The Nexcore GI Insufflator is designed to use CO2 as a distention media in the gastrointestinal tract when used in conjunction with a gastrointestinal endoscope. The Nexcore GI Insufflator allows the physician to modulate and control the CO2 to the patient in an identical manner to that being done currently with room air which is supplied by an air pump in the light source of the endoscopic equipment.
Device Story
Reusable electronic medical device; delivers CO2 to GI tract via endoscope to improve visualization, patient comfort, and procedure efficiency. Inputs: CO2 from pressurized cylinder/hospital supply. Operation: Primary/secondary regulators reduce pressure; electronically controlled proportional valve regulates flow rate using differential pressure across in-line orifice. Output: CO2 gas at user-adjustable flow rates (high, medium, low, ultra-low). Used in hospitals/outpatient endoscopy suites; operated by physicians. Safety features: Redundant pressure relief valves; audible/visual alarms; automatic shutdown. Benefits: Improved visualization and patient comfort compared to room air. Healthcare providers use touchscreen interface to control delivery; output affects distention levels during endoscopy.
Clinical Evidence
Bench testing only. Performance testing compared output to predicate device, defined software control algorithms, and verified distal pressure limits against anatomical maximums. Software verification and validation conducted per FDA guidance (moderate level of concern). Electrical safety and EMC testing performed per IEC 60601-1 and IEC 60601-1-2. Biocompatibility evaluation performed per ISO 10993-1.
Technological Characteristics
Reusable electronic insufflator; mains-powered. Materials: Compliant with ISO 10993-1. Sensing: Differential pressure across in-line orifice. Actuation: Electronically controlled proportional valve. Safety: Redundant pressure relief valves (mechanical/electronic). Standards: IEC 60601-1 (3.1), EN 60601-1-2 (3.0), IEC 60601-1-8 (2.1), AAMI/ANSI/IEC 62304, ISO 14971, IEC 62366-1.
Indications for Use
Indicated for use as a CO2 distention media source during gastrointestinal endoscopic procedures in patients requiring GI tract insufflation.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:
1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an emblem featuring three stylized human profiles facing to the right, representing health and human services.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 29, 2016
Nexcore Technology, LLC % Olaf Teichert Responsible Third Party Official TUV SUD AMERICA, INC. 1775 Old Highway 8 NW New Brighton, MN 55112-1891
Re: K162332
> Trade/Device Name: Nexcore GI Insufflator Regulation Number: 21 CFR§ 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: FCX Dated: August 30, 2016 Received: August 31, 2016
Dear Olaf Teichert,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
For Division
# Douglas Silverstein -S 2016.09.29 14:07:51 -04'00'
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K162332
Device Name Nexcore GI Insufflator
Indications for Use (Describe)
The Nexcore GI Insufflator is designed to use CO2 as a distention media in the gastrointestinal tract when used in conjunction with a gastrointestinal endoscope.
The Nexcore GI Insufflator allows the physician to modulate and control the CO2 to the patient in an identical manner to that being done currently with room air which is supplied by an air pump in the light source of the endoscopic equipment.
Type of Use (Select one or both, as applicable)
| <span style="font-size:16px">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|------------------------------------------------------------------------------------|
| <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Nexcore Technology, LLC. The logo consists of a teal square with a white "X" inside of it, followed by the company name in black, sans-serif font. The word "NEXCORE" is in a larger font size than the words "TECHNOLOGY, LLC", which are stacked below it.
### 510(k) Summary
| Submitter: | Nexcore Technology, LLC |
|-----------------------------|----------------------------------------------------------------------------------------------------|
| Address: | 150 Hopper Avenue<br>Waldwick, NJ 07463 USA |
| Submitter Contact: | Milton Frank<br>CEO<br>mfrank@nexcoretech.com<br>Telephone: 201-968-9400 x202<br>Fax: 201-968-0597 |
| Trade Name: | Nexcore GI Insufflator |
| Common Name: | Endoscopic Insufflator |
| Classification Name: | Insufflator, automatic carbon-dioxide for endoscope |
| Product Code: | FCX |
| Regulatory Class: | II |
| Regulation Number: | 21 CFR Part 876.1500 |
| Classification Panel: | Gastroenterology and Urology |
| Date Prepared: | July 21, 2016 |
| Predicate Device: | Bracco CO2MPACT Endoscopic Insufflator (P/N: 710300) |
| Predicate<br>510(k) Number: | K111648 |
# Device Description:
The Nexcore GI Insufflator is designed to use CO2 as a distention media in the gastrointestinal tract when used in conjunction with a gastrointestinal endoscope.
{4}------------------------------------------------
The Nexcore GI Insufflator is a reusable electronic medical device that delivers CO2, via a gastrointestinal (GI) endoscope, to provide improved visualization, increase patient comfort, and increase procedure efficiency during GI endoscopic procedures. The insufflator operates by receiving CO2 from a pressurized cylinder or hospital infrastructure supply and regulating it down in stages to a lower pressure by means of primary and secondary regulators. Each stage of pressure, including the output pressure, has a pressure relief valve as a redundant safety backup for the regulator system. The flow rate is controlled by an electronically controlled proportional valve, which utilizes differential pressure across an in-line orifice to measure flow rate. The output flow rate is user adjustable with four pre-defined settings: high, medium, low and ultra-low.
The device contains software that receives inputs from a touchscreen user interface and controls the operation of the device, including the delivery of CO2 to the endoscope. There are no direct patient-contacting components in this device, other than the delivered CO2 gas. The device is reusable and is not sterile. The Nexcore GI Insufflator is used in hospitals and outpatient endoscopy suites. The device operates from mains power.
There is one model (NX-350) of this device. The accessories sold with the device are the line (power) cord and a pin-indexed high pressure hose for CO2 cylinder connection.
The Nexcore GI Insufflator is used with commercially available, separately cleared devices, including a GI endoscope and tube set. The appropriate compatible tube set is intended to provide CO2 or air (if CO2 is not used) to an endoscope for endoscopic procedures and must have a ≤0.2 micron hydrophobic filter, check valve, and luer fittings at both ends. The tubing ID must be sufficient to deliver 4L/min.
### Intended Use:
The Nexcore GI Insufflator is designed to use CO2 as a distention media in the gastrointestinal tract when used in conjunction with a gastrointestinal endoscope.
The Nexcore GI Insufflator allows the physician to modulate and control the CO2 to the patient in an identical manner to that being done currently with room air which is supplied by an air pump in the light source of the endoscopic equipment.
### Comparison to Predicate Device
The proposed and predicate devices have identical intended uses and identical principles of operation. The fundamental technological characteristics of the two devices are the same. Both are electronic medical devices that take in CO2 from a pressurized gas source and meter it to a tube set and GI endoscope in a controlled fashion. The proposed and predicate devices have substantially equivalent specifications, with only minor differences in a few specifications that do not raise different questions of safety and
{5}------------------------------------------------
effectiveness. Performance data demonstrate that the proposed device is safe and effective.
Many aspects of the designs were compared including: biocompatibility, inlet mains power, hose sets, inlet pressure, outlet overpressure, selectable flow rates, selectable shutdown times, form factor and gas supply configurations. All aspects are fundamentally equivalent, with only minor differences.
The minor differences between the proposed and the predicate devices include:
- The Nexcore GI Insufflator has a staged/redundant overpressure method as follows: an audible alarm with on-screen alert at 7.5psi; shutdown switch activated at 8psi; mechanical relief valve opens at 10psi. The predicate device has a mechanical relief valve that opens at 7.25psi. The Nexcore GI Insufflator limits output pressure to 5psi, by controlling flow rate. In order for these overpressure devices to react, the system would have to be in a fault mode where the 5psi limit is not functioning.
- The Nexcore GI Insufflator offers an "Ultra-Low" flow rate setting. The predicate device does not offer an "Ultra-Low" flow rate setting. The Ultra-Low setting was added to address market needs.
- The Nexcore GI Insufflator offers an intermediate 90 minute timer. The predicate o device does not offer an intermediate 90 minute timer. The 90 minute setting was added to address market needs.
The proposed and predicate devices are substantially equivalent.
### Summary of Non-Clinical Testing
### Performance (Bench) Testing
Non-clinical testing was performed on the Nexcore GI Insufflator. The purpose of the testing was to compare the output the predicate device against the Nexcore GI Insufflator. Follow up non-clinical testing was performed to further focus and define the software's control algorithms. This testing also included a distal pressure test. This was conducted to ensure the maximum delivered pressure does not exceed average anatomical maximum pressure for the GI tract.
In summary, the non-clinical testing proved equivalent performance to the existing FDA cleared product along with delivered pressure not exceeding maximum anatomical pressures. The Nexcore GI Insufflator meets its design inputs.
{6}------------------------------------------------
#### Software Testing
Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "moderate" level of concern, since a non-mitigated failure or latent flaw in the software could result in minor injury to the patient, perforated colon.
## Electrical Safety and Electromagnetic Compatibility (EMC) Testing
Electrical safety and EMC testing were conducted on Nexcore GI Insufflator. The device complies with the IEC 60601-1 standard for safety and the IEC 60601-1-2 standard for EMC.
### Biocompatibility
A biocompatibility evaluation was conducted on Nexcore GI Insufflator in comparison to a 510(k)-cleared reference device, the Xylog Insufflator (K945970). Based on this evaluation, the device complies with ISO 10993-1 for biocompatibility.
## Standards Applied
The following standards were applied to the development and design of the Nexcore GI Insufflator:
| Standard Code | Date Code | Edition/Revision | Description |
|------------------------|-----------|------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| IEC 60601-1 | 08-2012 | 3.1 | Medical Electrical Equipment Part 1: General<br>Requirements for Basic Safety and Essential<br>Performance |
| EN 60601-1-2 | 03-2007 | 3.0 | Medical Electrical Equipment Part 1-2: General<br>Requirements for Basic Safety and Essential<br>Performance- Collateral Standard: Electromagnetic<br>Disturbances- Requirements and Tests |
| IEC 60601-1-8 | 11-2012 | 2.1 | Medical Electrical Equipment Part 1-8: General<br>Requirements for Basic Safety and Essential<br>Performance- Collateral Standard: General<br>Requirements, Tests and Guidance for Alarm Systems<br>in Medical Electrical Equipment and Medical Electrical<br>Systems |
| AAMI/ANSI/IEC<br>62304 | 05-2006 | 1.0 | Medical Device Software- Software Life Cycle<br>Processes |
| ISO 14971 | 07-2012 | 2012 | Medical Devices- Application of Risk Management to<br>Medical Devices |
| IEC 62366-1 | 02-2015 | 1.0 | Medical Devices- Part 1: Application of Usability<br>Engineering to Medical Devices (General 1 (QS/RM)) |
{7}------------------------------------------------
| AAMI/ANSI/ISO<br>10-2009<br>10993-1 | 2009 (R)<br>2013 | Biological Evaluation of Medical Devices Part 1:<br>Evaluation and Testing Within a Risk Management<br>Process |
|-------------------------------------|------------------|----------------------------------------------------------------------------------------------------------------|
|-------------------------------------|------------------|----------------------------------------------------------------------------------------------------------------|
#### Conclusion
Non-clinical verification and validation activities were conducted to establish the performance and safety characteristics of the Nexcore GI Insufflator. The results of these activities demonstrate that the Nexcore GI Insufflator is safe and effective when used in accordance with its intended use and labeling. The Nexcore GI Insufflator has the same intended use and the same fundamental technological characteristics as the predicate device. The minor differences between the proposed and predicate devices do not raise different questions of safety and effectiveness. Performance data demonstrate that the proposed device is safe and effective. Therefore, the Nexcore GI Insufflator is considered substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.