K113310 · Byrne Medical, Inc. · FCX · Jun 20, 2012 · Gastroenterology, Urology
Device Facts
Record ID
K113310
Device Name
ENDOGATER AVANTAGE CO2 INSUFFLATOR
Applicant
Byrne Medical, Inc.
Product Code
FCX · Gastroenterology, Urology
Decision Date
Jun 20, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1500
Device Class
Class 2
Intended Use
The EndoGator Advantage CO2 Insufflator is designed to use CO2 as a distention media in the gastrointestinal tract when used in conjunction with a gastrointestinal endoscope. The EndoGator Advantage CO2 Insufflator contains a CO2 warming feature that allows the clinician to modulate and control delivery of CO2 to the patient that has been warmed to a target temperature 37°C.
Device Story
EndoGator Advantage CO2 Insufflator EGA-501 receives CO2 from pressurized source; regulates pressure/flow via two-stage regulator and electrical solenoid flow restrictors. Device delivers CO2 through endoscope to distend GI tract for improved visualization. Features include user-selectable flow settings (low, medium, high), pressure relief safety valve, and flow-through heating element (target 37°C) with dual redundant resistance temperature detector sensors and safety thermostat. Optional water bottle warmer accessory connection available. Used in clinical settings by physicians/clinicians during GI endoscopy. Output allows clinician to modulate CO2 delivery; warming feature aims to match tissue temperature, potentially improving patient comfort. Bench testing confirms performance, safety, and equivalence to predicate.
Clinical Evidence
Bench testing only. Data provided includes simulated insufflation performance, CO2 flow/pressure comparison, heater performance (with/without flow shut-off), pressure relief valve performance, electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), and water bottle warmer performance.
Indicated for use as a CO2 distention media in the gastrointestinal tract during endoscopic procedures, in conjunction with a gastrointestinal endoscope. Includes optional CO2 warming to 37°C.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
K111648 — BRACCO DIAGNOSTICS, INC. CO2MPACT ENDODCOPIC INSUFFLATOR · Bracco Diagnostic, Inc. · Nov 3, 2011
K161554 — ENDOFLATOR 40, ENDOFLATOR 50 · KARL STORZ Endoscopy-America, Inc. · Mar 2, 2017
K180711 — FUJIFILM Endoscopic CO2 Regulator GW-100 · Fujifilm Corporation · May 25, 2018
K053008 — E-Z-EM ENDOSCOPIC CO2 REGULATOR · E-Z-Em, Inc. · Dec 27, 2005
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for MEDIVATORS, a Cantel Medical Company. The logo consists of a circular graphic on the left, followed by the word "MEDIVATORS" in bold, sans-serif font. Below the word "MEDIVATORS" is the text "A Cantel Medical Company" in a smaller, regular font. The logo is simple and professional, conveying a sense of reliability and expertise.
# EndoGator Advantage CO2 Insufflator EGA-501
## JUN 2 0 2012 .
# 510(k) Summary
| Manufacturer: | Medivators<br>(formerly Byrne Medical, A Minntech Corporation Business Group) |
|----------------------|---------------------------------------------------------------------------------------------|
| Address: | 3150 Pollok Dr.<br>Conroe, TX 77303<br>USA |
| Official Contact: | Brent Geiger, MS, RAC<br>International RA Manager<br>bgeiger@medivators.com<br>763-553-3345 |
| Trade Name: | EndoGator Advantage CO2 Insufflator EGA-501 |
| Common Name: | Endoscopic Insufflator |
| Classification Name: | Insufflator, automatic carbon-dioxide for endoscope |
| Product Code: | FCX, Gastroenterology/Urology |
| Device Class: | II |
| Regulation No.: | 876.1500 |
| Date Prepared: | June 8, 2012 |
| Predicate Device: | E-Z-EM Endoscopic CO2 Regulator |
| 510(k) Number: | K053008 |
Medivators has provided the following information to the U.S. Food and Drug Administration to support the substantial equivalence determination of the subject EndoGator Advantage CO2 Insufflator to the predicate E-Z-EM Endoscopic CO2 Regulator device currently cleared for commercial distribution in the United States.
#### Device Description
The EndoGator Advantage CO2 Insufflator EGA-501 is indicated to use CO2 as a distention media in the gastrointestinal tract when used in conjunction with a gastrointestinal endoscope. It is intended to be used for delivery of CO2 via an endoscope system to provide improved visualization during gastrointestinal endoscopic procedures.
The EndoGator Advantage CO2 Insufflator operates by receiving CO2 from a pressurized source and requlating it down to a lower pressure and flow rate by means of a two-stage pressure regulator and flow restrictor system. The device controls delivery of CO2 to an endoscope system for eventual use as a distention media in endoscopic procedures.
The output flow rate is user adjustable and can be set at low, medium or high settings. Flow is controlled by a system of electrical solenoid flow restrictors. The device contains a pressure relief safety valve as a redundant safety backup to the pressure regulator system.
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for MEDIVATORS, a Cantel Medical Company. The logo consists of a circular graphic with several horizontal lines on the left, followed by the word "MEDIVATORS" in bold, sans-serif font. Below the word "MEDIVATORS" is the text "A Cantel Medical Company" in a smaller, regular font.
The EndoGator Advantage CO2 Insufflator includes a CO2 warming feature that allows the clinician to modulate and control delivery of CO2 that has been warmed to a target temperature of 37°C. The insufflator warms CO2 using a flow-through heating element with built-in dual redundant resistance temperature detector sensors and an added safety thermostat.
The device design also contains an optional feature that allows for connection of an externally attachable water bottle warmer accessory. The water bottle warmer system design is identical to the design used in the related Endogator Advantage Irrigation Pump EGA-500 (K113119). The feature was included in the insufflator design to provide the user with the option for warm water irrigation in the event that they choose to use an alternative irrigation pump to the EGA-500.
#### Indications for Use
The EndoGator Advantage CO2 Insufflator is designed to use CO2 as a distention media in the gastrointestinal tract when used in conjunction with a gastrointestinal endoscope.
The EndoGator Advantage CO2 Insufflator contains a CO2 warming feature that allows the clinician to modulate and control delivery of CO2 to the patient that has been warmed to a target temperature 37°C.
### Comparison to Other Devices in Commercial Distribution in the United States
Both the subject and predicate devices have the same intended use and general performance as endoscopic CO2 insufflators. They are both intended to regulate and administer CO2 as a distention media in the gastrointestinal tract when used in conjunction with a qastrointestinal endoscope. The EndoGator Advantage CO2 Insufflator has the same basic indications for use as the predicate device, but the indications do differ in that the subject device contains a CO2 warming feature that allows the clinician to modulate and control delivery of CO2 to the patient that has been warmed to a target temperature of 37°C while the predicate does not. The added CO2 warming feature does not present additional risks as it is simply being used to introduce CO2 that is heated to approximately the same target temperature as the surrounding tissues in qastrointestinal (GI) tract.
Certain other performance and functional characteristics of the subject device such as user selectable modes of operation, unit displays/controls, output pressure, pressure relief setting and optional water warming capability do differ slightly from the predicate, however evaluation and testing has been performed to demonstrate that the differences do not impact device safety or performance, and any risks associated with the differences are acceptable.
#### Summary of Non-Clinical Performance Data
Medivators has performed bench testing to support the substantial equivalence of the EndoGator Advantage CO2 Insufflator to the predicate E-Z-EM Endoscopic CO2 Regulator cleared via K053008 for commercial distribution in the United States. The following types of data were provided to FDA to support substantial equivalence and demonstrate that the device performs consistently, reliably and safely as intended:
- Simulated insufflation performance comparison .
- CO2 flow and pressure comparison testing .
- CO2 heater performance .
- CO2 heater performance with flow shut-off ◆
K113310 Response Page 2 of 3
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for Medivators, a Cantel Medical Company. The logo consists of a circular graphic on the left, followed by the word "MEDIVATORS" in bold, uppercase letters. Below the word "MEDIVATORS" is the text "A Cantel Medical Company" in a smaller font size. The logo is simple and professional, and it is likely used on the company's website, marketing materials, and products.
K113310
page 3 of 3
- CO2 temperature and flow through endoscope system .
- CO2 pressure relief valve performance .
- Risk analysis .
- Electrical safety IEC 60601-1 .
- Electromagnetic compatibility IEC 60601-1-2 .
- Water bottle warmer performance .
#### Conclusion
Medivators has provided the appropriate premarket notification and supporting safety and performance information in the form of a 510(k) submission. Based on the information provided, we believe that the EndoGator Advantage CO2 Insufflator EGA-501 is substantially equivalent to the E-Z-EM Endoscopic CO2 Regulator device.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized eagle emblem with three overlapping, curved shapes representing the eagle's head and wings.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Mr. Brent Geiger, MS, RAC International RA Manager Medivators (formerly Byrne Medical, A Minntech Corporation Business Group) 14605 28th Avenue North MINNEAPOLIS MN 55442
JUN 2 0 2012
Re: K113310
Trade/Device Name: EndoGator® Advantage CO2 Insufflator, Model EGA-501 Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: Class II Product Code: FCX Dated: June 11, 2012 Received: June 12, 2012
Dear Mr. Geiger:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{4}------------------------------------------------
Page 2 -
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerély wours,
Daymond K-torko
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Traditional 510(k) Submission Byrne Medical, Conroe, Texas 77303
# Indications for Use
510(k) Number (if known) KII 3310
Device Name: EndoGator® Advantage CO2 Insufflator
Indications for Use:
The EndoGator® Advantage CO2 Insufflator is designed to use CO2 as a distention media in the gastrointestinal tract when used in conjunction with a gastrointestinal endoscope.
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use 2.1 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Vogn. M. Whz
Division Sign-Off
Division of Reproductive, Gastro-Renal, and
Urological Devices
510(k) Number K113310
#### Section 6
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.