LAPAROTOMY BLADDER NECK SUSPENSION KIT WITH BONE ANCHORS
K971796 · Louisville Laboratories, Inc. · EZL · Aug 12, 1997 · Gastroenterology, Urology
Device Facts
Record ID
K971796
Device Name
LAPAROTOMY BLADDER NECK SUSPENSION KIT WITH BONE ANCHORS
Applicant
Louisville Laboratories, Inc.
Product Code
EZL · Gastroenterology, Urology
Decision Date
Aug 12, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5130
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Louisville Laboratories, Inc. Laparotomy Bladder Neck Suspension Kit With Bone Anchors is intended for bladder neck suspension for female stress incontinence due to urethral hypermobility.
Device Story
Kit for bladder neck suspension; used in laparotomy procedures to treat female stress incontinence caused by urethral hypermobility. Components include template assembly, suture retriever, suture cutter, bone anchors, and suture pulling loop. Surgeon uses template to position sutures at anatomical site; bone anchors secure sutures; standard cutter trims sutures. Device facilitates surgical correction of urethral hypermobility; benefits patient by restoring bladder neck support.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Manual surgical instrument kit. Components: template assembly, suture retriever, suture cutter, bone anchors, suture pulling loop. Delivery mechanism for bone anchors and suture positioning. Manual operation; no energy source. Non-electronic, non-software device.
Indications for Use
Indicated for female patients with stress incontinence due to urethral hypermobility requiring bladder neck suspension procedures.
Regulatory Classification
Identification
A urological catheter and accessories is a flexible tubular device that is inserted through the urethra and used to pass fluids to or from the urinary tract. This generic type of device includes radiopaque urological catheters, ureteral catheters, urethral catheters, coudé catheters, balloon retention type catheters, straight catheters, upper urinary tract catheters, double lumen female urethrographic catheters, disposable ureteral catheters, male urethrographic catheters, and urological catheter accessories including ureteral catheter stylets, ureteral catheter adapters, ureteral catheter holders, ureteral catheter stylets, ureteral catheterization trays, and the gastro-urological irrigation tray (for urological use).
Predicate Devices
Louisville Laboratories Laparoscopic Bone Anchored Urethropexy Instrument Tray (K963076)
{0}------------------------------------------------
K971796
# 510(k) Summary Louisville Laboratories, Inc Laparotomy Bladder Neck Suspension Kit With Bone Anchors
#### Sponsor/Applicant name, address, telephone number, 1.
Louisville Laboratories, Inc. 2400 Crittenden Drive Louisville, Kentucky 40217 (502) 634-5900 Telephone: Facsimile: (502) 634-5959
### Contact person:
Michael Campbell
## Date of summary preparation
May 14, 1997
- Device name 2.
| Trade/proprietary name: | Laparotomy Bladder Neck Suspension Kit With<br>Bone Anchors |
|-------------------------|-------------------------------------------------------------|
| Common/usual name: | Laparotomy Bladder Neck Suspension Kit With<br>Bone Anchors |
| Classification name: | Manual Surgical Instruments<br>Urological Catheter |
- Identification of the predicate or legally marketed device(s) to which equivalence 3. is being claimed
The Louisville Laboratories, Inc. Laparotomy Bladder Neck Suspension Kit With Bone Anchors is substantially equivalent to several devices on the market such as the Louisville laboratories Laparoscopic Bone Anchored Urethropexy Instrument Tray
{1}------------------------------------------------
(K963076), the Microvasive Percutaneous Bladder Neck Suspension Kit, (K931181, K931182, K932925) and several surgical techniques that have been performed for over forty years.
#### Device description 4.
The Laparotomy Bladder Neck Suspension Kit With Bone Anchors is comprised of a template assembly, suture retriever, suture cutter, bone anchors and suture pulling loop. The template assembly, suture retriever and suture cutter were cleared for marketing on April 4, 1997 under K963076, as part of the Louisville Laboratories Laparoscopic Bone Anchored Urethropexy Instrument Tray.
#### న. Intended use
The Louisville Laboratories, Inc. Laparotomy Bladder Neck Suspension Kit With Bone Anchors is intended for bladder neck suspension for female stress incontinence due to urethral hypermobility.
### A statement of how the technological characteristics of the device compare to 6. those of the predicate or legally marketed device(s) cited
The Laparotomy Bladder Neck Suspension Kit With Bone Anchors is substantially equivalent to the Louisville Laboratories Laparoscopic Bone Anchored Urethropexy Instrument Tray and the Microvasive Percutaneous Bladder Neck Suspension Kit. The technological characteristics are similar in that they include several kit components for delivering bone anchors for bladder neck suspension. Both the Laparotomy Bladder Neck Suspension Kit With Bone Anchors and predicate devices are similar in design in that they both use a method or a component for positioning the sutures to the appropriate anatomical site. Both the Laparotomy Bladder Neck Suspension Kit With Bone Anchors and predicate products are similar in that they use a standard scissors or a suture cutter for cutting the sutures.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for Louisville Laboratories. The logo consists of a black and white checkered cross-like symbol on the left, followed by the text "Louisville" on the top line and "Laboratories" on the bottom line. The text is in a serif font and is black in color.
# Premarket Notification Truthful and Accurate Statement for Laparotomy Bladder Neck Suspension Kit with Bone Anchors
I certify that, in my capacity as Regulatory Affairs Specialist of Louisville Laboratories, I believe to the best of my knowledge, that all data and information submitted in this premarket notification for the Laparotomy Bladder Neck Suspension Kit with Bone Anchors are truthful and accurate and that no material fact has been omitted.
---
Signature
4/28/97
Date
David Phelps, Regulatory Affairs Specialist Typed Name and Title
{3}------------------------------------------------
Image /page/3/Picture/2 description: The image shows the address of the Food and Drug Administration. The address is 9200 Corporate Boulevard, Rockville MD 20850. The text is black and the background is white.
Mary McNamara-Cullinane, RAC Staff Consultant AUG 1 2 1997 Medical Device Consultants, Inc. 49 Plain Street North Attleboro, Massachusetts 02760
K971796 Re : Laparotomy Bladder Neck Suspension Kit with Bone Anchors Requlatory Class: II Product Codes: EZL, MBI, HWC, HET Dated: May 14, 1996 Received: May 15, 1997
Dear Ms. McNamara-Cullinane:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to
{4}------------------------------------------------
Page 2 - Mary McNamara-Cullinane, RAC
comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations.
This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.qov/cdrh/dsmamain.html".
Sincerely yours,
Cellia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
510(k) Number (if known):
Device Name: Louisville Laboratories, Inc_
Indications For Use:
The Louisville Laboratories, Inc. Laparotomy Bladder Neck Suspension Kit With Bone I he Louisville Laboutered, fire: 25px crocedures for bladder neck suspension to correct female stress incontinence due to urethral hypermobility.
K971796
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NECESSARY)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
(Per 21 CFR 801.109)
OR
Over-The-Counter Use
(Division Sign-Off)
Division of General Restorative Devices K971726.
510(k) Number (Optional Format 1.2-96)
Louisville Laboratories, Inc. 5/14/97 Laparotomy Bladder Neck Suspension Kit With Bone Anchors - 510(k) Page v
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.