTrac) |
| Waveform, shape | dual phase, rectangular pulses | Biphasic symmetrical rectangular | asymmetric, balanced pulse | Balanced biphasic | Identical<br>(to Otto Bock,<br>Hollister) |
| Frequency<br>Mixed<br>Stress<br>Urge | 50Hz | 35 Hz<br>10 Hz | 12.5, 50, 100, 200 Hz | 10, 12.5, 20, 50, 100, 200<br>Hz | Identical<br>(to MyoTrac,<br>Hollister) |
| Pulse width<br>Mixed<br>Stress<br>Urge | 200 µs/phase | 300 µs/phase<br>500 µs/phase | 0.2 ms | 0.3 or 1 ms | Within range |
| Time<br>On<br>Off<br>Total Session Time | 20 secs<br>10 sccs<br>12 mins | 9 - 12 s for Stress, 13 s for Urge<br>7- 9 s for Stress, 3 s for Urge<br>5 - 25 mins | 2 - 20 sccs<br>2 - 50 secs<br>0 - 120 mins | 1 - 80 secs<br>0 - 80 secs<br>1 - 30 mins | Within range |
| Max output voltage<br>(500Ω) | 50 Vdc | 50 Vdc | 30 Vdc | 30 Vdc | Within range<br>Identical<br>(to Otto Bock) |
| Max output current<br>(500Ω) | 100 mA | 100 mA | 100 mA | 100 mA | Identical<br>(to Otto Bock) |
| Maximum phase<br>charge (500Ω) | 50 µC | 50 µC | 60 µC | Unknown | Identical |
| Electrode surface area | 10.5 cm² x 2 | 21.2 cm² | Various, 1.53 cm² to 10.5 cm² | unknown | Equivalent |
| Max current density | 4.7 mA/cm² | 4.7 mA/cm² | Various, 4.7 mA/ cm² to 32 mA/cm² | unknown | Identical |
| Maximum power<br>density (500Ω) | 14.3 µW/cm² | 23.5 µW/cm² | Various, 11.32 µW/ cm² to 22.84 µW/cm² | Unknown | Equivalent |
| Biofeedback | Air pressure, 0.2 psi | EMG | EMG | pressure, 0 - 350 cm H2O<br>or combination | Equivalent<br>(to Hollister) |
K110179
Pg 5 of 7
{5}------------------------------------------------
Section 5 - 510(K) Summary.doc
K110179
pg. 6 of 7
{6}------------------------------------------------
Section 5 - 510(k) Summary.doc
K110179
pg. 7 of 7
{7}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/7/Picture/1 description: The image is a seal for the Department of Health & Human Services USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. In the center of the seal is a stylized image of an eagle.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -W066-G609 Silver Spring, MD 20993-0002
InControl Medical, LLC % Mr. Michael J. Leigh Consultant 12715 Falcon Drive BROOKFIELD WI 53005
FEB 2 2 2012
Re: K110179
Trade/Device Name: InControl InTone Regulation Number: 21 CFR§ 876.5320 Regulation Name: Nonimplanted electrical continence device Regulatory Class: II Product Code: KPI Dated: January 24, 2011 Received: January 31, 2012
Dear Mr. Leigh:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for association to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The I ou may were revelsions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it your de roo additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean r load of act how a a determination that your device complies with other requirements of the Act that I DTPhas intatutes and regulations administered by other Federal agencies. You must or unf I edolul StatuteD and regurements, including, but not limited to: registration and listing confirs min an an and 110 room g (21 CFR Part 801); medical device reporting of medical
{8}------------------------------------------------
Page 2-
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Sincerely yours,
Benjamin R. Twichell
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{9}------------------------------------------------
# Indications for Use
510(k) Number (if known): K110179
InControl InTone Device Name:
Indications For Use:
The InControl InTone device is a non-implanted electrical stimulator indicated for use in the the incontrol in the devious is a non-mail.com income electrical stimulation to the pelvic floor treatment of lemale unnery incontinents. It is intended for acute and ongoing treatment of mixed urinary incontinence where the following results may improve urinary control: mixed "unlary" incontinence" whole the religition of the detrusor muscle through reflexive strengthening of powle feature can be used for muscle re-education purposes.
Federal (USA) law restricts this device to sale by or on the order of a physician.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(Please DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Helmut Kohl
(Division Sign-Off)
Division of Reproductive, Gastro-Renal, and
Urological Devices
510(k) Number K110179
Page 1 of _ 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.