ESOPHAGEAL TTS STENT

K113551 · Taewoong Medical Co., Ltd. · ESW · Feb 14, 2012 · General, Plastic Surgery

Device Facts

Record IDK113551
Device NameESOPHAGEAL TTS STENT
ApplicantTaewoong Medical Co., Ltd.
Product CodeESW · General, Plastic Surgery
Decision DateFeb 14, 2012
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.3610
Device ClassClass 2
AttributesTherapeutic

Intended Use

For use in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors.

Device Story

Esophageal TTS Stent is an implantable metallic stent with a flexible introducer system; used to treat esophageal strictures caused by malignant tumors. Stent consists of Nitinol wire covered with silicone; provides outward radial force on the esophageal luminal surface to establish patency. Device is deployed via a disposable introducer system. Used by clinicians in a medical setting to restore esophageal patency, potentially benefiting patients by alleviating strictures.

Clinical Evidence

Bench testing only. Testing included deployment, expansion force, compression force, dimensional, tensile strength, packaging adhesive, and MR compatibility testing per FDA guidance for esophageal and tracheal prostheses.

Technological Characteristics

Implantable tubular stent made of Nitinol wire covered with silicone; includes a disposable introducer system. Mechanical expansion via radial force. MR compatible.

Indications for Use

Indicated for patients with esophageal strictures resulting from intrinsic and/or extrinsic malignant tumors.

Regulatory Classification

Identification

An esophageal prosthesis is a rigid, flexible, or expandable tubular device made of a plastic, metal, or polymeric material that is intended to be implanted to restore the structure and/or function of the esophagus. The metal esophageal prosthesis may be uncovered or covered with a polymeric material. This device may also include a device delivery system.

Special Controls

The special control for this device is FDA's “Guidance for the Content of Premarket Notification Submissions for Esophageal and Tracheal Prostheses.”

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ 510(k) Premarket Notification Taewoong Medical Esophageal TT #### 5.0 510(k) Summary In accordance with 21 CFR 807.87(h) and (21 CFR 807.92) the 510(k) Summary for the TTS Esophageal Stent device is provided below. | Device Common Name: | Esophageal Prosthesis | |-------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Device Proprietary Name: | Esophageal TTS Stent | | Submitter: | Taewoong Medical Co., Ltd.<br>1-5, Gomak-ri, Wolgot-myeon, Gimpo-si, Gyeonggi-<br>do, Korea 415871<br>TEL: +82-31-996-0641<br>FAX: +82-31-996-0645 | | Contact: | Stephen P. Rhodes<br>Senior Consultant, Medical Devices<br>Biologics Consulting Group, Inc.<br>Phone: (301) 579-3700<br>Email: srhodes@bcg-usa.com<br>Fax: 720-293-0014 | | Classification<br>Regulation: | 21 CFR 878.3610 | | Panel: | General & Plastic Surgery Devices | | Product Code: | ESW | | Predicate Devices: | K080782 - Taewoong Niti-S Esophageal Stent<br>K073667 - Taewoong Niti-S Biliary Stent | # Indication for Use: For use in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors. ### Device Description: The Esophageal TTS Stent consists of an implantable metallic stent and a flexible introducer system. The stent is a flexible, expandable tubular device made of Nitinol wire and covered with silicone. This device also includes a disposable introducer. Upon deployment, the stent imparts an outward radial force on the luminal surface esophagus to establish patency. ### Comparison to the Predicates The Esophageal TTS Stent is substantially equivalent to the Niti-S Esophageal Stent cleared in K080782 with respect to the materials, design, and range of available sizes. {1}------------------------------------------------ K113331 page 2 of 2 The introducer used to deliver the Esophageal TTS Stent is the same as the introducer cleared for use with the Niti-S Biliary Stent (K073667), except that the diameter is slightly larger to match the diameter of the esophageal stent. The materials, length and manufacturing are the same as previously cleared introducer. # Performance Data: Per FDA "Guidance for Industry – Guidance for the Content of Premarket Notifications for Esophageal and Tracheal Prostheses," the following tests were conducted: - . Deployment Testing - Expansion Force Testing . - Compression Force Testing . - Dimensional Testing . - Tensile Strength Testing . The following testing was also conducted: - Packaging Adhesive Testing . - . MR Compatibility ## Summary of Substantial Equivalence: Based on the indications for use, technological characteristics and performance test results, the Esophageal TTS Stent is substantially equivalent to the predicate Niti-S Esophageal Stent and Niti-S Biliary Stent. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image is a black and white seal for the Department of Health & Human Services - USA. The seal is circular with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the seal is a stylized image of an eagle. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 Taewoong Medical Co., Ltd. % Mr. Stephen P. Rhodes Senior Consultant Biologics Consulting Group, Inc. 400 N. Washington St. Suite 100 ALEXANDRIA VA 22314 FEB 1 4 2012 Re: K113551 Trade/Device Name: Esophageal TTS Stent Regulation Number: 21 CFR§ 878.3610 Regulation Name: Esophageal prosthesis Regulatory Class: II Product Code: ESW Dated: December 1, 2011 Received: December 1, 2011 Dear Mr. Rhodes: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or 10 devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical {3}------------------------------------------------ Page 2 - device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, Sincerely yours, Benjamin K-tivko Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ #### Indications for Use Statement 4.0 KII3551 510(k) Number (if known): Device Name: Esophageal TTS Stent Indications For Use: For use in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Helen Remm Division Sign-Off) ivision of Reproductive, Gastro-Renal, and Urological Devices Page 1 of 1 510(k) Number
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...