PILLING WECK 2 PIECE TAKE-APART INSTRUMENTS

K965176 · Pilling Weck Group · OCZ · Mar 4, 1997 · Gastroenterology, Urology

Device Facts

Record IDK965176
Device NamePILLING WECK 2 PIECE TAKE-APART INSTRUMENTS
ApplicantPilling Weck Group
Product CodeOCZ · Gastroenterology, Urology
Decision DateMar 4, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Pilling Weck 2-Piece Take-Apart Instrumentation is designed to be used endoscopically through cannulae to perform cutting, grasping, dissecting, retracting, and manipulating functions.

Device Story

Reusable endoscopic instruments consisting of handles and interchangeable jaw/shaft assemblies; used for cutting, grasping, dissecting, retracting, and manipulating tissue. Operated by physicians during endoscopic procedures via cannulae. Features include 360-degree rotation of jaws via handle knob; insulated or non-insulated handle options; flushing port for cleaning. Assembly involves inserting working ends through handle shaft, secured by detented proximal end and push-button lock. Device facilitates cleaning through take-apart design.

Clinical Evidence

Bench testing only.

Technological Characteristics

Reusable endoscopic instruments; stainless steel/medical grade materials; mechanical actuation; modular design (handle + interchangeable jaw/shaft); 360-degree rotation; flushing port for cleaning; push-button locking mechanism.

Indications for Use

Indicated for use in endoscopic procedures requiring cutting, grasping, dissecting, retracting, and manipulating of tissues through cannulae.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} 1/3 Pilling Weck 510(k) Summary MAR - 4 1997 K65176 1. Submitter's Name, Address and Telephone Number, and Contact Person: Miss Karenann J. Brozowski Pilling Weck Group Tall Pines Park Jaffrey, NH 03452 Telephone Number: (603)532-7706 Contact Person: Miss Karenann J. Brozowski 2. Name of the Device, including the trade or proprietary name if applicable, and the common name, and the classification name, if known; Trade Name: Pilling Weck 2-Piece Take-Apart Instruments Common Name: Endoscopic Forceps/Graspers/Scissors Classification Name: Endoscopic and Accessories 3. Identification of the legally marketed device that the submitter claims equivalence. The devices in consideration are substantially equivalent to the devices currently marketed by Karl Storz Endoscopy America, Inc. Under 510(k) K952149. A Teleflex Company One Weck Drive, P.O. Box 12600 Research Triangle Park, North Carolina 27709 (919) 544-8000 {1} 2/3 Page 2 4. A description of the device that is the subject of the premarket notification submission, such as might be found in the labeling or the promotional material of the device. The Pilling Weck 2-Piece Take-Apart Instrumentation are endoscopic instruments, which have jaws designed to cut, grasp, dissect, retract, and manipulate. The reusable handles are offered in both insulated and noninsulated actuating rings that adapt to various interchangeable jaw configurations with integral shafts. The shaft of the jaws fits into the handle and can be taken apart for cleaning. The jaws are rotated by turning the knob on the handle. 5. A statement of the intended use of the device that this the subject of the premarket notification submission: The Pilling Weck 2-Piece Take-Apart Instrumentation is designed to be used endoscopically through cannulae to perform cutting, grasping, dissecting, retracting, and manipulating functions. 6. If the device has the same technological characteristics as the predicate device identified in paragraph (a)(3) of this section, a summary of the technological characteristics of the new device in comparison to those of the predicate device. If the device has different technological characteristics from the predicate device, a summary of how the technological characteristics of the device compare to the legally marketed device. The intended uses of the Pilling Weck 2-Piece Take-Apart Instrumentation and the Karl Storz Endoscopy America, Inc. Devices under K952149 have the same intended use, and that is for endoscopy procedures though cannulae for cutting, grasping, dissecting, retracting, and manipulating. {2} 3/3 Page 3 The devices have reusable handles that accommodate inserts, that can be rotated 360 Degrees during use. Both devices include insulated and noninsulated handles, and accommodating shafts. In the Pilling Weck device there is included a flushing port to facilitate cleaning. This port is not present on the Karl Storz devices. All Pilling Weck device handles adapt the interchangeable working ends by inserting them through the inner diameter of the shaft that is an integral part of the handle. The fit is secured by a detented proximal end that locks securely in place by encapsulation with the actuation of a push button. In some of the Storz instruments, security is achieved by a one-quarter turn into the outer tube and then connecting into the outer tube handle. The rest of the Storz instruments are the same as Pilling Weck.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...