K032930 · Boston Scientific Corp · ESW · Oct 23, 2003 · General, Plastic Surgery
Device Facts
Record ID
K032930
Device Name
ULTRAFLEX ESOPHAGEAL NG STENT SYSTEM
Applicant
Boston Scientific Corp
Product Code
ESW · General, Plastic Surgery
Decision Date
Oct 23, 2003
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.3610
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The UltraflexTM Esophageal NG Stent System is intended for maintaining esophageal luminal patency in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors and, for occlusion of concurrent esophageal fistula (for covered stents only).
Device Story
Self-expanding metal stent preloaded on flexible delivery catheter; available in bare or covered configurations with proximal or distal release mechanisms. Used for maintaining esophageal luminal patency in malignant strictures and occluding esophageal fistulas. Placed fluoroscopically using radiopaque markers or endoscopically using a 10mm visual marker on the delivery system. Deployment involves manual retraction of a finger ring, which unravels suture crochet knots to release the stent. Used by physicians in clinical settings. Stent provides mechanical support to the esophageal wall to prevent occlusion, improving patient ability to swallow and reducing fistula-related complications.
Clinical Evidence
Bench testing only. No clinical data provided.
Technological Characteristics
Self-expanding metal stent; flexible delivery catheter; proximal or distal release mechanisms; 10mm visual marker on delivery system; suture crochet deployment mechanism. No changes to materials or fundamental design compared to predicate.
Indications for Use
Indicated for patients with esophageal strictures caused by intrinsic and/or extrinsic malignant tumors; covered stents also indicated for occlusion of concurrent esophageal fistula.
Regulatory Classification
Identification
An esophageal prosthesis is a rigid, flexible, or expandable tubular device made of a plastic, metal, or polymeric material that is intended to be implanted to restore the structure and/or function of the esophagus. The metal esophageal prosthesis may be uncovered or covered with a polymeric material. This device may also include a device delivery system.
Special Controls
The special control for this device is FDA's “Guidance for the Content of Premarket Notification Submissions for Esophageal and Tracheal Prostheses.”
{0}------------------------------------------------
OCT 2 3 2003
Boston Scientific Corporation One Boston Scientific Place Natick, MA 01760-1537 508.650.8000 www.bsci.com
### 510(K) Summary
#### 1. Submitter:
Boston Scientific Corporation One Boston Scientific Place Natick, MA 01760-1537
Contact: Kathleen Morahan Regulatory Affairs Manager Date Prepared: September 17, 2003
#### 2. Device:
Trade Name: Ultraflex™ Esophageal NG Stent System Common Name: Esophageal Stent Classification Name: Esophageal Prosthesis
#### 3. Predicate Device:
BSC UltraflexTM Esophageal Stent System - K955347
#### 4. Device Description:
The proposed Ultraflex™ Esophageal NG Stent System consists of a self-expanding metal stent preloaded on a flexible delivery catheter. The stent is offered either bare or covered and with either a proximal release or distal release delivery system.
The stent may be placed flouroscopically using radiopaque markers as a guide or endoscopically using the visual marker on the delivery catheter. To deploy the stent, hold the handle hub in the palm of one hand, and grasp the finger ring with the other hand. As the finger ring is retracted, the suture crochet knots are unraveled in a circular manner along the length of the stent, gradually deploying the stent. When the stent is completely released, the delivery catheter is removed.
### 5. Intended Use:
The UltraflexTM Esophageal NG Stent System is intended for maintaining esophageal luminal patency in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors and, for occlusion of concurrent esophageal fistula (for covered stents only).
{1}------------------------------------------------
### 6. Technological Characteristics:
The technological characteristics of the proposed and predicate devices are unchanged. The crochet suture pattern on the delivery catheter was modified to create a 10mm "visual marker" on the delivery system (for proximal release product codes only) to assist with endoscopic stent placement. The suture material threaded around the circumference of each stent end was also modifications have not impacted the device performance as demonstrated by bench testing.
# 7. Performance Data:
Bench testing was performed to aid with the establishment of substantial equivalence of the proposed device with the predicate device.
# 8. Conclusion:
BSC has demonstrated that the Ultraflex™ Esophageal NG Stent System is substantially equivalent to BSC's currently marketed Ultraflex™ Esophageal Stent System.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo features a stylized eagle with three horizontal lines above it, representing the department's mission to protect the health of all Americans and provide essential human services. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
Public Health Service
OCT 2 3 2003
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Ms. Kathleen Morahan Manager, Regulatory Affairs Microvasive Endoscopy Boston Scientific Corporation One Boston Scientific Place NATICK MA 01760-1537
Re: K032930
Trade/Device Name: Ultraflex " Esophageal Next Generation (NG) Stent System Noncovered and Covered, Distal Release; Noncovered and Covered, Proximal Release Regulation Number: 21 CFR §878.3610 Regulation Name: Esophageal prosthesis Regulatory Class: II Product Code: 79 ESW Dated: September 18, 2003 Received: September 23, 2003
Dear Ms. Morahan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If vour device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter:
| 8xx. 1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
510(K) Number:
To Be Determined K032930.
Device Name:
Ultraflex™ Esophageal NG Stent System
Indications for Use:
The Ultraflex™ Esophageal NG Stent System (non-covered) is intended for maintaining esophageal luminal patency in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors only.
The Covered Ultraflex™ Esophageal NG Stent System is intended for maintaining esophageal luminal patency in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors only, and occlusion of concurrent esophageal fistula.
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
(Per 21 CFR 801.1091)
(Optional Format 1-2-96)
(Division Sign-Off)
OR
Over-The-Counter Use ***_
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number _*** K032930
Special 510(k) for the Ultraflex™ Esophageal NG Stent System, September 17, 2003 Proprietary and Confidential Information of Boston Scientific Corporation
000017
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.