ALL-SILICONE FOLEY BALLOON CATHETER (TWO WAY; THREE WAY, RADIOPAQUE)
K013276 · Sewoon Medical Co., Ltd. · EZL · Feb 15, 2002 · Gastroenterology, Urology
Device Facts
Record ID
K013276
Device Name
ALL-SILICONE FOLEY BALLOON CATHETER (TWO WAY; THREE WAY, RADIOPAQUE)
Applicant
Sewoon Medical Co., Ltd.
Product Code
EZL · Gastroenterology, Urology
Decision Date
Feb 15, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5130
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
For urological use only. Two-Way Catheter: Urethral catheterization for bladder drainage. Three -Way Catheter: Urethral catheterization for bladder drainage and bladder irrigation. Radiopaque Catheter: Urethral catheterization for bladder drainage with radiopaque substance for radiographic visualization.
Device Story
All-silicone Foley balloon catheter; available in two-way and three-way configurations; includes radiopaque options. Device consists of a tube with a drainage eye on the proximal tip and an additional eye for irrigation. Used for bladder drainage and irrigation; radiopaque version allows for radiographic visualization. Available in French sizes 6-26 and balloon capacities 1.5cc-30cc; lengths 29 and 41 cm. Operated by clinicians in clinical settings. Device functions as a passive drainage/irrigation conduit; provides patient benefit by facilitating bladder management.
Clinical Evidence
Bench testing only. Device tested against ASTM F623-99 Standard Performance Specifications for Foley Catheters, covering drainage lumen flow rate, balloon integrity/rupture resistance, inflated balloon response to pullout, balloon volume maintenance, balloon/shaft sizing, deflation reliability, and biocompatibility.
Technological Characteristics
Material: Silicone elastomer. Configuration: 2-way and 3-way lumens. Dimensions: 6-26 French; 1.5cc-30cc balloon; 29 and 41 cm lengths. Radiopaque option available. Sterilization: ETO. Compliance: ASTM F623-99.
Indications for Use
Indicated for urethral catheterization for bladder drainage, bladder irrigation, and radiographic visualization in pediatric and adult male and female patients requiring urological intervention.
Regulatory Classification
Identification
A urological catheter and accessories is a flexible tubular device that is inserted through the urethra and used to pass fluids to or from the urinary tract. This generic type of device includes radiopaque urological catheters, ureteral catheters, urethral catheters, coudé catheters, balloon retention type catheters, straight catheters, upper urinary tract catheters, double lumen female urethrographic catheters, disposable ureteral catheters, male urethrographic catheters, and urological catheter accessories including ureteral catheter stylets, ureteral catheter adapters, ureteral catheter holders, ureteral catheter stylets, ureteral catheterization trays, and the gastro-urological irrigation tray (for urological use).
Predicate Devices
All Silicone Foley Catheters, Rochester Medical Corp. (K981612)
K091767 — 3 WAY SILICONE FOLEY BALLOON CATHETER · Cook Urological, Inc. · Oct 9, 2009
K130908 — DISPOSABLE SILICONE FOLEY CATHETER · Guangdong Baihe Medical Technology Co., Ltd. · Oct 3, 2013
K140667 — WELL LEAD HYDROPHILIC SILICONE FOLEY CATHETER · Well Lead Medical Co., Ltd. · Dec 15, 2014
K201776 — Foley Balloon Catheter for Urology · Apollon Co., Ltd. · Dec 30, 2020
Submission Summary (Full Text)
{0}------------------------------------------------
#### FEB 1 5 2002
#### EXHIBIT 2
Sewoon Medical Device Corp. Ltd. 238-108, Yongdoo-dong Dongdaemoon-gu, Seoul, KOREA Telephone : 82-2-922-6555 Fax: 82-2-922-6558 Contact Person: Gil Whan Lee September 10, 2001 Rev February 8, 2002 510(k) Summary of Safety and Effectiveness
- 1. Identification of the Device: Proprietary-Trade Name: Sewoon All-silicone Foley Balloon Catheter. (Two way Three way, Radiopaque) Classification Name: Urological Catheter 78 EZL Common/Usual Name: Foley Balloon Catheter
- 2. Equivalent legally marketed device: All Silicone Foley Catheters, Rochester Medical Corp. K981612.
- 3. Indications for Use (intended use) For Urological Use only. Two-Way Catheter: Urethral catheterization for bladder drainage. Three -Way Catheter: Urethral catheterization for bladder drainage and bladder irrigation. Radiopaque Catheter: Urethral catheterization for bladder drainage with radiopaque substance for radiographic visualization.
- 4. Description of the Device: The Sewoon Medical Device Corp. Ltd. catheter has a tube with a single drainage eye on the proximal tip. The catheter has an additional eye for irrigation purposes. The catheters are available with a radiopaque option. The catheter is available in a combination of French sizes, capacities and lengths to accommodate pediatric and adult male and female applications. Available catheter lengths are 29 and 41 cm. French sizes from 6 to 26 and balloon sizes 1.5cc to 30cc.
{1}------------------------------------------------
| Element of Comparison | | |
|-----------------------|-----------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------|
| | All Silicone<br>Foley Catheters,<br>Rochester Medical Corp.<br>K 981612 | ALL SILICONE<br>FOLEY BALLOON<br>CATHETER<br>Sewoon Medical Corp., ltd |
| Intended Use | For urological only | For urological only (SAME) |
| Material | Silicone Elastomer | Silicone Elastomer (SAME) |
| Lumen | 2 way, 3way | 2 way, 3way (SAME) |
| Balloon size | 2 way: 1.5cc, 3cc, 5cc, 30cc<br>3 way: 5cc, 30cc | 2 way: 1.5cc, 3cc, 5cc, 30cc (SAME)<br>3 way: 5cc, 30cc (SAME) |
| French size | 2 way: 6Fr, 8Fr, 10Fr, 12Fr,<br>14Fr, 16Fr, 18Fr,<br>20Fr, 24Fr, 26Fr<br>3 way: 16Fr, 18Fr, 20Fr, 22Fr,<br>24Fr, 26Fr | 2 way: 6Fr, 8Fr, 10Fr, 12Fr,<br>14Fr, 16Fr, 18Fr,<br>20Fr, 22Fr 24Fr,<br>26Fr<br>3 way: 16Fr, 18Fr, 20Fr, 22Fr,<br>24Fr, 26Fr (SAME) |
| Lengths available | "Male and Female Lengths"<br>(Actual lengths not published) | SAME: 29 and 41 cm |
| Sterilization | ETO | SAME |
## 5. Safety and Effectiveness, comparison to predicate device:
6. Conclusion: In all respects, the All-Silicone Foley Balloon Catheters are substantially equivalent to one or more silicone balloon catheters that are legally marketed and/or were marketed prior to 1976. Testing and certifications demonstrate that the device meets the ASTM standard referenced above. All Silicone Foley Balloon catheters have been tested to and meet the following test requirements of ASTM F623-99 Standard Performance Specifications for Foley Catheters:
Clause 4.1 " Flow Rate Through the Drainage Lumen"
Clause 4.2 " Balloon Integrity, Resistance to Rupture
Clause 4.3 " Inflated Balloon Response to pullout"
Clause 4.4 "Balloon Volume Maintenance"
Clause 4.5 "Balloon size and Shaft size"
Clause 4.6 "Deflation Reliability"
Clause 4.7 "Biocompatibility"
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird with three curved lines representing its wings or feathers. The logo is black and white.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
## FEB 1 5 2002
Sewoon Medical Device Corp. Ltd. c/o Daniel Kamm, P.E. Regulatory Engineer Kamm & Associates P.O. Box 7007 DEERFIELD IL 60015
Re: K013276 Trade/Device Name: All-Silicone Foley Balloon Catheter (Two way; Three way, Radiopaque) Regulation Number: 21 CFR §876.5130 Regulation Name: Urological catheter and accessories Regulatory Class: II Product Code: 78 EZL Dated: December 24, 2001 Received: December 26, 2001
Dear Mr. Kamm:
Dour MI. Hallini.
We have reviewed your Section 510(k) premarket notification of intent to market the indication We have reviewed your Section 510(L) premaited is substantially equivalent (for the indications
referenced above and have determined the device is substantialy interstate referenced above and nave determined the devices marketed in interstate for use stated in the enclosure) to regally manced tecated Device Americal Device Amendments, or to
commerce prior to May 28, 1976, the enactment of the Rederal Food Drug commerce prior to May 28, 1976, the enaculiers with the provisions of the Federal Food, Drug, devices that have been recrassilled in accordance was as a provisions of the Act. The and Cosmetic Act (Act) that do not require approvate or aparal controls provisions of the Act. The You may, therefore, market the device, subject of annual registration, listing of
general controls provisions of the Act include requirements for annual registrating and general controls provisions of the Act monde roquilibitions against misbranding and and the production of the states adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), a
nd be If your device is classified (see above) more regulations affecting your device can be
it may be subject to additional controls. Existing major regulation EDA may it may be subject to additional controls. Latisting major 1990. In addition, FDA may
found in the Code of Federal Regulations, Title 21, Tarts 800 to 898. In addition, FDA found in the Code of I ederal Rogalations, Four device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean
ing Active and Please be advised that FDA s issualice of a subscunser of the requirements of the Act
that FDA has made a determination that your device Force agencies. You must that FDA has made a delemination und your areas by other Federal agencies. You must or any Federal statutes and regulations anninisco, by on limited to: registration and listing
comply with all the Act's requirements, including, but no arcestion requirement comply with all the Act s requirements, metall); good manufacturing practice requirements as set
(21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice th (2) CFR Part 807); labeling (21 CFR Pair 820); and 820); and if applicable, the electronic
forth in the quality systems (QS) regulation (21 CFR Patt 820); and if applicable, forth in the quality systems (Q5) regulations (S1-542 of the Act); 21 CFR 1000-1050.
product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed notification. The I Dri inding of successification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please II you desire specific advice for your de rice on one of the following numbers, based on the regulation number at the top of this letter:
| 8xx.1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Additionally, for questions on the promote the regulation entitled, "Misbranding Office of Compilance at (301) 377-10577 - 10577 Part 807.97). Other general information on by reletence to premaired nothers (on the Division of Small Manufacturers, your responsibilities and Consumer & its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Nancy C Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
j) Indications for Use
510(k) Number K013276
Device Name: Sewoon Medical Device Corp. Ltd. All-Silicone Foley Balloon Catheter (Two Way; Three Way, Radiopaque)
#### Indications for Use:
Indications for Use:
For urological use only. Two-Way Catheter: Urethral cathererization for bladder drainere and his For urological use only. Two-Way Catheterization for bladder drainage and bladder
drainage. Three -Way Catheter: Urethral catheterization for bladder drainage with drainage. Three -Way Catheter. Urethral catheterization for bladder drainage with radiopaque substance for radiographic visualization
# Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use v
Over the Counter Use_ (Per 21 CFR 801.109)
David C. Ingram
OR
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.